Retrospective clinical images were used to evaluate the accuracy of the Quantra software (Version 2.2) in segregating breast density into BI-RADS 5th edition categories compared to a consensus of radiologists.
Breast density; BI-RADS; Retrospective clinical images; Software validation
Radiologists' consensus assessment of BI-RADS 5th Edition categories
Accuracy of breast composition estimation (Fatty vs. Dense)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Breast composition categorization
—
—
Accuracy 92.9% (Fatty) and 99.1% (Dense) for 2D; 92.0% (Fatty) and 94.0% (Dense) for Tomo
—
—
Independent data set of 230 studies
5 (radiologists)
Indications for Use
The Quantra™ software application is intended for use with mammographic images acquired using digital breast x-ray systems. The Quantra software segregates breast density into categories, which may be useful in the reporting of consistent BI-RADS® breast composition categories as mandated by certain state regulations. The Quantra software reports a result for each subject, which is intended to aid radiologists in the assessment of breast tissue composition. The Quantra software produces adjunctive information; it is not a diagnostic aid.
Device Story
Quantra software analyzes digital mammographic images to estimate breast composition; inputs include distribution and texture of parenchymal tissue patterns. Device transforms these inputs to categorize breast density into BI-RADS® categories (a, b, c, d). Used in clinical settings by radiologists; output provides adjunctive information to assist in consistent reporting of breast composition. Device does not provide diagnostic aid; output affects clinical workflow by standardizing density assessment reporting.
Clinical Evidence
Bench testing only. Retrospective study of 230 screening mammography studies compared Quantra v2.2 output against 5-radiologist consensus BI-RADS 5th Edition categories. For 2D imaging, accuracy was 92.9% for fatty (a+b) and 99.1% for dense (c+d) categories. For Tomo (3D) imaging, accuracy was 92.0% for fatty and 94.0% for dense categories. No clinical or animal studies performed.
Technological Characteristics
Software-based image processing application. Operates on digital breast x-ray images. Aligned with ACR BI-RADS® 5th Edition. Complies with IEC 62304 (software life cycle) and ISO 14971 (risk management). Moderate level of concern.
Indications for Use
Indicated for use with digital mammographic images to aid radiologists in assessing breast tissue composition by segregating breast density into categories consistent with BI-RADS® 5th Edition standards.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 20, 2017
Hologic, Inc. Neil Kelly Principal Regulatory Affairs Specialist 36 Apple Ridge Road DANBURY, CT 06810
Re: K163623
Trade/Device Name: Quantra™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: December 20, 2016 Received: December 22, 2016
Dear Neil Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Robert Ocks
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163623
Device Name Quantra™M
#### Indications for Use (Describe)
The Quantra™ software application is intended for use with mammographic images acquired using digital breast x-ray systems. The Quantra software segregates breast density into categories, which may be useful in the reporting of consistent BI-RADS® breast composition categories as mandated by certain state regulations. The Quantra software reports a result for each subject, which is intended to aid radiologists in the assessment of breast tissue composition. The Quantra software produces adjunctive information; it is not a diagnostic aid.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <div><span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807.92
| Date Prepared: | October 05, 2017 |
|----------------------|--------------------------------------------------------------------|
| Manufacturer: | Hologic, Inc.<br>36 Apple Ridge Road<br>Danbury, CT 06810 USA |
| Est. Registration #: | 1220984 |
| Contact Person: | Neil Kelly<br>Manager, Regulatory Affairs<br>Phone: (508) 263-8634 |
#### Identification of the Device:
| Proprietary/Trade Name: | Quantra™ |
|-------------------------|-------------------------------------|
| Software Version: | 2.2 |
| Classification Name: | System, Image Processing, Radiology |
| Regulatory Number: | 21 CFR 892.2050 |
| Product Code: | LLZ |
| Device Class: | Class II |
| Review Panel: | Radiology |
#### Identification of the Legally Marketed Predicate Device:
| Trade name: | Quantra™ |
|--------------------------|-------------------------------------|
| Software Version: | 2.1 |
| Classification Name: | System, Image Processing, Radiology |
| Regulatory Number: | 21 CFR 892.2050 |
| Product Code: | LLZ |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Submitter/510(k) Holder: | Hologic, Inc. |
| Clearance: | K142037 (cleared October 10, 2014) |
## Device Description:
Quantra is a software application used to produce assessments of breast composition and categorize them. A breast consists of fat and fibroglandular tissue. Fibroglandular tissue, also referred to as dense tissue, contains a mixture of fibrous connective tissue (stroma) and glandular tissue (epithelial cells), and usually appears brighter than surrounding tissue on a digital
{4}------------------------------------------------
mammographic image. Abnormal lesions also appear bright on a mammogram and can be obscured or masked by fibroglandular tissue.
The Quantra software is designed to estimate breast composition categories by analyzing distribution and texture of parenchymal tissue patterns which can be responsible for the masking effect during mammographic reading.
### Indications for Use:
The Quantra™ software application is intended for use with mammographic images acquired using digital breast x-ray systems. The Quantra software segregates breast density into categories, which may be useful in the reporting of consistent BI-RADS® breast composition categories as mandated by certain state regulations. The Quantra software reports a result for each subject, which is intended to aid radiologists in the assessment of breast tissue composition. The Quantra software produces adjunctive information; it is not a diagnostic aid.
#### Standards:
IEC 62304:2006 - Medical Device Software - Software Life Cycle Process ISO 14971:2012 – Medical Devices – Application of Risk Management to Medical Devices
#### Comparison with Predicate Device:
The focus of this submission is to introduce improvements in the proposed Quantra software algorithm to align with ACR BI-RADS® Atlas (Fifth Edition, 2013), and to enhance the accuracy of breast composition estimation. To the end user the proposed Ouantra software is still used in the exact same way as the predicate Quantra software (K142037). The differences are inherent to the algorithm and reported calculations.
#### Substantial Equivalence:
| | Quantra<br>SW Version 2.1<br>Predicate (K142037) | Quantra<br>SW Version 2.2<br>(Proposed) | Comparison |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | Quantra™ is a software<br>application intended for use with<br>images acquired using digital<br>breast x-ray systems. Quantra<br>calculates volumetric breast<br>density as a ratio of<br>fibroglandular tissue and total<br>breast volume estimates.<br>Quantra also provides area<br>breast density as a ratio of | The Quantra™ software<br>application is intended for use<br>with mammographic images<br>acquired using digital breast x-<br>ray systems. The Quantra<br>software segregates breast<br>density into categories, which<br>may be useful in the reporting of<br>consistent BI-RADS® breast<br>composition categories as | Similar |
| | fibroglandular tissue area and<br>total breast area estimates.<br>Quantra segregates breast<br>density into categories, which<br>may be useful in the reporting of<br>consistent BI-RADS® breast<br>composition categories as<br>mandated by certain state<br>regulations. The Quantra results<br>for each image, breast, and<br>subject, are intended to aid<br>radiologists in the assessment of<br>breast tissue composition.<br>Quantra produces adjunctive<br>information; it is not an<br>interpretive or diagnostic aid. | mandated by certain state<br>regulations. The Quantra<br>software reports a result for each<br>subject, which is intended to aid<br>radiologists in the assessment of<br>breast tissue composition. The<br>Quantra software produces<br>adjunctive information; it is not<br>a diagnostic aid. | |
| Level of<br>Concern | Moderate | Moderate | Same |
| Input<br>Modality | Digital x-ray breast images | Digital x-ray breast images | Same |
| Software<br>Product | Yes | Yes | Same |
| Reportable<br>Information | Fibroglandular (dense) Tissue Volume Total Breast Volume Volumetric Breast Density Area Breast Density Volume of Fibroglandular Tissue Score Volumetric Breast Density Score Estimate of Breast Density (integer) Estimate of Breast Density (fractional) | Quantra Generated Breast Density Category | Similar. Uses<br>distribution of<br>glandular<br>tissue rather<br>than density of<br>glandular<br>tissue when<br>assessing<br>breast<br>composition. |
| Statistical<br>comparison<br>to a<br>reference<br>population | Yes | Yes | Same, with the<br>proposed<br>device also<br>being<br>compared<br>against a |
| | | | dataset scored to BI-RADS 5th edition. |
| BI-RADS® Atlas | Yes, 4th Edition 2003 | Yes, 5th Edition 2013 | Similar, with enhancements |
| Brest Composition Estimates | Yes, displayed as 1, 2, 3 and 4 | Yes, displayed as a, b, c, d | Similar, with enhanced accuracy |
{5}------------------------------------------------
{6}------------------------------------------------
## Summary of Testing:
Non-clinical (Bench) testing has been performed using retrospectively collected images of 4view screening negative cases. The proposed Quantra software (Version 2.2) was tested against an independent data set of 230 studies, for which the 5 radiologists' consensus assessment of BIRADS 5th Edition categories are known.
### Quantra 2.2 2D Fatty vs. Dense
| | Quantra 2.2- QDC 2D | | | | |
|---------------------|---------------------|----------------|----------------|-----------------------|----------|
| | | Fatty<br>(a+b) | Dense<br>(c+d) | Per category<br>total | Accuracy |
| BI-RADS- 5th<br>Ed. | Fatty<br>(a+b) | 105 | 8 | 113 | 92.9% |
| | Dense<br>(c+d) | 1 | 116 | 117 | 99.1% |
| | | | Total | 230 | |
# Quantra 2.2 Tomo Fatty vs. Dense
| | Quantra2.2 – QDC Tomo (3D Center Projection) | | | | |
|---------------------|----------------------------------------------|----------------|-----------------------|----------|-------|
| | Fatty<br>(a+b) | Dense<br>(c+d) | Per category<br>total | Accuracy | |
| BI-RADS- 5th<br>Ed. | Fatty<br>(a+b) | 104 | 9 | 113 | 92.0% |
| | Dense<br>(c+d) | 7 | 110 | 117 | 94.0% |
| | | Total | 230 | | |
No animal studies or clinical studies were performed.
## Conclusion:
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics and performance testing, the proposed Quantra software (version 2.2) is substantially equivalent to the predicate Quantra software (version 2.1).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.