K170124 · KLS Martin L.P. · HRS · Mar 9, 2017 · Orthopedic
Device Facts
Record ID
K170124
Device Name
Level One Hand Plating System
Applicant
KLS Martin L.P.
Product Code
HRS · Orthopedic
Decision Date
Mar 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Level One Hand Plating System is used for stabilization of fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, wrist, fingers, feet, ankles and toes.
Device Story
The Level One Hand Plating System consists of metallic plates, washers, and screws for small bone fixation. Plates are pre-contoured to match patient anatomy, ranging from 0.6mm to 3.0mm in thickness. Screws are self-tapping, available in standard or multidirectional locking configurations, with diameters from 1.0mm to 2.7mm and lengths from 2mm to 32mm. The system includes necessary surgical instrumentation. Used in clinical/hospital settings by qualified physicians for osteosynthesis. The device is provided sterile via gamma irradiation to facilitate immediate surgical use. The system stabilizes bone segments to support healing and reconstruction.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical performance data, including biocompatibility testing per ISO 10993 and bacterial endotoxin testing per ANSI/AAMI ST72:2011.
Technological Characteristics
Materials: CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136). Components: Plates, washers, screws. Sterilization: Gamma irradiation. Design: Pre-contoured plates, self-tapping screws. No software or electronic components.
Indications for Use
Indicated for stabilization and fixation of fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, wrist, fingers, feet, ankles, and toes in patients requiring surgical bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 24, 2017
KLS Martin LP Mr. Gary Moore Ouality Management and Regulatory Affairs Manager 11201 Saint Johns Industrial Pkwy S Jacksonville, Florida 32246
Re: K170124
Trade/Device Name: Level One Hand Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: January 10, 2017 Received: January 13, 2017
Dear Mr. Moore:
This letter corrects our substantially equivalent letter of March 9, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K170124
Device Name
Level One Hand Plating System
#### Indications for Use (Describe)
The Level One Hand Plating System is used for stabilization of fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, wrist, fingers, feet, ankles and toes.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for KLS Martin L.P., a medical technology company. The logo is in red and gray, with the letters "KLS" in red and "martin L.P." in gray. The image also includes the company's address, phone number, fax number, and website. The address is 11201 Saint Johns Industrial Pkwy S Jacksonville, FL 3224.
11201 Saint Johns Industrial Pkwy S ● Jacksonville, FL 32246 www.klsmartinnorthamerica.com
# Section 5
### 510(k) Summary 21 CFR 807.92
| Submitter: | KLS Martin LP<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gary Moore<br>Quality Management and Regulatory Affairs Manager<br>Phone: 800-625-1557<br>Email: gmoore@klsmartin.com |
| Date Prepared: | January 10, 2017 |
| Trade Name: | Level One Hand Plating System |
| Common Name: | Plate, Fixation, Bone |
| Classification Name: | Single/multiple component metallic bone fixation appliances<br>and accessories (21 CFR 888.3030) |
| Regulatory Class: | II |
| Product Code: | HRS |
| Predicate Device: | LINOS MOH Hand Plating System (K141489) |
| Device Description: | The Level One (L1) Hand Plating System includes metallic<br>plates, washers, and screws intended for small bone fixation.<br>Plates are pre-contoured to accommodate patient anatomy,<br>available in various shapes and range in thickness from<br>0.6mm - 3.0mm, and are compatible with the standard and<br>multidirectional locking screws offered in the system.<br>Screws are self-tapping, available in a standard or<br>multidirectional locking configuration and range in diameter<br>from 1.0mm - 2.7mm with lengths from 2mm - 32mm.<br>Standard screws may be used alone or in conjunction with<br>the washers or plates for small fragment osteosynthesis.<br>Implants are manufactured from CP Titanium (ASTM F67)<br>or Ti-6Al-4V (ASTM F136). The system includes the<br>necessary instrumentation to facilitate implantation.<br>The purpose of this submission is to offer the previously<br>cleared non-sterile device system in K141489 in sterile<br>packaging via gamma irradiation. |
| Indications for Use: | The Level One Hand Plating System is used for stabilization<br>and fixation of fractures, revision procedures, joint fusion<br>and reconstruction of small bones of the hand, wrist, fingers. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for KLS Martin L.P. The letters "KLS" are in red, while the word "martin" is in gray. The letters "L.P." are in black.
1201 Saint Johns Industrial Pkwy S . Jacksonville, FL 32246 904-641-7746 ● 800-625-1557 ● Fax: 904-641-7378 klsmartinnorthamerica.com
### Technological Characteristics & Substantial Equivalence Discussion
### Similarities to Predicate
The subject and predicate device systems are identical with respect to intended use, materials, principle of operation, design, range of sizes offered, and mechanical performance. Anatomical sites and fixation methods used for the subject and predicate devices are identical. The subject and predicate devices are for prescription use only by qualified and trained physicians in healthcare facilities/hospitals.
### Differences to Predicate
The only difference between the subject and predicate device system is the method of sterilization. The predicate device system includes components that must be sterilized by the end-user. The subject device components will be provided sterile in sterile packaging via gamma irradiation.
### Non-Clinical Performance Data
Biocompatibility endpoints were evaluated in accordance with ISO 10993. The results of the testing did not product unexpected findings compared to the predicate device, and therefore, adequately addresses biocompatibility for implants with a permanent duration of contact.
Testing was performed in accordance with ANSI/AAMI ST72:2011 for detection and quantification for the presence of bacterial endotoxins. The results of the testing demonstrate that the subject device in sterile conforms to the required EU/device for implants and meet pyrogen limit specifications.
The subject device encompasses the previously cleared predicate devices to offer them sterile using gamma irradiation. No changes have been made to the mechanical properties, design, mechanism of action, intended use, or manufacturing process.
#### Clinical Performance Data
Clinical testing was not necessary for the determination of substantial equivalence.
### Conclusion
The L1 Hand Plating System has the same intended use and technological features compared to the predicate device. The L1 Hand product line will be offered in sterile packaging and have different stock numbers from the originally cleared stock numbers to identify the product as sterile. The similarities in technological characteristics do not raise new issues of safety or effectiveness and the performance data provided in the submission demonstrate the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.