K163591 · Becton, Dickinson and Company · NZW · Feb 27, 2017 · General Hospital
Device Facts
Record ID
K163591
Device Name
BD PosiFlush Heparin Lock Flush Syringe
Applicant
Becton, Dickinson and Company
Product Code
NZW · General Hospital
Decision Date
Feb 27, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The Heparin Lock Flush Solution, USP, BD PosiFlush™ Heparin Lock Flush Syringes are intended for maintenance of patency of vascular access devices only.
Device Story
Three-piece, sterile, single-use syringe; pre-filled with heparin sodium (10U/mL or 100U/mL) in 0.9% sodium chloride; 6% (Luer) connector; sealed with tip cap. Used by clinicians to flush vascular access devices to maintain patency. Prevents catheter occlusion. Not for systemic anticoagulation therapy.
Clinical Evidence
Bench testing only. No clinical data. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, hemolysis, systemic toxicity, sensitization, pyrogenicity, genotoxicity, chemical extractables). Design verification testing included container closure integrity, break loose/break out/sustaining force, stopper separation, leakage, pump force, ship test, dead space, and syringe-induced reflux.
Technological Characteristics
Materials: Polypropylene barrel/plunger/tip cap, bromobutyl rubber stopper, silicone lubricant. Sterilization: Moist heat per ISO 17665 (SAL 10^-6). 3mL or 5mL configurations in 10mL barrel diameter. 6% Luer connector.
Indications for Use
Indicated for maintenance of patency of vascular access devices. Not for use in anticoagulation therapy.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is composed of three abstract human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 27, 2017
Becton Dickinson And Company Aakash Jain Senior Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07045
Re: K163591
Trade/Device Name: BD Posiflush Heparin Lock Flush Syringe Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: NGT Dated: December 16, 2016 Received: December 20, 2016
Dear Aakash Jain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
For Tina Kiang Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
The Heparin Lock Flush Solution, USP, BD PosiFlush™ Heparin Lock Flush Syringe
Indications for Use (Describe)
The Heparin Lock Flush Solution, USP, BD PosiFlush™ Heparin Lock Flush Syringes are intended for maintenance of patency of vascular access devices only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Special 510(k) Summary
| Submitted By: | Aakash Jain<br>Senior Regulatory Affairs Specialist<br>Becton, Dickinson and Company<br>1 Becton Drive<br>Franklin Lakes, NJ 07417<br>Phone: (201)847-5879<br>Fax: (201)847-5307 | |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | February 8, 2017 | |
| Subject Devices: | Trade Name: | Heparin Lock Flush Solution, USP-<br>BD PosiFlush™ Heparin Lock Flush Syringe |
| | Common Name: | Heparin, Vascular Access Flush |
| | Classification: | Class II, 21 CFR §880.5200 |
| | Product Code: | NZW |
| Predicate Devices: | Trade Name: | Heparin Lock Flush Solution, USP-<br>BD PosiFlush™ Heparin Lock Flush Syringe |
| | 510(k) Reference: | K090680 |
| | Common Name: | Heparin, Vascular Access Flush |
| | Classification: | Class II, 21 CFR §880.5200 |
| | Product Code: | NZW |
| Device<br>Characteristics | Subject Device<br>BD PosiFlush™ Heparin Lock Flush<br>Syringe<br>(K163591) | Predicate Device<br>BD PosiFlush™ Heparin Lock Flush<br>Syringe<br>(K090680) |
| Manufacturer | Becton, Dickinson and Company | Becton, Dickinson and Company |
| Intended Use | The Heparin Lock Flush Solution,<br>USP, BD PosiFlush™ Heparin Lock<br>Flush Syringes are intended for<br>maintenance of patency of vascular<br>access devices only. | The Heparin Lock Flush Solution,<br>USP, BD PosiFlush™ Heparin Lock<br>Flush Syringes are intended for<br>maintenance of patency of vascular<br>access devices only. |
| Operating<br>Principle | The BD PosiFlush™ Heparin Lock<br>Flush Syringe is a three-piece,<br>sterile, single use syringe with a 6%<br>(Luer) connector pre-filled with<br>heparin sodium, USP in either<br>10U/mL or 100U/mL concentration<br>0.9% Sodium Chloride Injection,<br>USP, and sealed with a tip cap. | The BD PosiFlush™ Heparin Lock<br>Flush Syringe is a three-piece, sterile,<br>single use syringe with a 6% (Luer)<br>connector pre-filled with heparin<br>sodium, USP in either 10U/mL or<br>100U/mL concentration 0.9%<br>Sodium Chloride Injection, USP, and<br>sealed with a tip cap. |
| Materials | Barrel: Polypropylene<br>Plunger Rod: Polypropylene<br>Tip Cap: Polypropylene w/ Blue (10<br>units/mL) or Yellow (100 units/mL)<br>Colorant<br>Stopper Lubricant: Silicone<br>Stopper Material: Bromobutyl<br>Rubber | Barrel: Polypropylene<br>Plunger Rod: Polypropylene<br>Tip Cap: Polypropylene w/ Blue (10<br>units/mL) or Yellow (100 units/mL)<br>Colorant<br>Stopper Lubricant: Silicone<br>Stopper Material: Styrene-butadiene<br>Rubber |
| Packaging | Flow wrap<br>Shelf Carton<br>Case Carton | Flow wrap<br>Shelf Carton<br>Case Carton |
| Specification | Provided in 3 mL and 5 mL in either<br>10U/mL or 100U/mL<br>configurations. All sizes utilize a<br>consistent 10 mL syringe barrel<br>diameter. | Provided in 3 mL and 5 mL in either<br>10U/mL or 100U/mL configurations.<br>All sizes utilize a consistent 10 mL<br>syringe barrel diameter. |
| Device<br>Characteristics | Subject Device | Predicate Device |
| | BD PosiFlush™ Heparin Lock Flush<br>Syringe<br>(K163591) | BD PosiFlush™ Heparin Lock Flush<br>Syringe<br>(K090680) |
| Shelf Life | 10U/mL- 18 Months<br>100U/mL- 24 Months | 10U/mL- 18 Months<br>100U/mL- 24 Months |
| Sterilization | Per ISO 17665 | Per ISO 17665 |
| SAL Level | SAL 10-6 | SAL 10-6 |
| Biocompatibility | Per ISO 10993-1 | Per ISO 10993-1 |
## Device Description
Heparin Lock Flush Solution, USP, BD PosiFlush™ Heparin Lock Flush Syringe (also referred to as BD PosiFlush" Heparin Lock Flush Syringe) is provided in either 10 or 100 Units/mL concentrations in 0.9% Sodium Chloride Injection, USP packaged in a plastic, disposable, single use syringe. The syringe is a three-piece, single use syringe with a 6% (luer) connector prefilled with Heparin Lock Flush Solution, and sealed with a tip cap. The BD PosiFlush"" Heparin Lock Flush Syringe is provided in a 3mL or 5mL size configuration. The BD PosiFlush™ Heparin Lock Flush Syringe is provided with a sterile fluid path, sterilized via moist heat. The solution is delivered through the vascular access device to maintain catheter patency and is not to be used for anticoagulation therapy.
The change described in this submission is to qualify a new stopper material used in the BD PosiFlush " Heparin Lock Flush Syringe.
#### Intended Use
The Heparin Lock Flush Solution, USP, BD PosiFlush™ Heparin Lock Flush Syringe are intended for maintenance of patency of vascular access devices only.
{4}------------------------------------------------
# Technological Characteristics
The subject BD PosiFlush™ Heparin Lock Flush Syringe are equivalent to that of the predicate BD PosiFlush™ Heparin Lock Flush Syringe in intended use, fundamental scientific technology, operating principles, product design, and performance characteristics.
{5}------------------------------------------------
## Performance Data
Modifications to BD PosiFlush™ Heparin Lock Flush Syringe were evaluated using a risk management process. This risk assessment process was performed in accordance with ISO14971. The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
Biocompatibility evaluation was conducted in accordance with the FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," June 16, 2016. The battery of testing included the following tests and assessments. Cytotoxicity (per ISO10993-5), Hemolysis (per ISO10993-4). Acute systemic toxicity (per ISO10993-11), Sub-Chronic Toxicity (Per ISO10993-11), Intracutaneous reactivity (per ISO10993-10), Ocular irritation (per ISO 10993-10), Partial thromboplastin Time (ISO10993-4). Sensitization (per ISO10993-10). Pvrogenicity (per ISO 10993-11), Genotoxicity (per ISO 10993-3) and Chemical extractable analysis (per ISO 10993-18).
Design Verification testing was also conducted to ensure the device met the predetermined acceptance criteria for the following tests; Container Closure Integrity, Break Loose/Break Out/Sustaining force, Stopper Separation, Leakage, Pump Force, Ship Test, Dead Space, Syringe Induced Reflux were performed to demonstrate that the subject device met predetermined acceptance criteria per BD's internal specification.
## Substantial Equivalence
The BD PosiFlush™ Heparin Lock Flush Syringes have been verified to meet the established performance criteria above. The results of design verification testing demonstrate that the modifications made to the subject device do not affect the safety and efficacy of the device and demonstrated the same intended use, fundamental scientific technology and operating principles.
## Conclusion
The subject BD PosiFlush™ Heparin Lock Flush Syringes are substantially equivalent to the legally marketed predicate device, BD PosiFlush™ Heparin Lock Flush Syringes (K090680)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.