LV Liberty Vision Model 1 90 Yttrium Brachytherapy Source
K163572 · Lv Liberty Vision Corp. · KXK · Mar 15, 2017 · Radiology
Device Facts
Record ID
K163572
Device Name
LV Liberty Vision Model 1 90 Yttrium Brachytherapy Source
Applicant
Lv Liberty Vision Corp.
Product Code
KXK · Radiology
Decision Date
Mar 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5730
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the LV Liberty Vision Model 1 90Yttrium Brachytherapy Source is for the treatment of tumors and benign growths by temporary episcleral irradiation.
Device Story
LV Liberty Vision Model 1 is a singly-encapsulated Yttrium-90 brachytherapy source; consists of solid radioactive Yttrium element hermetically sealed in titanium capsule. Device attaches to manual radionuclide applicator system for temporary episcleral irradiation of tumors and benign growths. Used in clinical settings by trained personnel following licensing authority conditions. Emits beta particles for therapeutic effect. Healthcare providers use device to deliver localized radiation; output affects clinical decision-making by providing targeted treatment for ocular/surface growths. Benefits patient through localized tumor/growth control.
Clinical Evidence
No clinical data. Evidence based on bench testing (ISO 2919/ANSI N43.6 mechanical safety) and Monte Carlo dosimetric simulations.
Technological Characteristics
Solid Yttrium-90 active element; titanium encapsulation. Max source diameter 9.6mm, thickness 0.75mm. Beta emission (end-point 2280 keV, average 934 keV). Mechanical safety meets ISO 2919/ANSI N43.6 Class C53211. Manual operation; no electrical/thermal components. Sterility not required as device does not directly contact patient.
Indications for Use
Indicated for episcleral brachytherapy of tumors and benign growths. Intended for use within a manual brachytherapy system.
Regulatory Classification
Identification
A radionuclide brachytherapy source is a device that consists of a radionuclide which may be enclosed in a sealed container made of gold, titanium, stainless steel, or platinum and intended for medical purposes to be placed onto a body surface or into a body cavity or tissue as a source of nuclear radiation for therapy.
Special Controls
*Classification.* Class II (special controls). A prostate seeding kit intended for use with a radionuclide brachytherapy source only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Salutaris Medical Devices, Inc. Smd Sr90-1 Radionuclide Brachytherapy Source (K142701)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 15, 2017
LV Liberty Vision Corp. % Paul T. Finger, M.D. Chief Executive Officer 155 Fleet Street PORTSMOUTH NH 03801
Re: K163572
Trade/Device Name: LV Liberty Vision Model 1 90 ttrium Brachytherapy Source Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: KXK Dated: December 11, 2016 Received: December 19, 2016
Dear Dr. Finger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Page 2- Paul T. Finger, M.D.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K163572
Device Name
LV Liberty Vision Model 1 90Yttrium Brachytherapy Source
#### Indications for Use (Describe)
LV Liberty Vision Model 1 90 Yttrium Brachytherapy Source with individual activity up to 20 mCi (740 MBq), is indicated for episcleral brachytherapy of tumors and benign growths. The Model 1 source is intended for use within a manual brachytherapy system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over=The=Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5 510(k) Summary
Section 807.92(a)
| (1) Submitter | LV Liberty Vision Corp.<br>155 Fleet Street<br>Portsmouth, New Hampshire 03801; USA; | t: 603.766.0451 |
|------------------|-----------------------------------------------------------------------------------------------|-------------------|
| | Establishment Registration No.: | To Be Applied For |
| | Contact Person: Paul T. Finger, MD<br>Chief Executive Officer<br>e: pfinger@libertyvision.com | |
| (2) Device Name: | | |
### Classification Name: Radionuclide Brachytherapy Source (892.5730) (90 KXK) Common or Usual Name: Brachytherapy Source
- LV Liberty Vision Model 1 90 Yttrium Brachytherapy Source Proprietary Name:
- (3) Legally Marketed Predicate Device:
Salutaris Medical Devices, Inc. Smd Sr90-1 Radionuclide Brachytherapy Source, cleared under 510(k) number K142701 dated 24 April 2015
- (4) Description of LV Liberty Vision Model 1 90Yttrium Brachytherapy Source:
LV Liberty Vision Model 1 ®Yttrium Brachytherapy Source is a singly-encapsulated ®º (ttrium Brachytherapy Source. It consists of a titanium capsule containing a solid radioactive ®Yttrium element. The radioactive element is hermetically sealed in the titanium capsule and, in use, will be attached to a manual radionuclide applicator system (21 CFR 892.5650).
- (5) Intended Use
The intended use of the LV Liberty Vision Model 1 90Yttrium Brachytherapy Source is for the treatment of tumors and benign growths by temporary episcleral irradiation.
- (6) Technological Characteristics:
LV Liberty Vision Model 1 90 ttrium Brachytherapy Source is similar to the predicate high dose rate brachytherapy source that utilizes beta particles from 90Yttrium.
{4}------------------------------------------------
| Technological<br>Characteristic | LV Liberty Vision<br>Model 190Yttrium Brachytherapy<br>Source | Salutaris Medical Devices, Inc.<br>Smd Sr90-1 Radionuclide<br>Brachytherapy Source |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | The source consists of a solid 90Yttrium<br>active element, with a maximum diameter<br>of 9.6 mm and a maximum thickness of<br>0.75 mm, singly encapsulated in metallic<br>titanium with a maximum diameter of 10<br>mm and a maximum thickness of 1.0 mm. | The source consists of a solid<br>90Strontium/90Yttrium pellet (0.6 mm dia x<br>~2 mm long) singly encapsulated in<br>metallic stainless steel (0.8 mm dia x 3.1<br>mm long) and attached to a capsule<br>leader (0.9 mm diameter x 112 mm long). |
| Materials<br>Radionuclide<br>Encapsulation | 90Yttrium<br>Titanium | 90Strontium/90Yttrium<br>Stainless Steel |
| Performance<br>Dosimetry: | | |
| Central Axis<br>Dose Rate at 0.6 mm | 6 mm dia source: 1.02 Gy/min-mCi<br>8 mm dia source: 0.62 Gy/min-mCi<br>10 mm dia source: 0.42 Gy/min-mCi | 0.81 Gy/min-mCi |
| Dose Rate at 1.0 mm | 6 mm dia source: 0.81 Gy/min-mCi<br>8 mm dia source: 0.50 Gy/min-mCi<br>10 mm dia source: 0.34 Gy/min-mCi | 0.52 Gy/min-mCi |
| Sterility | This source never directly contacts the<br>patient; sterility is not required. | This source assembly never directly<br>contacts the patient; sterility is not<br>required. |
| Biocompatibility | This source never directly contacts the<br>patient; biocompatibility assessment is not<br>applicable. The outside of the entire<br>assembly is fabricated from titanium,<br>which is a biocompatible material. | This source assembly never directly<br>contacts the patient; biocompatibility<br>assessment is not applicable. The<br>outside of the entire assembly is<br>fabricated from stainless steel, which is a<br>biocompatible material. |
| Mechanical Safety | ISO 2919/ANSI N43.6 Class C53211<br>Applied for New Hampshire Registration | ISO 2919/ANSI N43.6 Class C53X11<br>Massachusetts Sealed Source<br>Registration<br>LA-1390-S-101-S |
| Chemical Safety | This source assembly never directly<br>contacts the patient; chemical safety<br>assessment is not applicable. The<br>outside of the entire assembly is<br>fabricated from titanium, which will not<br>chemically react with body tissue. | This source assembly never directly<br>contacts the patient; chemical safety<br>assessment is not applicable. The<br>outside of the entire assembly is<br>fabricated from stainless steel, which will<br>not chemically react with body tissue. |
| Energy Delivered | 90Yttrium (half-life: 2.67 days) decays by<br>beta emission. The 90Yttrium beta has an<br>end-point energy of 2280 keV and an<br>average energy of 934 keV. | 90Strontium/90Yttrium (half-life: 28.79<br>years) decays by beta emission. The<br>90Yttrium is in secular equilibrium with the<br>90Strontium. The 90Yttrium beta has an<br>end-point energy of 2280 keV and an<br>average energy of 934 keV. The<br>90Strontium beta has an end-point energy<br>of 546 keV and an average energy of 196<br>keV. The range of 196 keV betas is less<br>than 0.5 mm of tissue, and therefore has<br>no therapeutic effect. The therapeutic<br>effect is virtually exclusively from the<br>90Yttrium betas. |
| | | |
| Compatibility with<br>Environment and<br>Other Devices | 90Yttrium is a radioactive material and<br>should be strictly controlled. | 90Strontium/90Yttrium is a radioactive<br>material and should be strictly controlled. |
| | The source should only be used following | The source should only be used following |
| | the conditions and limitations specified by | the conditions and limitations specified by |
| | the licensing authority (NRC or Agreement<br>State). | the licensing authority (NRC or Agreement<br>State). |
| | The source should be stored in a shielded | The source should be stored in a shielded |
| | container with which it is used or the<br>transport container in which it is delivered. | container with which it is used or the<br>transport container in which it is delivered. |
| | If any source cannot be accounted for, the<br>loss should be reported to the federal or<br>state licensing agency. | If any source cannot be accounted for, the<br>loss should be reported to the federal or<br>state licensing agency. |
| | Store at normal room temperature. | Store at normal room temperature. |
| | When disposal is indicated, sources<br>should be disposed of in accordance with<br>the requirements of the institution's<br>radioactive material license. In general,<br>these sources can be disposed of by<br>means of a "Decay in Storage" method<br>approved by the requlatory authority in<br>accordance with 10 CFR 35.92 or<br>equivalent state regulations. Because of<br>the short halflife, sources which have<br>been stored for 60 days may be checked<br>for radioactive content and, if less than 5<br>nCi, be disposed of in normal waste. | When disposal is indicated, radioactive<br>material should be transferred to an<br>authorized recipient, typically the source<br>supplier. Radioactive material should<br>never be disposed of in normal waste. |
| | | |
| | Alternatively, radioactive material should<br>be transferred to an authorized recipient,<br>typically the source supplier. | |
| Where Used | This source should only be used within a<br>properly designed room following the<br>conditions and limitations specified by the<br>licensing authority (NRC or Agreement<br>State). | This source should only be used within a<br>properly designed room following the<br>conditions and limitations specified by the<br>licensing authority (NRC or Agreement<br>State). |
| Standards Met<br>Mechanical | ISO 2919/ANSI N43.6 | ISO 2919/ANSI N43.6 |
| Electrical Safety | Not Applicable | Not Applicable |
| Thermal Safety | Not Applicable | Not Applicable |
| Radiation Safety | This 90Yttrium source is radioactive, and<br>appropriate precautions must be taken<br>during handling to minimize radiation<br>exposure to personnel. Personnel<br>monitoring is required. | This 90Strontium/90Yttrium source is<br>radioactive, and appropriate precautions<br>must be taken during handling to minimize<br>radiation exposure to personnel.<br>Personnel monitoring is required. |
| | This source should only be used within a<br>properly designed room following the<br>conditions and limitations specified by the<br>licensing authority (NRC or Agreement<br>State). | This source should only be used within a<br>properly designed room following the<br>conditions and limitations specified by the<br>licensing authority (NRC or Agreement<br>State). |
| | In circumstances where emergency | In circumstances where emergency |
| operations must be performed, the<br>operator should use proper applicators,<br>maintain safe working distances and work<br>as rapidly as safely possible to minimize<br>radiation exposure. | op<br>op<br>ma<br>as<br>rac | |
{5}------------------------------------------------
{6}------------------------------------------------
ions must be performed, the or should use proper applicators, ain safe working distances and work idly as safely possible to minimize on exposure.
### Section 807.92(b)
- (7) Nonclinical Tests
### Physical Testing
The LV Liberty Vision Model 1 90 Yttrium Brachytherapy Source has been subjected to the tests specified in American National Standard (ANSI) N43.6 and International Organization for Standardization (ISO) Standard 2919, as referenced in the FDA "Guidance for the Submission of Premarket Notifications for Photon-Emitting Brachytherapy Sources" dated 2 August 2000.
Prototype sources were subjected to the tests specified in ANSI N43.6-2007 and ISO 2919-2012 and have equaled or exceeded the requirements corresponding to a classification of C53211, which is the requirement for brachytherapy sources. This exceeds to the physical testing of the predicate device.
#### Dosimetry
The dose distribution around the LV Liberty Vision Model 1 90 fttrium Brachytherapy Source was calculated by Monte Carlo simulation. This is equivalent to the dosimetry of the predicate device.
- (8) Clinical Tests
Not Applicable
- (9) Conclusions
The results of the nonclinical physical testing and the dosimetric analysis, demonstrate that the LV Liberty Vision Model 1 ** Yttrium Brachytherapy Source is as safe, as effective, and performs as well or better than the legally marketed predicate device, Salutaris Medical Devices, Inc. Smd Sr90-1 Radionuclide Brachytherapy Source, cleared under 510(k) number K142701 dated 24 April 2015.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.