K163531 · Terumo Cardiovascular Systems Corporation · DTQ · Jun 7, 2017 · Cardiovascular
Device Facts
Record ID
K163531
Device Name
Advanced Perfusion System 1
Applicant
Terumo Cardiovascular Systems Corporation
Product Code
DTQ · Cardiovascular
Decision Date
Jun 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4220
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment. The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.
Device Story
Advanced Perfusion System 1 is a configurable heart-lung machine with distributed network architecture. System components include a base chassis with power/battery backup, Central Control Monitor (CCM) for configuration/control, roller pumps, centrifugal control units, and various pods (Air Bubble Detection, Level Detection, Pressure, Temperature, Flowmeter, Venous Line Occluder, Interface Pods). Used in clinical settings (e.g., OR) by perfusionists or qualified medical professionals. System monitors extracorporeal circuit parameters; provides automated control of venous line clamping and gas blending. CCM displays system status and alerts. Modifications include a power system daughterboard for diagnostics and updated CCM service messages/battery status displays. Benefits include customized perfusion configurations and enhanced power system monitoring to support safe operation during bypass procedures.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing confirmed performance within design requirements for power system diagnostics, battery backup switching, and status monitoring.
Technological Characteristics
Configurable heart-lung system; distributed network architecture. Components: roller/centrifugal pumps, pods for sensing (pressure, flow, temp, air, level), electronic gas blender. Power: 100/120V or 220/240V with battery backup. Connectivity: Networked pods and CCM. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for patients requiring extracorporeal circulation of blood for arterial perfusion, regional perfusion, or cardiopulmonary bypass procedures for up to 6 hours. Must be operated by a qualified medical professional experienced in such equipment.
Regulatory Classification
Identification
A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/13 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2017
Terumo Cardiovascular Systems Corporation Bryan Hann Senior Engineer-regulatory Affairs 6200 Jackson Road Ann Arbor, Michigan 48103
Re: K163531
Trade/Device Name: Advanced Perfusion System 1 Regulation Number: 21 CFR 870.4220 Regulation Name: Cardiopulmonary Bypass Heart-Lung Machine Console Regulatory Class: Class II Product Code: DTQ Dated: May 5, 2017 Received: Mav 8. 2017
Dear Bryan Hann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Aguel-S
Bram D. Zuckerman, M.D. for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163531
Device Name Advanced Perfusion System 1
Indications for Use (Describe)
The Terumo® Advanced Perfusion System 1 is indicated for use for up to 6 hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Attachment 1
#### Terumo Cardiovascular Systems
Advanced Perfusion System 1 Traditional 510(k)
Section 4: 510(k) Summary
### Premarket Notification K163531
Terumo® Advanced Perfusion System 1 510(k) Summary This section is prepared in accordance with 21 CFR Part 807.92
## General Information
| Submitter's Name and Address | Terumo Cardiovascular Systems Corporation<br>6200 Jackson Road<br>Ann Arbor MI, 48103 |
|------------------------------|---------------------------------------------------------------------------------------|
| Name of Contact Person | Bryan Hann |
| Phone number | Tel: (734) 741-5816 |
| Fax number | Fax: (734) 741-6069 |
| email | bryan.hann@terumomedical.com |
| Establishment Registration # | 1828100 |
| Date prepared | May 3, 2017 |
## Device Information
| Trade Name | Terumo® Advanced Perfusion System 1 |
|----------------------|--------------------------------------------------------|
| Common/Genric Name | Heart-Lung Machine |
| Classification name | Console, Heart-Lung Machine,<br>Cardiopulmonary Bypass |
| Classification panel | 74 Cardiovascular |
| Device Class | Class II |
| Regulation Number | 21 CFR §870.4220 |
| Product Code(s) | DTQ |
## Predicate Device Information
The modified Terumo® Advanced Perfusion System 1 is substantially equivalent in function and intended use to the Terumo® Advanced Perfusion System 1 cleared in Premarket Notification K153376.
Image /page/3/Picture/14 description: The image shows the Terumo logo. The logo consists of a red swoosh above the word "TERUMO" in green. The font is sans-serif and bolded.
{4}------------------------------------------------
Advanced Perfusion System 1 Traditional 510(k)
## Indications for Use and Device Description
#### Indications for Use
The Terumo® Advanced Perfusion System 1 is indicated for use for up to six hours in the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
The centrifugal pump is indicated for use in cardiopulmonary bypass procedures only.
#### Device Description
The Advanced Perfusion System 1) is a configurable heart-lung system with a distributed network architecture that allows the user to customize the number and types of system components, which can then be configured, displayed, and controlled from a central monitor. The system is designed to enable users to choose from the Terumo CVS supplied components to define and configure a heart-lung system to meet individual institution requirements.
The System 1 components are listed below.
- · System 1 Base:
- Chassis platform Provides operating power and back up battery power for all o system components (100/120V or 220/240V)
- Central Control Monitor (CCM) A touch screen display used for o configuration and control of system components
- o Two roller pump hand cranks and hand crank bracket
- Pump(s) and pump mounting hardware Up to eight pumps can be used with System 1, including the following:
- 6'' Roller Pump O
- o 4" Roller Pump
- Centrifugal Control Unit with Centrifugal Drive Motor (up to 2) O
- Pods
- Air Bubble Detection Pod Used to detect air bubbles in the extracorporeal O circuit, in conjunction with the air sensor
- o Level Detection Pod Used to monitor liquid levels within a hard shell reservoir.
- O Pressure Pod - Used to monitor the pressure in the extracorporeal circuit
- Temperature Pod Used to monitor the temperature in the extracorporeal O circuit and / or the patient
- Flowmeter Pod Used to monitor flow volume and generate an alarm if O backflow is detected
Image /page/4/Picture/28 description: The image shows the word "TERUMO" in green, with a red angled line above the word. The font is sans-serif and bold. The red line is angled upwards from left to right.
{5}------------------------------------------------
#### Section 4: 510(k) Summary
### Premarket Notification K163531
- Venous Line Occluder Pod Used with the Occluder Head to provide a o computer controlled tube clamping mechanism to regulate flow in the venous line
- Interface Pods to enable data transfer between cardiac monitoring and data o display systems (i.e., Terumo CDI™ 500 Monitor, and TLink™ Data Management System)
- · Electronic Gas Blender Provides control and monitoring of the gas output to the oxygenator
- Flexible Lamps (15 inch or 33 inch) for local illumination
- · Mounting hardware (e.g., center poles, crossbars, and brackets)
### Device Illustration
Image /page/5/Figure/12 description: This image shows a labeled diagram of a Centrifugal Control Unit. The diagram labels the Centrifugal Control Unit Local Display and Control and Centrifugal Motor, Small Roller Pump, Lamp, Large Roller Pump, Roller Pump Local Display and Control, Central Control Monitor, Occluder, Electronic Gas Blender Local Controls, Wheel Lock, Power Status Indicator LED, Hand Cranks, and On/Off Switch. The diagram provides a detailed overview of the different components and their locations within the Centrifugal Control Unit.
Image /page/5/Picture/13 description: The image shows the Terumo company logo. The logo features a red swoosh above the company name, which is written in green, block letters. The font is sans-serif and appears to be bolded.
{6}------------------------------------------------
# Comparison of Technological Characteristics to Predicate Device
Extracorporeal circulation of blood is the technological principle for both the subject and predicate System 1 devices. This is based on the use of centrifugal pumps as well as roller pumps for the circulation of blood in arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures. The subject and predicate devices are based on the following technological principles:
- · Central Control Monitor used for configuration and control of system components
- · Centrifugal and roller type pumps used for extracorporeal circulation of blood
- · The use of pods to connect peripheral components and devices for safety, monitoring, and data transfer
- · Electronic gas blender for control and monitoring of the gas output to the oxygenator
The following technological differences exist between the subject and predicate devices:
- · The addition of a daughterboard to the Power System to support software modifications that enable power system diagnostics that were not previously available
- · The addition of service messages to the Central Control Monitor to inform users for backup battery maintenance
- · Central Control Monitor update to the Power Status Display (battery icon) to provide additional information related to the power status of the battery backup system.
Image /page/6/Picture/12 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the company name, which is written in green, block letters. The red swoosh is curved and positioned above the left side of the word "TERUMO."
{7}------------------------------------------------
Advanced Perfusion System 1 Traditional 510(k)
### Section 4: 510(k) Summary
# Performance Data
### Biocompatibility Testing
The Terumo® System 1 does not come into contact with the patient. Therefore, Terumo CVS is not submitting biocompatibility testing as part of this premarket notification.
### Electrical Safety and Electromagnetic Compatibility (EMC)
Evaluation of the hardware modification for electrical safety and EMC testing was conducted on the System 1 with the modifications described in this premarket notification. The System 1 complies with the IEC 60601-1 standard for electrical safety and the IEC 60601-1-2 standard for EMC.
### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.
### Design Verification and Validation Testing
Design Verification and Validation testing were conducted and have resulted in data that demonstrates that the System 1 performs within the defined design input requirements for the proposed modifications. The following test characteristics were performed on the subject device:
- Battery Backup supply voltage and calculated charge .
- Battery Backup Health signal ●
- Battery Capacity ●
- AC Mains to Battery Backup power switching ●
- Battery Backup function status diagnostic ●
- . Simulated use case testing
## Animal Study
Animal testing was not required to demonstrate the substantial equivalence of the System 1 to the predicate device. Therefore, Terumo CVS is not submitting animal performance testing as part of this premarket notification.
## Clinical Studies
Clinical testing was not required to demonstrate the substantial equivalence of the System 1 to the predicate device. Therefore, Terumo CVS is not submitting clinical performance testing as part of this premarket notification.
Image /page/7/Picture/26 description: The image shows the logo for Terumo. The logo consists of a red swoosh above the word "TERUMO" in green, block letters. The swoosh is positioned above and to the left of the word.
{8}------------------------------------------------
Advanced Perfusion System 1 Traditional 510(k)
Section 4: 510(k) Summary
Premarket Notification K163531
### Conclusion
The hardware modification as well as the software modifications to the System 1 have not changed the device indications for use, technical characteristics, or fundamental scientific technology. Software verification and validation, electrical safety, electromagnetic compatibility, and design verification and validation testing have confirmed that the modifications do not adversely impact system performance. The modified System 1 is substantially equivalent to the currently marketed device cleared under K153376.
Image /page/8/Picture/8 description: The image shows the logo for Terumo. The logo consists of the word "TERUMO" in green, with a red swoosh above the letters "TE". The font is bold and sans-serif. The logo is simple and modern.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.