The PrisMax control unit is intended for: · Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload. All treatments administered via the PrisMax control unit must be prescribed by a physician.
Device Story
PrisMax is a hemodialysis delivery system for Continuous Renal Replacement Therapy (CRRT) in clinical settings. It processes blood and fluids to remove waste, restore acid-base balance, correct electrolyte imbalances, and manage fluid removal. The system utilizes a control unit, pumps (blood, effluent, dialysate, replacement, pre-blood pump), and scales to manage therapy. It supports SCUF, CVVH, CVVHD, and CVVHDF modalities. A physician prescribes the treatment; a clinician operates the device. The system monitors pressures (access, return, filter, effluent) and provides alarms. It improves upon the predicate Prismaflex by offering finer flow rate increments (10 ml/hr for dialysate/replacement, 5 ml/hr for fluid removal) and simplified setup (fixed TMP default). Benefits include precise therapy control and reduced risk of use-error.
Clinical Evidence
Bench testing only. Performance testing confirmed functional equivalence to the predicate. Testing included compliance with IEC 60601-2-16 (hemodialysis equipment), IEC 60601-1 Edition 3.1 (electrical safety), and EMC testing. Design validation, human factors, and software validation were performed to confirm the device meets user needs and safety requirements.
Technological Characteristics
Hemodialysis delivery system; includes control unit, pumps, and scales. Connectivity: Standalone. Software: Version 1.0.6.0. Standards: IEC 60601-1 Ed 3.1, IEC 60601-2-16. Sterilization: Not specified. Sensing: Pressure sensors (access, return, filter, effluent). Energy: Electrical.
Indications for Use
Indicated for Continuous Renal Replacement Therapy (CRRT) in patients weighing 20 kg or more with acute renal failure and/or fluid overload. Must be prescribed by a physician.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, all facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 3, 2017
Baxter Healthcare Corporation Chris Scavotto Senior Manager Regulatory Affairs 32650 N. Wilson Road Round Lake, IL 60073
K163530 Re:
> Trade/Device Name: PrisMax Regulation Number: 21 CFR§ 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI Dated: March 31, 2017 Received: April 4, 2017
Dear Chris Scavotto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K163530
Device Name PrisMax
Indications for Use (Describe)
The PrisMax control unit is intended for:
· Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.
All treatments administered via the PrisMax control unit must be prescribed by a physician.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 5. 510(k) Summary
December 15th, 2016
#### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
### CONTACT PERSON:
Chris Scavotto Senior Regulatory Manager, Global Regulatory Affairs 32650 N Wilson Road Round Lake, IL 60073 Telephone: 763-463-4621 Fax: 763-463-4606
### IDENTIFICATION OF THE DEVICE:
Common Name: Hemodialysis Delivery System Trade Name or Proprietary Name: PrisMax Classification Panel: 78 Gastroenterology and Urology Classification: High Permeability Hemodialysis System (876.5860) Class: Class II Product Code: 78KDI
| Table 1. Product Code(s) for PrisMax System | | | |
|---------------------------------------------|--|--|--|
|---------------------------------------------|--|--|--|
| Code Number | Name |
|-------------|--------------------------------------------|
| 115369 | PrisMax control unit |
| 115370 | PrisMax Accessory, Auto Effluent |
| 115485 | PrisMax Accessory, Auto Effluent Extension |
### PREDICATE DEVICE:
#### Table 2. Predicate Device(s)
| Device | Company | Predicate 510(k) | Clearance Date |
|------------|------------------|------------------|----------------|
| Prismaflex | Gambro Lundia AB | K131516 | 1/3/2014 |
{4}------------------------------------------------
## DESCRIPTION OF THE DEVICE:
The PrisMax System is intended for Continuous Renal Replacement Therapy (CRRT) for patients with acute renal failure and/or fluid overload. Reference the PrisMax control unit in Figure 1.
The goals of acute renal failure treatments are removal of waste products, restoration of acid-base balance; correction of electrolyte imbalances (e.g., hyperkalemia), patient fluid balance, nutritional support, and other conditions in which fluid removal is needed. PrisMax System offers four Continuous Renal Replacement Therapy (CRRT) options: Slow Continuous Ultrafiltration (SCUF), Continuous Veno-Venous Hemofiltration (CVVH), Continuous VenoVenous Hemodialysis (CVVHD), and Continuous Venovenous Hemodialfiltration (CVVHDF).
Image /page/4/Picture/5 description: The image shows a dialysis machine with a screen on top. The screen displays various parameters and readings related to the dialysis process. The machine has several tubes, pumps, and filters that are essential for the dialysis procedure. There are also bags hanging from the machine, which likely contain fluids used during the treatment.
Figure 1. PrisMax control unit
The proposed device PrisMax, which is the subject of this Traditional 510(k) premarket notification, consists of PrisMax Control Unit. Also included are accessories for removing effluent. The proposed device PrisMax uses the current marketed device
{5}------------------------------------------------
Prismaflex as the predicate. The Prismaflex was previously cleared under 510(k) premarket notification K131516 (cleared on January 3rd, 2014).
## INDICATIONS FOR USE:
The PrisMax control unit is intended for:
- . Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.
All treatments administered via the PrisMax control unit must be prescribed by a physician.
## TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The differences between the predicate Prismaflex 7.10 and the proposed device PrisMax as reflected in the Substantial Equivalence (SE) Table are considered substantially equivalent. Changes included in the SE table are marked and discussed in Table 3.
Identification of Predicate Device:
- Trade Name: Prismaflex, 7.10
- Manufacturer: Gambro Lundia AB, Sweden
- Common Name: Hemodialysis Delivery System.
- Clearance: K0131516 cleared in January 2014.
Identification of Proposed Device:
- Trade Name: PrisMax, 1.0.6.0.
- Manufacturer: Gambro UF Solutions, Minneapolis Minnesota
- Common Name: Hemodialysis Delivery System.
- Clearance: Pending.
PrisMax will not utilize the TPE2000 Blood Tubing Set. It will utilize the CRRT Blood Tubing sets referenced in the Substantial Equivalence Table (M60/M100/M150 + HF1000/HF1400) all which have been previously cleared. The predicate device Prismaflex was cleared in 2011 to include TPE2000 Blood Sets. The PrisMax is not
{6}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
requesting clearance for TPE Blood Sets. References to TPE have been removed from Table 3.
| Features | SE | Proposed<br>PrisMax 1.0.6.0 | Predicate<br>Prismaflex 7.10, K131516 |
|-----------------------------------------------------|----|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | SE | The PrisMax control unit is intended for:<br>• Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.<br><br>All treatments administered via the PrisMax control unit must be prescribed by a physician. | The Prismaflex control unit is intended for:<br>• Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.<br><br>All treatments administered via the Prismaflex control unit must be prescribed by a physician. |
| Dedicated Disposable Sets Available in U.S. | SE | For CRRT:<br>M60/M100/M150<br>HF1000 & HF1400 | For CRRT:<br>M60/M100/M150<br>HF1000 & HF1400 |
| Syringe Sizes A | SE | 20 & 50 ml | 20, 30 & 50 ml |
| Anticoagulation | SE | User-controllable as continuous or bolus | User-controllable as continuous or bolus |
| Dialysate Flow RateB | SE | CVVH & CVVHDF:<br>Range: 0 to 8000 ml/hr<br>Increment: 10 ml/hr | CVVH & CVVHDF:<br>Range: 0 to 8000 ml/hr<br>Increment: 50 ml/hr |
| Dialysate Flow Rate Accuracy | SE | ± 30 ml/hr | ± 30 ml/hr |
| Replacement solution / Fluid Flow RateC | SE | CVVH & CVVHDF:<br>Range: 0 to 8000 ml/hr<br>Increment: 10 ml/hr | CVVH & CVVHDF:<br>Range: 0 to 8000 ml/hr<br>Increment: 50 ml/hr |
| Replacement Flow Rate Accuracy | SE | ± 30 ml/hr | ± 30 ml/hr |
| Blood Flow Rate | SE | Range: 10-450 ml/min | Range: 10-450 ml/min |
| Blood Flow Rate Accuracy | SE | ±10 % of user-set rate at nominal blood flow of 450 ml/min or the highest achievable disposable blood flow, having 37 °C, at an access pressure of -200 mmHg and without any PBP flow | ±10 % of user-set rate at nominal blood flow of 450 ml/min or the highest achievable disposable blood flow, having 37 °C, at an access pressure of -200 mmHg and without any PBP flow |
| | | Proposed | Predicate |
| Features | SE | PrisMax 1.0.6.0 | Prismaflex 7.10, K131516 |
| Pre-Blood Pump<br>Flow RateD | SE | SCUF:<br>Range: 0 to 2000 ml/hr | SCUF:<br>Range: 0 to 1000 ml/hr |
| | SE | CVVH, CVVHD, CVVHDF:<br>Range: 0 to 4000 ml/hr | CVVH, CVVHD, CVVHDF:<br>Range: 0 to 4000 ml/hr |
| Pre-Blood Pump<br>Accuracy | SE | ± 30 ml/hr | ± 30 ml/hr |
| Effluent Pump<br>Flow Rate | SE | Range: 0 to 10,000 ml/h<br>Depending on the therapy selected. | Range: 0 to 10,000 ml/h<br>Depending on the therapy selected. |
| ECG Discharger | SE | YES | YES |
| Therapies | SE | SCUF<br>CVVH<br>CVVHD<br>CVVHDF | SCUF<br>CVVH<br>CVVHD<br>CVVHDF |
| Pumps | SE | PBP solution<br>Replacement solution<br>Dialysate solution<br>Effluent<br>Blood | PBP solution<br>Replacement solution<br>Dialysate solution<br>Effluent<br>Blood |
| Scales | SE | Dialysate<br>Replacement<br>Effluent<br>Pre blood (PBP) | Dialysate<br>Replacement<br>Effluent<br>Pre blood (PBP) |
| Trans Membrane<br>Pressure Alarms<br>TMP (CRRT)E | SE | CRRT TMP:<br>Default: +300 mmHg | CRRT TMP:<br>User settable: +70 to +350 mmHg<br>Default: +350 mmHg |
| Dialysate<br>Conductivity and<br>Temperature | SE | Dialysate Conductivity and<br>Temperature are not controlled by<br>PrisMax | Dialysate Conductivity and<br>Temperature are not controlled by<br>Prismaflex |
| Patient FluidF<br>Removal<br>Performance<br>(Range) | SE | 0 to 2000 ml/hr<br>Increment: 5 ml/hr | 0 to 2000 ml/hr<br>Increment: 10 ml/hr |
## Table 3. Substantial Equivalence Table Device Comparison
{7}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
{8}------------------------------------------------
| | | Proposed | Predicate |
|-------------------------------------------------------|----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features | SE | PrisMax 1.0.6.0 | Prismaflex 7.10, K131516 |
| Patient Fluid<br>Removal<br>Performance<br>(Accuracy) | SE | ± 30 ml/hr<br>± 70 ml/3hr<br>± 300 ml/24hr<br>Scales calibrated at ambient<br>temperature at which they will be<br>used. Ambient temperature change<br>less than ±3 ℃<br>(5.4 °F) during treatment. | ± 30 ml/hr<br>± 70 ml/3hr<br>± 300 ml/24hr<br>Scales calibrated at ambient<br>temperature at which they will be<br>used. Ambient temperature change<br>less than ±3 °C<br>(5.4 °F) during treatment. |
| Access Pressure<br>Sensor | SE | Range:<br>-250 to +450 mmHg<br>Accuracy:<br>±15 mmHg | Range:<br>-250 to +450 mmHg<br>Accuracy:<br>±15 mmHg |
| Return Pressure<br>Sensor | SE | Range:<br>-50 to +350 mmHg<br>Accuracy:<br>±5 mmHg | Range:<br>-50 to +350 mmHg<br>Accuracy:<br>±5 mmHg |
| Filter Pressure<br>Sensor | SE | Range:<br>-50 to +450 mmHg<br>Accuracy:<br>±15 mmHg | Range:<br>-50 to +450 mmHg<br>Accuracy:<br>±15 mmHg |
| Effluent Pressure<br>Sensor | SE | Range:<br>-350 to +400 mmHg (CRRT)<br>Accuracy:<br>±15 mmHg | Range:<br>-350 to +400 mmHg (CRRT)<br>Accuracy:<br>±15 mmHg |
| Control Unit<br>Software G | SE | Version 1.0.6.0 | Version 7.10 |
### Table 3: Foot notes:
A. Syringe Sizes: 20, 30 & 50 ml. Removed 30 ml.
- . The 30ml syringe is removed due to infrequent use in the market. The device behavior in conjunction with the specified syringes remains strictly unchanged.
- . The updated specification does not affect safety or effectiveness of the device.
{9}------------------------------------------------
B. Dialysate Flow Rate: CVVH & CVVHDF: Range: 0 to 8000 ml/hr, Increment: 10 ml/hr. Updated the increment from 50ml/hr to 10ml/hr.
- . The setting range was increased from 50 to 10ml/hr. This allows the practitioner to select a more precise setting and reduces limitation existing with a 50ml/hr increment.
The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.
C. Replacement solution / Fluid Flow Rate: CVVH & CVVHDF: Range: 0 to 8000 ml/hr, Increment: 10 ml/hr. Updated the increment from 50ml/hr to 10ml/hr.
- . The range was increased from 50 to 10ml/hr. This allows the practitioner to select a more precise setting and reduces limitation existing with a 50ml/hr increment.
The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.
D. Pre-Blood Pump Flow Rate: SCUF: Range: 0 to 2000 ml/hr. Align with Predicate device Prismaflex from clearances K110823 and K131516.
- . The Prismaflex SE table incorrectly states a range of 0 to 1000 ml/hr vs. the actual of 0 to 2000 ml/hr. As per Prismaflex requirement sss 1283, the predicate device has operated at the Range: 0 to 2000 ml/hr since the original clearance in 2005. The PrisMax is equivalent in practice and requirements. The Prismaflex SE table will be updated on the next submission.
The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.
{10}------------------------------------------------
E. Trans Membrane Pressure Alarms TMP (CRRT): User settable: +70 to +350 mmHg, default +350 mmHg. Implemented fixed default of +300 mmHg.
- . User Settable TMP was removed due to infrequent use in the predicate device and to reduce risk of use-error handling during set up.
- Increase usability by removing steps during set up.
- Implement lowered default to reduce risk.
The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of the PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.
F. Patient Fluid Removal Performance (Range): 0 to 2000 ml/hr Increment: 5 ml/hr.
Updated the increment from 10ml/hr to 5ml/hr.
- . Increase setting range by improving increment range of setting from 10 to 5ml/hr. This allows the practitioner to select a more precise setting and reduces limitation existing with a 10ml/hr increment.
The updated specification does not introduce new or increased risks to the system and does not introduce new questions of safety and effectiveness. Compliance of PrisMax System to the updated specifications has been verified successfully. Successful validation has substantiated that the system does not raise new questions of safety and effectiveness.
## G. Control Unit Software: Software version 1.0.6.0 or Rev. A, from Prismaflex 7.10.
The device software on the PrisMax has been implemented correctly. The software has been verified and validated subsequent to risk analysis. The verification and validation tests including Human Factors and Software Validation. The software does not raise questions of safety and effectiveness. PrisMax is considered substantially equivalent to the predicate device Prismaflex.
## Substantial Equivalence Summary
The differences between the PrisMax System and its predicate device do not introduce new questions of safety and effectiveness. All modifications have been verified and
{11}------------------------------------------------
validated per Design Controls Activities. As shown through successful verification and validation testing, the PrisMax Control Unit is considered substantially equivalent to its predicate.
## DISCUSSION OF NONCLINICAL TESTS:
Performance testing was conducted on the PrisMax System to evaluate the functional performance of the system. The performance testing confirms PrisMax remains as safe and effective as Prismaflex and is substantially equivalent.
In summary, the PrisMax Control Unit has successfully implemented performance requirements and subsequent outputs verifying and validating:
- . The PrisMax design validation meets the user needs and intended use and is substantially equivalent to the predicate.
- . The device complies with IEC60601-2-16 Hemodialysis Equipment. Testing was confirmed by CSA, the recognized test laboratory as part of the testing to 60601-1 Edition3.1. The testing specifically confirms the device meets the requirements for Essential Performance according to the particular standard.
- . Electrical safety testing according to the most recent IEC60601-1 Edition 3.1 standard. The standard includes reports for software, alarms, usability, safety and performance.
- Electromagnetic compatibility (EMC) testing.
- Risk Assessment and risk control measures. A therapy level, product level and process level hazard analysis confirms the device doesn't perform in an unexpected or unsafe manor.
- Labeling, Software including cybersecurity, Human Factors, have been successfully implemented.
# SUBSTANTIAL EQUIVALENCE DECISION
Based on the the information provided in this premarket notification, Baxter Healthcare Corporation believes the proposed PrisMax is substantially equivalent, for purposes of section 510(k) of the Federal Food, Drug and Cosmetic Act only, to the predicate device identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.