K170056 · Tianjin Zhengtian Medical Instrument Co., Ltd. · HWC · Oct 23, 2017 · Orthopedic
Device Facts
Record ID
K170056
Device Name
IRENE Cannulated Screw System
Applicant
Tianjin Zhengtian Medical Instrument Co., Ltd.
Product Code
HWC · Orthopedic
Decision Date
Oct 23, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IRENE Cannulated Screw System is intended for fracture fixation, fusion and osteotomies of bones appropriate for the size of the device. For specific screw indications please see below. Acutec Headless Compression Screw: Used for fusion, fractures, or osteotomies of the clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus and other small bones. Cannulated Screw: Intended for bone fracture fixation and bone fragment fixation. Washers may be used with the Cannulated Screw in cases where the patient has poor bone quality.
Device Story
IRENE Cannulated Screw System consists of Acutec Headless Compression Screws and Cannulated Screws with optional washers. Device used for bone fracture fixation, fusion, and osteotomies. Screws are cannulated to allow placement over guide wires. Operated by orthopedic surgeons in clinical settings. Output is mechanical fixation of bone fragments. Benefits include stabilization of fractures and support for bone healing. Device is non-sterile and requires steam sterilization before use.
Clinical Evidence
No clinical data is presented in this submission. Bench testing only.
Technological Characteristics
Materials: Titanium Alloy (ASTM F 136, ISO 5832-3). Design: Cannulated, headless or headed bone screws. Energy source: N/A (mechanical). Connectivity: N/A. Sterilization: Steam sterilization. Software: None.
Indications for Use
Indicated for fracture fixation, fusion, and osteotomies of bones appropriate for device size, including clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus, and other small bones. Cannulated screws indicated for bone fracture and fragment fixation; washers indicated for patients with poor bone quality.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
OsteoMed Headless Cannulated Screw System (k063298)
DARCO® Headed Cannulated Screws (k100359)
Reference Devices
IRENE Locking Compression Plate System (k151508)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a symbolic representation of a parent cradling a child.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Tianjin ZhengTian Medical Instrument Co., Ltd Wang Qi Regulatory Affairs Specialist No.318, Jingyi Road, Airport Economic Zone Tianjin, CN 100082 P.R. China
October 23, 2017
Re: K170056
Trade/Device Name: IRENE Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: September 26, 2017 Received: 09/26/2017
Dear Wang Qi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K170056
Device Name IRENE Cannulated Screw System
Indications for Use (Describe)
IRENE Cannulated Screw System is intended for fracture fixation, fusion and osteotomies of bones appropriate for the size of the device. For specific screw indications please see below.
Acutec Headless Compression Screw:
Used for fusion, fractures, or osteotomies of the clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus and other small bones.
Cannulated Screw:
Intended for bone fracture fixation and bone fragment fixation. Washers may be used with the Cannulated Screw in cases where the patient has poor bone quality.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR\$807.92 and the Safe Medical Devices Act of 1990, the following information is provided for IRENE Cannulated Screw System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, "Format for Traditional and Abbreviated 510(k)", issued on August 12, 2005.
# 1. 510(k) Submitter and Owner
Tianjin ZhengTian Medical Instrument Co., Ltd Address: No.318, Jingyi Road, Airport Economic Zone, Tianjin, P.R. China Official Correspondent: Xiang, Zhaojun Position: Regulatory Affairs Manager Tel: +86 10 82292929 Fax: +86 10 82290740 E-mail: xiangzhaojun@naton.cn Date of preparation of the Summary: Oct 17, 2017
# 2. Submission Contact
Wang, Qi Position: Regulatory Affairs Specialist Tel: +86 10 82292929 E-mail: wangqi@naton.cn
# 3. Proposed Device
Trade Name: IRENE Cannulated Screw System Common Name: Screw, Fixation, Bone Class: Class 2 Regulation: 21 CFR 888.3040 Product Code: HWC (Bone Fixation Screw) Review Panel: Orthopedic
# Device description:
IRENE Cannulated Screw System includes two kinds of screws, Acutec Headless Compression Screw and Cannulated Screw. Acutec Headless Compression Screw has six different diameters in various lengths and is
{4}------------------------------------------------
manufactured from Titanium Alloy (ASTM F 136 and ISO 5832-3). Cannulated Screw has four different diameters in various lengths and contains two optional washers. Cannulated Screw and washers are manufactured from Titanium Alloy (ISO 5832-3).
Indication for use:
IRENE Cannulated Screw System is intended for fracture fixation, fusion and osteotomies of bones appropriate for the size of the device. For specific screw indications please see below.
Acutec Headless Compression Screw:
Used for fusion, fractures, or osteotomies of the clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus and other small bones.
Cannulated Screw:
Intended for bone fracture fixation and bone fragment fixation. Washers may be used with the Cannulated Screw in cases where the patient has poor bone quality.
| | Predicate Device 1 | Predicate Device 2 | Reference Device |
|--------------------------|-------------------------------------------------|------------------------------------|----------------------------------------------|
| 510(k)<br>Number | k063298 | k100359 | k151508 |
| Predicate<br>Device Name | OsteoMed Headless<br>Cannulated Screw<br>System | DARCO® Headed<br>Cannulated Screws | IRENE Locking<br>Compression Plate<br>System |
| Manufacture | OsteoMed | Wright | Zhengtian |
| Classification | 888.3040(HWC) | 888.3040(HWC) | 888.3040(HWC)<br>888.3030(HRS) |
#### 4. Predicate Device and Reference Device
#### 5. Preclinical Data
Mechanical testing is conducted to demonstrate that differences in technological characteristics of the proposed device to predicate device do not raise new questions regarding safety and effectiveness, and Substantially Equivalent to the predicate device is thus concluded. The testing including torsional properties, driving torque test and axial pullout test incorporates the following standard as test method to evaluate the performance of both proposed and predicate device: ASTM F543-13, Standard
{5}------------------------------------------------
Specification and Test Methods for Metallic Medical Bone Screws.
# 6. Clinical Data
No clinical data is presented in this submission.
# 7. Substantial Equivalence Comparison
Table 1: comparison between Acutec Headless Compression Screw and predicate devices 1
| Element | Proposed Device<br>Acutec Headless Compression<br>Screw | Predicate Device 1<br>OsteoMed Headless Cannulated<br>Screw System |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Class | 2 | 2 |
| Regulation<br>number | 21 CFR 888.3040 | 21 CFR 888.3040 |
| Indication<br>for use | Used for fusion, fractures, or<br>osteotomies of the clavicle,<br>humerus, radius, ulna, ilium,<br>femur, patella, fibula, tibia, talus,<br>malleolus, calcaneus and other<br>small bones. | The OsteoMed Headless Screw<br>System is indicated for use in<br>bone reconstruction, osteotomy,<br>arthrodesis, joint fusion, fracture<br>repair, and fracture fixation of<br>bone appropriate for the size of<br>the device. Screws are intended<br>for single use only. |
| Material | Ti6Al4V | Ti6Al4V |
| Design | Fully threaded, headless,<br>cannulated | Headless, cannulated |
| Diameter | 2.5~7.5mm | 2.0~6.5mm |
| Length | 8~120mm | 10~120mm |
| Supplied | Non-sterile | Non-sterile |
| Single-use<br>only | Yes | Yes |
| Sterilization | Steam sterilization | Steam sterilization |
Table 2: comparison between Cannulated Screw and predicate devices 2
| Element | Proposed Device | Predicate Device 2 |
|-------------------|------------------|----------------------|
| | Cannulated Screw | Wright DARCO® Headed |
| | | Cannulated Screws |
| Class | 2 | 2 |
| Regulation number | 21 CFR 888.3040 | 21 CFR 888.3040 |
{6}------------------------------------------------
| Indication for use | | | |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| | Intended for bone fracture<br>fixation and bone fragment<br>fixation. Washers may be<br>used with the Cannulated<br>Screw in cases where the<br>patient has poor bone<br>quality. | Intended for use over a guide pin<br>or wire for bone fracture fixation<br>and bone fragment fixation.<br>Washers may be used with the<br>screws in cases where the<br>patients has poor bone quality. | |
| | Material | Ti6Al4V | Ti6Al4V |
| | Design | self-tapping,<br>cannulated | self-tapping/self-drilling,<br>cannulated |
| | Diameter | 3.5~7.3mm | 2.7~7.5mm |
| | Length | 20~110mm | 8~120mm |
| | Supplied | Non-sterile | Non-sterile |
| | Single-use only | Yes | Yes |
| Sterilization | Steam sterilization | Steam sterilization | |
#### 8. Substantial Equivalence Conclusion
The proposed device is found to possess similar technological characteristics on design under the premises of sharing same intended use. Device performance is compared through the testing of mechanical properties of the proposed and predicate device by methods stipulated in published standards in order to evaluate whether differences existing in the design raise safety issue by compromising performance. Results of testing indicates no difference between the level of performance of proposed device and the predicate device, through which the differences are deemed not relevant to device performance and has been justified, and the conclusion of Substantial Equivalence is reached.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.