neuroinspire is a stereotactic software application intended to allow trained clinicians to plan surgical trajectories for neurological procedures. neuroinspire defines a stereotactic region for this planning based on fiducial markers in CT images. neuroinspire reports specific surgical trajectories intended to be manually set on head frame systems validated for use with neuroinspire. neuroinspire has been validated with the Leksell Stereotactic System Type G (comprising of head ring, head frame and CT Localizer). No other components of the Leksell Stereotactic System should be used. No other stereotactic system should be used. The user should consult Section 1.7 Accuracy of these Instructions for Use to assess if the accuracy of neuroinspire is suitable for their needs.
Device Story
neuroinspire is a PC-based surgical planning software for image-guided neurosurgery. It ingests patient MRI and CT DICOM images; allows neurosurgeons to reconstruct, slice, zoom, and rotate 3D brain visualizations; and performs manual segmentation of anatomical structures. The software defines a stereotactic region using fiducial markers in CT images to calculate target coordinates and plan surgical trajectories. Outputs include specific trajectory coordinates intended for manual entry into the Leksell Stereotactic System Type G. Used in clinical settings by neurosurgeons to facilitate pre-surgical planning; the software does not modify source hospital data. By providing precise trajectory planning, it assists surgeons in navigating neurological procedures, potentially improving surgical accuracy and patient outcomes.
Clinical Evidence
No clinical data. Bench testing only. Accuracy validated using a CT phantom mimicking a human skull and a portable CMM system. Positional error: mean 0.96 mm (99% CI upper bound 1.090 mm). Angular error: mean 0.45° (99% CI upper bound 0.539°).
Technological Characteristics
PC-based software application. Supports DICOM MRI and CT image integration. Features include manual segmentation, multi-planar reconstruction, and trajectory calculation. Validated for use with Leksell Stereotactic System Type G. Software developed per IEC 62304. Major level of concern.
Indications for Use
Indicated for trained clinicians to plan surgical trajectories for neurological procedures requiring stereotactic image-guided neurosurgery.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2017
Renishaw plc % Roberto Liddi Head of Regulatory Affairs and Quality Assurance New Mills Wotton-under-edge, Gloucestershire, GL12 8JR GB
Re: K163666 Trade/Device Name: neuro | inspire Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: August 21, 2017 Received: September 11, 2017
Dear Dr. Liddi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163666
Device Name neuro | inspire
#### Indications for Use (Describe)
neuroinspire is a stereotactic software application intended to allow trained clinicians to plan surgical trajectories for neurological procedures. neuroinspire a stereotactic region for this planning based on fiducial markers in CT images. neuroinspire reports specific surgical trajectories intended to be manually set on head frame systems validated for use with neuroinspire has been validated with the Leksell Stereotactic System Type G (comprising of head ring, head frame and CT Localizer). No other components of the Leksell Stereotactic System should be used. No other stereotactic system should be used. The user should consult Section 1.7 Accuracy of these Instructions for Use to assess if the accuracy of neuroinspire is suitable for their needs.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is in orange, and below it, the words "apply innovation" are in black. The Renishaw logo is a registered trademark.
#### 510(k) Summary 4:
# 1. Submitter:
2.
3.
Renishaw plc.
| New Mills | | | | | |
|---------------------------------------|------------------------------------------------|----------------------------|------------------------|------------------------|---------------------|
| Wotton-under-Edge | | | | | |
| Gloucestershire | | | | | |
| GL12 8JR | | | | | |
| United Kingdom | | | | | |
| | Phone: | 0044 1453 524600 | | | |
| Contact person: | Roberto Liddi | | | | |
| | Head of Regulatory Affairs & Quality Assurance | | | | |
| | Renishaw plc. | | | | |
| | New Mills | | | | |
| | Wotton-under-Edge | | | | |
| | Gloucestershire | | | | |
| | GL12 8JR | | | | |
| | United Kingdom | | | | |
| | Phone: | 0044 1453 524600 | | | |
| | Email: | roberto.liddi@renishaw.com | | | |
| Preparation date: 14th September 2017 | | | | | |
| Device: | | | | | |
| Trade name: | neuro inspireTM | | | Common name: | <i>neuroinspire</i> |
| Classification regulation: | 21 CFR 882.4560 | | | Stereotaxic instrument | |
| Product code: | HAW | | | | |
| Predicate device: | | | | | |
| Brainlab AG | iPlan Stereotaxy | | | K101627 | |
| Device Classification Name: | | | Stereotaxic Instrument | | |
| Regulatory Class: | Class II | | | | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is in large, orange, sans-serif font. Below the company name is the slogan "apply innovation" in a smaller, black, sans-serif font. To the right of the company name is a small, orange square with a white line through it.
#### 4. Device description:
neuroinspire surgical planning software is an application that enables the viewing of medical images to support the image based neurosurgery. It may be used by neurosurgeons to reconstruct and display multiple image sets to facilitate 2D and 3D targeting and trajectory planning in relation to an intended neurological surgical procedure. It allows the neurosurgeon to download and integrate MRI and CT DICOM (digital imaging communication) images of a patient's brain onto a PC-based platform. The neurosurgeon can then virtually slice the brain into sections, zoom in for close-ups and rotate images in order to visualize and analyze the patient's brain prior to surgery.
neuroinspire has been validated with the Leksell Stereotactic System Type G (comprising of head ring, head frame and CT Localizer).
neuroinspire surgical planning software allows a neurosurgeon to:
- . load data acquired from an MRI scanner
- . load data acquired from a CT scanner;
- manually reconstruct target anatomy using multi-planer views; .
- calculate target coordinates;
- plan trajectories; ●
- print a surgical plan.
#### 5. Indications for Use:
neuroinspire is a stereotactic software application intended to allow trained clinicians to plan surgical trajectories for neurological procedures. neuroinspire defines a stereotactic region for this planning based on fiducial markers in CT images. neuroinspire reports specific surgical trajectories intended to be manually set on head frame systems validated for use with neuroinspire. neuroinspire has been validated with the Leksell Stereotactic System Type G (comprising of head frame and CT Localizer). No other components of the Leksell Stereotactic System should be used. No other stereotactic system should be used. The user should consult Section 1.7 Accuracy of these Instructions for Use to assess if the accuracy of neuroinspire is suitable for their needs.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is in orange, and below it, the words "apply innovation" are in black. The Renishaw logo is a registered trademark.
| Function | Predicate Description of<br>BrainLab iPlan Stereotaxy<br>K101627 | Comparison of Predicate to neuroinspire | Equivalence |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Intended Use | The viewing, presentation and documentation of medical imaging, including different modules for image processing, image fusion, atlas assisted visualization and segmentation, intraoperative functional planning where the output can be used e.g. with stereotactic image guided surgery or other devices for further processing and visualization | Software application running on a standard computer workstation for viewing and presentation of medical images. It provides the capability for processing of patient images and the planning of surgical procedures for any medical condition in which the use of stereotactic image guide brain based neurosurgery is considered appropriate | Same |
| AC/PC<br>Localization | This planning task allows the definition of AC/PC co-ordinate system | This planning task allows the definition of AC/PC co-ordinate system | Same |
| Localization | Assign a localizer frame for CT or MRI localization Perform automatic detection of localizer rods | Assign a localizer frame for CT localization Perform automatic detection of localizer rods | Substantially<br>Equivalent |
| Trajectory<br>Planning | Plan pathways for surgical instruments or resection, definition of entry, target points and diameter for trajectories | Plan pathways for surgical instruments or resection, definition of entry, target points and diameter for trajectories | Same |
| Stereotactic<br>planning | Planning of stereotactic trajectories Usage of AC/PC coordinates and Schaltenbrandt-Wahren atlas. Calculation of stereotactic arc settings for planned trajectories | Planning of stereotactic trajectories Usage of AC/PC coordinates and Shaltenbrandt-Wahren atlas. Calculation of stereotactic arc settings for planned trajectories | Same |
| Function | Predicate Description<br>of<br>BrainLab iPlan<br>Stereotaxy<br>K101627 | Comparison of<br>Predicate to<br>neuroinspire | Equivalence |
| Load and<br>Import | Load existing treatment<br>data from different data<br>sources, import patient<br>data from DICOM or<br>other archive types,<br>manage<br>(delete/copy/move)<br>patient folders | neuroinspire imports<br>patient data in DICOM<br>form. It manages<br>patient data locally (on<br>the PC) but does not<br>modify or delete the<br>source hospital data.<br>This is a safety<br>measure as the<br>original data cannot be<br>altered through the<br>use of neuroinspire. | Substantially<br>Equivalent |
| View and<br>Adjustment | Review patient data in<br>various reconstructions<br>or overlay, side-by-side<br>comparison of different<br>modalities, aligning the<br>data set orientation,<br>import or export<br>screenshot images | neuroinspire allows the<br>review of patient data<br>in various<br>reconstructions or<br>overlay, side-by-side<br>comparison of MRI<br>and CT modalities,<br>alignment of datasets<br>and export of<br>screenshots, but there<br>is no import of<br>screenshot data. | Substantially<br>Equivalent |
| Image<br>Fusion | Align available image<br>sets automatically,<br>manually or using<br>landmarks for various<br>combinations of images<br>sets and modalities such<br>as CT MRI, PET and<br>SPECT. Visual<br>verification of alignment. | neuroinspire aligns<br>available image sets<br>automatically,<br>manually or using<br>landmarks for various<br>combinations of<br>images sets and<br>modalities but only<br>MRI and CT modalities<br>are supported. MRI<br>and CT are the most<br>common imaging<br>modalities used in<br>neurosurgery. | Substantially<br>Equivalent |
| Function | Predicate Description<br>of<br>BrainLab iPlan<br>Stereotaxy | Comparison of<br>Predicate to<br>neuroinspire | Equivalence |
| | K101627 | | |
| Object<br>Creation | Outline anatomical<br>structures using manual<br>or automatic<br>segmentation methods.<br>Advanced manipulation<br>for 3D objects with<br>scaling, logical<br>operations and object<br>splitting. Volumetric<br>measurements based<br>on the created 3D<br>objects | With neuroinspire the<br>user can manually<br>segment anatomical<br>structures using<br>manual segmentation<br>methods. There are no<br>automatic<br>segmentation<br>functions but the<br>volume of a<br>segmented features is<br>reported. | Substantially<br>Equivalent |
| Save and<br>Export | Save the current<br>treatment plan to the<br>patient folder. Export the<br>results to the navigation,<br>as DICOM or STL<br>format | The neuroinspire user<br>can save and export<br>the treatment plans in<br>a proprietary format,<br>not in DICOM or STL<br>and does not export<br>the results to a<br>navigation system. | Substantially<br>Equivalent |
| Registration<br>Points | Automatic detection of<br>CT or MR registration<br>markers for navigation,<br>manual placement of<br>markers and anatomical<br>landmarks | No equivalence is<br>claimed | No equivalence is<br>claimed |
| Advanced<br>Object<br>Planning | Mirror and split<br>segmented structures | No equivalence is<br>claimed | No equivalence is<br>claimed |
| BOLD MRI<br>mapping | Processing of blood<br>oxygen level dependent<br>(BOLD) MRI data.<br>Definition of block<br>design functional task,<br>calculation of activation<br>areas based on BOLD<br>MRI data, time series<br>view for activation<br>signal, creation of 3D<br>objects from activation | No equivalence is<br>claimed | No equivalence is<br>claimed |
| Function | Predicate Description<br>of<br>BrainLab iPlan<br>Stereotaxy | Comparison of<br>Predicate to<br>neuroinspire | Equivalence |
| | K101627 | | |
| Fiber<br>Tracking | Processing of diffusion<br>tensor imaging (DTI)<br>using various ways to<br>define and combine<br>seed regions of interest.<br>Definition of multiple<br>fiber bundles and<br>creation of 3D objects<br>from fiber bundles.<br>Volumetric<br>measurements and<br>detailed fiber information | No equivalence is<br>claimed | No equivalence is<br>claimed |
| Electrode<br>recording | Planning of parallel<br>electrode tracks. Enter<br>and display<br>microelectrode<br>recording and<br>stimulation results,<br>display information<br>stepwise along tracks | No equivalence is<br>claimed | No equivalence is<br>claimed |
### 6. Comparison to Technological Characteristics with Predicate Device:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is written in orange, with the tagline "apply innovation" written in black below it. To the right of the word "RENISHAW" is a symbol that is partially obscured.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is in orange, and below it, the words "apply innovation" are in black. The Renishaw logo is a registered trademark.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is in large, orange, sans-serif font. Below the word "RENISHAW" is the phrase "apply innovation" in a smaller, black, sans-serif font. To the right of the word "RENISHAW" is a small, stylized image that is also orange.
#### a. Conclusion
Except for the differences stated above, the proposed neuroinspire software is similar in design, intended use and operation to the predicate, K101627. The last five functions in the above table are specific areas of functionality that extend the scope of BrainLab's iPlan Stereotaxy device beyond the scope of neuroinspire.
#### 7. Performance Data:
#### a. Nonclinical Data
Verification and validation of this device was conducted and documentary evidence has been provided as part of this submission. The software is considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
Software was developed in compliance with FDA software guidance and IEC 62304.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the Renishaw logo. The word "RENISHAW" is in orange, and below it is the phrase "apply innovation" in black. To the right of the word "RENISHAW" is a small, light-colored graphic.
Human factors studies have been carried out to validate the safe and effective user experience, the documentary evidence of these studies has been provided as part of this submission.
A ground truth measurement of a CT phantom mimicking a human skull and containing a number of representative brain targets was established using a portable CMM system. Testing was performed and demonstrated that the positional and angular accuracy of transferring the trajectory information from neuroinspire to the Leksell stereotactic system is <2 mm and <2° respectively. The documentary evidence of this measurement has been provided as part of this submission, results are summarized below:
Specification for maximum Euclidean positional error: <2 mm
Mean: 0.96 mm ST Dev: 0.42 mm 99% Cl: 0.130 mm 99% CI upper bound: 1.090mm
Specification for maximum angular error: <2°
Mean: 0.45°
ST Dev: 0.29°
99% Cl: 0.089°
99% CI upper bound: 0.539°
#### b. Clinical Data
No clinical data has been submitted.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the Renishaw logo. The logo consists of the word "RENISHAW" in orange, with the tagline "apply innovation" in black underneath. To the right of the word "RENISHAW" is a small graphic element, also in orange.
## 8. Conclusions:
The conclusions drawn from the nonclinical tests demonstrate that the device, when used as intended, is as safe, as effective, and performs as well as or better, than the legally marketed device identified in section 3 of this summary.
4-8 Device name: neuro | inspire™ Traditional 510(k) submission Submitter: Renishaw plc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.