New Deantronics E Green™ Electrosurgical Monopolar Pencils
K170161 · New Deantronics Taiwan , Ltd. · GEI · Mar 29, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170161
Device Name
New Deantronics E Green™ Electrosurgical Monopolar Pencils
Applicant
New Deantronics Taiwan , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The New Deantronics E Green™ Electrosurgical Monopolar Pencils are intended to remove tissue and control bleeding by use of high-frequency electrical current.
Device Story
Electrosurgical monopolar pencil used in open surgeries to remove tissue and control bleeding; utilizes high-frequency RF electrical current. Device consists of two parts: a gamma-sterilized disposable electrosurgical pencil and a reusable monopolar cord (autoclavable up to 100 times). Operated by surgeons or clinical staff in hospital or clinic settings. The device delivers RF energy to the surgical site via a stainless steel electrode. The healthcare provider manually controls the device via a button switch on the pencil body. The output (tissue cutting/coagulation) is immediate and visual, allowing the surgeon to manage bleeding and perform tissue resection during procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through electrical safety (IEC 60601-1, 60601-2-2), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1, 5, 10, 12), sterilization validation (ISO 11137, ISO 17665), and usability engineering (IEC 62366-1).
Technological Characteristics
Monopolar electrosurgical pencil; RF energy source. Materials: stainless steel electrode, HDPE overmold, ABS/Nylon pen body, Silicone/TPV/PPSU cord. Dimensions: 12.3 cm body, 10-15 ft cord. Rated voltage: 4600 V. Sterilization: Gamma (pencil), Steam (cord). Standards: IEC 60601-1, IEC 60601-2-2, ISO 10993, ISO 11137, ISO 17665, ASTM F1980.
Indications for Use
Indicated for use in surgical procedures to remove tissue and control bleeding via high-frequency electrical current. Intended for use in hospital and clinic environments.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
New Deantronics Disposable Hand Switching Pencil, Push Button, 2 Piece With and Without Holster (K982744)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2017
New Deantronics Taiwan, Ltd % Mr. Lewis Ward Consultant L.W. Ward and Associates Incorporated 4655 Kirkwood Court Boulder, Colorado 80301
Re: K170161
Trade/Device Name: New Deantronics E Green™ Electrosurgical Monopolar Pencils Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: March 16, 2017 Received: March 22, 2017
Dear Mr. Ward:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K170161
Device Name
New Deantronics E Green™ Electrosurgical Monopolar Pencils
Indications for Use (Describe)
The New Deantronics E Green™ Electrosurgical Monopolar Pencils are intended to remove tissue and control bleeding by use of high-frequency electrical current.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* |
| of this information collection, including suggestions for reducing this burden, to: | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect |
| Food and Druq Administration<br>Office of Chief Information Officer<br>PRAStaff(@fda.hhs.gov | Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Staff |
| | |
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 510(k) Summary
### Submitter:
New Deantronics Taiwan Ltd. 12F., No.51, Sec. 4, Chong Yang Rd., Tu Cheng Dist, New Taipei City 23675, Taiwan R.O.C. (886) 2-2268-1726 Phone
大瓏企業股份有限公司
New Deantronics Taiwan, Ltd.
新北市土城區中央路4段51號12樓
12F., No.51, Sec. 4,
Chong Yang Rd., Tu Cheng Dist.,
New Taipei City 23675,
Taiwan R.O.C.
Tel: 886 2 2268-1726
Fax: 886 2 2268-3800
### US
I.
II.
(886) 2-2268-3800 Fax
### Contact Person:
Lewis Ward L.W. Ward and Associates, Inc. 4655 Kirkwood Court Boulder, CO 80301 USA 303-530-3279 lwward@qwest.net Date Prepared: Jan. 11, 2017
Trade Name: New Deantronics E Green™ Electrosurgical Monopolar Pencil
Common Name: Electrosurgical accessory, Electrosurgical Monopolar Pencil
Classification: 21 CFR 878.4400, electrosurgical cutting and coagulation device and III. accessories. Class 2, General and Plastic Surgery Panel 79
IV. Product Code: GEI
### V. Indications for Use:
The New Deantronics E Green™ Electrosurgical Monopolar Pencils are intended to remove tissue and control bleeding by use of high-frequency electrical current.
### VI. Predicate Device
510(k) 982744. New Deantronics Disposable Hand Switching Pencil, Push Button, 2 Piece With and Without Holster.
Device Description and Technological Characteristics VII.
The E Green™ Reusable Monopolar Cord is an electrosurgical accessory used with E Green™ Disposable Electrosurgical Pencil. Electrosurgical Pencils are intended to remove tissue and control bleeding by use of high-frequency electrical current.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for New Deantronics Group. The logo features a blue letter 'D' with a red circle inside, followed by the text 'DEANTRONICS' in blue, bold letters. Below the main logo, the words 'New Deantronics Group' are written in a smaller, less bold font.
This product is a monopolar, electrosurgical pencil which is used for most open surgeries. The pencil can be divided in two parts, one is a gamma sterilized disposable electrosurgical pencil and another is a reusable cord which needs to be autoclaved by the user before use in a standard operating room environment. The E Green™ Reusable Monopolar Cord is used ONLY in combination with E Green™ Disposable Electrosurgical Pencil. The number of uses of the Reusable Monopolar Cord following steam sterilization is 100 times.
This device and the predicate devices have the similar technological characteristics, operating principles and performance characteristics. The proposed device is equivalent to the identified
大瓏企業股份有限公司
New Deantronics Taiwan, Ltd.
新北市土城區中央路4段51號12樓
12F., No.51, Sec. 4,
Chong Yang Rd., Tu Cheng Dist.,
New Taipei City 23675,
Taiwan R.O.C.
Tel: 886 2 2268-1726
Table 1: New Deantronics E Green™ Reusable Monopolar Cord
| Item | Catalog No. | ND P/N | Description |
|------|-------------|----------|----------------------------------------|
| 1 | GMC01 | P0001NE1 | E Green™ Reusable Monopolar Cord, 10ft |
| 2 | GMC015 | P0001NE2 | E Green™ Reusable Monopolar Cord, 15ft |
Table 2: New Deantronics E Green™ Disposable Electrosurgical Pencil
predicate device with respect to technological characteristics and function.
| Item | Catalog No. | ND P/N | Description |
|------|-------------|----------|--------------------------------------------|
| 1 | GPEN01 | P0101SE1 | E Green™ Disposable Electrosurgical Pencil |
### VIII. Non-Clinical Testing
IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2:2014: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
IEC 60601-2-2: 2009 + C1:2014, Medical electrical equipment = Part 2-2: Particular requirements for the safety of high frequency surgical equipment.
ISO 14971:2007, Medical Device- Application of Risk Management to Medical Devices
ISO 10993-1: 2009. Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process.
ISO 10993-5: 2009, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity.
ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
ISO 10993-12: 2012, Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference Materials.
ISO 11137-2:2013, Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose.
U.S.
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image contains the logo for New Deantronics. The logo features the word "NEW" in a light gray outlined font, followed by a blue circle with a red dot in the center. To the right of the circle is the word "DEANTRONICS" in a bold, dark blue font. Below the logo is the text "New Deantronics Group" in a smaller, dark blue font.
ISO 11607:2006, Packaging for terminally sterilized medical devices -Part 1: Requirements for materials, sterile barrier systems and packaging systems [Including: Amendment 1 (2014)]; Part 2: Validation requirements for forming, sealing and assembly processes [Including: Amendment 1 (2014)].
ISTA 2A Packaged-Products 150 lb (68 kg) or Less
U.S
大瓏企業股份有限公司
New Deantronics Taiwan, Ltd.
新北市土城區中央路4段51號12樓
12F., No.51, Sec. 4,
Chong Yang Rd., Tu Cheng Dist.,
New Taipei City 23675,
Taiwan R.O.C.
Tel: 886 2 2268-1726
Fax: 886 2 2268-3800
criteria for cleaning reusable medical devices. ISO 17665-1:2006/(R) 2013, Sterilization Of Health Care Products -- Moist Heat -- Part 1:
AAMI TIR30:2011, A compendium of processes, materials, test methods, and acceptance
Requirements For The Development, Validation, And Routine Control Of A Sterilization Process For Medical Devices.
ANSVAAMI ST81:2004/(R) 2010, Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices.
IEC 62366-1:2015, Medical devices -- Part 1: Application of usability engineering to medical devices
ASTM F1980-07:2011, Standard guide for accelerated aging of sterile barrier systems for medical devices. (Sterility)
### Non-clinical testing conclusion:
All the test results demonstrate the performance of New Deantronics E Green™ Electrosurgical Monopolar Pencil meet the requirements of its pre-defined acceptance criteria and intended uses. The results confirm that the New Deantronics E Green™ Electrosurgical Monopolar Pencil is as safe and effective as the predicate devices.
## IX. Substantial Equivalence
The New Deantronics E Green™ Electrosurgical Monopolar Pencil is substantially equivalent to the New Deantronics Disposable Hand Switching Pencil, Push Button, 2 Piece With and Without Holster (K982744) based on intended use, technology, materials, and performance. The claim of substantial equivalence is supported by information the information provided in this submission.
| X. Key Features Comparison | | |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| FEATURE | New Deantronics E Green™<br>Electrosurgical Monopolar Pencil | New Deantronics Disposable Hand<br>Switching Pencil, Push Button<br>(K982744) |
| Intended Use | E Green™ Electrosurgical Monopolar<br>Pencils are intended to remove tissue and<br>control bleeding by use of high-frequency<br>electrical current. | Electrosurgical Pencils are intended to<br>remove tissue and control bleeding by use of<br>high-frequency electrical current. |
| Use | E Green™ Disposable Electrosurgical<br>Pencil is single use.<br>E Green™ Reusable Monopolar Cord is<br>reusable. | Single Use |
### X. Kev Features Comparison
{6}------------------------------------------------
| FEATURE | New Deantronics E Green™<br>Electrosurgical Monopolar Pencil | New Deantronics Disposable Hand<br>Switching Pencil, Push Button<br>(K982744) |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Use Environment | Hospital, clinic | Hospital clinic |
| Dimensions | Pen body length : 12.3 cm<br>Cord length : 10 ft and 15 ft | Pen body length : 16.2 cm<br>Cord length : 10 ft and 15 ft |
| Rated Voltage | 4600 V | 4600 V |
| Material | Electrode: stainless steel, HDPE overmold<br>Pen body: ABS, Nylon<br>Cord: Silicone, TPV, PPSU | Electrode: stainless steel, HDPE overmold<br>Pen body: ABS, Nylon<br>Cord: PVC |
| Specification | Button switch Two piece body | Button switch Two piece body |
| Energy Source<br>Type | RF Energy | RF Energy |
| Sterile method<br>and SAL | E Green™ Disposable Electrosurgical<br>Pencil: Gamma irradiation, SAL 10-6.<br>E Green™ Reusable Monopolar Cord:<br>Steam, SAL 10-6. | Gamma irradiation, SAL 10-6. |
# Taiwan
大瑞企業股份有限公司 New Deantronics Taiwan, Ltd. 新北市土城區中央路 4段 51 號 12 樓 12F., No.51, Sec. 4, Chong Yang Rd., Tu Cheng Dist., New Taipei City 23675, Taiwan R.O.C. Tel: 886 2 2268-1726 Fax: 886 2 2268-3800
-------------------------
### U.S.
New Deantronics Ltd. 1990 North California Blvd. Suite 1040 Walnut Creek, CA. 94596 Tel: (925) 280 8388 Fax: (925) 280 1788
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.