K163657 · Spectrum Medical , Ltd. · DRY · Jan 18, 2017 · Cardiovascular
Device Facts
Record ID
K163657
Device Name
Quantum Workstation
Applicant
Spectrum Medical , Ltd.
Product Code
DRY · Cardiovascular
Decision Date
Jan 18, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4330
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The intended use of the Quantum Workstation is for the non-invasive continuous monitoring of oxygen saturation and hematocrit / hemoglobin concentration of the blood in an extracorporeal circuit. When using its range of accessories, the Quantum Workstation is configured to measure and display the following measurements: SaO2 Arterial Saturation (%) Venous Saturation (%) Sv02 Hb Hemoglobin (g/L and gm/dl) Calculated Hematocrit (%) Hct The Workstation provides monitoring information to trained clinicians and can be configured by them to set parameter specific alarms. The Workstation's monitoring and alarm functionality does not directly control patient care. The User makes clinical judgments regarding the treatment of the patient as a result of information displayed by the Workstation.
Device Story
Quantum Workstation is an online cardiopulmonary bypass blood gas monitor; performs non-invasive continuous monitoring of arterial/venous oxygen saturation (SO2), hemoglobin (Hb), and calculated hematocrit (Hct) in extracorporeal circuits. Input: optical signals from Hb/SO2 sensors attached to external surface of blood tubing; principle of operation: comparison of visible light absorbance patterns for oxy/deoxy-hemoglobin and infrared light transmission for Hb concentration. Output: real-time individual/trend readings and parameter-specific alarms on a 14.75" portrait touch screen. Used in hospital surgical environments by perfusionists. Device features Wi-Fi, Ethernet, USB ports, and SD card storage; powered by AC mains with lithium-ion battery backup. Clinicians use displayed data to make treatment decisions; device does not control patient care. Benefits include continuous monitoring of blood parameters during extracorporeal circulation.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by bench performance testing, including electrical safety (AAMI ANSI ES60601-1), EMC (IEC 60601-1-2), mechanical testing, and in vitro performance testing confirming SO2, Hb, and Hct measurement accuracy is unchanged from the predicate M2 Monitor.
Technological Characteristics
Pole-mounted monitor with 14.75" portrait touch screen. Sensing principle: optical (visible/infrared light). Connectivity: Wi-Fi, Ethernet, USB, and Spectrum Accessory Ports (SAPs). Power: AC mains with dual lithium-ion battery backup (2-hour life). Standards: AAMI ANSI ES60601-1:2005/(R)2012, IEC 60601-1-2:2014, IEC 60601-2-49:2011. Software class: Class B per IEC 62304:2006.
Indications for Use
Indicated for patients of all ages and genders undergoing extracorporeal circulation requiring non-invasive continuous monitoring of blood oxygen saturation, hematocrit, and hemoglobin levels.
Regulatory Classification
Identification
A cardiopulmonary bypass on-line blood gas monitor is a device used in conjunction with a blood gas sensor to measure the level of gases in the blood.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 18, 2017
Spectrum Medical, Ltd. Mark Drain Chief Financial Officer Harrier 4, Meteor Business Park Cheltenham Road East Gloucester, GL2 9QL England
Re: K163657
Trade/Device Name: Quantum Workstation Regulation Number: 21 CFR 870.4330 Regulation Name: Cardiopulmonary bypass on-line blood gas monitor Regulatory Class: Class II Product Code: DRY Dated: December 16, 2016 Received: December 23, 2016
Dear Mr. Drain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Willemsen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K163657
Device Name Quantum Workstation
#### Indications for Use (Describe)
The intended use of the Quantum Workstation is for the non-invasive continuous monitoring of oxygen saturation and hematocrit / hemoglobin concentration of the blood in an extracorporeal circuit. When using its range of accessories, the Quantum Workstation is configured to measure and display the following measurements:
- SaO2 Arterial Saturation (%)
- Venous Saturation (%) Sv02
- Hb Hemoglobin (g/L and gm/dl)
- Calculated Hematocrit (%) Hct
The Workstation provides monitoring information to trained clinicians and can be configured by them to set parameter specific alarms.
The Workstation's monitoring and alarm functionality does not directly control patient care. The User makes clinical judgments regarding the treatment of the patient as a result of information displayed by the Workstation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21 CFR 807.92.
510(k) Number: K163657
### I. SUBMITTER
Spectrum Medical Ltd Harrier 4, Meteor Business Park, Cheltenham Road East, Gloucester. GL2 9QL England
Phone: +44 (0) 1242 650 120 Fax: +44 (0) 8452 808 127
Contact Person: Mr. Mark Drain, Chief Financial Officer
Date Summary Prepared: January 17, 2017
# II. DEVICE
| Proprietary Name: | Quantum Workstation |
|----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Blood Gas Monitor |
| Classification Name: | Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass<br>(21 CFR 870.4330) |
| Regulatory Class: | II |
| Product Code: | DRY |
| Panel: | Office of Device Evaluation (ODE) /<br>Division of Cardiovascular Devices (DCD)<br>Circulatory Support Devices Branch (CSDB) |
# III. PREDICATE DEVICE
Spectrum Medical Ltd's M2 Monitor (K071725)
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
{4}------------------------------------------------
# IV. DEVICE DESCRIPTION
The Quantum Workstation is an online, cardiopulmonary bypass, blood gas monitor. It is used for extracorporeal monitoring of blood oxygen (arterial and venous) saturation, hematocrit, and hemoglobin levels.
The Quantum Workstation consists of a pole-mounted 14 ¾" portrait high definition touch screen. The touch screen displays individual and trend readings with alarm settings. Third party data may also be displayed (without alarms). The Quantum Workstation has a Wi-Fi adapter and provides memory storage via an SD (Secure Digital) card. The Quantum Workstation is powered from the AC Mains supply and also incorporates a battery backup that automatically switches on in the event of an interruption to the mains power supply. The battery backup is provided via two (2) lithium-ion batteries with a two-hour minimum life.
The Quantum Workstation includes the following ports / connections:
- . One (1) sensor port for the Hb / SO2 sensor
- One (1) LAN / Ethernet port
- . Three (3) USB 2.0 ports
- Eight (8) additional LAN ports described as SAP (Spectrum Accessory Ports) to . support a range of Spectrum Medical manufactured modules - these are for future use
Accessories for the Quantum Workstation include the power supply, mounting arm (long or short), and Hb / SO2 sensor. Different Hb / SO2 sensors are available based on the diameter and thickness of the extracorporeal tubing.
# V. INTENDED USE / INDICATIONS FOR USE
The intended use of the Quantum Workstation is for the non-invasive continuous monitoring of oxygen saturation and hematocrit / hemoglobin concentration of the blood in an extracorporeal circuit. When using its range of accessories, the Quantum Workstation is configured to measure and display the following measurements:
| SaO2 | Arterial Saturation (%) |
|------|----------------------------|
| SvO2 | Venous Saturation (%) |
| Hb | Hemoglobin (g/L and gm/dl) |
| Hct | Calculated Hematocrit (%) |
The Workstation provides monitoring information to trained clinicians and can be configured by them to set parameter specific alarms.
The Workstation's monitoring and alarm functionality does not directly control patient care. The User makes clinical judgments regarding the treatment of the patient as a result of information displayed by the Workstation.
{5}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
As shown in Table A, Spectrum Medical Ltd's M2 Monitor and Quantum Workstation have the same manufacturer, intended use, clinical application, clinical setting / site of user, target patient population, and principle of operation / mechanism of operation / fundamental scientific technology. They also have the same performance for SO2, Hb, and Hct. They both display individual and trend readings with alarm settings and don't require recalibration due to negligible drift. They both have a Wi-Fi adapter, Ethernet port, and SD card memory storage. They both run on line power with battery backup available. They both are also compatible with the intended environment and with other devices.
Spectrum Medical Ltd's M2 Monitor and Quantum Workstation differ in that the Quantum Workstation has a larger touch screen display, has additional USB and LAN ports, allows for display of 3%-party data without alarms, and provides additional graphs with a pinch action feature for graph resizing. The M2 Monitor uses three nickel hydride batteries to provide an ~1hour battery life whereas the Quantum Workstation uses two lithium-ion batteries to provide a 2hour minimum battery life.
{6}------------------------------------------------
| Device | Predicate Device | Proposed Device |
|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | M2 Monitor | Quantum Workstation |
| 510(k) Number | K071725 | To be assigned |
| Manufacturer | Spectrum Medical Ltd<br>Gloucester, England | Same |
| Intended Use | Extracorporeal monitoring of<br>blood oxygen saturation,<br>haematocrit and haemoglobin<br>levels | Same |
| Clinical Application | Online monitoring of<br>extracorporeal arterial and<br>venous blood lines, and<br>hematocrit and hemoglobin levels<br>using sensor probes attached to<br>the external surface of blood<br>tubing | Same |
| Clinical Setting /<br>Sites of Use | Hospital | Same |
| Target user | Perfusionist | Same |
| Target patient<br>population | Patients (all ages, both genders)<br>undergoing extracorporeal<br>circulation | Same |
| Principal of Operation<br>/ Mechanism of<br>Action | Comparison of the different<br>patterns of absorbance of visible<br>light reflected from oxy and deoxy<br>forms of hemoglobin | Same |
| Blood oxygen<br>saturation<br>(SO2 %) | Single synchronization with<br>reference blood gas analyzer;<br>independent of blood flow, blood<br>temperature, and hemodilution | Device uses equivalent Hb /<br>SO2 sensor to M2 Monitor (has<br>improved electrical isolation and<br>different colored cable boot) |
| Principal of Operation<br>/ Mechanism of<br>Action<br>Hemoglobin (Hb) /<br>Hematocrit (Hct) | Measures hemoglobin<br>concentration by transmitting<br>infrared light through the blood<br>tube and quantifying the level of<br>signal attention with a photodiode<br>Calculates hematocrit from<br>hemoglobin measurement | Device uses equivalent Hb /<br>SO2 sensor to M2 Monitor (has<br>improved electrical isolation and<br>different colored cable boot) |
| Device | Predicate Device | Proposed Device |
| Performance | SO2<br>Range = 20-100%<br>Range Temperature = 15-37°C<br>Mean Offset = 0.48<br>Standard Deviation = ±1.90 | Same |
| | Hb / Hct<br>Range:<br>5-15 g/dL / 15-45%<br>(for 9/16" OD tube size)<br>5-16.6 g/dL / 15-50%<br>(for all other sensors: 5/16",<br>3/8", 7/16" OD tube size)<br>Range Temperature = 15-37°C<br>Mean Offset = 0.03<br>Standard Deviation = ±0.60 | |
| Human Factors | Touch screen display showing<br>individual and trend readings with<br>alarm settings<br>Flash memory storage for<br>recording case history<br>No recalibration required due to<br>negligible drift | Same general human factors<br>features<br>Also allows for display of 3rd-<br>party data without alarms |
| Touchscreen | 10.4" landscape | 15" portrait |
| Design | Compact, lightweight, pole<br>mounted on clamp | Same general design features<br>Small differences in weight and<br>pole clamp design |
| Compatibility with<br>Intended<br>Environments | Used in surgical environments<br>and conforms to 60601 electrical<br>safety and EMC requirements<br>Compliant with 2nd edition IEC<br>60601 (and 3rd edition in a later<br>update) | Same general electrical<br>compatibility features<br>Compliant with 3rd edition IEC<br>60601 |
| Battery Backup | One-hour life via three Nickel<br>Hydride batteries | Two-hour minimum life via two<br>Lithium-Ion batteries |
| Wireless Connectivity | Wi-Fi and Bluetooth | Wi-Fi |
| Wired Connectivity | 2 USB / 3 Accessory Ports | 3 USB / 8 Spectrum Accessory<br>Ports (SAPs)<br>(Allows connectivity to 3rd-party<br>devices for data display /<br>graphing; alarms are not<br>associated with these data)<br>SAPs are for future use |
| Disposable /<br>Reusable | Not applicable<br>(No patient-contacting parts) | Same |
| Device | Predicate Device | Proposed Device |
| FDA-Recognized<br>Standards Met | Electrical Safety:<br>With 510(k) submission:<br>IEC 60601-1 2nd Edition 1988 with<br>A1:1991 and A2:1995 including<br>relevant clauses of IEC 60601-2-<br>49:2001 clauses 5, 6, 14, 17, 19,<br>20, 44, 49 & 56 | Electrical Safety:<br>AAMI ANSI ES60601-<br>1:2005/(R)2012 and A1:2012<br>including relevant clauses of<br>IEC 60601-2-49:2011 |
| | Updated in 2013:<br>IEC 60601-1 3rd Edition;<br>UL 60601-1/CAN/CSA C22.2 No.<br>601.1 | |
| | Electromagnetic Compatibility:<br>With 510(k) submission:<br>IEC 60601-1-2:2001 Clause 36<br>only (EMC) | Electromagnetic Compatibility:<br>IEC 60601-1-2:2014 |
| | Updated in 2013:<br>IEC 60601-1-2:2007 | |
| | Other:<br>IEC 60601-1-8:2006<br>IEC 62133:2002<br>AAMI ANSI IEC 62304:2006<br>ISO 14971:2000 + A1:2003<br>Updated in 2013:<br>ISO 14971:2007 | Other:<br>IEC 60601-1-6:2010 + A1:2013<br>IEC 60601-1-8:2006 + A1:2012<br>IEC 62133:2012<br>AAMI ANSI IEC 62304:2006<br>IEC 62366:2007+A1:2014<br>ISO 14971:2007 |
# Table A: Comparison Table, M2 Monitor vs. Quantum Workstation
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
# VII. PERFORMANCE DATA - NON-CLINICAL TESTING
No animal testing was submitted to support the substantial equivalence of the Quantum Workstation to the M2 Monitor.
The following bench performance testing was performed to support the substantial equivalence of the Quantum Workstation to the M2 Monitor:
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing, including defibrillator protection and electrosurgery interference requirements, were conducted on the Quantum Workstation and its accessories. The system complies with the following safety and emissions standards:
- . AAMI ANSI ES60601-1:2005/(R)2012 + A1:2012
- IEC 60601-1-2:2014 .
- IEC 60601-1-6:2010 + A1:2013 .
- IEC 60601-1-8:2006 + A1:2012 .
- . IEC 60601-2-49:2011
# Mechanical testing
In addition to the mechanical tests included in the 60601 tests, the following mechanical tests were performed on the Quantum Workstation:
- . screen impact test for the touch screen,
- . effects of cleaning products on the enclosure material and labels / markings, and
- . product function after exposure to non-operational (storage and transport) environmental requirement extremes of temperature and humidity.
# Software verification and validation testing
The software for the Quantum Workstation and its predicate, M2 Monitor, were determined to be Class B software (non-serious injury is possible) per IEC 62304:2006 because, if the M2 Monitor or Quantum Workstation display faulty or incorrect data, clinicians will rely on other medical devices in the operating room to confirm the diagnostic values. Therefore, there are no risks of death or serious injury possible from the device's software.
System-level and subcomponent-level software verification and validation testing were conducted for the Quantum Workstation.
A usability validation test was also performed for the Quantum Workstation user interface.
# SO2 and Hb performance testing
In vitro testing was performed to confirm, as shown in Table A, that the SO2 and Hb (with calculated Hct) functional performance for the Quantum Workstation was unchanged from the M2 Monitor.
{10}------------------------------------------------
#### VIII. PERFORMANCE DATA - CLINICAL TESTING
No clinical data were submitted to support the substantial equivalence of the Quantum Workstation to the M2 Monitor.
### IX. CONCLUSIONS
There are minor differences between the Quantum Workstation and the predicate device (M2 Monitor) in technological characteristics with regards to battery type and life, touch screen display size, number of USB and LAN ports, pinch action for graph resizing, and display of 3°party data without alarms. However, these differences do not raise new questions of safety or effectiveness. Thus, the device characteristics compared in Table A and the results of the bench performance tests confirm that the Quantum Workstation is substantially equivalent to the M2 Monitor.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.