This device is intended to establish and maintain nephrostomy and internal drainage following pyeloplasty for pediatic patients. This device is intended for patients 6 months and older.
Device Story
The Salle Intraoperative Pyeloplasty Stent Set is a single-lumen, double-pigtail ureteral stent used for internal and external drainage following pediatric pyeloplasty. The device is inserted by a physician to bridge the renal pelvis and bladder, with a flexible extension exiting percutaneously. It features sideports and ink marks for visualization during placement and includes an internal positioning stylet. The set comprises the stent, wire guide, connecting tube, adaptor, and retention disc. It is an indwelling device intended for short-term use (up to 4 weeks). The physician monitors placement via fluoroscopy. By providing a conduit for urine, the device facilitates healing and prevents obstruction post-surgery, benefiting the patient by maintaining renal drainage.
Clinical Evidence
Bench testing only. Testing included retention/break strength (time zero, post-aging, and after 30-day artificial urine soak), wire guide compatibility, kink radius, flow rate/leakage, radiopacity (ASTM F640-12), and MR compatibility (ASTM F2052-15, F2213-06, F2119-07). Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, material-mediated pyrogenicity, 2-week and 4-week implantation, subacute/subchronic toxicity, and genotoxicity.
Technological Characteristics
Single-lumen, double-pigtail ureteral stent; 4.0 or 4.7 French diameters; 12-18 cm lengths. Materials evaluated per ISO 10993-1. Radiopaque for fluoroscopic visualization per ASTM F640-12. MR conditional per ASTM F2052-15, F2213-06, and F2119-07. Supplied sterile for single-use; 3-year shelf life.
Indications for Use
Indicated for pediatric patients 6 months and older requiring nephrostomy and internal drainage following pyeloplasty.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Expel™ Nephroureteral Drainage Stent with Twist-Loc™ Hub System (K141344)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2017
Cook Incorporated Minjin Choi Regulatory Affairs Specialist 750 Daniels Wav. P.O. Box 489 Bloomington, IN 47404
Re: K170010
> Trade/Device Name: Salle Intraoperative Pyeloplasty Stent Set Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: July 14, 2017 Received: July 14, 2017
Dear Minjin Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be on a white background. The letters are clearly legible and evenly spaced.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170010
#### Device Name Salle Intraoperative Pyeloplasty Stent Set
Indications for Use (Describe)
This device is intended to establish and maintain nephrostomy and internal drainage following pyeloplasty for pediatic patients. This device is intended for patients 6 months and older.
Type of Use (Select one or both, as applicable)
√ Over-The-Counter Use (21 CFR 801 Subpart C) Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif letters. Below "COOK" is the word "MEDICAL" in smaller, white, sans-serif letters, set against a red background that is angled slightly.
COOK INCORPO 750 DANIELS BLOOMINGTON, IN 4740 9.2235 TOLL FREE: 800.45 WWW.COOKMEDICAL.COM
# 2.0 510(k) Summary
# Salle Intraoperative Pyeloplasty Stent Set 21 CFR §807.92 Date Prepared: December 30, 2016
## Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|--------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Minjin Choi |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone: | (812) 339-2235 x104901 |
| Contact Fax: | (812) 332-0281 |
### Device Information:
| Trade Name: | Salle Intraoperative Pyeloplasty Stent Set |
|------------------------------------|--------------------------------------------|
| Common Name: | Stent, Ureteral |
| Classification Name: | Ureteral Stent |
| Classification Regulation: | 21 CFR §876.4620, Product Code FAD |
| Device Class/Classification Panel: | Class II, Gastroenterology/Urology |
# Predicate Device:
- " Primary predicate device:
Expel™ Nephroureteral Drainage Stent with Twist-Loc™ Hub System (K141344)
- י Secondary predicate device:
Pediatric Ureteral Stent (K060673)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in smaller, white block letters.
K1700
Page 2
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
## Device Description:
The Salle Intraoperative Pyeloplasty Stent is a single-lumen ureteral stent inserted following pyeloplasty in order to provide internal and external drainage. It is a prolonged indwelling device not to exceed 4 weeks in the body.
The Salle Intraoperative Pyeloplasty Stent Set is comprised of a stent, wire guide, connecting tube, adaptor, and retention disc with pull tie. The Salle Intraoperative Pyeloplasty Stent is a double pigtail stent that advances from the renal pelvis to the bladder with a flexible extended portion that is brought out percutaneously through the kidney. The Salle stent is available in outer diameters of 4.0 or 4.7 French with working lengths of 12 to 18 cm. The stent's proximal pigtail forms in the renal pelvis, while the distal pigtail forms in the bladder. Drainage can occur internally (from the kidney to the bladder) or externally (from renal pelvis to the outside of the patient). The distal (bladder) pigtail coil is available in a multi-length stent configuration with drainage holes. The proximal (kidney) pigtail coil also contains sideports and ink marks to facilitate the visualization of accurate placement. The proximal end of the stent has an internal stylet which acts as a positioning stylet.
The set will be supplied sterile and is intended for one-time use. The set is packaged in a peel-open pouch with a three-year shelf life.
### Indications for Use:
The Salle Intraoperative Pyeloplasty Stent Set is intended to establish and maintain nephrostomy and internal drainage following pyeloplasty for pediatric patients. This device is intended for patients 6 months and older.
### Comparison to Predicate Devices:
The Salle Intraoperative Pyeloplasty Stent Set and its primary predicate device, the Expel Nephroureteral Drainage Stent with Twist-Loc Hub System (K141344), are substantially equivalent in that these devices have similar intended uses, fundamental technological characteristics, and design. The modifications from the predicate device include:
- I Indications for Use
- I Stent material
- I Placement method
- I External end of stent
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is written in white, block letters. Below "COOK" is a smaller rectangle with the word "MEDICAL" written in white letters.
K1700
Page 3
COOK INCORPORATED
750 DANIELS WAY
BLOOMINGTON, IN 47404 USA
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
The proposed Salle Intraoperative Pyeloplasty Stent Set is also similar in indications for use, dimensions, and methods of operation to the Pediatric Ureteral Stent (K060673). Differences between the proposed device and the predicate device include:
- 트 Stent sizes
- 트 Stent material
- 트 Placement method
- I External end of stent
Differences between the characteristics of the proposed device sets and the predicate devices are supported by testing.
# Performance Data:
The biocompatibility evaluation for the Salle Intraoperative Pyeloplasty Stent Set was conducted in accordance with the FDA's Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" June 16, 2016 and International Standard ISO 10993-1 "Biological evaluation of medical devices – Part 1 Evaluation and testing within a risk management process" (ISO 10993-1:2009).
The following biocompatibility testing was successfully completed;
- Cytotoxicity
- Sensitization
- I Intracutaneous
- 트 Acute Systemic Toxicity
- Material-Mediated Pyrogenicity
- Implantation - 2 Week
- 트 Implantation - 4 Week
- I Subacute/Subchronic Toxicity
- 트 Genotoxicity
- Toxicological Risk Assessment of Extractable Chemicals
A summary of the bench testing completed is detailed in Table 1.
Testing was successfully performed per Cook Inc.'s design control system in line with 21 CFR 820.30. The results from performance testing demonstrated that the subject device will remain sterile and will perform as intended for its 3 year shelf life.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font on a red background.
Table 1: Summary of the Bench Testing Performed
| Test | Acceptance Criteria | Result |
|--------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Retention Strength and<br>Break Strength Testing<br>(Simulated use – Time<br>Zero and post aging) | Retention Strength<br>The minimum retention strength shall be greater than<br>or equal to 0.03 N, but less than or equal to 3.9 N<br>(90% coverage, 90% confidence).<br>Break Strength<br>The minimum break strength shall be greater than<br>3.9 N (90% coverage, 90% confidence). | All acceptance<br>criteria were met.<br>Pass |
| Retention Strength and<br>Break Strength Testing<br>Following a 30 Day<br>Artificial urine Soak<br>(Simulated use – Time<br>Zero and post aging) | Retention Strength<br>The minimum retention strength shall be greater than<br>or equal to 0.03 N, but less than or equal to 3.9 N<br>(90% coverage, 90% confidence).<br>Break Strength<br>The minimum break strength shall be greater than<br>3.9 N (90% coverage, 90% confidence). | All acceptance<br>criteria were met.<br>Pass |
| Wire Guide Compatibility,<br>Kink Radius, and Flow<br>Rate and Leakage Post<br>Kinking Testing | Wire Guide Compatibility<br>The appropriate size wire guide shall pass through<br>the inner lumen of each test article.<br>Kink Radius<br>The kink radius of each test article shall be less than<br>30 mm based on worst-case analysis.<br>Flow Rate and Leakage Post Kinking<br>Flow rate performed for characterization purposes<br>only. If leakage detected, test article deemed a<br>failure. | All acceptance<br>criteria were met.<br>Pass |
| Radiopacity Testing | Testing Per ASTM F640-121<br>The Salle Stent is visible under fluoroscopy.<br>Visibility of the Salle Stent is equal to or greater than<br>the visibility of the user-defined standard. | All acceptance<br>criteria were met.<br>Pass |
| MR Testing | Magnetically Induced Displacement Force<br>Testing per ASTM F2052-152<br>Deflection Angle <45°<br>Magnetically Induced Torque<br>Testing per ASTM F2213-063 (2011)<br>$τmag< τgrav$<br>Electrical Conductivity<br><1 S/m<br>MR Image Artifacts<br>Testing per ASTM F2119-074 (2013)<br>For information only | All acceptance<br>criteria were met.<br>Pass |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a dark red background.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA HE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
### Conclusion:
The results of these tests support a conclusion that the Salle Intraoperative Pyeloplasty Stent Set will perform as intended. The proposed device set does not raise new questions of safety or effectiveness as compared to the predicate devices.
1 ASTM F640-12 Standard Test Methods for Determining Radiopacity for Medical Use
2 ASTM F2052-15 Standard test method for measurement of magnetical devices in the magnetic resonance environment
3 ASTM F2213-06 (2011) Standard test method for magnetically induced torque on passive implants in the magnetic resonance environment.
4 ASTM F2119-07 (2013) Standard test method for evaluation of MR image artifacts from passive implants
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.