K163526 · Neomedlight · LBI · Oct 6, 2017 · General Hospital
Device Facts
Record ID
K163526
Device Name
Bilicocoon™ Phototherapy System
Applicant
Neomedlight
Product Code
LBI · General Hospital
Decision Date
Oct 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The BiliCocoon™ Phototherapy System is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates and infants under 3 months old and weighing less than 10kg. It can be used in the clinical setting or in the home.
Device Story
BiliCocoon™ Phototherapy System treats unconjugated hyperbilirubinemia in neonates/infants (<3 months, <10kg). System comprises blue light electronic generator (Light Box) and light-emitting fabric (Pad). Light Box directs 430-490nm light into fiberoptic Pad; Pad transmits light to patient. Disposable non-woven cover interfaces with patient skin. Used in clinical or home settings; operated by clinicians or caregivers. User interface on Light Box allows session timer setup; other parameters (irradiance, wavelength) are fixed. Device provides phototherapy to transform bilirubin into water-soluble isomers for elimination. Benefits include effective bilirubin reduction via AAP-recommended blue light therapy; larger treatment surface area compared to predicate; potential for above/underneath illumination.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993-5, -10), spectral output, irradiance, effective surface area, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), phototherapy performance (IEC 60601-2-50), home use safety (IEC 60601-1-11), and photobiological safety (IEC 62471).
Technological Characteristics
Blue LED phototherapy (430-490nm); irradiance 30-40 μW/cm²/nm. Components: Light Box generator, fiberoptic Pad (Nest/Bag versions), non-woven polyester-viscose disposable cover. Class II electrical safety (double insulation). Dimensions: 21.5x19.8x16.0 cm; weight 1.4 kg. Connectivity: None (standalone). Sterilization: Not specified (cleaning validation performed).
Indications for Use
Indicated for treatment of unconjugated hyperbilirubinemia in neonates and infants <3 months old and <10kg. Used in clinical or home settings.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
Predicate Devices
neoBLUE® blanket LED Phototherapy System (K103589)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an eagle-like design with three stylized human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2017
Neomedlight Pierre Girons CEO 89-90 Rue Frederic Fays Villeurbanne, 69100 FRANCE
Re: K163526
Trade/Device Name: Bilicocoon™ Phototherapy System Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: Class II Product Code: LBI Dated: August 29, 2017 Received: September 1, 2017
Dear Pierre Girons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image contains a signature and a name. The name is "Tina Kiang-S". The signature is a complex series of loops and lines to the left of the name. The text is black and the background is white.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163526
Device Name BiliCocoon™ Phototherapy System
Indications for Use (Describe)
The BiliCocoon™ Phototherapy System is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates and infants under 3 months old and weighing less than 10kg. It can be used in the clinical setting or in the home
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K163526 510(K) SUMMARY
| Manufacturer's Name: | NEOMEDLIGHT<br>89-90 Rue Frederic FAYS<br>69100 Villeurbanne<br>France<br>Ph.: +33 6 33 902 900 |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Corresponding Official: | Pierre SAINT-GIRONS<br>CEO |
| Telephone Number: | +33 6 33 902 900 |
| E-mail: | pierre.saintgirons@neomedlight.com |
| Preparation Date: | October 3, 2017 |
| Trade Name: | BiliCocoon™ Phototherapy System |
| Common or Usual Name: | Neonatal phototherapy unit |
| Regulation Name:<br>Regulation Number:<br>Product Code: | Unit, Neonatal Phototherapy<br>21 CFR 880.5700<br>LBI : unit, neonatal phototherapy<br>PDH : blanket, neonatal phototherapy |
| Device Class: | Class II |
| Primary Predicate Device: | K103589, neoBLUE® blanket LED Phototherapy System |
#### Device Description
The BiliCocoon™ Phototherapy System is a phototherapy system designed to treat unconjugated hyperbilirubinemia in newborns and infants under 3 months old and weighing less than 10kg. It is a phototherapy device, which emits light in the absorption spectrum of the bilirubin, from 430 to 490 nm. A custom optic directs light from the LEDs into the fiberoptic Pad. The system is composed of a blue light electronic generator - the Light Box - and of a light emitting fabric – the Pad – which transmits the blue light to the newborn. The Pad is provided in two versions: Bag Pad. The Light Box has a user interface to set the session time and inform the user about eventual errors. The other parameters of the therapy (irradiance level, light wavelength, etc.) are fixed.
The BiliCocoon™ Phototherapy System is used with the BiliCocoon™ Disposable, a non-woven disposable designed to fit the Pad which interfaces between the newborn. The disposable is the only part which contacts the newborn skin.
The BiliCocoon™ Phototherapy System can be fixed on the BiliCocoon™ Fixation system.
{4}------------------------------------------------
#### Indications for Use
The BiliCocoon™ Phototherapy System is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates and infants under 3 months old and weighing less than 10kg. It can be used in the clinical setting or in the home.
The device is prescription only.
#### Substantial Equivalence Discussion
The table below includes a comparison of the technological characteristics between the new device and those of the predicate device:
| Characteristic | Predicate Device<br>neoBLUE blanket LED Phototherapy<br>System, K103589 | Subject Device<br>BiliCocoon™ Phototherapy System<br>K163526 | Comments |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Indications for Use | The neoBLUE blanket LED<br>Phototherapy System is intended for<br>the treatment of neonatal<br>hyperbilirubinemia. It can be used in<br>the clinical setting or in the home. | The BiliCocoon™ Phototherapy<br>System is intended for the treatment<br>of unconjugated hyperbilirubinemia<br>in the population of neonates and<br>infants under 3 months old and<br>weighing less than 10kg. It can be<br>used in the clinical setting or in the<br>home. | Different – see<br>Comment 1 |
| Population | Neonates | Neonates and infants under 3 months<br>and 10kg | Different – see<br>comment 1 |
| Treatment<br>Method | Underneath blue light | Nest Pad : Underneath blue light<br>Bag Pad : Underneath and above blue<br>light | Different – See<br>Comments 2, 4 |
| Sites of use | Clinical, Home Use | Clinical, Home Use | Same |
| Type of Device | Portable | Portable | Same |
| Technological Features | | | |
| Type of Blue<br>light | Blue Led Light with peak emission<br>between 450 and 475 nm | Blue Led Light with peak emission<br>between 430 and 490 nm | Different – See<br>Comment 2 |
| Intensity<br>(irradiance)<br>average | 30 - 35 μW/cm²/nm | 30 - 40 μW/cm²/nm | Different – See<br>Comment 2 |
| Connection | 1 bundle of optical<br>fiber (diameter 1.8 cm) | 6 bundle of optical fiber (diameter<br>0.75cm) | Different – See<br>Comment 3 |
| Power Supply | 12 VCC / 8.3A 100W | 12 VCC / 7.5A 90W | Different – See<br>Comment 3 |
| Power Security | double insulation class I | double insulation class II | Different – See<br>Comment 3 |
| Leakage Current | <100μA | <100μA | Same |
| Light Box | | | |
| Dimensions | 11.4 cm x 22.9 cm x 14.0 cm | 21.5 cm x 19.8 cm x 16.0 cm | Similar |
| Weight | 1.5 kg | 1.4 kg | Similar |
| PAD Treatment Area | | | |
| light emitting area | Small Blanket : 819 cm²<br>Large Blanket : 542 cm² | Nest Pad : 1200 cm²<br>Bag Pad : 1200 cm² | Different - See<br>Comment 4 |
| effective treatment area<br>according to IEC<br>60601-2-50 | Small Blanket : 504 cm²<br>Large Blanket : 296 cm² | Nest Pad : 1200 cm²<br>Bag Pad : 1200 cm² | Different - See<br>Comment 4 |
| PAD Materials | Polyurethane (mattress covering the<br>fiber optics PAD) | Polyurethane (mattress covering the<br>fiber optics PAD) | Same |
| Use Intent | Multiple Patients | Multiple Patients | Same |
| Disposable Blanket | | | |
| Use Intent | Single Patient | Single Patient | Same |
| patient contact | Non-woven polypropylene | Nonwoven Polypropylene/Polyester-<br>Viscose | Different - See<br>Comment 5 |
| Alarms | | | |
| Bad optical connection | Yes | Yes | Same |
| LED overheating | Yes | Yes | Same |
| LED Control | Yes | Yes | Same |
| Noise | 25dB | 35dB | Similar |
| Timer | No | Yes | Different - See<br>Comment 6 |
{5}------------------------------------------------
#### Comment 1
The predicate device is intended for the treatment of hyperbilirubinemia in the neonate population. Hyperbilirubinemia in neonates is a common pathology which is characterized by an augmentation bilirubin. Hyperbilirubinemia is commonly associated to unconjugated hyperbilirubinemia because it is the most common pathology creating an augmentation of bilirubin. Phototherapy is a method to treat unconjugated hyperbilirubinemia by transforming bilirubin into water-soluble isomers that can be eliminated without conjugation in the liver.
The subject devices restricts the intended use to unconjugated hyperbilirubinemia which is the most common form of hyperbilirubinemia. Conjugated hyperbilirubinemia is rare and not treated by phototherapy device intended to treat hyperbilirubinemia is essentially referring to unconjugated hyperbilirubinemia. Intended use of Bilicocoon™ Phototherapy System is thus equivalent to the intended use of the predicate device.
The claimed targeted population is also different but this difference does not bring any additional risk. Neonates are defined by the Agency as birth to 28 days. However, the chosen the targeted population is up to 3 months and 10kg. This will not change the risk, as an older and heavier population is less critical.
#### Comment 2
Both devices use the same principle - blue light, peak emission range 430-490nm and irradiance >30uW/cm²/nm - recommended by the American Academy of Pediatrics for intensive phototherapy treatment. The difference in the light peak emission, irradiance and surface area for treatment does not raise new or different questions of safety or effectiveness.
{6}------------------------------------------------
## Comment 3
The connector of the BiliCocoon™ Phototherapy System is based on the same technologies than the predicate, except that the subject device employs solutions to prevent overheating and fiber degradation due to heat. The same electrical safety and performance testing was performed for both the subject and predicate device.
## Comment 4
There are no relevant differences in the therapy, except that the subject device enables to cover a larger portion of the body surface of the infant, by providing larger fiber optic pads and the possibility of illuminating from underneath and above the patient. The American Academy of Pediatrics recommends covering as much as possible the body surface of the infant for phototherapy treatment. Therefore the differences in treatment area do not raise new or different questions of safety or effectiveness.
## Comment 5
The predicate device disposable component material is non-woven polypropylene. The BiliCocoon™ Phototherapy System disposable is non-woven polyester viscose. Both devices underwent the same biocompatbility tests per ISO 10993 for the patient contacting materials and met the same requirements; thereby these are substantially equivalent.
#### Comment 6
●
The predicate software and the BiliCocon™ Phototherapy System software remain linked to the microcontroller, without any communication, interpretation or data exchange feature. The BiliCocoon™ Phototherapy System adds a screen and navigation button to set up a timer, but is not adding any risks as both the predicate device and the subject device can be switched off manually and safely at any time.
#### Performance Testing
The following bench testing was performed and reviewed to support the substantial equivalence of the subject device:
- Biocompatibility tests performed per ISO 10993 for patient contacting materials (disposable cover).
- O The disposable cover is considered a skin-contacting device with prolonged exposure
- Cytotoxicity per ISO 10993-5, USP 87
- . Intracutaneous irritation Study per ISO 10993-10
- . Sensitization per ISO 10993-10, USP 88
- Bench tests performed to measure spectral output. light irradiance and effective surface treatment area. ●
- Electrical safety tests performed per IEC 60601-1 (3ra edition A1:2012) ●
- Electromagnetic Compatibility tests (EMI/EMC) performed per IEC 60601-1-2 (4th edition). ●
- Phototherapy safety and performance tests performed per IEC 60601-2-50 (2na edition). ●
- Home Use has been assessed per IEC 60601-1-11 (2w edition:2015) and the FDA Guidance, "Applying Human ● Factors and Usability Engineering to Medical Devices"
- . Photobiological safety assessed with IEC 62471.
- Cleaning Validation Testing .
There is no difference of non-clinical testing performed between the device under submission and the predicate device.
#### Clinical Tests
Not Applicable
#### Conclusions
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The BiliCocoon™ Phototherapy System is substantially equivalent to the NeoBLUE blanket LED Phototherapy System cleared under K103589 with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.