K163579 · KLS Martin L.P. · JEY · Nov 21, 2017 · Dental
Device Facts
Record ID
K163579
Device Name
KLS Martin Individual Patient Solutions
Applicant
KLS Martin L.P.
Product Code
JEY · Dental
Decision Date
Nov 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
KLS Martin Individual Patient Solutions implant devices are in the stabilization and fixation of mandibular fractures and mandibular reconstruction.
Device Story
Device consists of patient-specific metallic bone plates and screws for internal fixation of the mandible. Input data derived from patient CT scans; physician provides input during virtual planning to modify digital models. Trained engineers finalize designs; verification performed by virtually fitting device model over 3D patient anatomy model. Plates manufactured via traditional (subtractive) or additive (laser sintering) methods. Used in clinical settings by physicians. Output is a custom-fit implant that stabilizes bone fragments, facilitating reconstruction and fracture healing. Benefits include improved anatomical fit compared to standard pre-contoured plates.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical testing per ASTM F382 (bending resistance and fatigue life) and biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, irritation, chemical/material characterization). Sterilization validated per ISO 17665-1:2006.
Technological Characteristics
Materials: CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136). Manufacturing: Subtractive or additive (laser sintering). Plate thickness: 1.0-3.0 mm. Screw diameter: 2.0-3.2 mm. Sterilization: Non-sterile, steam sterilization by end-user. Patient-specific design based on CT imaging.
Indications for Use
Indicated for the stabilization and fixation of mandibular fractures and mandibular reconstruction in patients requiring surgical intervention.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
Predicate Devices
KLS Martin Mandibular Reconstruction System II (K032442)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 21, 2017
KLS Martin LP Gary Moore Quality Management And Regulatory Affairs Manager 11201 Saint Johns Industrial Pkwy S Jacksonville, Florida 32246
Re: K163579
Trade/Device Name: KLS Martin Individual Patient Solutions Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY Dated: October 20, 2017 Received: October 23, 2017
Dear Gary Moore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mary S. Runner -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163579
Device Name
KLS Martin Individual Patient Solutions
#### Indications for Use (Describe)
KLS Martin Individual Patient Solutions implant devices are in the stabilization and fixation of mandibular fractures and mandibular reconstruction.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for KLS Martin L.P., a company located in Jacksonville, FL. The logo is in red and gray, with the letters "KLS" in red and "martin" in gray. The company's address is 11201 Saint Johns Industrial Pkwy S, Jacksonville, FL 32246. The phone number is 904-641-7746, the toll-free number is 800-625-1557, and the fax number is 904-641-7378. The website address is www.klsmartinnorthamerica.com.
11201 Saint Johns Industrial Pkwy S • Jacksonville, FL 32246 www.klsmartinnorthamerica.com
K163579
Section 5
510(k) Summary 21 CFR 807.92
| Submitter: | KLS Martin LP<br>11201 Saint Johns Industrial Pkwy S<br>Jacksonville, FL 32246 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gary Moore<br>Quality Management and Regulatory Affairs Manager<br>Phone: 800-625-1557<br>Email: gmoore@klsmartin.com |
| Date Prepared: | November 20, 2017 |
| Trade Name: | KLS Martin Individual Patient Solutions |
| Common Name: | Plate, Bone |
| Classification Name: | Bone plate (21 CFR 872.4760) |
| Regulatory Class: | II |
| Product Code: | JEY |
| Predicate Device: | KLS Martin Mandibular Reconstruction System II (K032442) |
| Reference Devices: | Patient Contoured Mesh (K062570)<br>KLS Mini Osteosynthesis System (K943347)<br>Centre-Drive Drill-Free Screw (K971297)<br>KLS Martin Thoracic Plating System (K153482) |
| Device Description: | KLS Martin Individual Patient Solutions is comprised of<br>patient-specific models and metallic bone plates used in<br>conjunction with metallic bone screws for internal fixation of<br>mandibular bone. The devices are manufactured based on<br>medical imaging (CT scan) of the patient's anatomy with<br>input from the physician during virtual planning and prior to<br>finalization and production of the device. The physician only<br>provides input for model manipulation and interactive<br>feedback by viewing digital models of planned outputs that<br>are modified by trained KLS Martin engineers during the<br>planning session. For each design iteration, verification is<br>performed by virtually fitting the generated device model<br>over a 3D model of the patient's anatomy to ensure its<br>dimensional properties allow an adequate fit. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for KLS Martin L.P. The text "KLS" is in red, while "martin" is in gray, and "L.P." is in black. The font is sans-serif and modern.
11201 Saint Johns Industrial Pkwy S • Jacksonville, FL 32246 Phone: 904-641-7746 • 800-625-1557 • Fax: 904-641-7378 www.klsmartinnorthamerica.com
Implants are provided non-sterile, range in thickness from 1.0 - 3.0 mm, and are manufactured using traditional (subtractive) or rapid prototyping (additive) methods from either CP Titanium (ASTM F67) or Ti-6Al-4V (ASTM F136) materials. These patient-specific devices are fixated with previously cleared KLS Martin screws.
- KLS Martin Individual Patient Solutions implant devices are Indications for Use: intended for use in the stabilization and fixation of mandibular fractures and mandibular reconstruction.
### Technological Characteristics & Substantial Equivalence Discussion
The intended use and technological characteristics of the subject device, KLS Martin Individual Patient Solutions, are either identical or have been demonstrated to be substantially equivalent to the primary predicate device, KLS Martin Mandibular Reconstruction System II (K032442):
The subject and primary predicate devices are intended for reconstructive surgery in the facial skeleton and share the same fundamental principle of operation - metallic plates used in conjunction with metallic screws for mandibular reconstructive surgery. The subject device of this submission is specific for mandibular fractures and reconstruction, which is identical to the predicate device (K032442), and similar to the reference device (K062570) cleared for use in the maxillofacial, mandibular, and craniofacial skeleton. Additional reference devices (K943347, K971297, and K153482) intended for bone fixation have been included to cover the range of screw characteristics described in this submission.
The design and materials used for the mandibular reconstruction plates in the subject device is similar to those cleared in KLS Martin Mandibular Reconstruction System II (K032442). The design and dimensions of the plates will be based on the patient-specific data, using similar methods described in Patient Contoured Mesh (K062570), made from either CP Titanium (ASTM F67) using traditional manufacturing methods or Ti-6Al-4V powder (ASTM F136) via additive, laser sintering manufacturing methods, and fall within previously cleared specifications for plate thicknesses ranging from 1.0 - 3.0 mm.
Plates will be fixated with previously cleared KLS Martin titanium screws ranging in diameter from 2.0 - 3.2 mm in lengths from 5 - 22 mm (K943347, K971297, K032442, and K153482). The screw characteristics (head style, type, range of size) have been cleared in the reference devices.
All devices are provided non-sterile for steam sterilization by the end-user.
The angles and dimensions of each KLS Martin Individual Implant Solutions implant devices is determined based on the patient anatomy and surgical plan and should not be further contoured intraoperatively, whereas the plates in the KLS Martin Mandibular Reconstruction System II (K032442) are angled and transversely pre-curved to follow the natural curves of the mandible, but are not designed based on patient specific data as described in the reference device K062570.
{5}------------------------------------------------
K163579 Page 3 of 6
# KUS ITOCIO CIT CII LEIT LE PE 1100 Satustial Play S - Jacksoville, FL 3220
12-11-737
11201 Saint Johns Industrial Pkwy S • Jacksonville, FL 32246 www.klsmartinnorthamerica.com
| Device Comparison Table | | | | | | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Individual Patient<br>Solutions | KLS Martin<br>Mandibular<br>Reconstruction<br>System II<br>K032442 | Patient Contoured<br>Mesh<br>K062570 | KLS Mini<br>Osteosynthesis<br>System<br>K943347 | Centre-Drive<br>Drill-Free Screw<br>K971297 | KLS Martin Thoracic<br>Plating System<br>K153482 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | (Reference Device) |
| Indications for Use | KLS Martin Individual<br>Patient Solutions<br>implant devices are<br>intended for use in the<br>stabilization and<br>fixation of mandibular<br>fractures and<br>mandibular<br>reconstruction. | The KLS Martin<br>Mandibular<br>Reconstruction System<br>II is intended for use in<br>the stabilization and<br>fixation of mandibular<br>fractures and<br>mandibular<br>reconstruction. | Patient Contoured Mesh<br>(PCM) is intended to<br>replace bony voids in<br>mandibular,<br>maxillofacial, or<br>craniofacial skeleton. | The device system has<br>the same intended use<br>as its predicate<br>(K854886). | KLS Martin<br>Centre-Drive Drill-Free<br>Screws are intended for<br>use in rigid internal<br>fixation<br>of the oral-maxillo-<br>cranid-facial bones. The<br>bone screws are used to<br>anchor plates<br>which are contoured to<br>fit the bony surface and<br>stabilize the bone<br>fragments. The<br>addition of the self<br>drilling feature is the<br>only difference between<br>the submitted device<br>and the predicate device<br>referenced. | The KLS Martin<br>Thoracic Plating<br>System is indicated for<br>use in the stabilization<br>and fixation of<br>fractures in the chest<br>wall including sternal<br>reconstructive surgical<br>procedures, trauma, or<br>planned osteotomies. |
| Patient-specific<br>configuration? | Yes. Devices are<br>manufactured | No. Devices are<br>provided in a standard | Yes. Devices are<br>manufactured | No. Devices are<br>provided in a standard | No. Devices are<br>provided in a standard | No. Devices are<br>provided in a standard |
| | patient-specific, based<br>on a CT scan of the<br>patient. | shape and matched to<br>the patient's anatomy<br>intraoperatively. | patient-specific, based on<br>a CT scan of the patient. | shape and matched to<br>the patient's anatomy<br>intraoperatively. | shape and matched to<br>the patient's anatomy<br>intraoperatively. | shape and matched to<br>the patient's anatomy<br>intraoperatively. |
| Classification | 21 CFR 872.4760.<br>Class II | 21 CFR 872.4760.<br>Class II | 21 CFR 882.5330,<br>Class II | 21 CFR 872.4760.<br>Class II | 21 CFR 888.3030.<br>Class II | 21 CFR 888.3030.<br>Class II |
| Device Comparison Table | | | | | | |
| | Individual Patient<br>Solutions | KLS Martin<br>Mandibular<br>Reconstruction<br>System II<br>K032442 | Patient Contoured<br>Mesh<br>K062570 | KLS Mini<br>Osteosynthesis<br>System<br>K943347 | Centre-Drive<br>Drill-Free Screw<br>K971297 | KLS Martin Thoracic<br>Plating System<br>K153482 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | (Reference Device) |
| Product Code | JEY | MQN | GXN | JEY | HRS | HRS |
| Material | CP Titanium or<br>Ti-6A1-4V | CP Titanium or<br>Ti-6Al-4V | CP Titanium or<br>Ti-6A1-4V | Plates: CP Titanium<br>Screws: Ti-6A1-4V | Ti-6Al-4V | CP Titanium and<br>Ti-6Al-4V |
| Manufacturing<br>Method | Traditional<br>(Subtractive)<br>3D (Additive) | Traditional<br>(Subtractive) | Traditional (Subtractive)<br>3D (Additive - Skull<br>Models) | Traditional<br>(Subtractive) | Traditional<br>(Subtractive) | Traditional<br>(Subtractive) |
| Sterilization | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) | Non-sterile (Steam) or<br>Sterile (gamma<br>irradiation) |
| Anatomical Sites | Mandible | Mandible | Craniofacial,<br>Maxillofacial,<br>Mandibular | Craniofacial,<br>Maxillofacial,<br>Mandibular | Craniofacial,<br>Maxillofacial,<br>Mandibular | Thoracic |
| Plates | | | | | | |
| Thickness | 1.0 mm - 3.0 mm | 1.0 mm - 3.0 mm | 0.6 mm - 1.0 mm | 0.6 - 1.0 mm | Not applicable | 1.0 mm - 3.0 mm |
| Style | Non-compression<br>Compression<br>Threaded | Non-compression<br>Compression<br>Threaded | Mesh and solid with or<br>without holes | Standard<br>Low Profile | Not applicable | Sternal: Standard linear<br>shapes (locking)<br>Rib: Bi-linear,<br>pre-contoured X-&<br>Z-shaped (locking) |
| Width<br>(screw-hole<br>dependent) | Min: 7 mm<br>Max: 8.5 mm | Min: 6 mm<br>Max: 8 mm | Not applicable | Not applicable | Not applicable | Not applicable |
| Device Comparison Table | | | | | | |
| | Individual Patient<br>Solutions | KLS Martin<br>Mandibular<br>Reconstruction<br>System II<br>K032442 | Patient Contoured<br>Mesh<br>K062570 | KLS Mini<br>Osteosynthesis<br>System<br>K943347 | Centre-Drive<br>Drill-Free Screw<br>K971297 | KLS Martin Thoracic<br>Plating System<br>K153482 |
| | (Subject Device) | (Primary Predicate) | (Reference Device) | (Reference Device) | (Reference Device) | (Reference Device) |
| Length | Min: 31 mm<br>Max: 320 mm | Min: 28.5 mm<br>Max: 240.65 mm | Not applicable | Not applicable | Not applicable | Not applicable |
| Degree of curvature<br>(in-plane) | Min: 90°<br>Max: 180° | Min: 125°<br>Max: 180° | Not applicable | Not applicable | Not applicable | Not applicable |
| Degree of curvature<br>(out-of- plane) | Min: 60°<br>Max: 180° | Min: N/A<br>Max: 180° | Not applicable | Not applicable | Not applicable | Not applicable |
| Hole spacing | Greater than or equal to<br>8 mm | 8 mm | Not applicable | Not applicable | Not applicable | Not applicable |
| Number of Holes | Min: 4<br>Max: Dependent on<br>length & hole spacing | Min: 4<br>Max: 27 | Not applicable | Not applicable | Not applicable | Not applicable |
| Screw Diameter | 2.0 mm - 3.2 mm | 2.0 mm - 3.2 mm | 1.5 mm | 1.5 mm - 2.3 mm | 1.0 mm - 2.0 mm | 2.3 mm - 3.2 mm |
| Screw Length | 5 mm - 22 mm | 8 mm - 22 mm | 3.5 mm - 15 mm | 4 mm - 19 mm | 2 mm - 7 mm | 7 mm - 17 mm |
| Screw Style | maxDrive & crossDrive<br>(Drill-Free, locking<br>[ThreadLock Taper<br>Screw -TLTS],<br>standard) | Drill-Free, locking<br>(ThreadLock Taper<br>Screw -TLTS) | standard, Drill-Free | crossDrive (slotted,<br>standard, low profile)<br>Centre-Drive<br>(standard) | Centre-Drive, Drill-<br>Free (standard) | maxDrive, Drill-Free<br>screw (locking,<br>standard) |
{6}------------------------------------------------
K163579 Page 4 of 6
# KUS ITOCICCIIN LINE 11201 Saint Iohns Industrial Play S - Jacksonille, FL 3220
11-61-737
11201 Saint Johns Industrial Pkwy S ● Jacksonville, FL 32246 www.k1smartinnorthamerica.com
{7}------------------------------------------------
K163579 Page 5 of 6
# KUS MOCIC CIM CHIN LINE PHONE: 00461-1714 6 800-62-1537 Fax: 904-61-737
11201 Saint Johns Industrial Pkwy S • Jacksonville, FL 32246 www.klsmartinnorthamerica.com
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for KLS Martin L.P. The logo is composed of the text "KLS martin L.P.", with "KLS" in red, "martin" in gray, and "L.P." in black. The font is sans-serif and bolded.
11201 Saint Johns Industrial Pkwy S · Jacksonville, FL 32246 Phone: 904-641-7746 • 800-625-1557 • Fax: 904-641-7378 www.klsmartinnorthamerica.com
### Performance Testing - Non-clinical
Bench testing was used to demonstrate that any differences between the subject and predicate devices do not raise any new safety and effectiveness questions. Mechanical testing was conducted in accordance with ASTM F382 to compare the bending properties of the additive manufactured subject plates against the previously cleared subtractive manufactured predicate plates. The bending resistance and fatigue life of the subject devices was determined to be equivalent or better than the predicate devices.
Steam sterilization validation was performed for the dynamic-air-removal cycle in accordance with ISO 17665-1:2006 to a sterility assurance level (SAL) of 106 using the biological indicator (BI) overkill method. All test method acceptance criteria were met.
Biocompatibility endpoints were evaluated in accordance with ISO 10993. The battery of cytotoxicity, chemical analysis, sensitization and irritation, and chemical/material characterization testing conducted on the subject device were within the pre-defined acceptance criteria, and therefore, adequately addresses biocompatibility for implants with a permanent duration of contact.
### Performance Testing - Clinical
Clinical testing was not necessary for the substantial equivalence determination.
### Substantial Equivalence Conclusion:
The proposed device has the same intended used as the predicate devices. The performance testing included in the submission demonstrates that the differences in technological characteristics do not raise any new safety and effectiveness questions. The information submitted supports substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.