← Product Code [QEW](/productcode/QEW) · K163549

# ClotTriever Thrombectomy System (K163549)

_Inari Medical, Inc. · QEW · Feb 16, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163549

## Device Facts

- **Applicant:** Inari Medical, Inc.
- **Product Code:** [QEW](/productcode/QEW.md)
- **Decision Date:** Feb 16, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5150
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and ClotTriever Thrombectomy System is indicated for: - The non-surgical removal of soft thrombi and emboli from blood vessels. - Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature.

## Device Story

ClotTriever Thrombectomy System is a single-use, sterile device for peripheral vascular thrombus removal. System comprises a ClotTriever Sheath (with self-expanding distal mesh funnel, hemostatic valve) and a ClotTriever Catheter (three coaxial shafts with expandable member and tissue collection net). Operated by physician under fluoroscopic guidance. Principle of operation: expandable nitinol structure is deployed within vessel, drawn through obstruction to capture clot, and retracted into sheath for removal. Radiopaque markers and tips aid visualization. Device restores blood flow by mechanical extraction. Benefits include non-surgical intervention for thrombus/emboli. No software or AI components.

## Clinical Evidence

No clinical data. Evidence includes bench testing (verification/validation) and acute bovine animal study. Bench tests: ISO 10555-1, ISO 594-1/2, ISO 10993-1 (biocompatibility), ISO 11607-1/2 (packaging). Tests covered mechanical integrity (tensile, torque, kink radius), leakage, flow, and simulated use. Shelf life validated to 12 months via accelerated aging.

## Technological Characteristics

Materials: Nitinol metal structures, thermoplastic polymer catheter shafts. Components: Sheath with self-expanding distal mesh funnel, catheter with expandable member and tissue collection net. Radiopaque markers/tips for fluoroscopy. Sterilization: Ethylene oxide. Connectivity: None (mechanical). Standards: ISO 10555-1, ISO 594-1/2, ISO 10993-1, ISO 11607-1/2.

## Regulatory Identification

An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.

## Predicate Devices

- Infusion Aspiration Catheter System ([K143563](/device/K143563.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

Inari Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313

Re: K163549

Trade/Device Name: ClotTriever Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE Dated: February 8, 2017 Received: February 10, 2017

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando Aguel-S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K163549

Device Name ClotTriever Thrombectomy System

Indications for Use (Describe)

The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and ClotTriever Thrombectomy System is indicated for:

- · The non-surgical removal of soft thrombi and emboli from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.

The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Inari Medical. The logo consists of a purple circle with orange lines inside, followed by the word "INARI" in large, purple letters. Below "INARI" is the word "MEDICAL" in smaller, gray letters.

# 510(K) SUMMARY

| Date prepared                         | November 8, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                  | Inari Medical, Inc.<br>9272 Jeronimo Road, Suite 124<br>Irvine, CA 92618<br>949.600.8433 x114                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Contact person                        | Eben Gordon<br>Vice President, Regulatory Affairs & Quality Assurance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Trade name                            | ClotTriever Thrombectomy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common name                           | Embolectomy catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation Name                       | Embolectomy catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification number                 | 21 CFR 870.5150                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Product code                          | DXE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Regulatory class                      | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate device                      | Infusion Aspiration Catheter System (K143563)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Description                           | The ClotTriever Thrombectomy System is a single-use, sterile medical device<br>system designed for use in the peripheral vasculature.  The ClotTriever<br>Thrombectomy System consists of the ClotTriever Sheath and the ClotTriever<br>Catheter. The ClotTriever Sheath consists of a polymeric sheath equipped with a<br>self-expanding distal mesh funnel, a flush/aspiration port with tubing clamp, and a<br>proximal hemostatic valve. An obturator is provided to aid insertion. Other<br>provided accessories include a clot reservoir, a flush port adapter, an aspiration<br>insert and a 60 ml syringe. The ClotTriever Catheter consists of three (3) pre-<br>assembled polymeric coaxial catheters terminating in an expandable member and<br>tissue collection net. At the proximal end of the catheter is a handle used to<br>enable expansion of the expandable member and net. Two ports terminating in<br>stopcocks are provided for de-airing the catheter shafts. To aid in fluoroscopic<br>visualization, the ClotTriever Sheath obturator and ClotTriever Catheter distal tips<br>are radiopaque, and radiopaque marker bands are located on the intermediate shaft<br>at the proximal end of the expandable member, and at the distal ends of the<br>ClotTriever Sheath shaft and ClotTriever Catheter outer shaft. |
| Indications for Use                   | The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and<br>ClotTriever Sheath. The ClotTriever Thrombectomy System is indicated for:<br>The non-surgical removal of soft thrombi and emboli from blood vessels.<br>●                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                       | Injection, infusion, and/or aspiration of contrast media and other fluids<br>●<br>into or from a blood vessel.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                       | The ClotTriever Thrombectomy System is intended for use in the peripheral<br>vasculature.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Summary of<br>substantial equivalence | The ClotTriever Thrombectomy System and the predicate device have the same<br>intended use: removal of obstructing material (including emboli and thrombi)<br>from blood vessels.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                       | The principle of operation is the same for ClotTriever Thrombectomy System and<br>the Infusion Aspiration Catheter System - expanding nitinol structures are drawn<br>through the vessel obstruction to capture clot and restore blood flow.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                       | The ClotTriever Thrombectomy System and Infusion Aspiration Catheter System<br>have similar materials of construction.  Both use nitinol metal structures to<br>remove the obstruction and have thermoplastic polymer catheter shafts.  The<br>ClotTriever Thrombectomy System and Infusion Aspiration Catheter System<br>share the same hazard of a broken or protruding nitinol wire causing vessel<br>puncture. To mitigate this possibility, the nitinol structures are 100% inspected<br>twice for broken and protruding nitinol braids during the manufacturing process.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                       | The expanded diameter for the ClotTriever Catheter is 16 mm which is within the<br>range for the Infusion Aspiration Catheters of 12.5 mm to 22.3 mm for (models<br>10-101 and 10-103).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                       | The ClotTriever Catheter is compatible with the 13 Fr ClotTriever Sheath versus<br>the Aspiration Guide Catheter which is compatible with a 20 Fr sheath.  The<br>ClotTriever's smaller profile is inherently safer than the larger predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                       | Non-Clinical Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                       | In accordance with the Design Failure Modes and Effects Analysis, verification<br>and validation testing was identified to support the substantial equivalence of the<br>ClotTriever Thrombectomy System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                       | This testing demonstrated compliance with relevant standards (e.g. ISO 10555-1,<br>ISO 594-1/2, etc.) and product specifications.  These tests included:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                       | Pouch seal and dye penetration<br>●<br>Pouch peel, seal strength<br>●<br>Visual and dimensional inspections<br>●<br>Guidewire compatibility<br>●<br>Deployment Force of ClotTriever Sheath from Dilator<br>●<br>Retraction Force of Dilator through ClotTriever Sheath<br>●<br>Insertion of Delivery Catheter through ClotTriever Sheath<br>●<br>Deployment Force of ClotTriever from Delivery Catheter<br>●<br>Retraction Force of Handle to Expand/Collapse ClotTriever<br>●<br>Retraction Force of ClotTriever into Delivery Catheter<br>●<br>Retraction Force of ClotTriever Catheter through ClotTriever Sheath<br>●<br>ClotTriever Sheath/Dilator Kink Radius<br>●<br>ClotTriever Device Kink Radius<br>●                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

{4}------------------------------------------------

- .

{5}------------------------------------------------

- Leakage Testing, ClotTriever Sheath and Accessories ●
- . Leakage Testing, ClotTriever Sheath and Flush Port
- Leakage Testing, ClotTriever Device Hemostasis Valves .
- Air Leakage, ClotTriever Sheath
- Leakage Testing, ID of ClotTriever Device w/Guidewire in Place
- Vacuum Testing ClotTriever Sheath ●
- Test Conical Fittings with 6% Luer Taper, Dilator .
- Flow Test Through ClotTriever Sheath/Dilator ●
- Corrosion Resistance
- Simulated Use Track & Tensile, ClotTriever Sheath, Dilator, Clot ● Reservoir
- Simulated Use Track & Tensile ClotTriever Catheter
- Simulated Use Track & Turn-to-Failure, ClotTriever Sheath and ● Accessories
- . Simulated Use Track & Turn-to-Failure, ClotTriever Catheter
- Torque Testing, ClotTriever Handle ●
- Torque Testing, Clot Reservoir Luer Activated Valve ●
- Particulate Matter .

Biocompatibility testing in accordance with ISO 10993-1:

- Cvtotoxicitv
- Guinea pig maximization sensitization ●
- . Intracutaneous irritation
- Acute systemic toxicity ●
- Material mediated pyrogen ●
- Bacterial mutagenicity ●
- Hemolysis, direct contact and extract method .
- Complement activation .
- Thromboresistance .
- USP Physicochemical

The shelf life of the ClotTriever Thrombectomy System is 12 months from the date of manufacture based on accelerated aging studies. Verification testing was conducted on sterilized (ethylene oxide), accelerated-aged devices to support the 12 months shelf life.

Package integrity testing was conducted according to ISO 11607-1/2 guidelines. These tests included:

- . Pouch seal and dye penetration
- Pouch peel, seal strength .

Acute evaluation of the safety and performance of the ClotTriever Thrombectomy System was successfully performed in a bovine model.

Clinical testing was not required for the determination of substantial equivalence.

### Conclusion

Test results demonstrated that all acceptance criteria were met, and, therefore, the device conforms to established product specifications and intended use.

Based upon the technology, materials, intended use, non-clinical testing, and

{6}------------------------------------------------

animal study results, it is concluded that the ClotTriever Thrombectomy System is substantially equivalent to the Infusion Aspiration Catheter System. These results demonstrate that the ClotTriever Thrombectomy System is as safe, as effective, and performs as well as or better than the legally marketed predicate device identified above.

---

**Source:** [https://fda-staging.innolitics.com/device/K163549](https://fda-staging.innolitics.com/device/K163549)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
