K163549 · Inari Medical, Inc. · QEW · Feb 16, 2017 · Cardiovascular
Device Facts
Record ID
K163549
Device Name
ClotTriever Thrombectomy System
Applicant
Inari Medical, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Feb 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and ClotTriever Thrombectomy System is indicated for: - The non-surgical removal of soft thrombi and emboli from blood vessels. - Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel. The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature.
Device Story
ClotTriever Thrombectomy System is a single-use, sterile device for peripheral vascular thrombus removal. System comprises a ClotTriever Sheath (with self-expanding distal mesh funnel, hemostatic valve) and a ClotTriever Catheter (three coaxial shafts with expandable member and tissue collection net). Operated by physician under fluoroscopic guidance. Principle of operation: expandable nitinol structure is deployed within vessel, drawn through obstruction to capture clot, and retracted into sheath for removal. Radiopaque markers and tips aid visualization. Device restores blood flow by mechanical extraction. Benefits include non-surgical intervention for thrombus/emboli. No software or AI components.
Clinical Evidence
No clinical data. Evidence includes bench testing (verification/validation) and acute bovine animal study. Bench tests: ISO 10555-1, ISO 594-1/2, ISO 10993-1 (biocompatibility), ISO 11607-1/2 (packaging). Tests covered mechanical integrity (tensile, torque, kink radius), leakage, flow, and simulated use. Shelf life validated to 12 months via accelerated aging.
Technological Characteristics
Materials: Nitinol metal structures, thermoplastic polymer catheter shafts. Components: Sheath with self-expanding distal mesh funnel, catheter with expandable member and tissue collection net. Radiopaque markers/tips for fluoroscopy. Sterilization: Ethylene oxide. Connectivity: None (mechanical). Standards: ISO 10555-1, ISO 594-1/2, ISO 10993-1, ISO 11607-1/2.
Indications for Use
Indicated for non-surgical removal of soft thrombi and emboli from peripheral blood vessels and for injection, infusion, or aspiration of fluids/contrast media.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 16, 2017
Inari Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K163549
Trade/Device Name: ClotTriever Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE Dated: February 8, 2017 Received: February 10, 2017
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K163549
Device Name ClotTriever Thrombectomy System
Indications for Use (Describe)
The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and ClotTriever Thrombectomy System is indicated for:
- · The non-surgical removal of soft thrombi and emboli from blood vessels.
- · Injection, infusion, and/or aspiration of contrast media and other fluids into or from a blood vessel.
The ClotTriever Thrombectomy System is intended for use in the peripheral vasculature.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Inari Medical. The logo consists of a purple circle with orange lines inside, followed by the word "INARI" in large, purple letters. Below "INARI" is the word "MEDICAL" in smaller, gray letters.
# 510(K) SUMMARY
| Date prepared | November 8, 2016 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Inari Medical, Inc.<br>9272 Jeronimo Road, Suite 124<br>Irvine, CA 92618<br>949.600.8433 x114 |
| Contact person | Eben Gordon<br>Vice President, Regulatory Affairs & Quality Assurance |
| Trade name | ClotTriever Thrombectomy System |
| Common name | Embolectomy catheter |
| Regulation Name | Embolectomy catheter |
| Classification number | 21 CFR 870.5150 |
| Product code | DXE |
| Regulatory class | II |
| Predicate device | Infusion Aspiration Catheter System (K143563) |
| Description | The ClotTriever Thrombectomy System is a single-use, sterile medical device<br>system designed for use in the peripheral vasculature. The ClotTriever<br>Thrombectomy System consists of the ClotTriever Sheath and the ClotTriever<br>Catheter. The ClotTriever Sheath consists of a polymeric sheath equipped with a<br>self-expanding distal mesh funnel, a flush/aspiration port with tubing clamp, and a<br>proximal hemostatic valve. An obturator is provided to aid insertion. Other<br>provided accessories include a clot reservoir, a flush port adapter, an aspiration<br>insert and a 60 ml syringe. The ClotTriever Catheter consists of three (3) pre-<br>assembled polymeric coaxial catheters terminating in an expandable member and<br>tissue collection net. At the proximal end of the catheter is a handle used to<br>enable expansion of the expandable member and net. Two ports terminating in<br>stopcocks are provided for de-airing the catheter shafts. To aid in fluoroscopic<br>visualization, the ClotTriever Sheath obturator and ClotTriever Catheter distal tips<br>are radiopaque, and radiopaque marker bands are located on the intermediate shaft<br>at the proximal end of the expandable member, and at the distal ends of the<br>ClotTriever Sheath shaft and ClotTriever Catheter outer shaft. |
| Indications for Use | The ClotTriever Thrombectomy System consists of the ClotTriever Catheter and<br>ClotTriever Sheath. The ClotTriever Thrombectomy System is indicated for:<br>The non-surgical removal of soft thrombi and emboli from blood vessels.<br>● |
| | Injection, infusion, and/or aspiration of contrast media and other fluids<br>●<br>into or from a blood vessel. |
| | The ClotTriever Thrombectomy System is intended for use in the peripheral<br>vasculature. |
| Summary of<br>substantial equivalence | The ClotTriever Thrombectomy System and the predicate device have the same<br>intended use: removal of obstructing material (including emboli and thrombi)<br>from blood vessels. |
| | The principle of operation is the same for ClotTriever Thrombectomy System and<br>the Infusion Aspiration Catheter System - expanding nitinol structures are drawn<br>through the vessel obstruction to capture clot and restore blood flow. |
| | The ClotTriever Thrombectomy System and Infusion Aspiration Catheter System<br>have similar materials of construction. Both use nitinol metal structures to<br>remove the obstruction and have thermoplastic polymer catheter shafts. The<br>ClotTriever Thrombectomy System and Infusion Aspiration Catheter System<br>share the same hazard of a broken or protruding nitinol wire causing vessel<br>puncture. To mitigate this possibility, the nitinol structures are 100% inspected<br>twice for broken and protruding nitinol braids during the manufacturing process. |
| | The expanded diameter for the ClotTriever Catheter is 16 mm which is within the<br>range for the Infusion Aspiration Catheters of 12.5 mm to 22.3 mm for (models<br>10-101 and 10-103). |
| | The ClotTriever Catheter is compatible with the 13 Fr ClotTriever Sheath versus<br>the Aspiration Guide Catheter which is compatible with a 20 Fr sheath. The<br>ClotTriever's smaller profile is inherently safer than the larger predicate device. |
| | Non-Clinical Testing |
| | In accordance with the Design Failure Modes and Effects Analysis, verification<br>and validation testing was identified to support the substantial equivalence of the<br>ClotTriever Thrombectomy System. |
| | This testing demonstrated compliance with relevant standards (e.g. ISO 10555-1,<br>ISO 594-1/2, etc.) and product specifications. These tests included: |
| | Pouch seal and dye penetration<br>●<br>Pouch peel, seal strength<br>●<br>Visual and dimensional inspections<br>●<br>Guidewire compatibility<br>●<br>Deployment Force of ClotTriever Sheath from Dilator<br>●<br>Retraction Force of Dilator through ClotTriever Sheath<br>●<br>Insertion of Delivery Catheter through ClotTriever Sheath<br>●<br>Deployment Force of ClotTriever from Delivery Catheter<br>●<br>Retraction Force of Handle to Expand/Collapse ClotTriever<br>●<br>Retraction Force of ClotTriever into Delivery Catheter<br>●<br>Retraction Force of ClotTriever Catheter through ClotTriever Sheath<br>●<br>ClotTriever Sheath/Dilator Kink Radius<br>●<br>ClotTriever Device Kink Radius<br>● |
{4}------------------------------------------------
- .
{5}------------------------------------------------
- Leakage Testing, ClotTriever Sheath and Accessories ●
- . Leakage Testing, ClotTriever Sheath and Flush Port
- Leakage Testing, ClotTriever Device Hemostasis Valves .
- Air Leakage, ClotTriever Sheath
- Leakage Testing, ID of ClotTriever Device w/Guidewire in Place
- Vacuum Testing ClotTriever Sheath ●
- Test Conical Fittings with 6% Luer Taper, Dilator .
- Flow Test Through ClotTriever Sheath/Dilator ●
- Corrosion Resistance
- Simulated Use Track & Tensile, ClotTriever Sheath, Dilator, Clot ● Reservoir
- Simulated Use Track & Tensile ClotTriever Catheter
- Simulated Use Track & Turn-to-Failure, ClotTriever Sheath and ● Accessories
- . Simulated Use Track & Turn-to-Failure, ClotTriever Catheter
- Torque Testing, ClotTriever Handle ●
- Torque Testing, Clot Reservoir Luer Activated Valve ●
- Particulate Matter .
Biocompatibility testing in accordance with ISO 10993-1:
- Cvtotoxicitv
- Guinea pig maximization sensitization ●
- . Intracutaneous irritation
- Acute systemic toxicity ●
- Material mediated pyrogen ●
- Bacterial mutagenicity ●
- Hemolysis, direct contact and extract method .
- Complement activation .
- Thromboresistance .
- USP Physicochemical
The shelf life of the ClotTriever Thrombectomy System is 12 months from the date of manufacture based on accelerated aging studies. Verification testing was conducted on sterilized (ethylene oxide), accelerated-aged devices to support the 12 months shelf life.
Package integrity testing was conducted according to ISO 11607-1/2 guidelines. These tests included:
- . Pouch seal and dye penetration
- Pouch peel, seal strength .
Acute evaluation of the safety and performance of the ClotTriever Thrombectomy System was successfully performed in a bovine model.
Clinical testing was not required for the determination of substantial equivalence.
### Conclusion
Test results demonstrated that all acceptance criteria were met, and, therefore, the device conforms to established product specifications and intended use.
Based upon the technology, materials, intended use, non-clinical testing, and
{6}------------------------------------------------
animal study results, it is concluded that the ClotTriever Thrombectomy System is substantially equivalent to the Infusion Aspiration Catheter System. These results demonstrate that the ClotTriever Thrombectomy System is as safe, as effective, and performs as well as or better than the legally marketed predicate device identified above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.