CARESORB® - Polyglactin 910 Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures; but not for use in cardiovascular or neurological tissues. CARESORB® RAPID Polyglactin 910 (fast absorbing) Surgical Sutures are indicated for use in superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures.
Device Story
Multifilament, braided, synthetic absorbable surgical sutures composed of 90% glycolide and 10% L-lactide copolymer. Coated with poly(glycolide-co-L-lactide) and calcium stearate. Used by surgeons for soft tissue approximation and ligation. Sutures provide temporary wound support via tensile strength retention before undergoing hydrolysis and absorption. CARESORB RAPID is designed for faster absorption (42 days) for superficial skin/mucosa closure. Available in various USP sizes with or without stainless steel needles. Sterilized by ethylene oxide. Benefits include secure tissue approximation during healing and predictable absorption profiles, reducing need for suture removal.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), physical performance (USP <861>, <871>, <881>), and in-vivo/in-vitro resorption/breaking strength retention studies comparing subject devices to predicates.
Technological Characteristics
Multifilament braided suture; 90% glycolide/10% L-lactide copolymer; coated with poly(glycolide-co-L-lactide) and calcium stearate. USP sizes 6-0 to 1 (CARESORB) and 4-0 to 2-0 (CARESORB RAPID). Sterilized by ethylene oxide. Complies with USP <861> (diameter), <871> (needle attachment), and <881> (tensile strength).
Indications for Use
Indicated for general soft tissue approximation/ligation (CARESORB) or superficial soft tissue approximation of skin/mucosa requiring short-term support (CARESORB RAPID). Contraindicated for cardiovascular or neurological tissues; CARESORB RAPID also contraindicated for ligation and ophthalmic procedures.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Ethicon's Coated VICRYL™ Rapide (Polyglactin 910) Synthetic Absorbable Suture (K033746)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, arranged in a way that they also resemble an abstract bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 29, 2017
CPT Sutures Co., Ltd. % Ms. Natalya Valerio Official Correspondent for CPT Sutures Co., Ltd. MDI Consultants. Inc. 55 Northern Blvd.. Suite 200 Great Neck, New York 11021
Re: K170166
Trade/Device Name: CARESORB® - Polyglactin 910 Surgical Suture, CARESORB RAPID® - Polyglactin 910 (fast Absorbing) Surgical Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: August 31, 2017 Received: September 1, 2017
Dear Ms. Valerio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170166
Device Name CARESORB® - Polyglactin 910 Surgical Suture
Indications for Use (Describe)
CARESORB® - Polyglactin 910 Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures; but not for use in cardiovascular or neurological tissues.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170166
Device Name
CARESORB® RAPID - Polyglactin 910 (fast absorbing) Surgical Suture
### Indications for Use (Describe)
CARESORB® RAPID Polyglactin 910 (fast absorbing) Surgical Sutures are indicated for use in superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
K170166 p. 1 of 7
# 510(k) SUMMARY
The assigned 510(k) number is: K170166
#### 1. Submitter's Identification:
CPT Sutures Co., Ltd. 8 Dao Tri Street Phu Thuan Ward District 7, Ho Chi Minh, Vietnam
Date Summary Prepared: August 31, 2017
Mr. Khoa Do Contact: Vice Director CPT Sutures Co., Ltd. e-mail: khoa.do@cpt-medical.com phone: (+84) 28 38 222 228
#### 2. Name of the Device:
| Trade Name: | CARESORB® - Polyglactin 910 Surgical Suture and<br>CARESORB® RAPID - Polyglactin 910 fast absorbing<br>Surgical Suture |
|--------------------|------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Suture, Absorbable, Synthetic, Polyglycolic Acid |
| Regulation Number: | 21CFR 878.4493 |
| Regulation Name: | Absorbable Poly(glycolide/L-lactide) Surgical Suture |
| Regulatory Class: | Class II |
| Product Code: | GAM |
#### 3. Information for the 510(k) Cleared Device (Predicate Device):
Ethicon's Coated VICRYL™ (Polyglactin 910) Synthetic Absorbable Suture, 510(k) # K022269
Ethicon's Coated VICRYL™ Rapide (Polyglactin 910) Synthetic Absorbable Suture, 510(k) # K033746
#### 4. Device Description:
The CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 are multifilament, braided, sterile synthetic absorbable surgical sutures composed of a copolymer of 90% glycolide and 10% L-lactide. The CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 are coated with copolymer of Poly(glycolide-co-L-lactide) (30/70) and calcium stearate.
The CARESORB® - Polyglactin 910 suture is available dyed with FDA-approved color additive D&C Violet No. 2 - Cl 60725 or undyed in the natural beige color.
{5}------------------------------------------------
The CARESORB® RAPID - Polyglactin 910 suture is available undyed in the natural beige color only.
The CARESORB® sutures are available in USP sizes 6-0 through 1 and CARESORB® RAPID sutures in USP sizes 4-0 through 2-0, in different lengths, with or without a standard needle attached.
The CARESORB® and CARESORB® RAPID sutures meet USP Monograph for Synthetic Absorbable Sutures, except for diameter.
#### 5. Indications for Use:
CARESORB® - Polyglactin 910 Surgical Sutures are indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures; but not for use in cardiovascular or neurological tissues.
CARESORB® RAPID - Polyglactin 910 (fast absorbing) Surgical Sutures are indicated for use in superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required, but not for use in ligation, ophthalmic, cardiovascular or neurological procedures.
#### 6. Comparison to the 510(k) Cleared Devices (Predicate Devices):
| Comparison<br>Criteria | New Device<br>CARESORB® -<br>Polyglactin 910<br>Surgical Suture | New Device<br>CARESORB® RAPID -<br>Polyglactin 910<br>Surgical Suture | Predicate<br>Device(s) |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device | Ethicon's Coated<br>VICRYL™ (Polyglactin<br>910) Synthetic<br>Absorbable Suture,<br>510(k) # K022269 | Ethicon's Coated<br>VICRYL™ Rapide<br>(Polyglactin 910)<br>Synthetic<br>Absorbable Suture,<br>510(k) # K033746 | Image: checkerboard pattern |
| Classification | Synthetic<br>Absorbable<br>Surgical Suture | Synthetic<br>Absorbable<br>Surgical Suture | Same |
| Indications<br>for Use | General soft tissue approximation and / or ligation, including use in ophthalmic procedures, but not in cardiovascular or neurological tissues | Superficial soft tissue approximation of the skin and mucosa, where only short-term wound support (7-10 days) is required but not for use in ligation, ophthalmic, cardiovascular or neurological procedures | Same for each<br>respective<br>predicate |
| Suture Material | Copolymer of 90%<br>glycolide and 10%<br>L-lactide | Copolymer of 90%<br>glycolide and 10%<br>L-lactide | Same |
| Material<br>Presentation | Multifilament, braided | Multifilament, braided | Same for each<br>respective<br>predicate |
| Coating | Coated with copolymer<br>of Poly(glycolide-co-L-<br>lactide) (30/70) and<br>calcium stearate | Coated with copolymer<br>of Poly(glycolide-co-L-<br>lactide) (30/70) and<br>calcium stearate | Same |
| Color Additive | Available dyed with FDA-<br>approved color additive<br>D&C Violet No. 2 - Cl<br>60725 as well as undyed<br>in natural beige color | Available undyed only | Same for each<br>respective<br>predicate |
| Sterilization | Sterilized by<br>Ethylene Oxide | Sterilized by<br>Ethylene Oxide | VICRYL™ is<br>sterilized by<br>Ethylene Oxide<br>VICRYL™ Rapide<br>is sterilized by<br>Gamma Irradiation |
| Single Use | Yes | Yes | Same |
| How Supplied | Sutures of various<br>lengths and gauge sizes<br>of USP 6-0 to 1, with or<br>without the standard<br>stainless steel needle<br>(series 302 with silicon<br>coating) attached.<br>Available as one-dozen<br>and three-dozen boxes. | Sutures of various<br>lengths and gauge sizes<br>of USP 4-0 to 2-0, with<br>or without the standard<br>stainless steel needle<br>(series 302 with silicon<br>coating) attached.<br>Available as one-dozen<br>and three-dozen boxes. | Sutures of various<br>gauge sizes and<br>lengths, with or<br>without standard<br>stainless steel<br>needle attached.<br>Available as one-<br>dozen and three-<br>dozen boxes. |
| Absorption<br>Profile | Suture absorption begins<br>as a loss of tensile<br>strength followed by<br>eventual absorption by<br>hydrolysis. Complete<br>absorption occurs<br>between 56-70 days. | Suture absorption begins<br>as a loss of tensile<br>strength followed by<br>eventual absorption by<br>hydrolysis. Complete<br>absorption occurs in 42<br>days. | Same for each<br>respective<br>predicate |
| Breaking<br>Strength<br>Retention Profile | Percent Breaking<br>Strength Retention (%<br>BSR): > 65% at 14 days;<br>> 40% at 21 days. | Percent Breaking<br>Strength Retention (%<br>BSR): 50% at 5 days;<br>0% at 14 days. | Same for each<br>respective<br>predicate |
| Suture Diameter | Complies with<br>USP <861> except for<br>slight oversize in the<br>suture diameter | Complies with<br>USP <861> except for<br>slight oversize in the<br>suture diameter | Same |
| Suture Length | ≥ 95% of the claimed<br>label length as required<br>by USP | ≥ 95% of the claimed<br>label length as required<br>by USP | Same |
| Needle<br>Attachment | Complies with<br>USP <871> | Complies with<br>USP <871> | Same |
| Tensile Strength | Complies with<br>USP <881> | Complies with<br>USP <881> | Same |
| Biocompatibility | Complies with all ISO<br>10993-1 required tests<br>for absorbable surgical<br>sutures | Complies with all ISO<br>10993-1 required tests<br>for absorbable surgical<br>sutures | Same |
| Packaging | Validated sterile barrier<br>packaging system | Validated sterile barrier<br>packaging system | Same |
{6}------------------------------------------------
{7}------------------------------------------------
### Discussion of Non-Clinical Tests Performed for Determination of Substantial 7. Equivalence:
The below referenced testing demonstrated substantial equivalence of the CARESORB® - Polyglactin 910 Surgical Suture and CARESORB® RAPID -Polyglactin 910 (fast absorbing) Surgical Suture to the Ethicon's Coated VICRYL™ and Ethicon's Coated VICRYL™ Rapide predicate devices respectively. The testing was conducted on the final sterilized devices in accordance with the FDA Class II Special Controls Guidance Document: Surgical Sutures.
# Performance Testing:
- 1) ISO 10993-3 Biological Evaluation of Medical Devices Part 3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity, 2014
- 2) ISO 10993-5 Biological Evaluation of Medical Devices Part 5: Tests for In Vitro Cytotoxicity, 2009 (R) 2014
- 3) ISO 10993-6 Biological Evaluation of Medical Devices Part 6: Tests for Local Effects after Implantation, 2007 (R) 2014
- 4) ISO 10993-7 Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Residuals, 2008 (R) 2012
- 5) ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for Irritation and Skin Sensitization, 2010 (R) 2014
{8}------------------------------------------------
- 6) ISO 10993-11 Biological Evaluation of Medical Devices Part 11: Tests for Systemic Toxicity, 2006 (R) 2010
- 7) USP 39-NF33 Absorbable Surgical Suture, 2016
- 8) USP 39-NF33 <861> Sutures Diameter, 2016
- 9) USP 39-NF33 <871> Sutures Needle Attachment, 2016
- 10) USP 39-NF33 <881> Tensile Strength, 2016
- 11) ISO 11607-1 Section 5.2 Whole package microbial aerosol challenge test, 2014
- 12) ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials, 2015
- 13) ASTM F1140/F1140M-13 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages, 2013
- 14) ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test), 2011
- 15) USP39-NF 34 <71> Sterility Test, 2016
- 16) USP39-NF 34 <85> Bacterial Endotoxins Test, 2016
- 17) ISO 15223-1 Third Edition 2016-11-01, Medical Devices Symbols To Be Used With Medical Device Labels, Labelling, And Information To Be Supplied - Part 1: General Requirements. (General I (QS/RM))
The biocompatibility test results demonstrated no evidence of cytotoxicity, irritation, delayed dermal sensitization or acute systemic toxicity for CARESORB® Surgical Suture. In the short-term, mid-term and long-term Intramuscular Implantation Tests with Histopathology, the CARESORB® Surgical Suture was considered a non-irritant as compared to the predicate VICRYL™ used as a control. In the mid-term Intramuscular Implantation Test with Histopathology, the CARESORB® Rapid Surgical Suture was considered a non-irritant as compared to the predicate VICRYL™ Rapide used as a control.
The CARESORB® Suture was also found to be non-pyrogenic, non-mutagenic and non-genotoxic.
{9}------------------------------------------------
# Resorption Profile:
The resorption profiles of the final sterilized sutures were demonstrated in vivo.
The CARESORB® sutures were compared to the predicate VICRYL™ sutures in the in-vivo implantation study. The study results demonstrated that CARESORB® sutures have similar tensile strength retention profile in comparison to the marketed predicate VICRYL™ sutures.
The in-vivo breaking strength retention profile of CARESORB® RAPID was based on the data provided by the suture material supplier from the in-vivo implantation study. The resorption profile of CARESORB® RAPID was determined to be similar to that of the marketed predicate VICRYL™ Rapide.
The resorption profile of the final sterilized CARESORB® RAPID suture was also demonstrated in vitro.
The CARESORB® RAPID. sterilized by EtO. were compared to the predicate VICRYL™ Rapide sutures, sterilized by gamma irradiation, and to the raw material supplier's sutures, sterilized by gamma irradiation. All sutures were immersed in the phosphate buffer solution (pH 7.4) for 5. 14. 21 and 28 days and then tested for breaking strength retention at each time point. The study results demonstrated that all three sutures had essentially identical resorption profiles under the same study conditions and that tensile strength and tensile strength retention of CARESORB® RAPID was comparable to those of VICRYL™ Rapide and the raw material supplier's sutures.
Additionally, the rate of absorption of CARESORB® sutures was compared to that of predicate VICRYL™ sutures in the in-vivo implantation study. The study results demonstrated that CARESORB® sutures and VICRYL™ sutures have very similar absorption in tissue.
The rate of absorption of CARESORB® RAPID was based on the data provided by the suture material supplier from the in-vivo implantation study. The CARESORB RAPID rate of absorption in tissue was determined to be similar to that of the marketed predicate VICRYL™ Rapide.
## Shelf Life:
CARESORB® and CARESORB® RAPID sutures have been tested in real time stability studies as well as in the accelerated aging studies.
The real time stability studies for both sutures were performed at 18°C to 26°C temperature and relative humidity of no more than 65%. The accelerated aging stability studies were performed at 68°C to 72°C temperature and relative humidity of no more than 65% for CARESORB® and at 53°C to 57°C temperature and relative humidity of no more than 65% for CARESORB® RAPID.
The aged sutures were tested for parameters specified in the USP Monograph for Absorbable Surgical Sutures and other applicable requirements:
{10}------------------------------------------------
- Diameter, USP 39 <861>
- Needle Attachment, USP 39 <871> ●
- Tensile Strength, USP 39 <881> ●
- In vitro Breaking Strength Retention (BSR)
- Visual Inspection ●
- . Moisture content
- Sterility Testing, USP 39-NF34 <71> ●
The study results demonstrated conformance to the specified acceptance criteria and supported the 5 year expiration date assignment for CARESORB® and CARESORB® RAPID sutures.
#### 8. Discussion of Clinical Tests Performed:
Not Applicable
#### 9. Conclusions:
The CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 Surgical Sutures are composed of the same material, have the same intended use and main technological characteristics as their respective predicate devices Ethicon's Coated VICRYL™ and Ethicon's Coated VICRYL™ Rapide Surgical Sutures.
Based on the results of completed performance testing inclusive of physical testing, biocompatibility testing and absorption profile, it can be concluded that CARESORB® - Polyglactin 910 and CARESORB® RAPID - Polyglactin 910 are substantially equivalent in terms of safety and effectiveness to Ethicon's Coated VICRYL™ and Ethicon's Coated VICRYL™ Rapide respectively.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.