← Product Code [NKB](/productcode/NKB) · K170108

# Xultan 5.5 Pedicle Screw System (K170108)

_Met One Technologies · NKB · May 2, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170108

## Device Facts

- **Applicant:** Met One Technologies
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** May 2, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Xultan 5.5 Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients. The Xultan 5.5 Pedicle Screw System is intended for posterior, pedicle fixation as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine (T1-S2/ilium); degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvature (i.e., scoliosis and/or lordosis); tumor; and failed previous fusion (i.e., pseudoarthrosis).

## Device Story

The Xultan 5.5 Pedicle Screw System is a spinal implant set used for posterior, pedicle fixation to immobilize and stabilize spinal segments. The system consists of titanium and silicone screws (available in cannulated or non-cannulated versions), straight and curved rods, crosslinks, and associated surgical instrumentation. It is used by surgeons in an operating room setting to provide structural support as an adjunct to spinal fusion. The device is implanted to treat various spinal instabilities and deformities. Clinical benefit is derived from the mechanical stabilization of the spine, which facilitates fusion and addresses conditions such as DDD, trauma, and scoliosis.

## Clinical Evidence

Bench testing only. The device underwent static axial compression bending, static torsion, and dynamic axial compression bending testing in accordance with ASTM F1717-15. Results confirmed that the mechanical strength is sufficient for the intended use and equivalent to predicate devices.

## Technological Characteristics

Materials: Titanium and silicone. Components: Cannulated/non-cannulated screws, straight/curved rods, crosslinks, instrumentation. Mechanical testing: ASTM F1717-15 (static axial compression bending, static torsion, dynamic axial compression bending).

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- International Synergy™ VLS ([K000236](/device/K000236.md))
- S 100 PSS ([K111940](/device/K111940.md))

## Reference Devices

- AVS® Cage Spacer ([K143163](/device/K143163.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be connected.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 2, 2017

Met One Technologies % Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Testing Corp. 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K170108

Trade/Device Name: Xultan 5.5 Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: NKB Dated: January 12, 2017 Received: April 6, 2017

Dear Mr. Maxwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Vincent J. Devlin -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170108

Device Name

Xultan 5.5 Pedicle Screw System

#### Indications for Use (Describe)

The Xultan 5.5 Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients. The Xultan 5.5 Pedicle Screw System is intended for posterior, pedicle fixation as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar and sacral spine (T1-S2/ilium); degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvature (i.e., scoliosis and/or lordosis); tumor; and failed previous fusion (i.e., pseudoarthrosis).

| Type of Use (Select one or both, as applicable)                                                                                         |                                                                                                                                        |
|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;"> </span> <span style="font-size: 10pt;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size: 10pt;"> </span> <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Submitter's Name:          | Met One Technologies                                             |
|----------------------------|------------------------------------------------------------------|
| Submitter's Address:       | 154 N. Festival Dr., Ste. F<br>El Paso, TX 79912                 |
| Submitter's Telephone:     | 915.301.0834                                                     |
| Contact Person:            | Kenneth C. Maxwell II<br>Empirical Testing Corp.<br>719.337.7579 |
| Date Summary was Prepared: | 05 April 2017                                                    |
| Trade or Proprietary Name: | Xultan 5.5 Pedicle Screw System                                  |
| Common or Usual Name:      | Thoracolumbosacral Pedicle Screw System                          |
| Classification:            | Class II per 21 CFR §888.3070 Device Classification              |
| Product Code:              | NKB                                                              |
| Classification Panel:      | Division of Orthopedic Devices                                   |

#### 510(K) SUMMARY

## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Xultan 5.5 Pedicle Screw System is an implant device manufactured from titanium and silicone. The screws are available cannulated or non-cannulated in various diameters and lengths to accommodate various patient anatomies. The system includes straight rods, curved rods, crosslinks, and associated instrumentation.

## INDICATIONS FOR USE

The Xultan 5.5 Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. The Xultan 5.5 Pedicle Screw System is intended for posterior, pedicle fixation as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine (T1-S2/ilium): degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvature (i.e. scoliosis, kyphosis and/or lordosis); tumor; and failed previous fusion (i.e. pseudoarthrosis).

### TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates:

- Indications for use
- Structural support mechanism
- Principles of operation

Table 5-1: Predicate Devices

{4}------------------------------------------------

| 510k<br>Number | Trade or Proprietary<br>or Model Name | Manufacturer       | Predicate<br>Type |
|----------------|---------------------------------------|--------------------|-------------------|
| K000236        | International<br>Synergy™ VLS         | INTERPORE<br>CROSS | Primary           |
| K111940        | S 100 PSS                             | Renovis            | Additional        |
| K143163        | AVS® Cage Spacer                      | Stryker            | Reference         |

# PERFORMANCE DATA

The Xultan 5.5 Pedicle Screw System has been tested in the following test modes:

- Static axial compression bending per ASTM F1717-15
- Static torsion per ASTM F1717-15 ●
- Dynamic axial compression bending per ASTM F1717-15 o

The results of this non-clinical testing show that the strength of the Xultan 5.5 Pedicle Screw System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

# CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Xultan 5.5 Pedicle Screw System is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K170108](https://fda-staging.innolitics.com/device/K170108)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
