OptiFix Open Absorbable Fixation System - 20 absorbable fasteners
K163698 · C.R. Bard, Inc. · GDW · Apr 11, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163698
Device Name
OptiFix Open Absorbable Fixation System - 20 absorbable fasteners
Applicant
C.R. Bard, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Apr 11, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OptiFix™ Open Absorbable Fixation System is indicated for the fixation of surgical mesh to tissues during open surgical procedures, such as hernia repair.
Device Story
OptiFix Open is a sterile, single-use surgical instrument designed for fixating surgical mesh to tissue during open procedures like hernia repair. The device consists of a deployment component with an ergonomic handle, a trigger mechanism, and a 27cm curved shaft (35° angle) that delivers 20 synthetic absorbable fasteners. Fasteners are composed of Poly (D, L)-lactide dyed with D & C Violet No. 2. The surgeon operates the device manually; squeezing the trigger drives the fastener into the tissue. The device provides mechanical fixation of the mesh, which is intended to benefit the patient by securing the mesh in place during the healing process. The device is intended for use by healthcare professionals in an open surgical setting.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), actuation torque, fastener deployment, gap height, ball burst testing, pullout force, and resorption profile assessment. Preclinical safety was established by referencing animal and histological data from previously cleared reference devices (K082396, K132134).
Technological Characteristics
Single-use manual deployment system. Fasteners: Poly (D, L)-lactide, 6.7mm length, dyed with D & C Violet No. 2. Shaft: 27cm length, 35° angle. Sterilization: Gamma irradiation. Biocompatibility: ISO 10993-1 compliant. No electrical or software components.
Indications for Use
Indicated for the fixation of surgical mesh to tissues during open surgical procedures, such as hernia repair.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, composed of three overlapping human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 11, 2017
C.R. Bard Incorporated Ms. Katherine Earle Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K163698
Trade/Device Name: Optifix Open Absorbable Fixation System - 20 Absorbable Fasteners Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: December 27, 2016 Received: December 29, 2016
Dear Ms. Earle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163698
Device Name
OptiFix™ Open Absorbable Fixation System
Indications for Use (Describe)
The OptiFix™ Open Absorbable Fixation System is indicated for the fixation of surgical mesh to tissues during open surgical procedures, such as hernia repair.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Section 5 510(k) Summary
#### I. SUBMITTER
Davol Inc. 100 Crossings Boulevard Warwick, RI 02886
| Contact Person: | Katherine Earle |
|-----------------|-------------------------------|
| | Regulatory Affairs Specialist |
| Phone: | (401) 825-8450 |
| Fax: | (401) 825-8765 |
| E-mail: | katherine.earle@crbard.com |
Date Prepared: December 27, 2016
#### II. DEVICE
| Name of Device: | OptiFix™ Open Absorbable Fixation System |
|-----------------------|------------------------------------------|
| Product Code: | 0113320 |
| Common or Usual Name: | Implantable Staple |
| Classification Name: | Implantable Staple (21 CFR §878.4750) |
| Regulatory Class: | II |
| Product Code: | GDW |
#### III. PREDICATE DEVICE
The predicate device for this submission is the OptiFix™ Absorbable Fixation System (K142873), cleared March 12, 2015; marketed by Davol, Inc. This predicate has not been subject to a design-related recall.
The reference devices used in this submission are the following:
- Davol Absorbable Fastener System SorbaFix™ (K082396), cleared January 2, 2009 .
- OptiFix™ Absorbable Fixation System (K132134), cleared December 2, 2013 ●
## IV. DEVICE DESCRIPTION
The OptiFix" Open Absorbable Fixation System, hereinafter referred to as "OptiFix" Open", is a sterile (via gamma) single use device that is comprised of a deployment component and an absorbable fastener component. The device delivers 20 synthetic absorbable fasteners via a curved shaft. The shaft of the OptiFix™ Open Absorbable Fixation System is 27cm in length. The fasteners are 6.7mm in length, are manufactured from Poly (D, L)-lactide and are dyed with D & C Violet No. 2.
{4}------------------------------------------------
#### V. INDICATIONS FOR USE
The OptiFix" Open Absorbable Fixation System is indicated for the fixation of surgical mesh to tissues during open surgical procedures, such as hernia repair.
The Indications for Use statement for the proposed device falls within the indications for use of the predicate device. The proposed and predicate devices have the same intended use.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The fixation of surgical mesh to tissues during open surgical procedures, such as hernia repair using the mechanical actuation (trigger) force to guide the fastener and launch the implant is the operating principle for both the proposed and predicate devices.
The proposed and predicate devices are based on the following similar elements:
- Deployment component - mechanical component used to position and launch the fastener (implant) into tissue, consisting of a shaft connected to an ergonomic handle and integrated user-controller trigger.
- Fastener component -absorbable component composed of Poly (D, L)-lactide, used to ● fixate surgical mesh to tissue.
The following differences exist between the proposed and predicate devices:
- The handle configuration of the proposed device is rotated 180° compared to the predicate device
- The deployment shaft of the proposed device is shorter in length (27 cm) compared to the deployment shaft of the predicate device (39cm).
- The deployment shaft of the proposed device has a 35° permanent angle, whereas the ● predicate device has a straight deployment shaft.
## VII. PERFORMANCE DATA
The following performance data is provided in support of the substantial equivalence determination.
## Performance standards
No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
## Biocompatibility testing
The biocompatibility evaluation for the OptiFix™ Open device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", June 16, 2016; and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
{5}------------------------------------------------
The absorbable fasteners are determined to be long term implants (>30 days) and are supported by the biocompatibility testing listed below.
- Cytotoxicity ●
- Sensitization ●
- Intracutaneous Reactivity ●
- Single-does Systemic Toxicity ●
- Rabbit Pyrogenicity
- . 14-Day Repeat Dose Intravenous Toxicity
- Genotoxicity
- . Chronic Toxicity, 13 weeks
- Implantation, 4 weeks
- . Implantation, 8 weeks
- Implantation, 12 weeks
- Implantation, 26 weeks
- Implantation, 52 weeks ●
The deployment component of the proposed OptiFix™ Open device is determined to be tissue contacting for a duration of less than 24 hours, and has been supported by the biocompatibility testing as listed below.
- Cytotoxicity ●
All samples tested met the acceptance criteria and the proposed OptiFix™ Open device is biocompatible for its intended use.
## Electrical safety and electromagnetic compatibility (EMC)
The proposed OptiFix "Open device is not an electro-mechanical medical device nor is it a medical system, therefore this section does not apply.
## Software Verification and Validation Testing
The proposed OptiFix" Open device does not contain software; therefore this section does not apply.
## Product Testing
The following non-clinical tests were completed for the proposed and predicate device. The OptiFix "Open device passed all the test requirements and demonstrated substantial equivalence to the test results of the predicate device.
- · Performance and Functional testing of the OptiFix™ Open device
- Actuation (trigger) Torque ●
- Fastener Deployment
- Fastener Gap Height
- Ball Burst Testing
- Single Fastener Pullout (Pluck) Force
{6}------------------------------------------------
- · Resorption Profile Assessment of the OptiFix™ Open fastener
- Physical Properties
- Mechanical Shear Force
All samples tested met the established acceptance criteria.
#### Animal Study
Animal studies were not performed on the proposed device for this submission. Based on identical materials and substantially equivalent mechanical strengths measured at the bench level, the in vivo safety and performance of the proposed device was evaluated via the animal and histological studies conducted on the reference devices: Davol Absorbable Fastener System - SorbaFix" (K082396) and OptiFix " Absorbable Fixation System (K132134), which were used to support the predicate OptiFix " Absorbable Fixation System (K142873).
#### Clinical Studies
Clinical studies were not performed for the submission of this proposed device nor were clinical studies performed for the predicate device.
## VIII. CONCLUSIONS
The proposed OptiFix™ Open device is substantially equivalent to the legally marketed predicate device for the following reasons:
- A) The same intended use as the predicate device.
- B) Similar technological characteristics as the predicate device such as: the fastener delivery system, deployment device trigger, and the fastener dimensions, shape/design and material.
- C) The same principle of operation.
The comparative analysis, see Table 5-1 below, as well as the bench and preclinical testing results demonstrate that the proposed OptiFix™ Open device performs as intended and is substantially equivalent to the predicate device that is currently marketed for the same intended use.
| Device<br>Features | Proposed device<br>OptiFix™ Open | Predicate device<br>OptiFix™ (K142873) |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same as predicate | Soft tissue fixation |
| Indication for<br>Use | Indicated for the fixation of<br>surgical mesh to tissues during<br>open surgical procedures, such as<br>hernia repair. | Indicated for the approximation of<br>soft tissue and fixation of surgical<br>mesh to tissues during open or<br>laparoscopic surgical procedures,<br>such as hernia repair |
| Table 5-1: Device Substantial Equivalence - General Characteristics | | |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Features | Proposed deviceTM<br>OptiFixTM Open | Predicate deviceTM<br>OptiFixTM (K142873) |
| Fastener<br>Material | Same as predicate | Poly (D,L) – Lactide |
| Fastener Violet<br>Dye | Same as predicate | D & C Violet No. 2 conforms to 21<br>CFR § 74.1602(a)(1) and (b) |
| Fastener Body<br>Contact | Same as predicate | Long term implant (>30 days)<br>contacting tissue and/or bone |
| Fastener<br>Shape/Design | Same as predicate<br>Image: Purple fastener pointing down | Push Tack with retention features<br>on end<br>Image: Purple fastener pointing down |
| Overall<br>Fastener<br>Dimensions | Same as predicate | 6.7mm overall fastener length<br>Fastener head: 3.7 mm length; 3.5<br>mm thickness |
| Fastener<br>Manufacturing<br>Method | Same as predicate | Injection - Molded |
| Fastener<br>Absorption<br>Time | Same as predicate | 360 days |
| Fastener<br>Quantity per<br>Device | 20 fasteners | 15 & 30 fasteners |
| Fastener<br>Delivery<br>System | Same as predicate | Push – Impact tube pushes<br>fasteners forward over a guidewire |
| Deployment<br>Component<br>Configuration | Image: Deployment Component Configuration | Image: Deployment Component Configuration |
| Deployment<br>Component<br>Shaft Length | 27cm in length | 39cm in length |
| Deployment<br>Component<br>Shaft Angle | 35° | None |
| Deployment<br>Component<br>Body Contact | Same as predicate | Transient use – Less than 24 hour<br>duration |
| Device<br>Sterilization | Same as predicate | Gamma Irradiation<br>(25 - 40 kGy) |
| Device<br>Features | Proposed device<br>OptiFix™ Open | Predicate device<br>OptiFix™ (K142873) |
| Device<br>Packaging | Device is placed into a PETG<br>(Polyethylene terephthalate<br>Glycol-modified) thermoformed<br>tray which is inserted into a foil<br>pouch, sealed and placed into an<br>SBS (Solid Bleached Sulfate)<br>paperboard printed shelf carton. | Device is placed into an SBS<br>(Solid Bleached Sulfate) cardboard<br>insert which is inserted into a foil<br>pouch, sealed and then placed into<br>an SBS (Solid Bleached Sulfate)<br>paperboard printed shelf carton. |
{7}------------------------------------------------
{8}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.