RFID Localization System- Localizer Reader, RFID Localization System- Localizer Surgical Probe, RFID Localization System- Localizer Surgical Probe (5 pack), RFID Localization System- Tag Applicator, RFID Localization System -Tag Applicator (10 pack)
K163667 · Health Beacons, Inc. · PBY · Apr 28, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163667
Device Name
RFID Localization System- Localizer Reader, RFID Localization System- Localizer Surgical Probe, RFID Localization System- Localizer Surgical Probe (5 pack), RFID Localization System- Tag Applicator, RFID Localization System -Tag Applicator (10 pack)
Applicant
Health Beacons, Inc.
Product Code
PBY · General, Plastic Surgery
Decision Date
Apr 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.
Device Story
RFID Localization System (RFLS) uses miniature RFID tags as markers for breast lesion localization. System components include a Tag, Tag Applicator, hand-held Reader, and Surgical Probe. The Tag is placed percutaneously under image guidance. During surgery, the surgeon uses the Reader and Surgical Probe to detect the RFID Tag; the system provides audible tones (increasing pitch/volume with proximity), a visual bar indicator, and a distance measurement readout. This guides the surgeon to the target tissue for excision. Used in clinical settings by breast surgeons and radiologists. Benefits include precise intra-operative localization of non-palpable lesions.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included system verification, reader/probe design verification, accuracy and repeatability evaluation, tissue marker migration, usability validation, electrosurgical tool compatibility, EMC/safety testing, biocompatibility, and magnetic resonance testing (ASTM F2119-07).
Indicated for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal in patients requiring breast surgery.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
SAVI Scout Reflector and SAVI Scout System (K161507)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 28, 2017
Health Beacons, Inc. Ms. Nancy Confrey Chief Executive Officer 34 Walden Street, #753 Concord, Massachusetts 01742
Re: K163667
Trade/Device Name: RFID Localization System- Localizer Reader, RFID Localization System- Localizer Surgical Probe, RFID Localization System-Localizer Surgical Probe (5 Pack), RFID Localization System- Tag Applicator, RFIDd Localization System -tag Applicator (10 Pack) Regulation Number: 21 CFR 878.4300
Regulation Name: Implantable clip Regulatory Class: Class II Product Code: PBY Dated: April 21, 2017 Received: April 24, 2017
Dear Ms. Confrey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K163667
Device Name RFID Localization System
Indications for Use (Describe)
The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue.
The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------|-------------------------------------------------------------|
| <div> </div> | <div> </div> |
| <div> </div> | <div> </div> |
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### l. SUBMITTER
Health Beacons, Inc. 34 Walden Street #753 Concord, MA 01742 Phone: (978) 287-4635 Fax: (978) 246-6019 Establishment Registration No.: TBD
Contact Person: Nancy Confrey Chief Executive Officer Phone: (978) 287-4635 nconfrey@healthbeacons.com
Alternate Contact: Diana DeGregorio Lincé Consulting, LLC Regulatory Affairs Consultant (925) 980-8047 ddegregorio@linceconsulting.com
Date Prepared December 23, 2016
#### ll. DEVICE
Trade Name: RFID Localization System (RFLS) Common Name: Temporary Tissue Marker Classification Name: Implantable clip Classification: 21 CFR§ 878.4300 Product Code: PBY Class II Device Class:
{4}------------------------------------------------
#### lll. PREDICATE
Cianna Medical, Inc. SAVI Scout Reflector and SAVI Scout System (K161507)
#### IV. DEVICE DESCRIPTION
The RFID Localization System (RFLS) is a marker-with-detector localization device that employs miniature RFID tags as markers and a hand-held reader that can measure distance to the tag. The RFLS is comprised of a Tag, Tag Applicator, Reader, and Surgical Probe. The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.
#### V. INDICATIONS FOR USE
The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image quidance (such as ultrasound or radiography) or aided by non-imaging quidance (RFLS), the RFID Tag is located and surgically removed with the target tissue.
The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.
#### VI. COMPARISON CHARACTERISTICS THE OF TECHNOLOGICAL WITH PREDICATE DEVICE
The RFLS has similar features as compared to the predicate device as shown in the following table:
| Manufacturer | Cianna Medical, Inc. | Health Beacons, Inc. |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Device Name | SAVI Scout Reflector and SAVI<br>Scout System | RFID Localization System (RFLS) |
| 510(k) Number | K161507 | TBD |
| Anatomical<br>Locations | Breast Tissue | Same |
| Primary Device<br>Components | Electromagnetic wave reflective<br>device (Reflector) pre-loaded in<br>Delivery System, Handpiece, and<br>Console | RFID Tag pre-loaded in Tag<br>Applicator, Reader, and Surgical<br>Probe |
| Breast Tissue<br>Penetration /<br>Delivery Device | Needle Implanter | Same |
| Visibility | X-ray, Ultrasound | Same |
| Principles of<br>Operation | electromagnetic wave technology to<br>detect the tissue marker | radiofrequency wave technology to<br>detect tissue marker |
| Localization<br>Accuracy | Unknown | ±7 mm |
{5}------------------------------------------------
| Manufacturer | Cianna Medical, Inc. | Health Beacons, Inc. |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SAVI Scout Reflector and SAVI<br>Scout System | RFID Localization System (RFLS) |
| 510(k) Number | K161507 | TBD |
| Indicators | Visual and Audible | Same |
| Implant<br>Delivery Device<br>Size/Working<br>Length | 16 Ga/130mm | 12 Ga/100mm |
| Implant<br>Delivery Device<br>Needle Marking | 1 cm marker increments | Same |
| Implant<br>Dimensions | Approximately<br>0.05" (1.27 mm) x 0.49" (12.4 mm)<br>long | Approximately<br>0.09" (2.3 mm) diameter x 0.37"<br>(10.6 mm) long |
| Patient Contact<br>Materials | Reflector: 3922 Loctite, Nitinol<br>•<br>SE508, light oxide<br>Delivery System: 304 Stainless<br>•<br>Steel<br>Handpiece: Acrylonitrile butadiene<br>•<br>styrene, Polyester film, Loctite<br>4011, NPO (COG) K20 (Dielectric<br>material)<br>• Console: N/A (no patient contact) | Tag: Soda lime bioglass,<br>polypropylene homopolymer<br>• Tag Applicator: 304 Stainless<br>Steel<br>• Surgical Probe: Acrylonitrile<br>butadiene styrene, 304 Stainless<br>•<br>Steel, TPU, Silicone Elastomer<br>Reader: Polycarbonate, Polyester<br>•<br>film (label) |
| Sterilization<br>Method | Single-use, sterilized components:<br>• Implanter and reflector (Ethylene<br>Oxide)<br>• Handpiece (Ethylene Oxide) | Single-use, sterilized components:<br>• Tag Applicator and RFID Tag<br>(Ethylene Oxide)<br>• Surgical Probe (Gamma) |
| Tissue Marker<br>Locator | Reusable, non-sterile Console | Reusable, non-sterile Reader |
| Conformance<br>to Standards | ISO 10993<br>IEC 60601-1<br>IEC 60601-1-2<br>ASTM F2119-07 | ISO 10993<br>IEC 60601-1 (3rd edition)<br>IEC 60601-1-2 (4th edition)<br>FCC 47 CFR Part 15 subpart C<br>ASTM F2119-07 (2013) |
The technological characteristics and principals of operation of the RFLS is substantially equivalent to the named predicate device. The primary difference between the two systems is the Savi Scout device uses electromagnetic wave technology to detect the tissue marker, while the subject device tissue marker is located using RFID technology (radiofrequency wave). However, this minor difference in technology does not raise new or different questions of safety and effectiveness. The performance testing included in the 510k demonstrates the safety of the system for its intended use and confirms the performance of the system for use in intra-operative surgical guidance. Specifically, the performance testing demonstrates the safe use of the system to facilitate intra-operative guided surgical procedures by detecting the presence of the Tag by providing an audible tone which
{6}------------------------------------------------
increases pitch and volume with decreased distance, a visible bar indicator, and a measurement readout.
#### VII. PERFORMANCE DATA
Performance testing was conducted to evaluate and characterize the performance of the RFLS. The following performance testing was conducted on the RFLS to support a determination of substantial equivalence to the predicate device.
- RFLS System Verification ●
- Reader/Probe Design Verification
- Localizer System and Component Accuracy, Repeatability Evaluation ●
- . Test Media Evaluation
- Reader and Surgical Probe Operating Environment Evaluation ●
- Tissue Marker Migration Evaluation
- Usability Verification and Validation ●
- Electrosurgical tool compatibility
- EMC/Safety Testing
- Biocompatibility ●
- Magnetic Resonance Testing ●
- Packaging Validation ●
- Sterilization Validation
#### VIII. CONCLUSIONS
The RFLS has been carefully compared to the legally marketed predicate device with respect to intended use/indications for use, technological characteristics, anatomical sites, performance, safety characteristics, and labeling. In addition, non-clinical testing was conducted to verify and validate the performance of the device and ensure the RFLS functions as intended and meets design specifications. The comparison and non-clinical performance testing results demonstrate that the device is substantially equivalent to the predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.