K170068 · Wellray Medical Laser, Inc. · ILY · May 5, 2017 · Physical Medicine
Device Facts
Record ID
K170068
Device Name
Softcure 96
Applicant
Wellray Medical Laser, Inc.
Product Code
ILY · Physical Medicine
Decision Date
May 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Softcure™ 96 is intended to emit energy in the visible spectrum (red 635nm) to provide topical heating for the purpose of elevating tissue temperature for the temporary relief/reduction of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and temporary relaxation of muscles.
Device Story
Softcure 96 is a portable, wearable, battery-powered therapeutic device; uses semiconductor laser diodes to emit 635nm red light. Flexible pad conforms to body contours (back, legs, chest, arms). Microprocessor monitors laser temperature/output; variable speed fans provide cooling. Failsafe systems (capacitive proximity, body temperature, ambient light, reference voltage) ensure lasers energize only upon skin contact. Used in home or clinical settings by patients or providers. Output is topical heat (40–45°C) to elevate tissue temperature. Benefits include pain relief, muscle relaxation, and increased local circulation.
Clinical Evidence
Bench testing only. Performance testing on 8 subjects demonstrated the device achieves and maintains therapeutic topical heat (40–45°C) within 1 minute, sustained for at least 90 minutes. Compliance with IEC 60601-1, IEC 60601-1-2, and FDA 21 CFR 1040.10/1040.11 standards confirmed.
Technological Characteristics
Semiconductor laser diodes (635nm); flexible pad form factor; battery-powered; microprocessor-controlled; variable speed fan cooling. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, FDA 21 CFR 1040.10/1040.11.
Indications for Use
Indicated for patients requiring temporary relief of minor muscular/joint pain, stiffness, arthritis pain, muscle spasms, and local blood circulation increase. Suitable for OTC and prescription use.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2017
Wellray Medical Laser, Inc. Mr. Richard DeRoberts Chief Executive Officer 216 Moores Run Rd. Wardensville, West Virginia 26851
Re: K170068
Trade/Device Name: Softcure 96 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: February 8, 2017 Received: February 8, 2017
Dear Mr. DeRoberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170068
Device Name Softcure TM 96
Indications for Use (Describe)
The Softcure TM 96 is intended to emit energy in the visible spectrum (red 635mm) to provide topical heating for the purpose of elevating tissue temperature for the temporary relief/reduction of minor muscular and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and temporary relaxation of muscles.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:inline-block;"><span></span><span style="text-decoration:overline;">X</span></div> Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block;"><span></span><span style="text-decoration:overline;">X</span></div> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92.
A. Submitter: Wellray Medical Laser, Inc. 216 Moores Run Rd. Wardensville, WV 26851 USA
Phone: 304-874-4209 Fax: 304-874-4209
Contact: Mr. Richard O. DeRoberts CEO
Date prepared: January 4, 2017
B. Device Names:
Proprietary Name: Common/Usual Name: Classification Name:
Softcure™ 96 Infrared Therapeutic Heating Lamp Infrared Lamp
C. Predicate Devices Biocare™ Lumiwave® K051816 Sunetics Laser Brush, LHB 12 K121920 Anodyne® Therapy 480 Pro K931261 Biophotas Celluma ™ K131113
Device Description: D.
The Wellray® Medical Laser Softcure™ 96 is a therapeutic device using wavelengths of monochromatic light produced by semiconductor laser diodes to treat a variety of musculoskeletal concerns. The Softcure ™ 96 is a portable and wearable, battery powered, software controlled device intended for use on the body such as the back, legs, chest, arms or other areas where therapeutic light may be beneficial. A flexible pad, drawing power through an electrical cable from a carry able battery pack, emits red (635nm) light by means of semiconductor laser diodes to the desired body area. The flexible pad is designed to conform to the contour of the treatment site providing uniform treatment dosage over nonplanar body physique. The semiconductor laser diodes are monitored for temperature and light output by a microprocessor and cooled by variable speed fans to ensure the optimum light output. The lasers are allowed to energize only when four separate failsafe monitoring systems sense that the flexible pad is brought into close contact with the user's body. These failsafe systems include capacitive proximity sensors, body temperature sensors, an ambient light sensor and reference voltage monitoring.
E. Intended Use:
The Softcure™ 96 is intended to emit energy in the visible spectrum (red 635nm) to provide topical heating for the purpose of elevating tissue temperature for the temporary relief/reduction of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm, the temporary increase in local blood circulation, and temporary relaxation of muscles.
{4}------------------------------------------------
#### ட் Substantial Equivalence:
The Softcure™ 96 is substantially equivalent to the Biocare ™ Lumiwave® K051816, the Sunetics Laser Brush, LHB 12 - K121920, the Anodyne® Therapy 480 Pro - K931261 and the Biophotas Celluma ™ - K131113 as summarized in the following table. The Softcure ™ 96 and the predicate devices have the same intended use, the same diffused light characteristic, the same software control, the same use of a treatment timer, the same over-the-counter (OTC) use, the same method of administration (topical), the same configuration (pads/pods), and the same electrical safety.
| Predicate Device | Manufacturer | 510(k) Number |
|---------------------|------------------|---------------|
| Lumiwave® | Biocare TM | K051816 |
| Laser Brush, LHB 12 | Sunetics | K121920 |
| 480 Pro | Anodyne® Therapy | K931261 |
| Celluma TM | Biophotas | K131113 |
Where the Softcure ™ 96 differs is that it is wearable and is powered by a battery so that the user can move around the treatment site or home and perform light tasks. It also differs in that it has multiple failure detection/shut off controls and multiple failsafe monitoring systems.
The differences identified above do not impact adversely the Safety and Effectiveness of the Softcure TM 96 device.
Performance data have been included to demonstrate that the Softcure ™ 96 device meets its specifications, and functions safely.
#### Performance Testing Bench G.
Electrical safety and functional performance testing were conducted on the Softcure™ 96 demonstrating that the device is compliant with FDA 21 CFR J 1040.10 and 1040.11 standards, AAMI/ANSI ES 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012), IEC 60601-1-2 Medical Electrical Equipment -Part 1-2: General Requirements for Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and tests, 2007-03.
#### H. Performance Testing Non-Clinical
The use of light energy to generate heat for therapeutic use has been well documented and is a generally accepted alternative treatment modality for the temporary relief of pain and tissue repair.
The Softcure™ 96 is capable of achieving therapeutic heat temperature of 40 – 45 degrees centigrade as accepted by the FDA. An increase in topical heating of the tissue level by at least 5 degrees centigrade was reached within one (1) minute. The therapeutic temperature range was maintained for at least ninety (90) minutes.
The temperature versus time measurements were conducted on 8 subjects at various physical locations, i.e., leg, neck, back and shoulder. The pre-exposed topical skin temperature ranged from 36 to 39 degrees centigrade. These data demonstrate that the Softcure™ 96 meets the generally accepted topical
{5}------------------------------------------------
temperature range for therapeutic heat of 40-45 degrees centigrade during the recommended treatment time.
l. Conclusion
The Softcure ™ 96 device is substantially equivalent to the predicate devices K051816, K121920, K931261 and K131113.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.