K163676 · Silhouette Lift, Inc. · GAM · Jun 16, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163676
Device Name
Silhouette InstaLift
Applicant
Silhouette Lift, Inc.
Product Code
GAM · General, Plastic Surgery
Decision Date
Jun 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Silhouette Instalift device is indicated for use in mid-face suspension surgery to fixate the cheek sub dermis in an elevated position.
Device Story
Silhouette Instalift is a sterile, single-use surgical implant consisting of two stainless steel needles and an absorbable monofilament thread featuring bidirectional sets of 4, 6, or 8 cones. Used in mid-face suspension surgery, the device is implanted by a clinician to mechanically fixate the cheek sub dermis in an elevated position. The device provides a minimally invasive alternative for facial tissue repositioning. Clinical outcomes are assessed via patient-reported FACE-Q measures and 3D imaging to quantify facial lift. The device benefits patients by providing temporary aesthetic improvement in cheek contour and youthful fullness without requiring permanent anchor sutures.
Clinical Evidence
Prospective clinical study (N=20) evaluated treatment without permanent anchor sutures. Primary endpoints included patient-reported FACE-Q satisfaction scores and 3D imaging (Canfield Vectra M3) to measure facial lift at 1, 8, and 12 weeks. Results showed positive improvements in cheek symmetry, smoothness, attractiveness, contour, and youthful fullness. Age appraisal improved by a mean of 2.4 to 2.9 years younger post-treatment. Facial lift measurements showed 80% of subjects maintained ≥ 0.5 mm lift at 12 weeks. No serious adverse events reported.
Technological Characteristics
Materials: L-lactide/Glycolide bioresorbable resin (cones and monofilament), stainless steel (needles). Form factor: Monofilament with bidirectional cones. Sterilization: Sterile, single-use. Connectivity: None. Software: None.
Indications for Use
Indicated for mid-face suspension surgery to temporarily fixate cheek sub dermis in an elevated position.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2017
Silhouette Lift, Inc. % Mr. Brian Young Health Policy Associates, Inc 690 Canton Street Suite 302 Westwood, Massachusetts 02090
Re: K163676
Trade/Device Name: Silhouette Instalift Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: May 18, 2017 Received: May 19, 2017
Dear Mr. Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163676
Device Name Silhouette Instalift
Indications for Use (Describe)
The Silhouette Instalift device is indicated for use in mid-face suspension surgery to temporarily fixate the cheek sub dermis in an elevated position.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Submitted by: | Silhouette Lift, Inc.<br>A Sinclair Pharma Company<br>1 Technology Drive, Suite F211<br>Irvine, CA 92618 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Brian Young<br>Vice-President Regulatory Affairs/Quality Assurance<br>Health Policy Associates, Inc.<br>Phone: (781) 329-2993<br>Fax: (781) 329-2968<br>Email: byoung@healthpolicyassociates.com |
| Date Prepared | June 15, 2017 |
| Trade Name: | SILHOUETTE INSTALIFT |
| Common Name: | Absorbable surgical suture |
| Classification Name: | Absorbable poly(glycolide/l-lactide) surgical suture |
| Regulation Number: | 21 CFR 878.4493 |
| Classification Product Code: | GAM |
| Device Classification: | II |
#### Predicate Device:
The predicate is the original Silhouette Instalift device cleared under 510(k) number K142061 on April 24, 2015.
#### Device Description:
The Silhouette Instalift device consists of two Stainless Steel Surgical needles and an absorbable implant. The absorbable implant is available in three configurations, each consisting of a single monofilament that holds two bidirectional sets of 4, 6 or 8 cones. Both the monofilament and the cones are made from L- lactide/Glycolide bioresorbable resin material.
This device is supplied sterile for single use only.
{4}------------------------------------------------
#### Indications for Use:
The Silhouette Instalift device is indicated for use in mid-face suspension surgery to temporarily fixate the cheek sub dermis in an elevated position.
## Intended Use:
The Silhouette Instalift device is indicated for use in mid-face suspension surgery to fixate the cheek sub dermis in an elevated position.
## Technological Characteristics (comparison to Predicate Device):
The subject Silhouette Instalift device is identical to the original 510(k) cleared predicate except that the instructions for use have been modified to include an additional optional implantation technique. The additional technique is less invasive and provides clinicians greater flexibility in the treatment of their patients.
#### Performance Testing Summary:
Nonclinical Testing:
The design, materials, manufacturing and sterilization processes are the same as the original predicate version of the device. No additional bench testing was required to support substantial equivalence of this labeling change.
Clinical Testing:
A prospective, clinical study was performed to assess treatment with Silhouette Instalift without use of a permanent, monofilament, anchor suture. In this study, 20 subjects were assessed pretreatment, and compared to 1. 8. and 12 weeks following treatment.
Subjects responded positively to subject-reported outcomes questions regarding their satisfaction with treatment. There was an improvement in cheek satisfaction and age appraisal through 12 weeks post-treatment using the validated patient-reported outcome FACE-O measure. Additional details are provided in the Tables below:
{5}------------------------------------------------
| 1 = Very Dissatisfied; 2 = Somewhat Dissatisfied; 3 = Somewhat Satisfied; 4 = Very Satisfied | | | | | |
|----------------------------------------------------------------------------------------------|--------------------|----------------|----------------|-----------------|----------------|
| Question | Mean (SD) | | | | |
| | Enrollment<br>N=20 | Week 8<br>N=20 | Change | Week 12<br>N=20 | Change |
| a. How symmetric (similar)<br>your cheeks look? | 2.45<br>(1.05) | 3.20<br>(0.83) | 0.75<br>(1.29) | 3.30<br>(0.80) | 0.85<br>(1.23) |
| b. How smooth your<br>cheeks look? | 2.40<br>(0.99) | 3.00<br>(0.92) | 0.6<br>(1.54) | 3.00<br>(0.86) | 0.60<br>(1.31) |
| c. How attractive your<br>cheeks look? | 2.15<br>(0.99) | 3.00<br>(0.79) | 0.85<br>(1.46) | 3.20<br>(0.70) | 1.05<br>(1.15) |
| d. The contour (outline) of<br>your cheeks? | 1.90<br>(0.91) | 2.95<br>(0.83) | 1.05<br>(1.23) | 3.15<br>(0.75) | 1.25<br>(1.12) |
| e. The youthful fullness of<br>your cheeks? | 1.95<br>(1.00) | 3.00<br>(0.97) | 1.05<br>(1.39) | 3.15<br>(0.75) | 1.20<br>(1.20) |
## Table 1: FACE-Q Satisfaction with Cheeks (N=20 Subjects)
* Compared to enrollment. Wilcoxon Signed Rank test.
#### Table 2: FACE-Q Age Appraisal-VAS (N=20 Subjects)
| Question | Years<br>Mean (SD) | | | | |
|---------------------------------------------------------------------------------------------|-------------------------|----------------|---------------|-----------------|---------------|
| | Enroll-<br>ment<br>N=20 | Week 8<br>N=20 | Change | Week 12<br>N=20 | Change |
| How many years younger<br>or older do you think you<br>look compared to your<br>actual age? | +1<br>(6.9) | -2.9<br>(4.1) | -3.9<br>(5.3) | -2.4<br>(3.5) | -3.4<br>(5.3) |
* Compared to enrollment. Wilcoxon Signed Rank test.
Canfield 3D Vectra M3 Face and Neck System was used in the post-treatment period to measure clinical results.
{6}------------------------------------------------
| Timepoint | N (%); N=20 | | | | |
|--------------------------------------|-------------|----------|----------|----------|----------|
| | ≥ 0.5 mm | ≥ 1.0 mm | ≥ 1.5 mm | ≥ 2.0 mm | ≥ 2.5 mm |
| Pre-Treatment<br>(Visit 2) | 10 (50%) | 2 (10%) | 0 (0.0%) | 0 (0.0%) | 0 (0.0%) |
| 1 Week Post-Treatment<br>(Visit 4) | 20 (100%) | 16 (80%) | 9 (45%) | 1 (5%) | 0 (0.0%) |
| 8 Weeks Post-Treatment<br>(Visit 6) | 17 (85%) | 10 (50%) | 6 (30%) | 2 (10%) | 0 (0.0%) |
| 12 Weeks Post-Treatment<br>(Visit 7) | 16 (80%) | 7 (35%) | 5 (25%) | 4 (20%) | 3 (15%) |
Table 3: Amount of Facial Lift in Both Cheeks (N=20 Subjects)
The adverse events reported during the study were generally mild, of short duration, and resolved without sequelae. There were no serious adverse events, unanticipated problems, unanticipated adverse device effects, or deaths.
#### Substantial Equivalence:
The subject device utilizes the same design, materials, manufacturing and sterilization processes as the original predicate version of the device. The modified labeling technique was evaluated in a clinical study which demonstrated acceptable performance and safety thereby supporting a finding of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.