A&D Medical UA-1200BLE UltraConnect Digital Blood Pressure Monitor
Applicant
A&D Company, Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 3, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Pediatric
Indications for Use
A&D LifeSource UA-1200BLE digital blood pressure monitor is designed for adult at home or clinic use. It measures systolic, diastolic, and pulse rate. UA-1200BLE uses the oscillometric method to determine blood pressure. If the user has a massive body movement or irregular heartbeats, a heart symbol will be shown on the app display. At the end of the measurement, the blood pressure results are compared with JNC VII hypertension classification. The rating is also displayed. UA-1200BLE has Bluetooth® low energy wireless (BLE) communication capability with Smart devices. UA-1200BLE fits arm size from 22 to 42 cm.
Device Story
Oscillometric blood pressure monitor; measures systolic/diastolic BP and pulse rate. Device features integrated hoseless cuff design; automatic inflation via internal pump; solenoid valve for rapid exhaust. Input: oscillometric pressure signals from upper arm. Output: BP/pulse values displayed on OLED and transmitted via Bluetooth Low Energy (BLE) to connected app. Used in home or clinic by adult patients. App displays results, compares against JNC VII hypertension classification, and indicates irregular heartbeats. Benefits: enables convenient, wireless tracking of BP trends for hypertension management.
Clinical Evidence
Clinical validation performed per AAMI/ANSI/ISO 81060-2:2013. Study included 85 subjects (age >12, 69.4% male, 30.6% female). Primary endpoints: mean difference and standard deviation of BP measurements compared to reference. Results: SBP mean difference -2.66 mmHg (SD 6.90 mmHg), DBP mean difference +0.90 mmHg (SD 6.63 mmHg). All criteria met.
Indicated for adults (age >12) for the measurement of systolic and diastolic blood pressure and pulse rate in home or clinical settings. Fits arm circumferences 22-42 cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
A&D Medical UA-651 Digital Blood Pressure Monitor (K141160)
A&D Medical UA-767PBT Digital Blood Pressure Monitor (K043217)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 03, 2017
A&D Company, Ltd % Jerry Wang Director of Engineering 1756 Automation Parkway San Jose, California 95131
Re: K163683
Trade/Device Name: A&D Medical UA-1200BLE Ultraconnect Digital Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 31, 2017 Received: April 3, 2017
Dear Jerry Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Hillemann
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K163683
Device Name
UA-1200BLE UltraConnect Digital Blood Pressure Monitor
Indications for Use (Describe)
Measure blood pressure (systolic and diastolic) and pulse rate.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
മ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
Page 1 of 1.
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effective information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.
#### 1. Date Prepared
December 23, 2016
#### 2. Submitter's Information
A&D Engineering, Inc. Mr. Jerry Wang 1756 Automation Parkway, San Jose, CA 95131 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andonline.com
#### 3. Device Information
Proprietary Name:
Regulation Number: Regulation Name: Regulatory Class: Product Code:
A&D Medical UA-1200BLE UltraConnect Digital Blood Pressure Monitor 21 CFR 870.1130 Noninvasive Blood Pressure Measurement System Class II DNX
#### 4. Information for he 510(k) Cleared Devices (Predicate Devices)
- A&D Medical UA-651 Digital Blood Pressure Monitor with 510(k) number K141160 .
- A&D Medical UA-767PBT Digital Blood Pressure Monitor with 510(k) number . K043217
Both predicate devices are designed and manufactured by the same company and facilities as the modified device.
#### ട. Indications for Use
Measure blood pressure (systolic and diastolic) and pulse rate
#### 6. Intended Use
A&D LifeSource UA-1200BLE digital blood pressure monitor is designed for adult at home or clinic use. It measures systolic, diastolic, and pulse rate. UA-1200BLE uses the oscillometric method to determine blood pressure. If the user has a massive body movement or irregular heartbeats, a heart symbol will be shown on the app display. At the end of the measurement, the blood pressure results are compared with JNC VII hypertension classification. The rating is also displayed. UA-1200BLE has Bluetooth® low energy wireless (BLE) communication capability with Smart devices. UA-1200BLE fits arm size from 22 to 42 cm.
#### 7. Device Description - Technological and Operational Characteristics Comparison
UA-1200BLE of digital blood pressure monitor has the same design as the predicate devices with an inflatable cuff which is wrapped around the patient's upper arm. After the user starts the blood pressure measurement, the cuff is inflated automatically by an internal pump. The systolic
{4}------------------------------------------------
### K163683
and diastolic blood pressures are also determined by oscillometric method. The deflation rate is controlled by an internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released immediately. There is a maximum pressure safety setting at 299 mmHg. UA-1200BLE UltraConnect Digital Blood Pressure Monitor will not inflate the cuff higher than 299 mmHg. UA-1200BLE will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the OLED and transmitted via Bluetooth wireless module to the connected app. The cuff is also deflated automatically to 0 mmHg at the same time. If the monitor receives no further action from the user for 1 minute, it will automatically turn off by itself.
#### Summary of Substantial Equivalence 8.
### Modifications made from the predicate devices:
- · Remove the plastic host between the cuff and main BP monitor. Basically, UA-1200BLE is a hoseless BP monitor comparing with the predicate devices. We also changed the plastic molds so the UA-1200BLE digital blood pressure monitors have a new appearance.
- · Enable user ID so up to six users can use the UA-1200BLE.
- · Increase the memory size to 100 per user except in the visitor mode.
- · Change upper limit of the arm size from 17.7 inches (45 cm) to 16.5 inches (42 cm).
- · Change pressure sensor from static electricity capacity type to semiconductor type
| | Predicate Devices | | Modified Device |
|------------------------------------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------|
| Model | UA-767PBT | UA-651 | UA-1200BLE |
| Measurement<br>Method | Oscillometric Method | Oscillometric Method | No change |
| Measurement<br>Range | BP : 20 - 280mmHg<br>Pulse : 40 - 200beats per<br>minute | BP : 0 - 299mmHg<br>Pulse : 40 - 180beats<br>per minute | No change from UA-651 |
| Measurement<br>Accuracy | BP : +/- 3mmHg or +/-<br>2% of measured value,<br>whichever is greater<br>Pulse : +/- 5% (pulse) | BP : +/- 3mmHg<br>Pulse : +/- 5% (pulse) | No change from UA-651 |
| Minimum<br>Display<br>Resolution | 1 mmHg | 1 mmHg | No change |
| Pressure<br>sensor | Static electricity capacity<br>type sensor | Static electricity<br>capacity type sensor | Semiconducting type<br>sensor |
| Pressurizatio<br>n Source | Automatic internal pump | Automatic internal<br>pump | No change |
| Cuff<br>Deflation<br>Method | Automatic constant speed<br>mechanical exhaust valve | Automatic constant<br>speed mechanical<br>exhaust valve | Solenoid valve for rapid<br>exhaust |
| Display Type | Liquid crystal display | Liquid crystal display | OLED display |
| Cuff<br>Attachment<br>Method | By plastic hose connected<br>to monitor | By plastic hose<br>connected to monitor | Integrated (hoseless) |
| IHB | Enabled | Enabled | No change |
| %IHB | No | Enabled. %IHB is<br>displayed as frequency<br>of IHB detected | No change from UA-651 |
| Power Source | 6V DC, 4x1.5V AA<br>batteries or AC adaptor<br>as an option | 6V DC, 4x1.5V AA<br>batteries or AC<br>adaptor as an option | 3.7V Li-ion batteries<br>or AC adaptor as an<br>option |
| Protection<br>against<br>electrical<br>shock | Class I, Type B | Class I, Type B | No change |
| Battery Life | 6 months with one<br>measurement dairy | 6 months with one<br>measurement dairy | 3 months with one<br>measurement daily |
| Operating<br>Environment | 50°F(10°C) to<br>104°F(40°C)<br>at less than 85%RH | 50°F(10°C) to<br>104°F(40°C)<br>15%RH to 85%RH | No change |
| Storage<br>Environment | -4°F(-20°C) to<br>140°F(60°C)<br>at less than 85%RH | -4°F(-20°C) to<br>140°F(60°C)<br>10%RH to 95%RH | No change |
| Data<br>Memory Size | Last 40 measurements of<br>systolic, diastolic, and<br>pulse rate | Last 60 measurements<br>for each of the four<br>user | Last 100 measurements<br>for each of the five user |
| Dimensions | 163.7(H)x111(W)x66.7(<br>L) mm | 140(W)x60(H)x105(D<br>) mm | 40(W)x126(H)x26(D)<br>mm |
| Weight | 320g without batteries | 255g without batteries | 205g without batteries |
| Cuff Design | D-ring cuff with plastic<br>hose | D-ring cuff with<br>plastic hose | D-ring cuff(Integrated)<br>Hoseless |
| Arm Size | 13 to 45 cm | 22 to 45 cm | 22 to 42 cm |
| Clock<br>(Time/Date) | Yes (By App) | Yes | No change from UA-<br>767PBT |
| AM/PM | No | Yes | No change from UA-651 |
| Pressure<br>Indicator | Yes | Yes | No change |
| Wireless<br>Radio<br>Connectivity | Bluetooth wireless | No | No change from UA-<br>767PBT |
| Field service | Not allowed | Not allowed | No change |
| Automatic<br>Zero at<br>"START" | Yes | Yes | No change |
| Manual Zero<br>Adjustment | Not allowed | Not allowed | No change |
| Calibration | Not allowed in the field | Not allowed in the<br>field | No change |
| Sterilization | Not needed | Not needed | No change |
| Blood<br>Pressure<br>Reading<br>Classification<br>Criteria | USA JNC VII | WHO Classification<br>Criteria | No change from UA-<br>767PBT |
### Product Specification Comparison Table:
{5}------------------------------------------------
{6}------------------------------------------------
### Key Features Comparison
| Parameter | Predicate Devices<br>(UA-651 & UA-767PBT) | Modified Device<br>(UA-1200BLE) |
|-------------------------------------------|-------------------------------------------|---------------------------------|
| Field service | Not allowed | No change - the same |
| Automatic zero after<br>"START" operation | Yes | No change - the same |
| Manual zero adjustment | Not allowed | No change - the same |
| Calibration | Not allowed in the field | No change - the same |
| Sterilization | Not needed | No change - the same |
#### 9. Discussion of standards used in the design verification and design validation
A&D Medical follows FDA recognized consensus standards and guidance documents in our medical device development and manufacturing processes. The following standards were used to for the design verification and valuation of UA-1200BLE UltraConnect Digital Blood Pressure Monitor. These standards include three major groups. First group is the general quality system requirements. Second group is the special requirements for CDRH (870.1130). The third group is related to software and app life cycle processes requirements.
- AAMI/ANSI/ISO 14971:2007/(R) 2010 (Corrected 4 October 2007) ● Medical Devices - Applications Of Risk Management To Medical Devices (FDA Recognized Number 5-70)
- IEC 60601-1-11 Edition 1.0 2010-04 Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment [Including: Technical Corrigendum 1 (2011)] (FDA Recognized Number 19-6)
- . IEC 60601-1-2 Edition 3: 2007-03 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (FDA Recognized Number 19-1)
- AAMI/ANSI/IEC 60601-1-2:2007/(R)2012 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests (Edition 3) (FDA Recognized Number 19-2)
{7}------------------------------------------------
- AAMI/ANSI/ISO 81060-2 Second Edition 2013-05-01 ● Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type (FDA Recognized Number 3-117)
- . AAMI/ANSI/IEC 80601-2-30:2009 & A1:2013 Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers (FDA Recognized Number 3-130)
- . AAMI/ANSI/IEC 62304:2006 Medical Device Software - Software Life Cycle Processes (FDA Recognized Number 13-32)
### 10. Substantial Equivalence Conclusion: Non-clinical Test Summary:
- Safety Tests: IEC 60601-11 Edition 1.0 2010-04
- . EMC Tests: IEC 60601-1-2 Edition 3: 2007-03
- Reliability Tests: ASMI/ANSI/IEC 80601-2-30:2009 & A1:2013 ●
- Risk Assessment: ISO 14971:2012 .
- Software Assessment: IEC Medical Life Cycle Process & FDA "Guidance for the Content of Premarket Submissions for Software contained in Medical Devices", November 2005. . Bluetooth Tests: FCC Part 15 Subpart C : 2014
### Clinical Test Summary based on AAMI/ANSI/ISO 81060-2 Second Edition 2013-05-01
| Item | Standard Requirements | Study Results | | Pass/Fail |
|--------------------------|--------------------------------------|-----------------|-----------|-----------|
| Number of Subject | 85 or more | 85 | 100% | Pass |
| Number of<br>Measurement | 3 times or more | 3 times | 100% | Pass |
| Observer | 2 | 2 | 100% | Pass |
| Observer Difference | +/-4 mmHg or less | +/- 4 mmHg less | 100% | Pass |
| Gender | Male 30% or more | 59 | 69.4% | Pass |
| | Female 40% or more | 26 | 30.6% | Pass |
| Age | Older than 12 years : 100% | 85 | 100.0% | Pass |
| Circumference of<br>Arm | 22.0-32.0 cm : 40% or more | 49 | 57.6% | Pass |
| | 32.1 - 42.0 cm : 40% or more | 36 | 42.4% | Pass |
| | 22.0-27.0cm: 20% or more | 28 | 32.9% | Pass |
| | 37.0-42.0 cm: 20% or more | 17 | 20.0% | Pass |
| BP Range | SBP<100mmHg 5% or more | 19 | 7.5% | Pass |
| | SBP>160mmHg 5% or more | 15 | 5.9% | Pass |
| | SBP<140mmHg 20% or more | 51 | 20.0% | Pass |
| | DBP<60mmHg 5% or more | 19 | 8.2% | Pass |
| | DBP>100mmHg 5% or more | 23 | 9.0% | Pass |
| | DBP >85mmHg 20% or more | 87 | 34.1% | Pass |
| Measurement Result | Average : +/-5mmHg or less | SBP | -2.66mmHg | Pass |
| | | DBP | +0.90mmHg | Pass |
| | S.D. : 8mmHg or less<br>(Criteria 1) | SBP | 6.90mmHg | Pass |
| | | DBP | 6.63mmHg | Pass |
{8}------------------------------------------------
| S.D.: 8 mmHg or less<br>(Criteria 2) | SBP | 5.35 mmHg | Pass |
|--------------------------------------|-----|-----------|------|
| | DBP | 6.12 mmHg | Pass |
UA-1200BLE of digital blood pressure monitors Digital blood pressure monitors have the following similarities to the predicate devices, UA-651 and UA-767PBT digital blood pressure monitors, which previously received the 510(k) clearance.
- Same intended use.
- Same oscillometric method to determine the blood pressure & pulse rate
- Same inflation method - automatic internal pump
- . Same fast safety deflation method - solenoid controlled exhaust valve
- Same materials, no new materials used
- Same manufacturing processes at the same facility
- . Same Indications For Use
As a conclusion. UA-1200BLE UltraConnect digital blood pressure monitor as described in its labeling and comparison analysis has not changed as a result of the modifications. The fundamental scientific technology of the modified device has not changed, either. There is no significant difference that affects the safety or effectiveness of the modified device as compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.