Tyson Bio HT100 Blood Glucose Monitoring System, Tyson Bio HT100-B Blood Glucose Monitoring System
K170079 · Tyson Bioresearch, Inc. · NBW · Oct 6, 2017 · Clinical Chemistry
Device Facts
Record ID
K170079
Device Name
Tyson Bio HT100 Blood Glucose Monitoring System, Tyson Bio HT100-B Blood Glucose Monitoring System
Applicant
Tyson Bioresearch, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Oct 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Indications for Use
The Tyson Bio HT100 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alterative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use. The Tyson Bio HT100 Blood Glucose Monitoring System is comprised of the Tyson Bio HT100 Blood Glucose Meter and Tyson Bio HT100 Blood Glucose Test Strip. The Tyson Bio HT100-B Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alterative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use. The Tyson Bio HT100 Blood Glucose Monitoring System is comprised of the Tyson Bio HT100 Blood Glucose Meter and Tyson Bio HT100 Blood Glucose Test Strip.
Device Story
System measures capillary whole blood glucose via amperometric biosensor; requires 0.7uL sample applied to test strip containing glucose dehydrogenase and potassium ferricyanide. Electrons generated during reaction produce current proportional to glucose concentration. HT100-B model includes Bluetooth for data transmission to mobile phones. Intended for home self-testing by patients with diabetes; not for clinical/assisted-use settings. Users apply blood to strip; meter provides digital readout. Results aid in monitoring diabetes control effectiveness. Device features auto-coding, HI/LO error flags, and insufficient sample volume detection. Robustness validated for 3 years of single-patient use with specified disinfectant wipes.
Clinical Evidence
User performance study conducted with 121 untrained lay users comparing self-obtained fingertip, palm, and forearm samples against YSI 2300 laboratory reference. Results showed 100% of fingertip samples within ±15% of reference for glucose ≥75 mg/dL. Palm and forearm samples also demonstrated high accuracy. Usability questionnaire confirmed successful device operation by lay users. Analytical performance supported by precision, linearity, interference, hematocrit, and altitude studies.
Technological Characteristics
Electrochemical biosensor using glucose dehydrogenase (FAD). Amperometric detection. Measuring range 20-600 mg/dL. Sample volume 0.7uL. Hematocrit range 10-65%. Operating temp 10-40°C. Humidity 10-90%. Auto-coding strips. Powered by two AAA batteries. HT100-B includes Bluetooth connectivity. Memory capacity 500 results.
Indications for Use
Indicated for quantitative blood glucose monitoring in people with diabetes for home self-testing. Samples: capillary whole blood (fingertip, forearm, palm). Contraindications: neonatal use, diagnosis/screening for diabetes, critically ill patients, severe hypotension/shock, hyperglycemic-hyperosmolar state, or non-steady-state glucose conditions.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
ACCU-CHEK Performa device (k133741)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k170079
B. Purpose for Submission:
New Device
C. Measurand:
Capillary whole blood glucose from the fingertips, forearm, and palm
D. Type of Test:
Quantitative amperometric assay (FAD-glucose dehydrogenase)
E. Applicant:
Tyson Bioresearch, Inc.
F. Proprietary and Established Names:
Tyson Bio HT100 Blood Glucose Monitoring System
Tyson Bio HT100-B Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345, Glucose test system
2. Classification:
Class II
3. Product code:
NBW- system, test, blood glucose, over the counter
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See Indication(s) for use below.
2. Indication(s) for use:
The Tyson Bio HT100 Blood Glucose Monitoring System:
The Tyson Bio HT100 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alterative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use.
The Tyson Bio HT100 Blood Glucose Monitoring System is comprised of the Tyson Bio HT100 Blood Glucose Meter and Tyson Bio HT100 Blood Glucose Test Strip.
The Tyson Bio HT100-B Blood Glucose Monitoring System:
The Tyson Bio HT100 Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips, forearm, or palm. Alterative site testing should be performed only during steady-state (when glucose is not changing rapidly). Testing is done outside the body (In Vitro diagnostic use). It is intended for self-testing by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should only be used by a single patient and it should not be shared. It is not indicated for the diagnosis of or screening for diabetes or for neonatal use.
The Tyson Bio HT100 Blood Glucose Monitoring System is comprised of the Tyson Bio HT100 Blood Glucose Meter and Tyson Bio HT100 Blood Glucose Test Strip.
3. Special conditions for use statement(s):
Special conditions for use statement(s) for Tyson Bio HT100 Blood Glucose Monitoring System and Tyson Bio HT100-B Blood Glucose Monitoring System
- Use only capillary whole blood from finger, palm, and forearm
- Do not use for testing neonatal blood
- Altitude higher than 10745 feet may cause inaccurate results.
2
{2}
- Severe dehydration and excessive water loss may lead to inaccurate blood glucose test results. Consult your physician immediately if you believe you are suffering from severe dehydration.
- For single-patient use only and should not be shared.
- Should not be used to test critically ill patients
- Inaccurate results may occur in severely hypotensive individuals or patients in shock
- Inaccurate low results may occur for individuals experiencing a hyperglycemic-hyperosmolar state, with or without ketosis
- For in vitro diagnostics use only
- For over-the-counter use
- Alternative site testing should be performed only during steady-state (when glucose is not changing rapidly)
- Alternative site measurements should never be used to calibrate continuous glucose monitoring (CGMs)
- Alternative site measurements should never be used for insulin dosing calculation
- This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures. Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
4. Special instrument requirements:
Tyson Bio HT100 Blood Glucose Meter
Tyson Bio HT100-B Blood Glucose Meter
I. Device Description:
The Tyson Bio HT100 Blood Glucose Monitoring System consists of the Tyson Bio HT100 Blood Glucose Meter, the Tyson Bio HT100 Blood Glucose Test Strips, and the Tyson Bio HT100 Control Solution.
The Tyson Bio HT100-B Blood Glucose Monitoring System consists of the Tyson Bio HT100-B Blood Glucose Meter, the Tyson Bio HT100-B Blood Glucose Test Strips and the Tyson Bio HT100-B Control Solution. The Tyson Bio HT100-B Blood Glucose Monitoring System has Bluetooth transmission capability and the results from the Tyson Bio HT100-B Blood Glucose meter can be transmitted to a Bluetooth-capable cellphone.
{3}
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACCU-CHEK Performa Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k133741
3. Comparison with predicate:
| Similarities | | | |
| --- | --- | --- | --- |
| Item | Predicate Device (k133741) | Candidate Device (k170079) | |
| | ACCU-CHEK Performa Blood Glucose Monitoring System | Tyson Bio HT100 Blood Glucose Monitoring System | Tyson Bio HT100-B Blood Glucose Monitoring System |
| Indications for Use | Intended for the quantitative measurement of glucose in fresh capillary whole blood samples | Same | Same |
| Test Principle | Amperometry | Same | Same |
| Methodology | Electrochemical biosensor with Glucose Dehydrogenase (FAD) | Same | Same |
| Hematocrit | 10%-65% | Same | Same |
| Measuring Range | 20-600 mg/dL | Same | Same |
| Memory capacity | 500 results with time and date | Same | Same |
{4}
| Differences | | | |
| --- | --- | --- | --- |
| Item | Predicate Device (k133741) | Candidate Device (k170079) | |
| | ACCU-CHEK Performa Blood Glucose Monitoring System | Tyson Bio HT100 Blood Glucose Monitoring System | Tyson Bio HT100-B Blood Glucose Monitoring System |
| Sample Type | Capillary whole blood samples drawn from the fingertips | Capillary whole blood samples drawn from the fingertips, forearm, or palm | Capillary whole blood samples drawn from the fingertips, forearm, or palm |
| Battery Type | One CR2032 | Two AAA batteries | Two AAA batteries |
| Sample volume | 0.6uL | 0.7uL | 0.7uL |
| Operating Temperature Range | 16 to 35 °C (61-95°F) | 10 to 40 °C (50-104°F) | 10 to 40 °C (50-104°F) |
| Operating Humidity Range | 10-80% | 10-90% | 10-90% |
| Coding | Code key required | Auto coding | Auto coding |
| Bluetooth | none | none | Bluetooth transmission capability to a Bluetooth-capable cellphone |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP07-A2, “Interference Testing in Clinical Chemistry”; Approved Guideline
L. Test Principle:
Quantitative measurement of glucose within whole blood using the Tyson Bio HT100/ Tyson Bio HT100-B Blood Glucose Monitoring Systems, an amperometric biosensor, begins with a small drop of blood. This blood sample is then placed onto the Tyson Bio HT100 Test Strip where it is diffused and solvated into the reagent layer, which contains glucose dehydrogenase, potassium ferricyanide and nonreactive ingredients. Glucose in the sample reacts with glucose dehydrogenase and potassium ferricyanide. Electrons are generated producing a current, which is proportional to the glucose in the sample. This current is then measured and correlated to the glucose concentration within the whole blood sample.
{5}
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
The two systems in this submission are identical except for the presence or absence of blue tooth functionality and the name; therefore, performance data provided in this section using the Tyson Bio HT100 Blood Glucose Monitoring System is applicable to both systems.
a. Precision/Reproducibility:
The sponsor performed repeatability precision studies (within-run precision) using pooled venous whole blood samples spiked with glucose to eight glucose - concentration ranges (20-29 mg/dL, 30-50 mg/dL, 51-110 mg/dL, 111-150 mg/dL, 151-250 mg/dL, 251-400 mg/dL, 401-500 mg/dL, and 501-600 mg/dL). Each glucose sample was measured in replicates of 10 using 3 test strip lots and 10 meters for a total of 300 tests for each glucose level. Results are summarized below:
| Glucose Concentration (mg/dL) | Strip lot | n | Mean (mg/dL) | SD (mg/dL) | % CV |
| --- | --- | --- | --- | --- | --- |
| 20-29 | 1 | 100 | 22.0 | 1.9 | 8.6 |
| | 2 | 100 | 23.3 | 1.9 | 8.2 |
| | 3 | 100 | 23.4 | 1.8 | 7.7 |
| 30-50 | 1 | 100 | 43.9 | 3.9 | 8.9 |
| | 2 | 100 | 44.3 | 3.9 | 8.9 |
| | 3 | 100 | 45.5 | 3.4 | 7.5 |
| 51-110 | 1 | 100 | 85.0 | 3.0 | 3.6 |
| | 2 | 100 | 84.8 | 2.9 | 3.4 |
| | 3 | 100 | 86.7 | 3.1 | 3.6 |
| 111-150 | 1 | 100 | 146.7 | 3.9 | 2.7 |
| | 2 | 100 | 145.0 | 3.8 | 2.6 |
| | 3 | 100 | 145.3 | 3.8 | 2.6 |
| 151-250 | 1 | 100 | 217.2 | 4.2 | 1.9 |
| | 2 | 100 | 219.3 | 5.0 | 2.3 |
| | 3 | 100 | 221.3 | 5.6 | 2.5 |
| 251-400 | 1 | 100 | 330.0 | 6.9 | 2.1 |
| | 2 | 100 | 341.0 | 5.8 | 1.7 |
| | 3 | 100 | 341.2 | 6.8 | 2.0 |
| 401-500 | 1 | 100 | 449.5 | 6.8 | 1.5 |
| | 2 | 100 | 449.8 | 6.9 | 1.5 |
| | 3 | 100 | 449.6 | 8.0 | 1.8 |
| 501-600 | 1 | 100 | 549.3 | 10.8 | 2.0 |
| | 2 | 100 | 552.7 | 13.8 | 2.5 |
| | 3 | 100 | 551.1 | 9.7 | 1.8 |
{6}
Intermediate (between day) precision was evaluated using 3 lots of test strips and 10 meters. Glucose control solutions in three concentration ranges were used (60-80 mg/dL, 96-114 mg/dL and 280-420 mg/dL). Each sample was tested over 10 days with each meter, for each test strip lot, resulting in a total of 100 measurements per lot and 300 measurements per glucose level. Results are summarized below:
| Glucose Concentration (mg/dL) | Strip lot | n | Mean (mg/dL) | SD (mg/dL) | % CV |
| --- | --- | --- | --- | --- | --- |
| 30-50 | 1 | 100 | 47.3 | 3.0 | 6.4 |
| | 2 | 100 | 48.2 | 3.1 | 6.5 |
| | 3 | 100 | 49.8 | 3.2 | 6.5 |
| 96-144 | 1 | 100 | 130.3 | 4.9 | 3.8 |
| | 2 | 100 | 131.2 | 4.8 | 3.7 |
| | 3 | 100 | 130.5 | 3.9 | 3.0 |
| 280-420 | 1 | 100 | 345.7 | 10.7 | 3.1 |
| | 2 | 100 | 347.5 | 10.2 | 2.9 |
| | 3 | 100 | 342.1 | 10.3 | 3.0 |
b. Linearity/assay reportable range:
Linearity testing was performed using venous whole blood adjusted to 12 different glucose concentrations ranging from 22.8 to 639.4 mg/dL (22.8 mg/dL; 52.9 mg/dL; 81.1 mg/dL; 120.6 mg/dL; 183.1 mg/dL; 255.5 mg/dL; 326.3 mg/dL; 404.3 mg/dL; 466.3 mg/dL; 546.5 mg/dL; 586.8 mg/dL; 639.4 mg/dL). Each sample was measured in replicates of 10 with 3 test strip lots and 30 Tyson Bio HT100 blood glucose meters. The values were compared to a laboratory comparator method (YSI 2300 analyzer). The evaluation yielded the following regression analysis:
| Lot | Slope | y-intercept | R² |
| --- | --- | --- | --- |
| 1 | 0.9996 | 4.5779 | 0.9968 |
| 2 | 0.9955 | 4.5691 | 0.9973 |
| 3 | 0.9930 | 4.6833 | 0.9961 |
The results of the study support the claimed glucose measurement range of 20 to 600 mg/dL. If a sample result is less than 20 mg/dL the result is flagged by the meter as "LO" and if a sample result exceeds 600 mg/dL the result is flagged by the meter as "HI". The 'HI/LO' error message feature was validated to demonstrate that the feature functions as intended.
{7}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The Tyson Bio HT100 Blood Glucose Monitoring System and the Tyson Bio HT100-B Blood Glucose Monitoring System are traceable to a certified glucose reference material (NIST 965b).
Test Strip Stability:
Test strip open vial and closed vial stability was assessed in real-time studies and is ongoing. The testing protocols and acceptance criteria were reviewed and found to be acceptable. The manufacturer claims vial shelf life stability of 1 month and open-vial stability of 1 month when stored at 39.2 to 86°F (4-30°C) and relative humidity (RH) of 10-90%. The labeling states: Do not freeze.
d. Detection limit:
The reportable range is 20-600 mg/dL. This range was verified by the linearity study (M.1.b above).
e. Analytical specificity:
Interference studies were performed by spiking endogenous and exogenous substances into venous whole blood. Each potential interferent was tested at 3 glucose levels (50-100 70 mg/dL, 110-140 mg/dL and 250-350 mg/dL as measured on the YSI 2300 Analyzer) using 3 test strip lots measuring each sample in replicates. The % difference between the test sample and the control sample was calculated. The substances tested and the highest concentrations tested with no significant interference (defined by the sponsor as ≤ 10% bias) are summarized in the table below:
| Interference substance | Highest tested concentration with no significant interference |
| --- | --- |
| Acetaminophen | 4.25 mg/dL |
| Ascorbate(Ascorbic acid) | 3 mg/dL |
| Bilirubin | 25 mg/dL |
| Cholesterol | 1200 mg/dL |
| Creatinine | 10 mg/dL |
| Dopamine | 2 mg/dL |
| Galactose | 500 mg/dL |
| Gentisic acid | 2.5 mg/dL |
{8}
| Interference substance | Highest tested concentration with no significant interference |
| --- | --- |
| Glutathione | 3.07 mg/dL |
| Haemoglobin-HUMAN | 3000 mg/dL |
| Heparin | 5 IU/mL |
| Hydrogenated starch hydrolysates (HSH) | 0.09 mg/dL |
| Ibuprofen | 50 mg/dL |
| Icodextrin | 750 mg/dL |
| Isomalt | 0.09 mg/dL |
| Lactitol | 0.09 mg/dL |
| L -DOPA(L-3-4-dihydroxyphenylalanine) | 0.5 mg/dL |
| Maltose | 2575 mg/dL |
| Maltitol | 0.09 mg/dL |
| Mannitol | 0.09 mg/dL |
| Methyldopa | 2.5 mg/dL |
| Pralidoxime iodide (PAM) | 5 mg/dL |
| Salicylate | 50 mg/dL |
| Sodium Carbonate | 37.5 mEq/L |
| Sorbitol | 0.09 mg/dL |
| Tolbutamide | 100 mg/dL |
| Tolazamide | 6 mg/dL |
| Triglycerides | 2000 mg/dL |
| Uric acid | 10.0 mg/dL |
| Xylose | 5 mg/dL |
| Xylitol | 0.09 mg/dL |
The sponsor included the following information in their labeling:
If you have certain conditions that may cause your blood level of uric acid to rise, such as gout or kidney disease (>10 mg/dL uric acid) then your blood glucose results may be
{9}
inaccurate. If you are unsure, then ask your health care professional.
Acetaminophen interferes with this meter. If you are taking acetaminophen containing drugs (Tylenol and other medicines containing acetaminophen) you should not use this meter.
If you are taking Vitamin C (ascorbic acid, blood concentrations >3 mg/dL) at doses higher than recommended, it may interfere with your glucose meter and cause you to get inaccurate results with this system.
Do not use this system during or shortly after receiving xylose absorption therapy; xylose may case inaccurate blood glucose results
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
See section M.2.c. below for the user performance study.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
User Performance Studies:
To assess the performance of the Tyson Bio HT100 Blood Glucose Monitoring System in the hands of the intended users, a study was performed at one clinical site using three test strip lots with 121 untrained lay user participants that obtained their own fingerstick, palm and forearm samples. Each subject performed testing on their own sample and results were compared to those obtained by a health care professional on the YSI 2300. The range of glucose values for the samples as measured by YSI was 35 to 348 mg/dL. The results relative to the YSI are summarized below:
10
{10}
Summary of Fingertip vs. YSI
| Accuracy for blood glucose levels <75 mg/dL | | | |
| --- | --- | --- | --- |
| Within ± 5mg/dL | Within ± 10mg/dL | Within ± 15mg/dL | |
| 4/4 (100%) | 4/4 (100%) | 4/4 (100%) | |
| Accuracy for blood glucose levels ≥75 mg/dL | | | |
| Within ± 5% | Within ± 10% | Within ± 15 % | Within ± 20% |
| 96/117 (82%) | 116/117 (99%) | 117/117 (100%) | 117/117 (100%) |
Summary of Palm vs. YSI
| Accuracy for blood glucose levels <75mg/dL | | | |
| --- | --- | --- | --- |
| Within ± 5mg/dL | Within ± 10mg/dL | Within ± 15mg/dL | |
| 3/4 (75%) | 4/4 (100%) | 4/4 (100%) | |
| Accuracy for blood glucose levels ≥75mg/dL | | | |
| Within ± 5% | Within ± 10% | Within ± 15 % | Within ± 20% |
| 88/117 (75%) | 114/117 (97%) | 117/117 (100%) | 117/117 (100%) |
Summary of Forearm vs. YSI
| Accuracy for blood glucose levels <75mg/dL | | | |
| --- | --- | --- | --- |
| Within ± 5mg/dL | Within ± 10mg/dL | Within ± 15mg/dL | |
| 3/4 (75%) | 4/4 (100%) | 4/4 (100%) | |
| Accuracy for blood glucose levels ≥75mg/dL | | | |
| Within ± 5% | Within ± 10% | Within ± 15 % | Within ± 20% |
| 79/117 (68%) | 115/117 (98%) | 117/117 (100%) | 117/117 (100%) |
Linear Regression:
| Site | Slope | Intercept | R² | N |
| --- | --- | --- | --- | --- |
| Fingertip | 0.9772 | 1.0513 | 0.9935 | 121 |
| Palm | 0.9913 | 0.0089 | 0.9921 | 121 |
| Forearm | 0.9691 | 3.0295 | 0.987 | 121 |
Prior to the lay-user study the readability of the labeling (user manual, test strip package insert and quick reference guide) was assessed using a Flesch-Kincaid analysis and was found to be written at the 8th grade level.
During the lay user study the participants were asked to complete a questionnaire to evaluate the usability of the use of the device and the clarity of user manual. Overall, the users indicated that they could successfully conduct the test.
4. Clinical cut-off:
Not applicable
{11}
12
5. Expected values/Reference range:
The labeling includes the following expected glucose values for people without diabetes: Before a meal (fasting): <100 mg/dL 2 h after a meal: <140mg/dL
These ranges were cited from the American Diabetes Association, Classification and Diagnosis of Diabetes, 2017. Diabetes Care 40 (Suppl. 1):S11-S24
N. Instrument Name:
Tyson Bio HT100 Blood Glucose Meter
Tyson Bio HT100-B Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ X or No ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
There is no sample identification function with the Tyson Bio HT100 Blood Glucose Meter and the Tyson Bio HT100-B Blood Glucose Meter. As the blood sample is collected from the fingertip, palm or forearm the user directly applies the blood sample to the test strip.
4. Specimen Sampling and Handling:
Tyson Bio HT100 Blood Glucose Meter and Tyson Bio HT100-B Blood Glucose Meter is intended to be used with capillary whole blood from the fingertip, palm and forearm which can be applied directly to the test strip by capillary action.
{12}
5. Calibration:
There is no calibration required by the user for the Tyson Bio HT100 Blood Glucose Meter and the Tyson Bio HT100-B Blood Glucose Meter. The meters are automatically coded.
6. Quality Control:
Three levels of controls are available for use with this meter. The control solutions are sold separately. The control solution range for each control level is printed on the test strip vial label. The user is cautioned not to use the meter if the control result falls outside these ranges. The Tyson Bio HT100/HT100-B Blood Glucose meters may be set in a Control Solution test mode by pressing a button; the control results will not be included in any day average calculation when the meter is set in the Control Solution test mode.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Hematocrit
The effect of different hematocrit levels on the Tyson Bio HT100 Blood Glucose Monitoring System was evaluated using venous whole blood samples with hematocrit levels of 9-71% (9, 14, 19, 25, 30, 35, 40, 45, 50, 55, 61, 66 and 71%) at 5 concentrations of glucose (27, 64, 134, 224 and 375 mg/dL). Three test strips lots were evaluated with 10 meters. Each sample was tested in replicates for each hematocrit level and glucose levels. For each blood sample, individual glucose readings taken on the Tyson Bio HT100 glucose meter were compared to YSI comparator values to calculate the bias. The % bias of the Tyson Bio HT100 Blood Glucose meter results relative to YSI demonstrated adequate performance to support the claimed hematocrit range of 10-65%. The meter has an Hct H and Hct L feature that was validated to demonstrate that the feature functions as intended.
2. Altitude Testing
To evaluate the effects of altitude on the Tyson Bio HT100 Blood Glucose Monitoring System, venous blood samples were altered to 9 glucose concentrations intervals (47-53, 75-85, 110-120, 175-195, 245-265, 315-335, 380-410, 450-480, and 520-555 mg/dL) and tested at 164, 5719 and 10745 feet above sea level. Results obtained were compared with those obtained with the comparator method (YSI). The results demonstrate acceptable bias to the comparator to support the claims in the labeling at altitudes up to 10745 feet have no significant effect on blood glucose measurements.
3. Sample volume study
To assess the minimum sample volume of the Tyson Bio HT100 Blood Glucose Monitoring System, venous whole blood samples from volunteers were collected with K3-EDTA vacutainer tubes and were spiked with B-D-Glucose to four glucose concentrations (54.0, 117.0, 234.0 and 331.0 mg/dL). Blood samples were test with 10 measurements obtained from 10 meter with each level of blood sample using 3 different lots. Different testing volumes (0.5, 0.7, 1.0 ul) of each level were evaluated for accuracy.
13
{13}
The study results support the sponsor’s claim that 0.7ul is the minimum sample volume required. The meter has an insufficient sample volume feature that was validated to demonstrate that the feature functions as intended.
4. Test System Operating Conditions:
Operating conditions studies were performed using three test strip lots and ten meters, with venous blood altered to glucose concentrations of 41 mg/dL, 123 mg/dL and 345 mg/dL. Testing with the Tyson Bio HT100 Blood Glucose Monitoring System was performed at the combination of extremes of claimed temperature and humidity operating conditions and results compared to an established laboratory method (YSI). Results support the claims in the labeling that the test system can be used at temperatures from 50° to 104°F (10-40°C) and relative humidity of 10-90%.
5. Infection control studies:
The device system is intended for single-patient use. Disinfection efficacy studies were performed on the external meter materials demonstrating removal of the HBsAg antigen with the chosen disinfectant, Clorox Bleach Germicidal Wipes (EPA Registration #67619-12). Robustness studies were performed by the sponsor demonstrating that there was no change in performance or external materials of the meters after 550 cleaning and disinfection cycles comprised of 1100 wipes. The robustness studies support 3 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
6. EMC Testing:
The sponsor provided appropriate documentation certifying that electromagnetic (EMC) testing was performed and the Tyson Bio HT100 Blood Glucose Monitoring System and the Tyson Bio HT100-B Blood Glucose Monitoring System were found to be compliant.
7. This device was cleared after the FDA issued final guidance documents for prescription use blood glucose monitoring systems (BGMS) and over-the-counter use blood glucose monitoring systems (SMBG). However, the recommendations in the guidance documents were not followed for this device since the studies were conducted prior to the finalization of the guidance documents.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.