Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women. Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This test is not intended for use in screening blood and plasma donors. The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.
Device Story
Elecsys CMV IgM is an in vitro diagnostic assay for detecting IgM antibodies to Cytomegalovirus (CMV). It utilizes electrochemiluminescence immunoassay (ECLIA) technology on the cobas e 801 analyzer. The device processes human serum or plasma samples using a μ-capture protocol; biotinylated monoclonal anti-h-IgM antibodies and ruthenium-labeled recombinant CMV-specific antigens form complexes with streptavidin-coated microparticles. The analyzer measures the resulting electrochemiluminescent signal to determine antibody concentration. Operated by laboratory professionals in clinical settings, the system automates sample handling, incubation, and signal processing. Results are reported as COI (Cut-off Index) values, aiding clinicians in diagnosing recent or current CMV infections. The cobas e 801 offers higher throughput and modified sample/reagent handling compared to the predicate cobas e 601, while maintaining the same core immunoassay principle and clinical performance characteristics.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP05-A3; LoB/LoD per CLSI EP17-A2. Method comparison study (n=223) between cobas e 801 and cobas e 601 showed 100% positive percent agreement and 100% negative percent agreement. Interference studies confirmed no interference from endogenous substances (biotin, intralipid, hemoglobin, bilirubin, RF) or common drugs. Matrix equivalency confirmed for serum and plasma types.
Indicated for qualitative detection of IgM antibodies to CMV in human serum and plasma (Li-heparin, K2-EDTA, K3-EDTA) to aid diagnosis of recent or current CMV infection. Includes pregnant women. Not for immunocompromised/immunosuppressed individuals, neonatal screening, point-of-care, or blood/plasma donor screening.
Regulatory Classification
Identification
Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number: K163569
B. Purpose for Submission: Adding a previously cleared assay on a new instrument platform
C. Measurand: IgM antibodies to CMV
D. Type of Test: Electrochemiluminescence immunoassay
E. Applicant: Roche Diagnostics, Inc.
F. Proprietary and Established Names: Elecsys CMV IgM
G. Regulatory Information:
1. Regulation section: 21 CFR 866.3175
2. Classification: Class II
3. Product code(s): LFZ – Cytomegalovirus serological reagents
JJE – Discrete photometric chemistry analyzer for clinical use
4. Panel: Microbiology
H. Intended Use:
1. Intended use(s):
Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women.
Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This test is not intended for use in screening blood and plasma donors.
The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers.
2. Indication(s) for use: Same as intended use.
{1}
3. Special conditions for use statement(s): For prescription use
4. Special instrument requirements: cobas e 801 Immunoassay Analyzer
I. Device Description:
The Elecsys CMV IgM is a $\mu$ -capture immunoassay that uses streptavidin microparticles, biotinylated recombinant CMV-specific antigen labeled with a ruthenium complex, and electrochemiluminescence detection. The test system contains human serum-based calibrators intended for use with the system.
The total duration of the assay is 18 minutes. The assay steps are as follows:
- During the first incubation, $6\mu \mathrm{L}$ of sample are automatically prediluted 1:20 with diluent (Diluent Universal). Biotinylated monoclonal anti-human IgM specific antibodies are also added.
- During the second incubation, CMV-specific recombinant antigen labeled with a ruthenium complex and streptavidin-coated microparticles are added. Anti-CMV IgM antibodies present in the sample react with the ruthenium-labeled CMV-specific recombinant antigen. The complex becomes bound to the solid phase via interaction of biotin with streptavidin.
- The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell II M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
- The analyzer automatically calculates the cutoff based on the measurement of Calibrator 1 and Calibrator 2. The result of the samples is given either as reactive, borderline, or non-reactive as well as in the form of a cutoff index (signal of sample/cutoff)(COI) as shown in Table 1.
Table 1: Interpretation of Results
| Numeric Result (COI) | Result Message | Interpretation/Further Steps |
| --- | --- | --- |
| COI < 0.7 | Non-reactive | CMV IgM-specific antibodies not detected. |
| 0.7 ≥ COI > 1.0 | Borderline | Re-test the sample. If the result is indeterminate (borderline), collect and test a second sample within the following 2 weeks. |
| COI ≥ 1.0 | Reactive | CMV IgM-specific antibodies detected. |
The magnitude of the measured result above the cutoff is not indicative of the total amount of antibody present in the sample
{2}
The following reagents are provided in the Elecsys CMV IgM assay kit:
1) The reagent rackpack (cobas e pack) consists of reagents M, R1, and R2, and is labeled as CMVIGM:
a. M: Streptavidin-coated microparticles, 1 bottle, 14.1 mL
b. R1: Anti-h-IgM-Ab-biotin, 1 bottle, 19.7 mL
c. R2: CMV-Ag-Ru(bpy), 1 bottle, 19.7 mL
CMV-specific antigen labeled with ruthenium complex
2) CMVIGM Cal1: Negative calibrator 1, 1 bottle of 1.0 mL
Human serum negative for anti-CMV IgM; preservative.
3) CMVIGM Cal2: Positive calibrator 2, 1 bottle of 1.0 mL
Anti-CMV IgM (human serum) in HEPES buffer, bovine albumin, and preservative
The following materials are required but not provided:
1) PreciControl CMV IgM 8 x 1.0 mL each of PreciControl CMV IgM 1 and 2
2) Diluent Universal, 45.2 mL sample diluent
3) CalSet Vials, 2 x 56 empty snap-cap bottles
4) General laboratory equipment
5) Cobas e 801 Immunoassay Analyzer
6) Accessories for the cobas e 801 analyzer
The cobas e 801 Immunoassay Analyzer is a fully automated, software controlled analyzer system for in vitro determination of analytes in human body fluids. It is part of the cobas 8000 modular analyzer series. It uses electrochemiluminescent technology for signal generation and measurement. It is a modified version of the cobas e 601 analyzer module, part of the cobas 6000 modular analyzer.
J. Substantial Equivalence Information:
1. Predicate device name(s): Elecsys CMV IgM (on the cobas e 601)
2. Predicate 510(k) number(s): K142133
{3}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: Elecsys CMV IgM (on the cobas e 801) K163569 | Predicate Device: Elecsys CMV IgM (on the cobas e 601) K142133 |
| Indications for Use | Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women. Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This test is not intended for use in screening blood and plasma donors. The electrochemiluminescence immunoassay “ECLIA” is intended for use on cobas e immunoassay analyzers. | Immunoassay for the in vitro qualitative detection of IgM antibodies to CMV in human serum, lithium-heparin plasma, K2-EDTA plasma, and K3-EDTA plasma. The test is intended as an aid in the diagnosis of recent or current CMV infection in individuals for which a CMV IgM test was ordered, including pregnant women. Performance characteristics have not been evaluated in immunocompromised or immunosuppressed individuals. This test is not intended for use in neonatal screening or for use at point of care facilities. This assay is not intended for use in screening blood and plasma donors. The electrochemiluminescence immunoassay “ECLIA” is intended for use on Elecsys immunoassay analyzers. |
| Sample Types | Serum, serum with separating gel, Li-heparin, K2EDTA, and K3EDTA plasma | Same |
| Detection Technology | Electrochemiluminescence immunoassay (ECLIA) | Same |
| Assay Protocol | μ-Capture | Same |
| Antibody | Biotinylated monoclonal anti-h-IgM antibody (mouse) CMV-specific antigen (recombinant, E. coli) labeled | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device: Elecsys CMV IgM (on the cobas e 801) K163569 | Predicate Device: Elecsys CMV IgM (on the cobas e 601) K142133 |
| | with ruthenium complex Streptavidin-coated microparticles | |
| Calibrators | CMV IgM Cal 1 and CMV IgM Cal 2 | Same |
| Controls | PreciControl CMV IgM 1 &2 | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device: Elecsys CMV IgM (on the cobas e 801) K163569 | Predicate Device: Elecsys CMV IgM (on the cobas e 601) K142133 |
| Throughput | 300 tests/hour/module | 170 tests/hour/module |
| Sample Volume | 6 μL | 10 μL |
| Sample Capacity Onboard | 300 | 150 |
| Instrument Platform | Part of the cobas 8000 modular analyzer series | Part of the cobas 6000 modular analyzer series |
| Onboard Storage Temperature | 5-10°C | 18-22°C |
# K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP5-A3: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline - $3^{\text{rd}}$ Edition
- CLSI EP15-A2: User Verification of Performance for Precision and Trueness; Approved Guideline - $2^{\text{nd}}$ Edition
- CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - $2^{\text{nd}}$ Edition
# L. Test Principle:
The Elecsys CMV IgM immunoassay is based on the $\mu$ -Capture test format. During the first incubation step, biotinylated monoclonal anti-h-IgM-specific antibodies binds specifically to IgM the diluted test specimen. CMV-specific recombinant antigen labeled with a ruthenium complex and streptavidin-coated microparticles are then added for the second incubation. Anti-CMV IgM antibodies present in the sample react with the ruthenium-labeled CMV-specific recombinant antigen. The complex becomes bound to the solid phase via interaction of biotin with streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell II M. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier.
{5}
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
Precision of the Elecsys CMV IgM assay was evaluated according to CLSI EP05-A3 on one cobas e 801 Immunoassay Analyzer with one reagent lot. Two replicates per run for each control [PreciControl (PC) CMV IgM, level 1 and 2] and five native human serum and plasma samples (HSP) from single donors were measured (except for plasma sample 5 which was pooled). Measurements were performed in 2 runs per day for 21 days. All results met the pre-defined acceptance criteria and are shown in Table 2:
Table 2: Precision Results
| | | | Repeatability | | Intermediate Precision | |
| --- | --- | --- | --- | --- | --- | --- |
| Sample | n | Mean [COI] | SD# [COI] | CV# [%] | SD [COI] | CV [%] |
| HSPa 1* | 84 | 0.202 | 0.002 | 1.2 | 0.006 | 2.8 |
| HSP 2# | 84 | 0.847 | 0.011 | 1.3 | 0.015 | 1.8 |
| HSP 3# | 84 | 1.09 | 0.018 | 1.6 | 0.020 | 1.8 |
| HSP 4# | 84 | 3.46 | 0.033 | 1.0 | 0.049 | 1.4 |
| HSP 5- | 84 | 1.28 | 0.013 | 1.0 | 0.023 | 1.8 |
| PCb CMV IgM 1 | 84 | 0.225 | 0.002 | 0.9 | 0.006 | 2.6 |
| PC CMV IgM 2 | 84 | 1.85 | 0.036 | 1.9 | 0.041 | 2.2 |
*serum sample, # plasma sample, ^plasma pool
$^{\mathrm{a}}$ HSP=human specimen
bPreciControl
cStandard Deviation
$^{\mathrm{d}}$ Coefficient of Variation
# b. Linearity/assay reportable range:
Not Applicable.
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Reagent Stability: Two studies were conducted to evaluate the stability of the Elecsys CMV IgM assay reagents on the cobas e 801: onboard reagent stability and reagent real-time stability. These studies were reviewed and found to be acceptable. The reagents are stable when stored unopened at $2^{\circ}\mathrm{C}$ to $8^{\circ}\mathrm{C}$ for 12 months or onboard (5 to $10^{\circ}\mathrm{C}$ ) for 16 weeks.
Sample Stability: Three studies were conducted to evaluate sample stability: sample stability at $2 - 8^{\circ}\mathrm{C}$ , sample stability at $20 - 25^{\circ}\mathrm{C}$ , and sample stability at $-20^{\circ}\mathrm{C}$ . These studies were reviewed and found to be acceptable. Specimens are stable for 4 weeks at $2^{\circ}\mathrm{C}$ to $8^{\circ}\mathrm{C}$ , 7 days at $20 - 25^{\circ}\mathrm{C}$ , or 3 months at $-20^{\circ}\mathrm{C}$ ( $\pm 5^{\circ}\mathrm{C}$ ).
{6}
d. Detection limit:
Limit of Blank: The Limit of Blank (LoB) of the Elecsys CMV IgM assay on the cobas e 801 Immunoassay Analyzer was determined according to CLSI EP17-A2 as the 95th percentile of measurements of blank samples. Five blank serum samples were assayed with two replicates per run for six runs, using three reagent lots on one instrument, for a total of 60 blank replicates per reagent lot. The LoB was calculated to be 0.243 COI.
Limit of Detection: The Limit of Detection (LoD) of the Elecsys CMV IgM assay on the cobas e 801 Immunoassay Analyzer was determined according to CLSI EP17-A2 as the lowest amount of analyte in a sample that can be detected with 95% probability. Five samples were prepared by spiking human serum samples from single donors with low levels of analyte. The samples were tested with two replicates per sample per run, six runs, with three reagent lots, on one instrument for a total of 60 replicates per sample per reagent lot. The LoD was calculated to be 0.267 COI using the following equation: LoD=LoB+1.653 x SD.
e. Analytical specificity:
Endogenous Interferences: To evaluate the effect of elevated levels of biotin (10 to 100 ng/mL), intralipid (150 to 1500 mg/dL), hemoglobin (50 to 500 mg/dL), and bilirubin (2 to 20 mg/dL) on the Elecsys CMV IgM assay, four CMV IgM samples (low negative, low positive, positive, and high positive) were tested on the cobas e 801. The recovery (% COI) was calculated for each sample and compared to the reference unspiked sample.
Acceptance criteria: Negative/borderline samples (COI<1.0), deviation ± 0.2 COI; Positive samples (COI ≥ 1.0), deviation ± 20% (Recovery 80-120%).
The results met the acceptance criteria for all concentrations of endogenous interferent tested. Roche will receive the following claims for no interference observed at the following concentrations: biotin ≤100 ng/mL, intralipid ≤1500 mg/dL, hemoglobin ≤500 mg/dL, and bilirubin ≤20 mg/dL.
To evaluate the effect of rheumatoid factor (RF), native high RF samples were diluted 1:1 with RF negative-CMV IgM positive and RF negative-CMV negative samples. The results of the CMV positive sample diluted with a RF high sample is compared to a 1:1 dilution of the same CMV sample with RF negative samples with similar total IGM levels as the RF high sample. Seven samples with a CMV IgM COI from 0.203 to 5.77 were diluted 1:1 with 3 RF high/CMV-IgM negative samples. As a reference, the sample samples were diluted with RF negative/CMV-IgM negative samples. RF concentrations of 462 to 899 IU/mL were tested and met the acceptance criteria. Roche will receive a claim for no interference for RF ≤ 899 IU/mL.
Exogenous Interfering Substances (Drugs): The effect of quantitation of analyte in the presence of drugs was determined by comparing values obtained from native
7
{7}
human serum and plasma samples from single donors on the cobas e 801. The samples were spiked with 17 commonly used and 2 special pharmaceutical compounds and tested on the Elecsys 2010 Immunoassay Analyzer. Acetylcysteine and phenylbutazone were measured on the cobas e 411 Immunoassay Analyzer. Two samples negative and positive for CMV IgM were divided into aliquots and spiked with the potential interferent. The concentration of CMV IgM in the samples was approximately 0.2 COI and 1.9 (2.4 for acetylcysteine and phenylbutazone)COI. The CMV IgM concentration of the spiked aliquots was determined in 9-fold determination and compared to the CMV IgM concentration determined for the reference aliquot on the Elecsys 2010 and the cobas e 411 Immunoassay Analyzers, respectively. Acceptance criteria: Negative/borderline samples (COI<1.0), deviation $\pm 0.2\mathrm{COI}$ ; Positive samples (COI>1.0), deviation $\pm 10\%$ . Each compound was found to be non-interfering at the concentration listed in the Table 3.
Table 3:Drug Interences: Common Therapeutic Drugs
| Potential Interfering Exogenous Substance | Concentration Tested [mg/mL] |
| --- | --- |
| Acetylcystein* | 1660 |
| Ampicillin-Na | 1000 |
| Ascorbic acid | 300 |
| Ca-Dobesilate | 200 |
| Cyclosporine | 5 |
| Heparin | 5000 U/L |
| Levodopa | 20 |
| Methyldopa +1.5 | 20 |
| Metronidazole | 200 |
| Phenylbutazone* | 400 |
| Doxycycline | 50 |
| Acetylsalicylic Acid | 1000 |
| Rifampicin | 60 |
| Acetaminophen | 200 |
| Ibuprofen | 500 |
| Theophyllin | 100 |
| Ganciclovir | 800 [mg/L] |
| Valganciclovir | 900 [mg/L] |
*Data measured on the cobas e 411 Immunoassay Analyzer
f. Assay cut-off:
Please refer to the Decision Summary for K142133 for a summary of the assay cutoff (performed on the Elecsys instrument family member Elecsys 2010). Validation of the assay cut-off was performed by external clinical studies on the Elecsys 2010 (K142133) and a Method Comparison study on the cobas e 801.
The analyzer automatically calculates the cut-off based on the measurement of CMVIGM Cal1 and CMVIGM Cal2. The results are in the form of a cut-off index (signal sample/cut-off) as previously shown in Table 1.
{8}
g. High-dose hook effect:
Testing with the Elecsys CMV IgM assay on the cobas e 801 demonstrated no high-dose hook effect. Three naturally positive human serum and plasma samples from single donors with high CMV IgM concentrations were diluted in a series of 11 dilution steps. Each dilution was tested in singlicate. No high-dose hook effect was observed up to 18.7 COI.
h. Verification of IgM Specificity:
Please refer to the Decision Summary for K142133 (performed on the Elecsys instrument family member Elecsys 2010).
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed to compare the Elecsys CMV IgM on the cobas e 801 and the Elecsys CMV IgM on the cobas e 601 analyzer (predicate). A total of 223 leftover deidentified native human plasma samples from single donors were measured in singlicate. Measured values ranged between 0.210 to 20.7 COI for the cobas e 601 and between 0.189 to 21.0 COI for the cobas e 801. The results are shown in Table 4.
Table 4: Method Comparison
| | | cobas e 601 (original) | | | Total cobas e 801 |
| --- | --- | --- | --- | --- | --- |
| | | Negative COI <0.7 | Borderline 0.7 ≥ COI > 1.0 | Positive COI ≥ 1.0 | |
| cobas e 801 (new) | Negative COI < 0.7 | 142 | 1 | 0 | 143 |
| | Borderline 0.7 ≥ COI > 1.0 | 0 | 6 | 0 | 6 |
| | Positive COI ≥ 1.0 | 0 | 1 | 73 | 74 |
| Total cobas e 601 | | 142 | 8 | 73 | 223 |
The positive percent agreement rate was $100\%$ (100/100), the negative percent agreement rate was $100\%$ (73/73), and the percent agreement for Borderline results was $75\%$ (6/8). None of the samples changed from positive to false negative or negative to false positive.
b. Matrix comparison:
The study was conducted to assess the effect of anticoagulants on quantitation of analyte by the Elecsys CMV IgM immunoassay on the cobas e 801. The following matrices were evaluated: Li-heparin plasma, $\mathrm{K}_2$ -EDTA plasma, $\mathrm{K}_3$ -EDTA plasma, and serum from separator tubes. Data from using the serum separator tubes was generated on the Elecsys instrument family member Elecsys 2010 (K142133) and the
{9}
data is summarized here. Samples were collected into matched serum and plasma collection tubes and assayed in duplicate. The study was conducted using negative (non-reactive) and positive (reactive) samples. The results are summarized in Table 5 and Table 6 and support the use of the following sample types: Li-heparin plasma, $\mathrm{K}_2$ -EDTA plasma, $\mathrm{K}_3$ -EDTA plasma, and serum separator tubes.
Table 5: Matrix Equivalency Negative Samples
| Sample Matrix | Mean COI | Percent of samples showing differences in recovery relative to serum (COI) in non-reactive specimens | | |
| --- | --- | --- | --- | --- |
| | | < 0.07 COI | 0.07 – 0.1 COI | > 0.1 COI |
| Li-Heparin Plasma | 0.238 | 100% | 0 | 0 |
| K2-EDTA Plasma | 0.237 | 100% | 0 | 0 |
| K3-EDTA Plasma | 0.232 | 100% | 0 | 0 |
| Serum Separator Tubes | 0.265 | 100% | 0 | 0 |
Table 6: Matrix Equivalency Positive Samples
| Sample Matrix | Mean COI | Percent of samples showing differences in recovery relative to serum (COI) in reactive specimens | | |
| --- | --- | --- | --- | --- |
| | | < 10 % | 10 – 20 % | > 20 % |
| Li-Heparin Plasma | 2.15 | 77 % | 23 % | 0 % |
| K2-EDTA Plasma | 2.27 | 100 % | 0 % | 0 % |
| K3-EDTA Plasma | 2.32 | 100 % | 0 % | 0 % |
| Serum Separator Tubes | 2.27 | 90 % | 10 % | 0 % |
3. Clinical studies:
a. Clinical Sensitivity: Not Applicable.
b. Clinical specificity: Not Applicable.
c. Other clinical supportive data (when a. and b. are not applicable): Please refer to the Decision Summary for K142133 for a summary of clinical study results (performed on the Elecsys instrument family member Elecsys 2010).
4. Clinical cut-off: Not Applicable.
5. Expected values/Reference range: Not Applicable.
N. Instrument Name:
cobas e801 Immunoassay Analyzer
{10}
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Barcode
4. Specimen Sampling and Handling:
Input and transport of samples using universal sample racks, modular sample buffer input, core/transportation unit and STAT port.
5. Calibration:
Calibration must be performed once per reagent lot using fresh reagent (i.e., not more than 24 hours since the reagent kit was registered on the analyzer. Renered calibration is recommended after 12 weeks when using the same reagent lot, after 28 days when using the same cobas e pack on the analyzer, and as required (e.g., quality control finding outside the defined limits).
6. Quality Control:
Controls should be run individually at least once every 24 hours when the test is in use, once per cobas e pack, and following each calibration. Values should fall within the defined limits. Each laboratory should establish corrective measures to be taken if values fall outside defined limits.
11
{11}
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
The software documentation was reviewed and found to be acceptable. The firm provided documentation to support that the device was designed, developed, and maintained under appropriate software lifecycle processes.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.