The P-140-Flex Coil, Na-140-Flex Coil and C-140-Flex Coil are indicated for use as diagnostic device extensions for Philips 3.0 T MR Systems to produce cross-sectional spectroscopy images and/or spectra in any orientation of the internal structure of the whole body, e.g. of the calf, liver or heart. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
Device Story
Single-tuned surface transmit/receive (T/R) coils for Philips 3.0 T MR systems; input: RF signals from 31P, 23Na, or 13C spins; output: MR signals for cross-sectional spectroscopy imaging/spectra. Used in clinical settings by trained physicians to visualize internal anatomy/function. Flexible housing improves anatomical fit and patient comfort compared to rigid predicate. Fixed-tuned and matched to human tissue load; no detuning/switching mechanism in loops; decoupled from 1H body coil. Physician interprets images to assist in diagnosis.
Clinical Evidence
Bench testing only. Conformance to IEC 60601-1 and IEC 60601-2-33 standards demonstrated safety and effectiveness.
Technological Characteristics
Single-loop (linear) surface coil; 14 cm diameter; flexible housing. Fixed-tuned/matched to 31P (51.7 MHz), 23Na (33.8 MHz), or 13C (32.1 MHz) at 3.0 T. Includes T/R switch, low noise preamplifier, and connector for Philips 3.0 T MR systems. Decoupled from 1H body coil. Standards: IEC 60601-1:2005, ANSI/AAMI ES60601-1:2005/A2:2010, CAN/CSA-C22.2 No. 60601-1:08, IEC 60601-2-33:2010+A1:2013.
Indications for Use
Indicated for use as diagnostic extensions for Philips 3.0 T MR systems to produce cross-sectional spectroscopy images/spectra of internal human body structures (e.g., calf, liver, heart) to assist in diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2018
Rapid Biomedical Gmbh Christian Zimmermann Official Correspondent Kettelerstr. 3-11 Rimpar, 97222 DE
Re: k163661
Trade/Device Name: P-140-flex Coil, Na-140-flex Coil, C-140-flex Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: February 12, 2018 Received: February 12, 2018
Dear Christian Zimmermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara
Robert A. Ochs, Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K163661
Device Name
P-140-Flex Coil, Na-140-Flex Coil, C-140-Flex Coil
Type of Use (Select one or both, as applicable)
Indications for Use (Describe)
The P-140-Flex Coil, Na-140-Flex Coil and C-140-Flex Coil are indicated for use as diagnostic device extensions for Philips 3.0 T MR Systems to produce cross-sectional spectroscopy images and/or spectra in any orientation of the internal structure of the whole body, e.g. of the calf, liver or heart. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
| <span> <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740
EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for Rapid Biomedical. The text "RAPID" is in large, bold, black letters at the top. Below that, "Biomedical" is written in a smaller, bold, black font, underlined with a black line. At the bottom, in a smaller font, the text "MR Coils - Made to Measure" is present.
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
### General Information
Date of summary preparation: 2017-06-30
#### Manufacturer
RAPID Biomedical GmbH
Kettelerstrasse 3-11, 97222 Rimpar, Bavaria, Germany
FDA FEI: 3005049692
#### Designed and Manufactured for
Philips Medical Systems Nederland BV
Veenpluis 4-6, 5684 PC Best, Netherlands
FDA FEI: 3003768277
#### Contact Person
Mr. Christian Zimmermann RAPID Biomedical GmbH Kettelerstrasse 3-11, 97222 Rimpar, Bavaria, Germany Phone: +49 (9365) 8826-44, Fax: +49 (9365) 8826-99 e-mail: christian.zimmermann@rapidbiomed.de
### Classification and Device Names
Classification Panel: Radiology Classification Name: Magnetic Resonance Diagnostic Device Accessory Device Class: Class II [21 CFR § 892.1000] Product Code: MOS Product Nomenclature: Coil, Magnetic Resonance, Specialty Common Name: Special Purpose Coil Trade Names: P-140-Flex Coil, Na-140-Flex Coil or C-140-Flex Coil
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the title of a document, which is a Traditional 510(k) Premarket Notification. The document number is K163661, supplement 001. The document is about P-140-Flex Coil, Na-140-Flex Coil, and C-140-Flex Coil.
## Indications for Use
The P-140-Flex Coil, Na-140-Flex Coil and C-140-Flex Coil are indicated for use as diagnostic device extensions for Philips 3.0 T MR Systems to produce crosssectional spectroscopy images and/or spectra in any orientation of the internal structure of the whole body, e.g. of the calf, liver or heart. These images when interpreted by a trained physician, yield information that may assist in diagnosis.
## Device Description
The P-140-Flex Coil, Na-140-Flex Coil and C-140-Flex Coil are single tuned surface type transmit and receive coils for usage in magnetic resonance systems, the operating frequency and interface being adapted for usage on Philips 3.0 T MR Systems. The coils are for transmit and receive (T/R) and provide the required magnetic RF field for excitation of the spins and receive the MR signal of 31P, 23Na or 13C spins. This is used for axial, sagittal, coronal and oblique imaging, spectroscopic imaging and/or spectroscopy, which display the internal structure and/or function of the human anatomy, e.g. of the calf, liver or heart.The coils are optimized for local examinations of human subjects. The housing is flexible for better adaption to the anatomical region of interest. The coils consist of a single loop (linear) with a diameter of 14 cm, RF routing (T/R switch), a low noise preamplifier and a connector to the Philips 3.0 T MR Systems. The coils are fixed tuned and matched to the typical load of human tissue at the Larmor frequency of 31P (51.7 MHz), 23Na (33.8 MHz) or 13C (32.1 MHz) at 3.0 T. There is no detuning or switching mechanism in the loops. The loops are tuned permanently to the working frequencies of 31P, 23Na or 13C while being decoupled from the 1H body coil at the 1H frequency.
# Equivalency Information
RAPID Biomedical believes that the subject devices are substantially equivalent to the Philips Phosphorous Spectroscopy Surface Coil P-140 which is an extension of the cleared Philips Magnetic Resonance Diagnostic Devices (MRDD) which are described in the submission:
| Predicate Device Name | Clearance Number | Date |
|-----------------------|------------------|------------|
| ACHIEVA family | K043147 | 2005-02-15 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Rapid Biomedical logo on the left side. On the right side, the text reads, "Traditional 510(k) Premarket Notification K163661, supplement 001 P-140-Flex Coil, Na-140-Flex Coil, C-140-Flex Coil". The Rapid Biomedical logo has the word "RAPID" in large blue letters above the word "Biomedical" in smaller blue letters. Below that, the text "MR Coils - Made to Measure" is written in small letters.
### Summary of Characteristics of the Subject Device as compared with the Predicate Device
The subject device (P-140-Flex Coil, Na-140-Flex Coil and C-140-Flex Coil) and the predicate device (Philips P-140 Coil) are made for the usage on Philips 3.0T MR Systems as transmit/receive coils. The Philips 3.0 T MR Systems allow 31P, 23Na and 13C spectroscopy imaging and spectroscopy, including simultaneous 1H MR examinations and are capable of driving the subject device as described in the indications for use. Compared to the predicate device, the subject device yields flexible housing parts for better fitting the anatomy of the patient and offering better patient comfort. This gain in performance does not yield detrimental effects on the safety, reliability or other performance topics as compared to the predicate device. The active coil diameter of 14 cm of the subject device is exactly the same as the diameter of the predicate device. Subject device and predicate device do not offer a 1H channel for direct 1H MR examinations. These 1H examinations are provided by using the standard 1H body coil of the Philips 3.0 T MR System. The subject device is fixed tuned and matched while the predicate device has to be tuned and matched manually per patient.
Overall, the subject device and the predicate device share the substantial coil technology. Improvements in performance rectify the modifications in design while the intended use, the indication for use and safety stay substantially equivalent.
### General Safety and Effectiveness
The safety and performance parameters specified by the FDA Guidance document for MR Diagnostic Devices are unaffected by the modifications described within this submission.
The subject devices are conforming to:
- · IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 (or IEC 60601-1: 2012 reprint)
- o with National Differences for United States (US), ANSI/AAMI ES60601-1: 2005 / A2:2010 and
- o Canada (CA), CAN/CSA-C22.2 No. 60601-1:08
- IEC 60601-2-33: 2010 (Third Edition) + A1: 2013 for use in conjunction ● with IEC 60601-1:2005 + A1: 2012.
This will assure that the performance is to be considered safe and effective when used with the Philips 3.0 T MR Systems.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Rapid Biomedical. The logo has the word "RAPID" in large, bold, black letters on the top line. Below that, the word "Biomedical" is in a smaller, bold, black font. Underneath that is the phrase "MR Coils - Made to Measure" in a smaller, regular font.
## Substantial Equivalence
Testing was performed to support this claim of substantial equivalence and to show that the modifications do not raise any new questions pertaining to safety and effectiveness.
The modifications did not affect the Indications for Use and did not alter the Fundamental Scientific Technology.
RAPID Biomedical therefore believes the subject devices and the predicate device to be substantially equivalent (SE).
Rimpar, 2017-06-30
Signature:
Name:
Christian Zimmermann (Quality Management & Official Correspondent)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.