K170060 · Lumenis, Ltd. · GEX · Aug 9, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170060
Device Name
M22 and ResurFx Systems
Applicant
Lumenis, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The subject Lumenis M22 System has connection capability with the following available treatment handpieces, for multiapplication treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows: The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200nm (with 10 different filters) is indicated for: Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, ephelides (freckles) and tattoos . Cutaneous lesions, including warts, scars and striae Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, . erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and venous malformations . Removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles Mild to moderate inflammatory Acne (Acne vulgaris) . The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is indicated for: The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of . superficial and deep telangiectasias) and reticular veins (0.1 - 4.0 mm. diameter) of the leg . The removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles . The non-ablative treatment of facial wrinkles ResurFX module and handpiece. with wavelength of 1565 nm. is indicated for: Use in dermatological procedures requiring fractional skin resurfacing and coaqulation of soft tissue . The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064nm is indicated for: . Removal of dark tattoos . Treatment of pigmented lesions The ResurFX System with wavelength of 1565 nm. is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.
Device Story
M22 and ResurFX are multi-application aesthetic platforms using IPL, Nd:YAG laser, and Er:Glass fiber laser technologies. Systems deliver light/laser energy to skin via interchangeable handpieces; IPL uses cut-off/notch filters (400-1200nm); Nd:YAG (1064nm) targets vascular lesions/hair; ResurFX (1565nm) performs fractional resurfacing. Operated by clinicians in dermatological/aesthetic settings. AOPT (Advanced Optimal Pulse Technology) allows tailored pulse sequencing. Output includes thermal coagulation/ablation of target tissues. Benefits include hair reduction, lesion clearance, and skin resurfacing. System includes software presets and scanning shapes (line, rectangle, etc.) to assist user control. Output affects clinical decision-making by enabling precise, customized treatment of dermatological conditions.
Clinical Evidence
Bench testing only. Testing included risk analysis (ISO 14971), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and non-laser light source safety (IEC 60601-2-57). Verification and validation confirmed software updates, new handpiece accessories, and scanning shapes perform as intended within specifications.
Indicated for aesthetic and dermatological skin procedures including benign epidermal/cutaneous lesions, vascular lesions, unwanted hair reduction (skin types I-V), inflammatory acne, coagulation/hemostasis of vascular lesions/soft tissue, non-ablative facial wrinkle treatment, fractional skin resurfacing, and tattoo/pigmented lesion removal.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 9, 2017
Lumenis, Ltd. Naama Jacoby Head Of RA Ophthalmic & M22 Platforms 6 Hakidma Street Po Box 240 Yokneam, 2069204 IL
Re: K170060
Trade/Device Name: M22 And Resurfx Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, ONF, ONG Dated: August 7, 2017 Received: July 17, 2017
Dear Naama Jacoby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K170060
Device Name
Lumenis Family of IPL and Laser Systems: M22 and ResurFx Systems
Indications for Use (Describe)
The subject Lumenis M22 System has connection capability with the following available treatment handpieces, for multiapplication treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
#### The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200nm (with 10 different filters) is indicated for: .
- Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, ephelides (freckles) and tattoos .
- . Cutaneous lesions, including warts, scars and striae
- Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, . erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and venous malformations
- . Removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- Mild to moderate inflammatory Acne (Acne vulgaris) .
- The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is indicated for: ●
- The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of . superficial and deep telangiectasias) and reticular veins (0.1 - 4.0 mm. diameter) of the leg
- . The removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- . The non-ablative treatment of facial wrinkles
- ResurFX module and handpiece. with wavelength of 1565 nm. is indicated for: .
- Use in dermatological procedures requiring fractional skin resurfacing and coaqulation of soft tissue .
- The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064nm is indicated for: .
- . Removal of dark tattoos
- . Treatment of pigmented lesions
*Note
Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
The ResurFX System with wavelength of 1565 nm. is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.
Type of Use (Select one or both, as applicable)
図 Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the tagline "Energy to Healthcare" also in blue.
### 510(k) SUMMARY
## Lumenis Family of IPL and Laser Systems: M22 & ResurFX Systems
| Applicant Name: | Lumenis Ltd.<br>6 Hakidma Street PO Box 240<br>Yokneam Industrial Park,<br>Yokneam 2069204, Israel<br>Tel: +972-4-9599000<br>Fax: +972-4-9599046 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Naama Jacoby - Head of RA, Ophthalmic & M22 Platforms<br>6 Hakidma Street PO Box 240<br>Yokneam Industrial Park,<br>Yokneam 2069204, Israel<br>Tel: +972-4-9599085<br>Fax: +972-4-9599198<br>Mail: Naama.Jacoby@lumenis.com |
| Date Prepared: | August 07, 2017 |
| Trade Name: | Lumenis Family of IPL and Laser Systems:<br>M22 & ResurFX Systems |
| Classification Name: | Powered laser surgical instrument |
| Product Code: | GEX, ONF, ONG |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4810 |
| Panel: | General & Plastic Surgery |
| Predicate Device: | M22 System, cleared under K142860<br>ResurFX module, cleared under K130028 |
| Indications for Use: | The complete phrasing of the indications for use statement for the<br>Lumenis Family of IPL and Laser Systems, combining both the M22<br>and ResurFX Systems, is provided in the formal Indications for Use<br>Statement (FDA Form 3881) |
From this point on, the 510(k) summary addresses the M22 System and the ResurFX System separately as follows:
{5}------------------------------------------------
### M22 System
#### Indications for Use:
The subject Lumenis M22 System has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:
- · The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200nm (with 10 different filters) is indicated for:
- -Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, ephelides (freckles) and tattoos
- Cutaneous lesions, including warts, scars and striae
- -Benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
- -Removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- Mild to moderate inflammatory Acne (Acne vulgaris) -
- The Nd:YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd:YAG) is indicated for:
- -The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1 - 4.0 mm. diameter) of the leg
- The removal of unwanted hair from all skin types, and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
- -The non-ablative treatment of facial wrinkles
- · ResurFX module and handpiece, with wavelength of 1565 nm, is indicated for:
- -Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue
- The O-Switched Nd:YAG Laser Handpiece with a wavelength of 1064nm is indicated for:
- -Removal of dark tattoos
- Treatment of pigmented lesions l
*Note: Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.
{6}------------------------------------------------
#### Device Description:
The subject Lumenis M22 System is a multi-application, multi-technology platform with four (4) available treatment handpieces:
- -Universal Intense Pulsed Light (IPL) handpiece:
- । Multi-Spot Nd:YAG laser handpiece;
- । ResurFX non-ablative laser handpiece;
- -Q-Switched Nd:YAG laser handpiece.
The following accessories are provided with and/or may be purchased independently for each of the four (4) available treatment handpieces of the M22 System:
- । The Universal IPL handpiece has ten (10) different filters available: Cut-off filters of 515, 560, 590,615, 640, 695 and 755 nm, Notch Filters of 400-600 & 800-1200 nm and 530-650 & 900-1200 nm, and a Narrow band filter of 525-585 nm. Further, the IPL handpiece has three (3) sapphire cool light guides available with sizes of: 15mm x 35mm, 8mm x 15 mm, 6mm diameter.
- The Multi-Spot Nd:YAG handpiece has four (4) different light guides/tips available in sizes of: 2mm x 4mm, 6 mm, 9 mm and 1.5 mm.
- -The ResurFX handpiece has two (2) different treatment tips available: SapphireCool and Precision tips.
- । The Q-Switched Nd: YAG handpiece has both disposable and gold plated metal treatment tips available. The disposable treatment tips are available in four (4) different sizes of: 2, 2.5, 3.5, and 5 mm. The metal treatment tips are available in seven (7) different sizes of: 2, 2.5, 3.5, 4, 5, 6 and 8 mm.
### Technological Characteristics and Substantial Equivalence:
The intended use and indications for use of the M22 System are the selected predicate devices. In addition, the same technological characteristics and principles of operation apply for the M22 System and the predicate device.
The Lumenis M22 System main modification that have been made are:
- Addition of the IPL AOPT (Advanced Optimal Pulse Technology) Operation Mode that । assists the physician in specifically tailoring the treatment to the patient's needs.
- Addition of new accessories to the IPL Handpiece a Vascular Filter and a KTP filter. ।
- Addition of a new 6mm diameter light guide to the IPL Handpiece ।
- Addition of 4mm and 8 mm tips to the Q-switched Nd: YAG handpiece. ।
- । Addition of a new 18mm Precision tip to the ResurFX handpiece.
{7}------------------------------------------------
- Addition of scanning shapes to the ResurFX handpiece Vertical line and rectangle. ।
- Additional Lumenis software presets for user convenience. -
- -Software Upgrade to support the specified modifications.
These modifications have been introduced in order to configure a system that improves user convenience or increase their control of the delivered treatment.
Comparison table of technological characteristics of the Lumenis M22 System compared to those of the predicate device is provided below. The new accessories and features are bolded.
| Device<br>Feature | Lumenis M22 System (K142860)<br>(Predicate Device) | Lumenis M22 System (K170060)<br>(Subject Device) |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IPL Handpiece<br>Wavelength | 400 -1200 nm | Same |
| Pulse Duration (msec) | Up to 20 msec - single pulse<br>40-100 msec - multiple pulse | Same |
| Operational Wavelengths | The IPL handpiece comes with 8 filters<br>for various wavelengths:<br>Cut-off filters: 515, 560, 590, 615, 640,<br>695, 755 nm<br>Acne Filter (Notch filter 400-600 and<br>800-1200 nm) | The IPL handpiece comes with 10 filters<br>for various wavelengths:<br>Cut-off filters: 515, 560, 590, 615, 640,<br>695, 755 nm<br>Acne Filter (Notch filter 400-600 and<br>800-1200 nm)<br>Vascular Filter (Notch filter 530-650 &<br>900-1200 nm)<br>KTP filter (525-585 nm) |
| Spot sizes (cm²) | • 8 mm x 15 mm<br>• 15 mm x 35 mm | • 8mm x 15 mm<br>• 15 mm x 35 mm<br>• 6 mm round |
| Max Fluence | Up to 35 J/cm² | Up to 35 or 56 J/cm², upon tip size |
| Pulse Rate [Hz] | Up to 1 Hz | Same |
| Multiple Sequential<br>Pulsing | 1, 2 and 3 pulses | 1, 2 and 3 pulses, varying fluence per<br>pulse (AOPT mode) |
| Multi-Spot Nd:YAG Handpiece<br>Operational Wavelengths | 1064 nm | Same |
| Spot sizes (mm) | 2 x 4, 1.5, 6, 9 | Same |
| Max Fluence | Up to 600 J/cm², upon tip size | Same |
| Pulse Rate [Hz] | Up to 1 Hz | Same |
| Multiple Sequential<br>Pulsing | 1, 2 and 3 pulses | Same |
| ResurFX Handpiece<br>Operational Wavelengths | 1565nm | Same |
| Max Energy | up to 70mJ per micro- beam | up to 40 or 70mJ per micro- beam, upon tip |
| Type of laser | Er:Glass Fiber-laser with scanner | Same |
| Tip treatment width | 18mm SapphireCool Tip | 18 mm SapphireCool Tip<br>18 mm Precision Tip |
| Scanner | Dual axis scanner | Same |
| Scanning shapes | Line, square, rectangle, circle,<br>donut, hexagon | Line, square, rectangle, circle, donut,<br>hexagon, vertical line and vertical |
| Device | Lumenis M22 System (K142860) | Lumenis M22 System (K170060) |
| Feature | (Predicate Device) | (Subject Device) |
| | | rectangle |
| Q-Switched Nd:YAG Handpiece | | |
| Operational Wavelengths | 1064 nm | Same |
| Spot sizes (mm, diameter) | 2, 2.5, 3.5, 5, 6 | 2, 2.5, 3.5, 4, 5, 6, 8 |
| Max Fluence | Up to 14 J/cm², upon tip size | Up to 14 J/cm², upon tip size |
| Pulse Duration(nsec) | 6-8 | Same |
| Pulse Rate [Hz] | 0.5-5.0 | Same |
{8}------------------------------------------------
#### Substantial Equivalent Discussion:
The physical components of the M22 System and handpieces were unchanged compared to the legally marketed predicate device. The handpieces remain unchanged, except for the addition of several accessories and features.
The IPL Handpiece accessories that have been added have the same wavelength, pulse rate, use of sequential pulsing, and use of sapphire cooling. The expected fluence with use of the additional filters and light guides showed lower fluence than that of previously cleared devices. The higher 56J/cm² is supported by Lumenis 510(k) number K020839.
The Q-Switched Nd: Y AG Handpiece two new treatment tips sizes have the same wavelength, pulse rate, pulse duration, and maximum fluence. The expected fluence with use of new tips showed lower fluence than that of the previously cleared device.
The ResurFX Handpiece addition of various treatment shapes and the new Precision Tip have the same wavelength, max energy per pulse and treatment tip width.
All features of the predicate and subject device were validated to work as intended within V&V activities.
### Performance Bench Testing:
Performance testing was conducted in order to demonstrate the performance of the M22 System and its substantial equivalence, with respect to the safety and effectiveness of the cleared predicate device. The following activities were performed:
- Risk analysis activities in compliance with the requirements of ISO 14971.
- . Electrical safety and electromagnetic compatibility testing as required to conform with the following performance standards:
- । IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- । IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility - requirements and tests.
{9}------------------------------------------------
- IPL compatibility testing as required to conform with the following performance ● standard:
- । IEC 60601-2-57 Medical Electrical Equipment - Part 2-57: Particular Requirements For The Basic Safety And Essential Performance Of Non-Laser Light Source Equipment.
- Verification and validation testing: ●
- All software changes have been validated to work as intended. |
- -The Handpiece accessories and features that have been added were validated.
### Conclusions:
Test results indicated that the subject M22 System performs in accordance with its requirements and specifications, in similarity to its predicate device. Consequently, the M22 System was found to perform as well as its predicate, to be as safe and effective for its intended use as its predicate, and is substantially equivalent to its predicate device without raising any new safety and/or effectiveness issues.
{10}------------------------------------------------
### ResurFX System
#### Indications for Use:
The ResurFX System with wavelength of 1565 nm. is indicated for use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue.
#### Device Description:
The ResurFX System is an advanced computer-controlled laser emission system operating at a wavelength of 1565 nm, and is intended for Fractional Non-Ablative Skin Resurfacing application when using a dedicated handpiece.
The following accessories are provided with and/or may be purchased independently for the ResurFX System:
- -Two (2) different treatment tips: SapphireCool and Precision tips.
- । Footswitch.
### Technological Characteristics and Substantial Equivalence:
This ResurFX module (K130028, K142860) was modified to be a standalone system to provide an additional member of the growing product family to meet customer needs.
The intended use and indications for use of the ResurFX System are the selected predicate devices. In addition, the same technological characteristics and principles of operation apply for the ResurFX System and the predicate devices.
Comparison table of technological characteristics of the Lumenis ResurFX System compared to those of the predicate devices is provided below. The new accessories and features are bolded.
| | ResurFX Module (K130028, K142860)<br>(Predicate Devices) | ResurFX System (Subject)<br>(Subject Device) |
|----------------------------|----------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Operational<br>Wavelengths | 1565nm | Same |
| Max Energy | up to 70mJ per micro- beam | up to 40 or 70mJ per micro- beam, upon<br>tip |
| Type of laser | Er:Glass Fiber-laser with scanner | Same |
| Tips | SapphireCool | SapphireCool, Precision |
| Spot size, mm | 5-18 | Same |
| Tip treatment<br>width | 18mm | Same |
| Scanner | Dual axis scanner | Same |
| Scanning shapes | Line, square, rectangle, circle,<br>donut, hexagon | Line, square, rectangle, circle,<br>donut, hexagon, vertical line and vertical<br>rectangle |
| Input Power | 100-240 VAC 12Amax, 50/60Hz | 100-240 VAC, 3A, 50/60 Hz |
{11}------------------------------------------------
#### Substantial Equivalent Discussion:
The standalone ResurFX module was modified to be a standalone system. The ResurFX handpieces remain unchanged, except for the addition of an accessory and feature.
The ResurFX Handpiece addition of various treatment scanning shapes and the new Precision Tip have the same wavelength, max energy per pulse, and treatment tip width.
All features of the predicate and subject device were validated to work as intended within V & V activities.
#### Performance Bench Testing:
Performance testing was conducted in order to demonstrate the performance of the ResurFX System and its substantial equivalence, with respect to the safety and effectiveness of the cleared predicate devices. The following activities were performed:
- Risk analysis activities in compliance with the requirements of ISO 14971. ●
- Electrical safety and electromagnetic compatibility testing as required to conform with . the following performance standards:
- IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic । safety and essential performance.
- IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for । basic safety and essential performance- Collateral standard: Electromagnetic compatibility - requirements and tests.
- Verification and validation testing: ●
- All software changes have been validated to work as intended. |
- The ResurFX Handpiece added treatment shapes and the Precision Tip were । validated. The wavelength, max energy per pulse, type of laser, and treatment tip width remain unchanged. The expected fluence with use of the new scan shapes and Precision tip showed same or lower fluence than that of previously cleared devices.
- Environmental testing demonstrating the ability of the subject device to withstand variant . operation, storage and transportation conditions.
#### Conclusions:
Test results indicated that the subject ResurFX System performs in accordance with its requirements and specifications, in similarity to its predicate devices. Consequently, the ResurFX System was found to perform as well as its predicates, to be as safe and effective for its intended use as its predicate, and is substantially equivalent to its predicate device without raising any new safety and/or effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.