← Product Code [PIF](/productcode/PIF) · K170162

# Vesco Medical Nasoenteric Feeding Tubes (K170162)

_Vesco Medical, LLC · PIF · Sep 26, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170162

## Device Facts

- **Applicant:** Vesco Medical, LLC
- **Product Code:** [PIF](/productcode/PIF.md)
- **Decision Date:** Sep 26, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Vesco Medical Nasoenteric Feeding Tubes (NE Tubes), with and without flow-through stylet, are for the administration of nutrition, fluids, and medications to the stomach or small bowel by the nasoenteric route in pediatric, adult or elderly patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition. The Vesco Medical Nasoenteric Feeding Tubes have not been tested for use longer than short-term use (<30 days).

## Device Story

Nasoenteric feeding tube; flexible, radiopaque polyurethane tubing; single-port ENFit connector; tethered cap; distal tip with four eyelets; hydrophilic coating on distal tip for insertion friction reduction; graduated centimeter markings for depth; optional flow-through stylet wire for stiffening during insertion. Used in clinical settings for enteral feeding; operated by healthcare professionals. Device delivers nutrition/fluids/meds to stomach or small bowel; ENFit connector prevents misconnections to non-enteral devices. Benefits patient by providing nutritional support when oral intake is impossible.

## Clinical Evidence

No clinical data. Performance demonstrated via non-clinical bench testing, including biocompatibility (ISO 10993), physical/visual inspections, patency/flow rate, leakage (EN1615/EN1618), tensile strength, kinking resistance, and ENFit connector performance (ISO 80369-3).

## Technological Characteristics

Radiopaque polyurethane tubing; ENFit connector; hydrophilic coating; graduated markings. Complies with ISO 80369-3 (small-bore connectors), EN1615:2000 (enteral catheters), EN1618:1997, and EN 13868:2002. Biocompatibility per ISO 10993-1/5. Non-sterile, single-use. Sizes 8-14 French.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Covidien Dobbhoff™ Dual Port Feeding Tubes ([K112511](/device/K112511.md))
- Ross Product Division Abbott Laboratories Flexiflo Enteral Feeding Tubes ([K992494](/device/K992494.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure with three profiles facing right, representing health and well-being. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 26, 2017

Vesco Medical, LLC Rachel Kunzweiler Product Development Engineer 692 N. High Street, Suite 205 Columbus. OH 43215

Re: K170162

Trade/Device Name: Vesco Medical Nasoenteric Feeding Tubes Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PIF Dated: August 31, 2017 Received: August 31, 2017

Dear Rachel Kunzweiler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Charles Viviano -S

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K170162

Device Name Vesco Medical Nasoenteric Feeding Tubes

Indications for Use (Describe)

The Vesco Medical Nasoenteric Feeding Tubes), with and without flow-through stylet, are for the administration of nutrition, fluids, and medications to the stomach or small bowel by the nasoenteric route in pediatric, adult or elderly patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition. The Vesco Medical Nasoenteric Feeding Tubes have not been tested for use longer than short-term use (<30 days).

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# TAB 5 510(k) Summary

#### Submitter I.

| Official Contact    | Name: Chris O'Keefe<br>Title: VP Product Development and Innovation<br>Email: cokeefe@vescomedical.com |
|---------------------|--------------------------------------------------------------------------------------------------------|
|                     | Vesco Medical<br>Address: 692 N. High Street, Suite 205ef<br>Columbus, Ohio 43215                      |
|                     | Phone: 614-914-5991<br>Fax: 614-902-3275                                                               |
| Date of Preparation | September 24, 2017                                                                                     |

#### II. Device

| Name of Device:                       | Vesco Medical Nasoenteric Feeding Tubes                        |
|---------------------------------------|----------------------------------------------------------------|
| Common/Usual Name:                    | Gastrointestinal Tubes                                         |
| Device Classification:                | Class II                                                       |
| Classification Name/<br>Product Code: | Gastrointestinal Tubes and Accessories (21 CFR 876.5980) / PIF |

#### Legally Marketed Predicate Devices III.

- Covidien Dobbhoff™ Dual Port Feeding Tubes were most recently cleared under notification K112511.
- Ross Product Division Abbott Laboratories Flexiflo Enteral Feeding Tubes were most recently cleared under notification K992494.

#### Device Description IV.

The Vesco Medical Nasoenteric Feeding Tubes (NE Tubes) are non-sterile, single use devices consisting of flexible, radiopaque, polyurethane tubing adhered to a plastic male ENFit single-port connector with a tethered, twist-on cap. The plastic ENFit port allows for connection to enteral feeding sets and enteralspecific syringes while reducing the likelihood of misconnections to non-enteral devices. The cap allows for port coverage when enteral devices are not in use. The distal tip of the tubing is open and has four (4) evenly spaced, alternating eyelet openings to allow for first three inches of the distal tip

{4}------------------------------------------------

are coated with a hydrophilic lubricant to aid in friction reduction during tube insertion. The tubing is printed with graduated markings in centimeters, the length from the distal tip in order to provide users with approximate insertion depth information. Certain models of the Vesco Medical Nasoenteric Feeding tube are provided with a stylet wire attached to a flow-through, ENFit hub. The stylet wire is intended to stiffen the tube during insertion and be removed after insertion is complete. The proposed

#### V. Intended Use

The Vesco Medical Nasoenteric Feeding Tubes (NE Tubes), with and without flow-through stylet, are for the administration of nutrition, fluids, and medications to the stomach or small bowel by the nasoenteric route in pediatric, adult or elderly patients who have an intact gastrointestinal tract but are physically unable to manage nutritional intake through normal mastication and deglutition. The Vesco Medical Nasoenteric Feeding Tubes have not been tested for use longer than short-term use (<30 days).

device will be offered in sizes ranging between 8 French and 14 French.

## Comparison of Technological Characteristics with the Predicate VI. Device

The proposed Vesco Medical Nasoenteric Feeding Tubes and the predicate devices are all intended for administering nutrition, fluids, and medications through the nastoenteric route to the stomach or small bowel of patients unable to manage nutritional intake through normal mastication and deducts have similar intended uses, technological characteristics, manufacturing methods, and operating principles. There are no significant differences that would affect the performance of the Vesco Medical Nasoenteric Feeding Tubes as compared to the predicate devices. Table 5.1 lists the comparisons of the proposed device to the predicate devices.

| Design Features/Function of<br>Proposed Device    | Proposed Device Compared to<br>Covidien Dobbhoff Dual Port<br>Feeding Tubes K112511<br>(Predicate)            | Proposed Device Compared to<br>Ross Products Division Abbott<br>Flexiflo Enteral Feeding Tubes<br>K992494 (Predicate) |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Intended Use                                      | Similar and SE                                                                                                | Similar and SE                                                                                                        |
| Environment of Use                                | Similar and SE                                                                                                | Similar and SE                                                                                                        |
| Intended Users                                    | Similar and SE                                                                                                | Similar and SE                                                                                                        |
| Patient Population                                | Similar                                                                                                       | Same                                                                                                                  |
| Single Use                                        | Same                                                                                                          | Same                                                                                                                  |
| Nonsterile                                        | Same                                                                                                          | Same                                                                                                                  |
| Stomach and Small Bowel<br>Placement              | Same                                                                                                          | Same                                                                                                                  |
| ENFit Connector                                   | Different (predicate has catheter<br>slip-tip connector); no decrease<br>in performance of proposed<br>device | Different (predicate has catheter<br>slip-tip connector)                                                              |
| Single Port                                       | Different (predicate has dual port)                                                                           | Different (predicate has dual port)                                                                                   |
| Flow-through Stylet Hub                           | Unknown                                                                                                       | Different (predicate's hub is non-<br>ENFit and of different material)                                                |
| No Bolus Weight                                   | Unknown                                                                                                       | Similar (with or without bolus)                                                                                       |
| Radiopaque Tube Material                          | Similar and SE                                                                                                | Same                                                                                                                  |
| Hydrophilic Coating                               | Similar and SE                                                                                                | Same                                                                                                                  |
| Markings on Tube for Measuring<br>Inserted Length | Similar and SE                                                                                                | Same                                                                                                                  |

Table 5.1: Comparison of Vesco Medical Nasoenteric Feeding Tubes to predicate devices reqarding substantial equivalence (SE), similarities, and differences.

{5}------------------------------------------------

| Traditional 510(k)          |                | Vesco Medical Nasoenteric Feeding Tubes |
|-----------------------------|----------------|-----------------------------------------|
| Used With or Without Stylet | Similar and SE | Same                                    |
| French Sizes                | Same           | Same                                    |
| Lengths                     | Similar and SE | Same                                    |

#### Performance Data: VII.

# Non-Clinical Tests

Verification and validation Testing was performed with the Vesco Medical Nasoenteric Feeding Tubes. It was found that Vesco Medical Nasoenteric Feeding Tubes are in compliance with design and performance requirements according to ISO 80369 "Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applications," and EN1615:2000 "Enteral feeding catheters and enteral giving sets for single use and their connectors: Design and Testing." Additionally, Vesco Medical Nasoenteric Feeding Tubes that underwent accelerated aging conditioning were tested in order to substantiate an eighteen (18) month shelf life.

The following testing was conducted on the Vesco Medical Nasoenteric Feeding Tubes:

- 1. Biocompatibility Testing per ISO-10993-1 and ISO 10993-5
	- Cytotoxicity performed per ISO-10993-5 on Vesco Medical Nasoenteric Feeding Tubes a.
	- Cytotoxicity, intracutaneous, and maximization sensitization testing reports provided b. by Cedic for component with same materials, manufacturer, and manufacturing process as stylet hub.
	- C. Additional biocompatibility testing for the ENFit connector material and colorant leveraged from Cedic's previously cleared K140581.
	- d. Remaining biocompatibility testing leveraged from K992494 due to material and manufacturing equivalence (stylet wire, tubing, ink, lubricant, and adhesive).
- 2. Visual and Physical Inspections
	- a. Visual inspection for bleeding of tube ink markings per Vesco Medical Design Requirement
	- Visual inspection for legibility of tube ink markings per Vesco Medical Design ﻘ Requirement
	- Visual inspection for cracks or delamination of the lubricous coating per Vesco Medical C. Design Requirement
	- Physical inspection for water activation of lubricious coating per Vesco MedicalDesign d. Requirement
	- Visual Inspection for adequate adhesion between extruded NE Tube and ENFit e. connector per Vesco Medical Design Requirement
	- Patency of stylet per Vesco Medical Design Requirement f.
	- q. Stylet removal force per Vesco Medical Design Requirement
- 3. Patency of NE Tubes/flow rate testing (compared against predicate)
	- a. Using water
	- Using nutritional fluid b.
- 4. Leakage Testing per EN1615:2000 & EN1618:1997
- 5. Tensile Testing per EN1615:2000 & EN1618:1997
- 6. Resistance to Kinking per EN1615:2000 & EN 13868:2002, against predicate
- 7. Fluid Leakage per ISO 80369-3:2016, performed by Cedic
- 8. Stress Cracking per ISO 80369-3:2016, performed by Cedic
- Resistance to separation from axial load ISO 80369-3:2016, performed by Cedic 9.
- 10. Resistance to separation from unscrewing per ISO 80369-3:2016, performed by Cedic
- 11. Resistance to overriding per ISO 80369-3:2016, performed by Cedic
- 12. Disconnection by unscrewing per ISO 80369-3:2016, performed by Cedic
- 13. ENFit dimensional verification per ISO 80369-3:2016, performed by Cedic

{6}------------------------------------------------

Page 4 nt 4

- 14. Risk Analysis per ISO 14971:2012
- 15. Design Failure Modes and Effects Analysis (DFMEA)

# Clinical Tests

Clinical tests were not required to demonstrate performance of Vesco Medical Nasoenteric Feeding Tubes. Product functionality has been adequately assessed by non-clinical tests.

# Animal Tests

Animal tests were not required to demonstrate the performance of Vesco Medcical Nasoenteric Feeding Tubes. Product functionality has been adequately assessed by non-animal tests.

# VIII. Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the Vesco Medical Nasoenteric Feeding tubes are as safe, as effective, and perform as well as or better than the legally marketed devices identified in part III, "Legally Marketed Predicate Devices" of this section.

---

**Source:** [https://fda-staging.innolitics.com/device/K170162](https://fda-staging.innolitics.com/device/K170162)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
