K170039 · Advanced Sterilization Products (Asp) · FRC · Jun 20, 2017 · General Hospital
Device Facts
Record ID
K170039
Device Name
STERRAD VELOCITY Biological Indictator and Reader
Applicant
Advanced Sterilization Products (Asp)
Product Code
FRC · General Hospital
Decision Date
Jun 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The STERRAD VELOCITY™ Biological Indicator, in conjunction with the STERRAD VELOCITY™ Reader, is intended to be used as a standard method for frequent monitoring STERRAD Sterilization Systems: · STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear™ Technology • STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear™ Technology · STERRAD® 100S
Device Story
System comprises self-contained biological indicator (BI) and automated reader; monitors hydrogen peroxide gas plasma sterilization. BI contains Geobacillus stearothermophilus spores and nutrient medium with non-fluorescent substrate (α-MUG). During sterilization, hydrogen peroxide exposure inactivates spores and reduces enzyme activity. Reader incubates BI at 57°C; uses UV LED to excite fluorescence from hydrolyzed α-MUG. Photodetector measures fluorescence; algorithm determines positive/negative result based on enzyme activity levels. Provides final result in 30 minutes. Used in clinical settings by healthcare personnel to verify sterilization efficacy. Features barcode scanner and network connectivity for record-keeping. Benefits patient by ensuring sterilization cycle success before instrument reuse.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via spore population determination, D-value verification (Holcomb-Spearman-Karber/Survivor Curve), reduced incubation time (RIT) study, growth inhibition, holding time assessment, and chemical indicator functionality. System performance validated across all specified STERRAD cycles, including those with ALLClear™ technology.
Technological Characteristics
Self-contained BI with glass fiber spore disc (≥10^6 CFU G. stearothermophilus) and thermoplastic vial/cap. Reader uses UV LED, optical filters, and photodiode for fluorescence detection. Incubation at 57 ± 2°C. Connectivity via barcode scanner and network interface. Electrical safety per IEC/EN 61010-1/2-010; EMC per FCC Part 15 Class A.
Indications for Use
Indicated for routine monitoring of STERRAD hydrogen peroxide gas plasma sterilization cycles in healthcare facilities. Compatible with STERRAD 100NX, STERRAD NX, and STERRAD 100S systems.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2017
Advanced Sterilization Products (asp) Sun Choi Senior Regulatory Affairs Program Lead 33 Technology Drive Irvine, California 92618
Re: K170039
Trade/Device Name: STERRAD VELOCITY™ Biological Indicator and Reader Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: May 19, 2017 Received: May 22, 2017
Dear Sun Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S6
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170039
Device Name
STERRAD VELOCITY™ Biological Indicator and Reader
#### Indications for Use (Describe)
The STERRAD VELOCITY™ Biological Indicator, in conjunction with the STERRAD VELOCITY™ Reader, is intended to be used as a standard method for frequent monitoring STERRAD Sterilization Systems: · STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without ALLClear™ Technology
• STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear™ Technology
· STERRAD® 100S
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <label><input type="checkbox"/> Registration Use (Part 81 CFR 801 Subpart D)</label> <label><input checked="" type="checkbox"/> On-The-Spot Use (81 CFR 801 Subpart G)</label> |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized font. The letters are white and are set against a rounded, light blue background. There is a registered trademark symbol in the upper right corner of the logo.
# 510(k) Summary
# Advanced Sterilization Products STERRAD VELOCITY™ Biological Indicator and Reader
### General Information
| Submitter Name: | Advanced Sterilization Products<br>Division of Ethicon, Inc., a Johnson & Johnson company |
|-----------------|--------------------------------------------------------------------------------------------------------|
| Address: | 33 Technology Drive<br>Irvine, CA 92618 |
| Contact Person: | Sun Choi<br>Senior Regulatory Affairs Program Lead<br>Phone: 949-453-6378<br>Email: sjchoi@its.jnj.com |
| Date Prepared: | June 8, 2017 |
| 510(k) Number: | K170039 |
## Device Name
| Proprietary Name: | STERRAD VELOCITY™ Biological Indicator and Reader |
|----------------------|---------------------------------------------------|
| Common Name: | Biological Indicator |
| Classification Name: | Biological Sterilization Process Indicator |
| Device Class: | Class II |
| Product Code: | FRC |
| CFR Section: | 21 CFR 880.2800 |
### Predicate Devices
- Primary Predicate: ● STERRAD® CYCLESURE® 24 Biological Indicator, K151971 cleared October 29, 2015
- 3M Attest™ Rapid Readout Biological Indicator 1295 and 3M Attest™ Auto-Reader 490H, K140392 cleared August 13, 2014, and K152060 cleared December 1, 2015
### Device Description
### STERRAD VELOCITY BIOLOGICAL INDICATOR
The STERRAD VELOCITY Biological Indicator is a self-contained biological indicator (BI), used in conjunction with the STERRAD VELOCITY Reader, that is intended for frequent monitoring of the STERRD Sterilization Cycles, using rapid readout technology that provides a final fluorescence result in 30 minutes at the incubation temperature of 57 ± 2℃.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The letters are white and are set against a blue-green background. The logo is enclosed in a rounded rectangle shape, and there is a small registered trademark symbol in the upper right corner.
The STERRAD VELOCITY BI can also be determined as growth-positive or growth-negative via an optional visual pH-based color change result (using bromocresol purple). When using this method, the biological indicator must be cultured in an incubator at 55-60°C for 5 to 7 days to get a final visual result.
The STERRAD VELOCITY BI consists of a glass fiber disc containing a minimum of 1 x 10° Geobacillus stearothermophilus (ATCC 7953) spores and a glass ampoule containing nutrient growth medium and non-fluorescent substrate, as well as a vial, cap, cap label, insert, and chemical indicator. The spore disc, growth media ampoule, and insert are contained in a clear plastic vial with a vented cap. The cap is designed with sterilant ingress openings which allow for penetration of hydrogen peroxide vapor into the vial during the sterilization process. The chemical indicator (CI), placed on the top of the cap, is a Type 1 process indicator that changes color from red/pink to yellow with some red/orange/brown dots when exposed to hydrogen peroxide.
The STERRAD VELOCITY BI has the same α-glucosidase enzyme system for the fundamental scientific technology as the predicate device, 3M Attest™ Rapid Readout Biological Indicator 1295. The a-glucosidase enzyme, which is generated naturally during growth of G. stearothermophilus and released during spore germination, hydrolyzes the bond between the glucose and 4methylumbelliferyl (4-MU) moieties of 4-methylumbelliferyl a-D-glucopyranoside (a-MUG). In the combined state, α-MUG is not fluorescent. Once the bond between the glucose and 4-MU is hydrolyzed, the 4-MU component becomes fluorescent when excited with UV light. Therefore, the a-glucosidase enzyme in its active state can be detected by measuring the fluorescence produced by the enzymatic hydrolysis of α-MUG.
The resultant fluorescent by-product (4-MU), is detected by the Reader and the fluorescent signal is used to determine the positive or negative result of the biological indicator. The measured enzyme activity is reduced upon exposure to hydrogen peroxide. As the enzyme activity is directly correlated with the spore outgrowth, the reduction of the enzyme activity below a certain level indicates that all spores have been inactivated. The level of the fluorescence response is determined using the algorithm developed for the STERRAD VELOCITY BI and is used to distinguish between the positive and negative responses.
# STERRAD VELOCITY READER
The STERRAD VELOCITY Reader is designed to automatically read the STERRAD VELOCITY BI to obtain the final fluorescence result in 30 minutes at the incubation temperature of 57 ± 2℃. The STERRAD VELOCITY Reader utilizes the fluorometric enzymatic assay method to detect the enzyme activity from the BI and the fluorescence emitted from the BI is converted into a voltage. This voltage reading is then used by the fluorescence algorithm in the Reader to determine the final fluorescence result.
There are eight individual BI incubation wells in the STERRAD VELOCITY Reader. Its heater system is designed to maintain the biological indicators at 57 ± 2℃ to promote the outgrowth of the indicator organisms. Each well contains an ultraviolet light source that excites fluorescence in the growth medium, and a photodetector to detect that fluorescence.
The STERRAD VELOCITY Reader features a touch screen for an effective user interface. Directly under each well is a well number illuminated by a well status indicator light. Three colors (white, green, and red) and two states (off and solid line) are used for the indicator light on the touch screen to show the status of the BI processing. The Reader has a thermoplastic exterior which makes it easy to clean and maintain. A built-in barcode scanner coupled with network connectivity makes maintaining sterilization records easy.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The letters are white and outlined in a thin white line, set against a light blue background. The logo is enclosed in a rounded rectangle shape, and there is a small circled "R" symbol in the upper right corner, indicating a registered trademark.
The STERRAD VELOCITY BI and Reader have intended capability for "ecosystem" connectivity and integration. This system is intended to allow communication among STERRAD Sterilizers with ALLClear™ Technology.
# Intended Use/Indications for Use
The STERRAD VELOCITY Biological Indicator, in conjunction with the STERRAD VELOCITY Reader, is intended to be used as a standard method for frequent monitoring of the following STERRAD Sterilization Systems:
- STERRAD® 100NX (STANDARD, FLEX, EXPRESS, and DUO Cycles) with and without . ALLClear™ Technology
- STERRAD NX® (STANDARD and ADVANCED Cycles) with and without ALLClear™ Technology
- STERRAD® 100S ●
The indications for use statement for the STERRAD VELOCITY Biological Indicator and Reader is identical to those of the predicate devices with minor exceptions related to the inclusion of the recently cleared STERRAD ALLClear Technology, and removal of obsoleted STERRAD 50 and 200 Sterilizers.
These exceptions do not affect the safety and effectiveness of the subject device relative to its predicates. Both the subject and predicate devices share intended use for monitoring of hydrogen peroxide gas plasma sterilization processes.
### Comparison of Technological Characteristics with Predicate Device
The subject and predicate devices have technological characteristics based on their use in the monitoring of hydrogen peroxide gas plasma sterilization processes. The following technological characteristics are the same for both the subject device and predicate devices:
- . BI design: Self-contained biological indicator
- . Test organism: Geobacillus stearothermophilus (ATCC 7953)
- Viable spore population: ≥ 106 CFU/BI ●
- . Rapid readout technology: The a-glucosidase enzyme, generated naturally during growth of Geobacillus stearothermophilus, is detected by measuring the fluorescence produced by the enzymatic hydrolysis of a non-fluorescent substrate (α-MUG). The resultant fluorescent byproduct (4-MU) is detected in the reader and the fluorescence signal is used to determine a positive or negative BI result.
- Carrier material for spore disc: Glass fiber ●
- . Materials used for BI components (vial, cap, and insert): Transparent thermoplastic polymer for vial and insert, and opaque thermoplastic polymer for cap
- . Method of fluorescence detection: UV LED, optical filters, with sensing by photodiode
Some technological differences exist between the subject and predicate devices as listed below. However, these differences were successfully assessed through performance testing and do not raise any questions of safety and effectiveness. Therefore, the subject device is as safe and effective as the predicate devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The letters are white and are set against a light blue background. The logo is enclosed in a rounded rectangle shape. There is a registered trademark symbol in the upper right corner of the logo.
- . Resistance characteristics: The use of 5 mg/L hydrogen peroxide that is within the range used for the two predicates (2.5 mg/L and 10 mg/L); and the choice of two out of three recommended test methods as specified in ISO 11138-1:2006, Section 6.4.1.
- Incubation temperature: The temperatures used by the subject and predicate devices are within the temperature range of the optimal growth of test organism, Geobacillus stearothermophilus (ATCC 7953).
- . Reduced incubation time (RIT): The RIT of the subject device (30 minutes) has been successfully verified through performance testing.
- Chemical indicator: The chemical indicator for the subject device has been verified to be a . Type 1 process indicator through performance testing.
Refer to the following Table for a comparison of predicate and proposed device characteristics.
| | Predicate Devices | | Subject device |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Characteristics | 3M Attest™ Rapid Readout<br>Biological Indicator 1295<br>(K140392 and K152060) | STERRAD<br>CYCLESURE 24 BI<br>(K151971) | STERRAD<br>VELOCITY BI |
| Intended Use | Same as STERRAD<br>CYCLESURE 24 BI | Monitoring of hydrogen peroxide<br>gas plasma sterilization processes | Same as STERRAD<br>CYCLESURE 24 BI |
| Indications for Use | Used in conjunction with the 3M<br>Attest Auto-Reader 490H as a<br>standard method of routine<br>monitoring of vaporized hydrogen<br>peroxide sterilization processes<br>in:<br>• STERRAD 100S<br>• STERRAD NX<br>(STANDARD and<br>ADVANCED Cycles)<br>• STERRAD 100NX<br>(STANDARD, FLEX,<br>EXPRESS and DUO Cycles) | Used as a standard method for<br>frequent monitoring of the<br>following STERRAD<br>Sterilization Systems and has<br>been validated for use in health<br>care facilities:<br>• STERRAD 100S<br>• STERRAD 50<br>• STERRAD 200<br>• STERRAD NX<br>• STERRAD 100NX<br>(STANDARD, FLEX and<br>EXPRESS Cycles) | Used as a standard method for<br>frequent monitoring of the<br>following STERRAD<br>Sterilization Systems:<br>• STERRAD® 100NX<br>(STANDARD, FLEX,<br>EXPRESS, and DUO Cycles)<br>with and without<br>ALLClear™ Technology<br>• STERRAD NX®<br>(STANDARD and<br>ADVANCED Cycles) with<br>and without ALLClear™<br>Technology<br>• STERRAD® 100S |
| Organism<br>(Spore, Species, Strain) | Geobacillus stearothermophilus<br>traceable to ATCC™ 7953 | Same | Same |
| Viable Spore Population | $≥ 1x106 CFU/BI$ | Same | Same |
| Carrier Material | PET (Polyethylene terephthalate) | Glass Fiber | Same as STERRAD<br>CYCLESURE 24 BI |
| Resistance<br>Characteristics: | Tested at 10 mg/L vaporized<br>hydrogen peroxide<br>D-value $10 mg/L ≥ 1 second$<br>Survival Time $≥ 5 seconds$<br>Kill Time: 7 minutes | Tested at 2.5 mg/L hydrogen<br>peroxide<br>D-value $2.5 mg/L: ≥ ~1 second$<br>Survival Time: calculated based<br>on D-value<br>Kill Time: 60 seconds | Tested at 5 mg/L hydrogen<br>peroxide<br>D-value $5mg/L: ≥ 1 second$ , using<br>two D-value methods (Survivor<br>Curve and Fraction Negative)<br>per ISO 11138-1 |
#### Biological Indicator Substantial Equivalence Comparison Table
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the letters "ASP" in a stylized, interconnected design. The letters are white and are set against a rounded, light blue background. A small "R" in a circle, indicating a registered trademark, is visible in the upper right corner of the logo.
| | Predicate Devices | | Subject device |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 3M Attest™ Rapid Readout<br>Biological Indicator 1295<br>(K140392 and K152060) | STERRAD<br>CYCLESURE 24 BI<br>(K151971) | STERRAD<br>VELOCITY BI |
| Device Characteristics | | | |
| Rapid Readout<br>Technology | The α-glucosidase enzyme system is generated naturally during growth of Geobacillus stearothermophilus. The α-glucosidase enzyme in its active state is detected by measuring the fluorescence produced by the enzymatic hydrolysis of a non-fluorescent substrate. The resultant fluorescent by-product is detected in the reader. The fluorescence signal is used to determine a positive or negative result. | N/A | Same as 3M predicate |
| Incubation Temperature | 60 +/- 2°C | 55 - 60°C | 57 +/- 2°C |
| Reduced Incubation<br>Time | 4-hour fluorescence results read<br>on 3M 490H Auto-reader | 24-hour visual pH color change<br>results | 30-minute fluorescence results<br>read in STERRAD VELOCITY<br>Reader and optional visual pH<br>color change results at 5 to 7<br>days |
| Chemical Indicator | H2O2 sensitive ink; appears blue<br>until processed, then appears<br>pink | Throughput indicator, changes<br>color from red to golden yellow<br>or bronze to indicate exposure to<br>hydrogen peroxide | Type 1 process indicator<br>changes color from red/pink to<br>yellow or yellow with some<br>red/orange/brown dots to<br>indicate exposure to hydrogen<br>peroxide |
| Shelf Life | 18 months | 12 months | 6 months |
# Reader Substantial Equivalence Table
| | Predicate Device | Subject device |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Device Characteristics | 3M Attest™ 490H Auto-Reader (K140392) | STERRAD VELOCITY Reader |
| Intended Use | Used to incubate and automatically read the 3M Attest™ 1295 Rapid Readout Biological Indicator at 60°C for a final fluorescence result at 4 hours. | Used to incubate and automatically read the STERRAD VELOCITY Biological Indicator at 57°C for a final fluorescence result in 30 minutes. |
| Incubation Temperature | 60 +/- 2°C | 57 +/- 2°C |
| Basis of Rapid Readout | Fluorescence of biological indicator medium | Same |
| Use with Biological Indicators | 3M Attest™ Rapid Readout Biological Indicator 1295 | STERRAD VELOCITY BI |
| Method of Fluorescence Detection | UV LED, optical filters, with sensing by photodiode | Same |
| Indicator of Adequate<br>Sterilization Cycle | (-) on LCD Display | Same ("Negative" displayed on LCD Display) |
| Indicators of Possible<br>Sterilization Cycle Failure | (+) on LCD Display, Audible Alarm | Same ("Positive" displayed on LCD Display, Audible Alarm) |
| Incubation Wells | 10 incubation/reader wells | 8 incubation/reader wells |
| Voltage Range | 100-240 Volts AC (12 Volt DC conversion for internal circuitry) | Same |
| Product Safety | UL/IEC 61010-1 | Same |
| EMC Compliance | FCC Part 15, Subpart B, Class A | Same |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image features a logo with the letters 'ASP' prominently displayed. The letters are stylized and interconnected, with the 'A' resembling a triangle and the 'S' and 'P' forming a continuous, flowing line. The entire logo is rendered in a light blue color, and a small registered trademark symbol is visible in the upper right corner.
## Non-Clinical Data
# STERRAD VELOCITY BIOLOGICAL INDICATOR
Performance testing was conducted to satisfy the requirements for the STERRAD VELOCITY BI that is the subject of this submission, as outlined in the Guidance for Industry and FDA Staff: Biological Indicator (BI) Premarket Notification [510(k)] Submissions Document issued on: October 4, 2007, and ISO 11138-1:2006, Sterilization of health care products – Biological indicators - Part 1: General requirements. All testing yielded passing results. This testing is summarized in the following Table.
| Performance Testing | Description | Pass/Fail |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Spore Population<br>Determination | Verification of performance characteristics for spore population. | Pass |
| D-value Determination | Verification of D-values using Holcomb-Spearman-Karber<br>Procedure (HSKP) and Survivor Curve method. | Pass |
| Test Organism<br>Characterization | Verification of Geobacillus stearothermophilus (ATCC 7953). | Pass |
| Reduced Incubation Time<br>(RIT) Study | Verification of reduced incubation time: 30 minutes for<br>fluorescence results and 5 to 7 days for spore growth results. | Pass |
| Growth Inhibition Study | Verification that carrier and primary packaging materials do not<br>have bacteriostatic effects that inhibit the growth of the indicator<br>microorganism, Geobacillus stearothermophilus. | Pass |
| Holding Time Assessment | Verification that labeled holding time does not have adverse<br>effects on the resistance characteristics for the BI growth results or<br>the fluorescence response characteristics for the fluorescence<br>results. | Pass |
| Chemical Indicator<br>Functionality Study | Verification of functionality of the chemical indicator (CI) as a<br>Type 1 process indicator. | Pass |
| BI Performance Study in<br>STERRAD 100NX | Verification of performance in monitoring STERRAD 100NX<br>STANDARD, FLEX, EXPRESS, and DUO Cycles. | Pass |
| BI Performance Study in<br>STERRAD NX | Verification of performance in monitoring STERRAD NX<br>STANDARD and ADVANCED Cycles. | Pass |
| BI Performance Study in<br>STERRAD 100S | Verification of performance in monitoring STERRAD 100S<br>Cycle. | Pass |
| Impact of Load Conditioning<br>(ALLClear Technology)<br>Feature on STERRAD<br>VELOCITY BI | Verification that the load conditioning feature of the STERRAD<br>100NX and NX with ALLClear Technology has no statistically<br>significant impact on the STERRAD VELOCITY BI. | Pass |
| Shelf Life Study (Real Time) | Verification that the STERRAD VELOCITY BI meets its end of<br>shelf life requirements. | Pass |
#### Summary of Performance Testing
### STERRAD VELOCITY READER
Electrical Safety testing was conducted to verify compliance to the standards listed below. The testing provides a standardized level of assurance that the system is electrically safe when operated and maintained in accordance with the STERRAD VELOCITY Reader User's Guide.
- . IEC/EN 61010-1:2010; Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with the letters ASP in a stylized font. The A is represented by a triangle, and the letters are white against a light blue background. The logo is enclosed in a rounded rectangle, and there is a small registered trademark symbol in the upper right corner.
- . IEC/EN 61010-2-010:2014: Safety requirements for electrical equipment for measurement. control, and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials
- . UL 61010-1:2012; UL Standard for Safety Electrical Equipment for Measurement, Control, and Laboratory Use; Part 1: General Requirements
- UL 61010-2-010:2015; UL Standard for Safety Requirements for Electrical Equipment for . Measurement, Control and Laboratory Use – Part 2-010: Particular Requirements for Laboratory Equipment for the Heating of Materials
- . CAN/CSA-C22.2 No. 61010-1-12 (2012); Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 1: General Requirements
- CAN/CSA-22.2 No. 61010-2-010:15 (2015); Safety requirements for electrical equipment for . measurement, control and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials
Electromagnetic Compatibility testing was completed to verify compliance of Radiated and Conducted Emissions per US Code of Federal Regulations (CFR), Title 47, Part 15, Subpart B (Class A). Test results met the requirements of the regulation and standards as shown in the following table:
| Test Description | Standards | Test Result |
|--------------------------------|-------------------------------------------------------|-------------|
| Radiated Emissions<br>Class A | FCC Part 15, Subpart B<br>ICES-003<br>ANSI C63.4:2014 | Compliant |
| Conducted Emissions<br>Class A | FCC Part 15, Subpart B<br>ICES-003<br>ANSI C63.4:2014 | Compliant |
Software Verification and Validation Testing was conducted and documentation was provided within the submission as recommended by Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005. Unit, Integration, and System level testing were successfully completed.
# Clinical Data
No clinical data was generated in support of this Premarket Notification.
# Summarv
The STERRAD VELOCITY BI and Reader and its predicate devices, the STERRAD CYCLESURE 24 Biological Indicator and the 3M Attest Rapid Readout Biological Indicator 1295 and 3M Attest Auto-Reader 490H, have the same intended use and indications for use with minor exceptions that are related to the incorporation of new product clearances and removal of obsolete products. Further, the subject device and predicate devices largely share the same technological characteristics, i.e., some technological differences exist: however, these differences were successfully assessed through performance testing, and do not raise any questions of safety and effectiveness. Therefore, the subject device is as safe and effective as the predicate devices.
# Conclusion
The subject devices (STERRAD VELOCITY Biological Indicator and Reader) are substantially equivalent to the claimed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.