K170080 · Jjgc Industria E Comercio DE Materiais Dentarios S.A. · NHA · Nov 2, 2017 · Dental
Device Facts
Record ID
K170080
Device Name
Neodent Implant System - CM Pro PEEK Abutment
Applicant
Jjgc Industria E Comercio DE Materiais Dentarios S.A.
Product Code
NHA · Dental
Decision Date
Nov 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Pro PEEK Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability.
Device Story
Neodent Implant System - CM Pro PEEK Abutments are temporary prosthetic components; installed onto CM dental implants to support provisional prostheses for up to 6 months. Device consists of customizable PEEK cylindrical body and titanium alloy base with Morse taper connection; includes coupled screw for fixation. Used by dentists in clinical/surgical settings; customization performed chairside using hand-milling tools under aseptic conditions. Device maintains/shapes soft tissue during healing phase. Benefits include temporary stabilization of prosthesis and soft tissue management prior to final restoration. No CAD/CAM or automated design allowed.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993-1, -5, -18), dynamic fatigue testing (ISO 14801), and sterilization validation (ISO 11135, ISO 10993-7). Shelf life validated via accelerated aging (ASTM F1980).
Technological Characteristics
Materials: PEEK (customizable body) and Titanium alloy Ti-6Al-4V ELI (implant-to-abutment interface). Connection: 11.5° Morse taper with hexagonal indexing. Dimensions: 4.5mm and 6.0mm diameters; 0.8mm to 5.5mm lengths. Sterilization: Ethylene Oxide (EO). Connectivity: None (mechanical).
Indications for Use
Indicated for patients requiring temporary support for dental prosthesis structures (single or multiple units) on Neodent implants for up to 6 months. Suitable for one or two-stage procedures and immediate loading when good primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
JJGC Industria e Comercio de Materiais Dentarios SA % Jennifer Jackson Senior Manager, Regulatory Affairs & Quality Management Straumann USA, LLC 60-100 Minuteman Rd Andover, Massachusetts 01810
November 2, 2017
Re: K170080
Trade/Device Name: Neodent Implant System - CM Pro PEEK Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: September 25, 2017 Received: September 26, 2017
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mary S. Runner -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K170080
Device Name
Neodent Implant System - CM Pro PEEK Abutments
Indications for Use (Describe)
The Pro PEEK Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 5px;"> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="padding-right: 5px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
PSC Publishing Services (301) 443-6740
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
EF
{3}------------------------------------------------
### 510(k) Summary
| Submitter | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315<br>Owner/Operator No.: 9005052 | | | | |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| | on behalf of: | | | | |
| | JJGC Indústria e Comércio de Materiais Dentários SA<br>Av. Juscelino Kubitschek de Olivera, 3291<br>Curtiba, Parana, BRAZIL 81270-200<br>Registration No.: 3008261720<br>Owner/Operator No.: 10031702 | | | | |
| Contact Person | Jennifer M. Jackson, MS<br>Director, Regulatory Affairs & Quality, Straumann USA<br>E-Mail: jennifer.jackson@straumann.com<br>Telephone (978) 747-2509 | | | | |
| Date Prepared | 02/November/2017 | | | | |
| Prepared by | Ana Carolina Martins Vianna<br>RA & Compliance Manager, Institut Straumann AG<br>ana.vianna@straumann.com | | | | |
| Product Code<br>Device Class<br>Classification Panel<br>Classification Name<br>Common Name<br>Proprietary name | NHA (21 CFR 872.3630)<br>II<br>Dental<br>Endosseous dental implant abutment (21 CFR 872.3630)<br>Endosseous dental implant abutment<br>Neodent Implant System - CM Pro PEEK Abutments | | | | |
| Primary Predicate<br>Devices<br>Reference Devices | K093027 - Straumann RC Temporary Abutments, Institut<br>Straumann AG<br>K101945 - Neodent Implant System, JJGC Indústria e Comércio de<br>Materiais Dentários SA<br>K133510 - Neodent Implant System, JJGC Indústria e Comércio de<br>Materiais Dentários SA | | | | |
{4}------------------------------------------------
#### 5.1 Device Description
CM Pro PEEK abutments are temporary intermediary prosthetic components to be installed onto CM Implants to support the provisional prosthesis up to 6 months. They are composed of a customizable cylindrical body with an internal channel for the screw access and of a base for anti-rotational implant connection. The customizable portion is manufactured in PEEK (high performance polymer – specific for dental use) and the implant-to-abutment interface is made of titanium alloy. They have a coupled screw for fixing them to the Implant.
They present a hexagonal indexing at the apical end of the Morse taper connection to provide anti-rotational feature to the implant-to-abutment connection, and facilitates proper alignment of the prosthesis to the implant.
They are available in diameters of 4.5 and 6.0 mm and in lengths of 0.8, 1.5, 2.5, 3.5, 4.5 and 5.5 mm.
Image /page/4/Picture/6 description: The image shows a dental implant abutment. The abutment is made of two parts: a white ceramic or composite material on top and a silver-colored metal base. The metal base has a threaded screw at the bottom for attachment to the implant fixture in the jawbone. The abutment is designed to connect the implant to a dental crown or other restoration.
Figure 1: Model of CM Pro PEEK abutment
The customization of the PEEK chimney must be done into the mouth of patient or chairside. Chairside customization means that the device preparation must be done in a controlled environment in the operating room, in the context of the prosthetic placement, in a surgical site's aseptic environment using sterilized tools. The dentist can wear the PEEK chimney with the aid of suitable techniques, high speed handpiece and cooling, according to the patient interocclusal space, respecting the minimum of 5mm of height. No customization is allowed for the diameter/ wall thickness of the abutment, as well as no angulation. No CAD/CAM design and fabrication is allowed for the CM Pro PEEK Abutments. Only hand-milling may be used for abutment modification.
The CM Pro PEEK abutments can be used before the installation of the final prosthesis to maintain, stabilize and shape the soft tissue (gum) during the healing phase.
{5}------------------------------------------------
#### 5.2 Indications for Use
The Pro PEEK Abutments are indicated to be used on Neodent implants to provide temporary support for prosthesis structure for up to 6 months. They can be used in one or two stage procedures and also immediate load when there is good primary stability.
#### 5.3 Technological Characteristics
The temporary abutments made of PEEK are designed to be substantially equivalent to the predicate devices previously cleared per K093027.
The subject and predicate devices are based on the following same technological elements:
- Endosseous dental abutments to provide support for temporary restorations (primary predicate device);
- -Possibility of use in single or two-stage procedures, for single or multiple-unit restorations (reference devices);
- -Same sterilization methods and packaging (reference devices).
The following technological differences exist between the subject and predicate devices:
- Design of Morse taper implant-to-abutment interfaces; -
- -The subject abutment is provided sterile by EO exposition whereas its primary predicate device is provided non-sterile for end-user sterilization.
In the end of this Section, a comparison between the features of subject device and its predicate devices is shown in a tabular format. The assessment of the differences is also included.
#### 5.4 Performance data
The following performance data supports the substantial equivalence determination:
### Biocompatibility testing
The raw material PEEK used in fabrication of the CM Pro PEEK Abutments is the same of that cleared for the predicate device under K093027. Any difference in the biocompatibility posed by differences in the manufacturing process has been addressed by chemical characterization and cytotoxicity assessments. No extracts of leachable substances of concern were identified and no cytotoxicity response was observed.
The Biological Assessment has been performed according to ISO 10993-1 Biological evaluation of medical devices. Evaluation and testing, ISO 10993-5 Biological evaluation of medical devices. Tests for in vitro cytotoxicity and ISO 10993-18 Biological evaluation of medical devices. Chemical characterization of materials.
{6}------------------------------------------------
### Mechanical testing
The strength of the CM Pro PEEK Abutment is demonstrated through fatigue testing performed according to ISO 14801 - Dentistry - Implants - Dynamic fatigue test for endosseous dental implants and FDA document Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments.
### Sterilization validation
The subject abutments are sterilized by exposure to ethylene oxide (EO). Sterilization has been validated by the bioburden method, according to ISO 11135 Sterilization of health care products - Ethylene oxide - Requirements for the development, validation, and routine control of a sterilization process for medical devices. EO sterilization residuals have been verified to be less than the maximum allowable limits as defined in ISO 10993-7 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals. All methods achieve a Sterility Assurance Level (SAL) of 10-6.
The subject devices are not represented to be non-pyrogenic.
Shelf life was determined through both real time and accelerated aging protocol, according to ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices. After aging, samples were submitted to sterility test, thermal seal integrity test and tensile strength of thermal seal according to ISO 11737 - Sterilization of medical devices -Microbiological methods - Part 2: Tests of sterility Performed in the definition, validation and maintenance of a sterilization process and US Pharmacopeia; ASTM E499/E499M Standard Practice for Leaks Using the Mass Spectrometer Leak Detector in the Detector Probe Mode, ASTM F1929 Standard Test Method for detecting Seal Leaks in Porous Medical Packaging by Dye Penetration and ASTM F88/F88M Standard Test Method for Seal Strength of Flexible Barrier Materials.
The sterile barrier shelf life for the subject devices is 2 years.
#### Clinical data
No clinical data has been submitted, referenced, or relied upon to demonstrate substantial equivalence.
{7}------------------------------------------------
### Table 1: Comparison between the subject and predicate devices.
| | SUBJECT DEVICES | PRIMARY PREDICATE<br>DEVICES | REFERENCE DEVICES | | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Neodent Implant System -<br>CM Pro PEEK Abutment | Straumann RC Temporary<br>Abutment (K093027) | Neodent Implant System<br>(K101945) | Neodent Implant System<br>(K133510) | EQUIVALENCE DISCUSSION |
| Indications for<br>Use | The Pro PEEK Abutments are<br>indicated to be used on<br>Neodent implants to provide<br>temporary support for<br>prosthesis structure for up<br>to 6 months. They can be<br>used in one or two stage<br>procedures and also<br>immediate load when there<br>is good primary stability. | The Straumann RC<br>Temporary Abutments are<br>indicated for use in<br>Straumann RC Bone Level<br>Implants for temporary<br>restorations of single crowns<br>and bridges for up to six<br>months | The Neodent Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or lower<br>jaw to provide support for<br>prosthetic devices such as<br>artificial teeth, to restore<br>chewing function. It may be<br>used with single-stage or<br>two-stage procedures, for<br>single or multiple unit<br>restorations, and may be<br>loaded immediately when<br>good primary stability is<br>achieved and with<br>appropriate occlusal<br>loading. Multiple tooth<br>applications may be rigidly<br>splinted. | The Neodent Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or lower<br>jaw to provide support for<br>prosthetic devices such as<br>artificial teeth, to restore<br>chewing function. It may be<br>used with single-stage or<br>two-stage procedures, for<br>single or multiple unit<br>restorations, and may be<br>loaded immediately when<br>good primary stability is<br>achieved and with<br>appropriate occlusal<br>loading. Multiple tooth<br>applications may be rigidly<br>splinted | Equivalent<br>The Indications for Use for<br>the subject and primary<br>predicate devices are similar<br>with the exception of minor<br>wording differences.<br>The intention to provide<br>support to temporary<br>restorations is the same.<br>The subject device<br>Indications for Use<br>references the term<br>"prosthesis structure" which<br>is an overarching term that<br>encompasses both crowns<br>and bridges.<br>The specific indication for<br>immediate load when there<br>is good primary stability is<br>dependent on the implant<br>Indications for Use. The<br>implants also need to be<br>indicated for immediate<br>loading when good primary<br>stability is achieved. |
| | SUBJECT DEVICES | PRIMARY PREDICATE<br>DEVICES | REFERENCE DEVICES | | EQUIVALENCE DISCUSSION |
| | Neodent Implant System –<br>CM Pro PEEK Abutment | Straumann RC Temporary<br>Abutment (K093027) | Neodent Implant System<br>(K101945) | Neodent Implant System<br>(K133510) | |
| Platform<br>Diameter(s) | 4.5 and<br>6 mm | 4.5 mm | NA<br>(only permanent<br>abutments) | 3.3 mm<br>4.1/4.3 mm<br>5.0 mm | Equivalent<br>Subject device diameters are<br>the same of the predicate<br>devices or larger. Larger<br>diameters do not represent a<br>worst case in terms of<br>performance. |
| Material | Titanium alloy Ti-6Al-4V ELI<br>PEEK (polyetheretherketone) | Titanium alloy (Ti-6Al-7Nb,<br>TAN)<br>PEEK (polyetheretherketone) | Titanium Grade 4<br>Titanium alloy Ti-6Al-4V ELI | Titanium alloy Ti-6Al-4V ELI | Equivalent<br>The subject and the<br>predicate devices are made<br>of the same type of raw<br>material; the adequacy of the<br>systems have been<br>demonstrated via dynamic<br>fatigue tests and<br>biocompatibility assessment. |
| Implant-to-<br>Abutment<br>Connection | CM interface; 11.5° Morse<br>taper with anti-rotational<br>features. | Morse taper interface with<br>anti-rotational features | CM interface; 11.5° Morse<br>taper with anti-rotational<br>features. | HE (external hexagon)<br>interface | Equivalent<br>The performance of each<br>interface has been assessed<br>by dynamic fatigue testing. |
{8}------------------------------------------------
{9}------------------------------------------------
| | SUBJECT DEVICES | PRIMARY PREDICATE<br>DEVICES | REFERENCE DEVICES | | |
|---------------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Neodent Implant System –<br>CM Pro PEEK Abutment | Straumann RC Temporary<br>Abutment (K093027) | Neodent Implant System<br>(K101945) | Neodent Implant System<br>(K133510) | EQUIVALENCE DISCUSSION |
| Compatibility | Compatible with Neodent<br>CM implants | Compatible with Straumann<br>implants | Compatible with Neodent<br>CM implants | Compatible with Neodent<br>HE implants | Equivalent<br>The interfaces of Neodent<br>and Straumann devices are<br>not interchangeable. Their<br>performances are assessed<br>respecting the fixtures<br>compatibility. |
| Sterility | Delivered sterile by EO<br>exposure. | Delivered non-sterile. To be<br>sterilized by user before<br>placed in patient mouth<br>(moist steam sterilization). | Delivered sterile by EO<br>exposure. | Delivered sterile by gamma<br>irradiation. | Equivalent<br>The sterilization method and<br>posterior handling of the<br>device is the same as the<br>reference devices. |
{10}------------------------------------------------
#### 5.5 Conclusions
The documentation submitted in this premarket notification demonstrates that the subject devices have comparable features and performance and, therefore, are substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.