K163578 · Mhc Medical Products, LLC · FMI · Aug 3, 2018 · General Hospital
Device Facts
Record ID
K163578
Device Name
EasyTouch Safety Pen Needle
Applicant
Mhc Medical Products, LLC
Product Code
FMI · General Hospital
Decision Date
Aug 3, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EasyTouch™ Safety Pen Needle is intended for use with pen injector devices for the injection of drugs, including insulin and exenatide. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.
Device Story
EasyTouch™ Safety Pen Needle is a sterile, single-use, disposable hypodermic needle for subcutaneous drug delivery via pen injectors. Device attaches to standard injector pens; protective sheath hides needle before use. During injection, sheath retracts; upon removal from skin, sheath automatically extends and locks over needle to prevent accidental sticks. Used by patients or healthcare providers in home or clinical settings. Benefits include reduced risk of sharps injury. Device is mechanical; no electronic or software components.
Clinical Evidence
No clinical trials performed. Substantial equivalence supported by bench testing and simulated clinical environment testing. Simulated use study evaluated safety feature functionality with both professional healthcare workers and non-clinician users. Additional testing included biocompatibility (cytotoxicity, skin irritation/sensitization, systemic toxicity, hemolysis), pen injector compatibility (ISO 11608-1/2), sterility (EO residue), and shelf-life/packaging integrity.
Technological Characteristics
Hub/Protector: Polypropylene. Needle: 304 Stainless steel with silicone lubricant. Safety feature: Stainless steel spring. Dimensions: 29G (5mm, 8mm). Connectivity: None. Sterilization: Ethylene Oxide (SAL 10⁻⁶). Standards: ISO 11608 (pen interface), ISO 10993 (biocompatibility), ISO 7864 (sterile needles), ISO 9626 (tubing).
Indications for Use
Indicated for patients requiring subcutaneous drug injections (e.g., insulin, exenatide) via pen injector devices. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
August 3, 2018
MHC Medical Products, LLC. % Daniel Kamm Kamm & Associates 8870 Ravello Ct. Naples, Florida 34114
Re: K163578
Trade/Device Name: EasyTouch™ Safety Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: July 3, 2018 Received: July 6, 2018
## Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/6 description: The image shows a signature and a name. The name is "Tina Kiang-S". The signature is a stylized, looping design to the left of the name.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163578
Device Name EasyTouch™ Safety Pen Needle
#### Indications for Use (Describe)
The EasyTouch™ Safety Pen Needle is intended for use with pen injection of drugs, including insulin and exenatide. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the shield also serves to hide the needle before and after injection.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <span style="white-space: nowrap;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="white-space: nowrap;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K163578 MHC Medical Products 8695 Seward Rd. Fairfield, OH 45011 Tel: 877-358-4342 Contact Person: Sean O'Keefe Date Prepared: August 1, 2018
### 1. Identification of the Proposed Device:
Trade Name: EasyTouch™ Safety Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI
- 2. Identification of the Predicate Device: BD Autoshield™ Pen Needle, K060007. Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI
- 3. Description of Device: EasyTouch™ Safety Pen Needle is designed to be used with pen injectors for subcutaneous injections of a desired dose of drugs approved for delivery using a pen needle. The EasyTouch™ Safety Pen Needle is sterile (Ethylene Oxide sterilization procedure), and non-pyrogenic. It is a disposable, single use device. Additionally, the EasyTouch™ Safety Pen Needle is designed to reduce the occurrence of accidental needle sticks by providing a sheath that locks of the needle after use. Prior to injection, the end user will attach the EasyTouch™ Safety Pen Needle to an injector pen. The protective sheath will hide the needle from the user prior to use. Upon application of the needle to the sheath will retract to allow for the needle to enter the body. After the injection is complete, and the needle is removed from the skin, the protective sheath will automatically extend over the needle and lock into place. The EasyTouch™ Safety Pen Needle should be removed from the pen and discarded properly.
- 4. Intended Use/Indications for Use: The EasyTouch™ Safety Pen Needle is intended for use with pen injector devices for the injection of drugs, including insulin and exenatide. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental needle sticks from the patient end of the needle. The shield also serves to hide the needle before and after injection.
- 5. Technical Characteristics: EasyTouch™ Safety Pen Needles and the predicate devices have identical technological characteristics and perform the same way as common pen needles. The EasyTouch™ Safety Pen Needle and the predicate device both include safety features that are identical in function and performance. A detailed comparison table follows:
{4}------------------------------------------------
## Comparison Table
| | BD Autoshield™ Pen Needle, K060007 | EasyTouch™ Safety Pen Needle |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The BD AutoShield™ Pen Needle is<br>intended for use with pen injector devices<br>for the injection of drugs, including insulin<br>and exenatide. Additionally, the attached<br>safety shield automatically locks in place<br>and reduces the occurrence of accidental<br>needle sticks from the patient end of the<br>needle. The shield also serves to hide the<br>needle before and after injection. | The EasyTouch™ Safety Pen Needle is<br>intended for use with pen injector devices<br>for the injection of drugs, including insulin<br>and exenatide. Additionally, the attached<br>safety shield automatically locks in place and<br>reduces the occurrence of accidental needle<br>sticks from the patient end of the needle. The<br>shield also serves to hide the needle before<br>and after injection. SAME. |
| USE | OTC | SAME |
| Principle of<br>operation | Serves as a single use pen needle. The<br>device is removed from its packaging and<br>screwed into a pen injector device. The<br>patient then injects the medicine by first<br>setting the dose on the pen, and then<br>inserting the needle into the skin and then<br>pressing the button on the pen. After the<br>injection the needle automatically retracts<br>into a shielded housing, thus preventing<br>an accidental needle stick. | SAME |
| Interface with pen<br>injector (e.g. luer<br>taper or others), | Employs standardized dimensions and luer<br>taper per ISO 11608, Needle-based<br>injection systems for medical use –<br>Requirements and test methods —The<br>dimensions of the attachment part of the<br>needle shall be such that the needle fits<br>and functions with pen injectors which are<br>in accordance with ISO 11608-1.<br>Unscrewing torque of the needle:<br>0.06~0.08 N.m. The standard 6% luer<br>taper is employed to help ensure a leak<br>free connection to the syringe or pen<br>assembly. | SAME |
| Material<br>information: Hub,<br>Hub protector,<br>Safety feature<br>Needle | Hub: Polypropylene<br>Hub Protector: Polypropylene<br>Safety Feature: Stainless steel spring<br>Needle: 304 Stainless steel needle with<br>silicone lubricant | Hub: Polypropylene<br>Hub Protector: Polypropylene<br>Safety Feature: Stainless steel spring<br>Needle: 304 Stainless steel needle with<br>silicone lubricant<br>SAME |
| Sizes/Lengths | 29G, 30G, and 31G<br>5 mm, 8 mm, 12.7 mm lengths | 29G: 3/16" (5mm) and 5/16" (8mm)<br>(29G only). This is a subset of the predicate. |
| Sterilization | SAL: 10⁻⁶<br>Gamma Sterilized | SAL: 10⁻⁶ SAME<br>Ethylene Oxide EQUIVALENT<br>See discussion below. |
| Shelf life | 5 Years | 5 Years SAME |
| Single use | YES | YES SAME |
| Biocompatibility | Complies with ISO10993 | Complies with ISO10993 |
| | BD Autoshield™ Pen Needle, K060007 | EasyTouch™ Safety Pen Needle |
| | series standards, (presumed, not stated in<br>510(k) summary) | series standards, and the following tests are<br>performed,<br>● Cytotoxicity: No cytotoxicity<br>● Skin Irritation: No evidence of skin<br>irritation<br>● Skin Sensitization: No evidence of<br>sensitization<br>● Acute Systemic Toxicity: No systemic<br>toxicity<br>● Hemolysis: No evidence of hemolysis<br>SAME |
| Photo | Image: BD Autoshield Pen Needle | Image: EasyTouch Safety Pen Needle<br>SIMILARITY IS OBVIOUS |
{5}------------------------------------------------
### 6. Non-clinical testing: Bench Tests were performed. Bench Testing included:
- a. Performance of a Risk Analysis to guide the bench testing regimens.
- b. Biocompatibility per the ISO10993 series of biocompatibility tests: In- Vitro Cytotoxicity report, ISO 10993-5 Skin Sensitization test report ISO-10993-10 Intracutaneous study , ISO-10993-10 Systemic Toxicity test report, ISO-10993-11 Pyrogen test report , ISO-10993-11 All tests passed.
- ﻥ Mechanical testing: To establish proper functioning of the safety feature, testing was performed to evaluate the function of the safety feature in a simulated clinical environment utilizing both professional health care workers and non-clinician pen users using gloved hands. Reference: FDA Guidance Document: Guidance for Industry and FDA Staff, Medical Devices with Sharps Injury Prevention Features Document Issued on: August 9, 2005
- ರ. Pen injector compatibility testing was performed to show that the pen needle works properly with multiple brands of pen injectors. Validation basis ISO11608-1 and ISO11608-2
- e. Sterility testing including EO residues. A sterility assurance level of 10* has been demonstrated. We use Ethylene oxide sterilization whereas the predicate device uses gamma sterilization. The SAL levels for both methods are the same.
- f. Packaging and shelf life testing: Accelerated packaging testing was performed to assure a 5 year shelf life: Packaging integrity, packaging permeability, compression resistance, resistance to bacteria.
- Testing was performed to assure compliance with ISO 7864, Sterile hypodermic needles for single oo use -- Requirements and test methods.
{6}------------------------------------------------
- h. Testing was performed to assure compliance with ISO 9626, Stainless steel needle tubing for the manufacture of medical devices -- Requirements and test methods
- 7. Clinical testing: Clinical testing was not required (per FDA guidance) to establish substantial equivalence. However SIMULATED CLINICAL TESTING was performed to meet the requirements of the FDA Guidance on Sharps Injury Prevention (2005) (see above)
- 8. Conclusion: Based on device comparison information and non-clinical bench testing, the proposed device is substantially equivalent to legally marketed predicate device BD Autoshield™ Pen Needle, K06007.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.