The Triton Sponge System is a software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges. The Triton Sponge System is intended to be used with surgical sponges, software, hardware and accessory devices which have been validated for use with the Triton Sponge System to estimate the hemoglobin (Hb) mass contained on used surgical sponge System is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient serum Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The Triton Sponge System is also indicated for use to aid in counting surgical sponges and may be used to record and display case-specific blood components infused over time. The Triton Sponge System is additionally indicated for use to aid in managing surgical sponges, including providing a visual record of sponge images, and to record the user-entered weight of used surgical sponges in order to calculate an estimate of fluid volume on the sponges.
Device Story
Triton Sponge System is a software application for estimating blood loss from surgical sponges. Input: images of used sponges captured via iPad Pro camera and 3D IR laser depth sensor (NUI Sensor); user-entered patient serum Hb value; user-entered sponge weight. Operation: NUI sensor provides depth maps for automatic sponge detection; calibration placard normalizes ambient light; cloud-based algorithm processes images to estimate hemoglobin mass (sHbL) and blood volume (sEBL). Output: estimated blood loss values and cumulative error estimates displayed on iPad interface. Used in OR by surgical personnel. Benefits: provides objective, real-time blood loss estimation to aid clinical decision-making regarding fluid management and transfusion.
Clinical Evidence
Bench-only testing. Performance validation compared Triton-generated Hb mass/blood volume estimates against known quantities on reconstituted blood-soaked sponges. Testing covered variable lighting, Hb concentrations, blood volumes, and saline dilution. Bland-Altman analysis demonstrated strong positive linear correlation and acceptable agreement with pre-measured Hb mass. Human factors testing confirmed intended users could successfully perform critical tasks using the new hardware/accessories.
Technological Characteristics
Mobile app running on iPad Pro (A1584) with iOS 10+. Uses 3D IR laser depth sensor (Structure Sensor ST01) for imaging. Cloud-based image processing algorithm. Includes calibration placard for scene normalization. Connectivity: wireless communication between app and cloud server. Software: moderate level of concern. EMC/wireless coexistence tested per IEC 60601-1-2:2007.
Indications for Use
Indicated for use as an adjunct in the estimation of blood loss and management of surgical sponges in surgical procedures. Used to estimate hemoglobin mass and blood volume on used surgical sponges, aid in sponge counting, record infused blood components, and manage sponge records via visual imaging and weight entry.
Regulatory Classification
Identification
An image processing device for estimation of external blood loss is a device to be used as an aid in estimation of patient external blood loss. The device may include software and/or hardware that is used to process images capturing externally lost blood to estimate the hemoglobin mass and/or the blood volume present in the images.
Special Controls
In combination with the general controls of the Food, Drug & Cosmetic Act, the Image processing device for estimation of external blood loss is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. Demonstration of the performance characteristics must include a comparison to a scientifically valid alternative method for measuring deposited hemoglobin mass. The following use conditions must be tested:
(i) Lighting conditions;
(ii) Range of expected hemoglobin concentrations;
(iii) Range of expected blood volume absorption; and
(iv) Presence of other non-sanguineous fluids (
*e.g.,* saline irrigation fluid).(2) Human factors testing and analysis must validate that the device design and labeling are sufficient for appropriate use by intended users of the device.
(3) Appropriate analysis and non-clinical testing must validate the electromagnetic compatibility (EMC) and wireless performance of the device.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Software display must include an estimate of the cumulative error associated with estimated blood loss values.
(6) Labeling must include:
(i) Warnings, cautions, and limitations needed for safe use of the device;
(ii) A detailed summary of the performance testing pertinent to use of the device, including a description of the bias and variance the device exhibited during testing;
(iii) The validated surgical materials, range of hemoglobin mass, software, hardware, and accessories that the device is intended to be used with; and
(iv) EMC and wireless technology instructions and information.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, forming a symbol that resembles a stylized caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 25, 2017
Gauss Surgical, Inc. Artie Kaushik Manager RA/QA 334 State St., Suite 201 Los Altos, California 94022
Re: K163507
Trade/Device Name: Triton Sponge System Regulation Number: 21 CFR 880.2750 Regulation Name: Image Processing Device for Estimation of External Blood Loss Regulatory Class: Class II Product Code: PBZ Dated: December 12, 2016 Received: December 14, 2016
Dear Artie Kaushik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/8 description: The image shows the name "Jennifer R. Stevenson -S" in a large, clear font. The text is black and stands out against a white background. In the background, the letters "FDA" are faintly visible in a light blue color, creating a subtle watermark effect.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163507
Device Name Triton Sponge System
#### Indications for Use (Describe)
The Triton Sponge System is a software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges. The Triton Sponge System is intended to be used with surgical sponges, software, hardware and accessory devices which have been validated for use with the Triton Sponge System to estimate the hemoglobin (Hb) mass contained on used surgical sponge System is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient serum Hb value. The validated surgical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The Triton Sponge System is also indicated for use to aid in counting surgical sponges and may be used to record and display case-specific blood components infused over time. The Triton Sponge System is additionally indicated for use to aid in managing surgical sponges, including providing a visual record of sponge images, and to record the user-entered weight of used surgical sponges in order to calculate an estimate of fluid volume on the sponges.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 6: 510(k) Summary (21 CFR § 807.92(c))
#### I. SUBMITTER INFORMATION
| Submitter: | Gauss Surgical, Inc.<br>334 State St., Suite 201<br>Los Altos, CA 94022 |
|------------|---------------------------------------------------------------------------------------------------------------------|
| Contact: | Artie Kaushik<br>Manager RA/QA<br>Gauss Surgical, Inc.<br>Phone: +1(408) 480 6017<br>Email: artie@gausssurgical.com |
Date Summary Prepared: March 31, 2016
#### II. SUBJECT DEVICE INFORMATION
Device Trade Name: Gauss Surgical Triton Sponge System
- Common Name: Triton Sponge System Image Processing Device for Estimation of External Blood Loss in used surgical sponges
- Classification Name: Image Processing Device for Estimation of External Blood Loss (21 CFR §880.2750)
- Product Code: PBZ
#### III. PREDICATE DEVICE INFORMATION
Equivalent Devices: Triton Svstem K160338, August 5, 2016. This predicate device has not been subject to a recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The Triton Sponge System is a software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges. The Triton Sponge System is intended to be used with surgical sponges, software, hardware and accessory devices which have been validated for use with Triton Sponge System to estimate the hemoglobin (Hb) mass contained on used surgical sponges. This version of the Triton Sponge System includes 4 updates from the predicate Triton System (K160338):
Qualified new hardware accessory: Apple's iPad Pro to be used with Triton Sponge System.
Qualified new accessory for imaging sponges: Users will use a commercially available 3D IR laser depth sensor (referred to as the Natural User Interface or NUI Sensor in the submission)
{4}------------------------------------------------
that allows for automatic detection of sponges as well as a touch-free interface with the device to facilitate imaging of the sponge on Apple's iPad Pro device.
- To allow the use of the NUI Sensor with the Triton Sponge System, a new algorithm O called the Sponge Recognition Algorithm (SRA) was added to the System. The SRA analyzes the depth maps provided by the NUI Sensor to determine whether or not a User is presenting a sponge for imaging.
- o To allow the NUI Sensor to securely connect to the iPad Pro, NUI mounting brackets are provided to connect the NUI Sensor to the iPad Pro.
Addition of a step to include imaging a calibration placard with Triton Sponge App to normalize ambient light settings. The calibration card is provided to standardize the image of each sponge.
Updates to the Hemoglobin Algorithm to improve hemoglobin mass estimates by performing scene normalization as well as utilizing new data provided by the calibration palette and NUI Sensor.
## V. INDICATIONS FOR USE
#### Intended Use / Indications for Use:
The Triton System is a software application intended to be used as an adjunct in the estimation of blood loss and management of surgical sponges. The Triton System is intended to be used with surqical sponges, software, hardware and accessory devices which have been validated for use with the Triton System to estimate the hemoglobin (Hb) mass contained on used surgical sponges. The Triton System is also intended to calculate an estimate of blood volume on used surgical sponges from the estimated Hb mass and a user-entered patient serum Hb value. The validated surqical sponges, hardware, software, accessory devices and Hb mass ranges are listed in the Instructions for Use. The Triton System is also indicated for use to aid in counting surgical sponges and may be used to record and display case-specific blood components infused over time.
The Triton System is additionally indicated for use to aid in managing surgical sponges, including providing a visual record of sponge images, and to record the user-entered weight of used surgical sponges in order to calculate an estimate of fluid volume on the sponges.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Gauss Surgical Triton Sponge System is substantially equivalent to the Triton System which was cleared for commercialization via K160338. Both systems have the same indication for use, have the same technology, namely image processing software, to estimate blood loss on used surgical sponges in surgical procedures and run on the same mobile platform to estimate blood loss in surgical procedures.
Triton Sponge System is an incremental improvement to the Triton System. The changes to the system and the rationale for change are discussed below.
{5}------------------------------------------------
| ITEM<br>NO. | CHARACTERISTIC | SUBJECT DEVICE<br>TRITON SPONGE | PREDICATE DEVICE<br>TRITON SYSTEM<br>(K160338) | RATIONALE |
|-------------|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-----------|
| i | Device<br>Manufacturer | Gauss Surgical, Inc. | Gauss Surgical, Inc. | N/A |
| ii | 510(k) Clearance | K163507 | K160338 | N/A |
| iii | Product Code | PBZ | PBZ | N/A |
| iv | Regulatory Class | II | II | N/A |
| v | Regulatory<br>Classification | 21 CFR § 880.2750 | 21 CFR § 880.2750 | N/A |
| vi | Regulation Name | Image Processing Device for<br>Estimation of External Blood Loss | Image Processing<br>Device for Estimation<br>of External Blood Loss | N/A |
| 1 | Indications for Use | | | |
| | Indications for Use<br>Statement | The Triton Sponge System is a<br>software application intended to be<br>used as an adjunct in the estimation of<br>blood loss and management of<br>surgical sponges.<br><br>The Triton Sponge System is intended<br>to be used with surgical sponges,<br>software, hardware and accessory<br>devices which have been validated for<br>use with the Triton Sponge System to<br>estimate the hemoglobin (Hb) mass | Same | No change |
{6}------------------------------------------------
| | | contained on surgical sponges. The<br>Triton Sponge System is also intended<br>to calculate an estimate of blood<br>volume on used surgical sponges from<br>the estimated Hb mass and a user-<br>entered patient serum Hb value. The<br>validated surgical sponges, hardware,<br>software accessory devices and Hb<br>mass ranges are listed in<br>the<br>Instructions for Use. | | | | |
|-----|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|----|---------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Technological Characteristics | | | | | |
| 2 | Identical<br>Features | | | | | |
| 2.1 | Fundamental<br>Scientific<br>Technology | Mobile App for use on mobile platform<br>to estimate intraoperative Hb loss and<br>blood loss on sponges | Same<br>device | as | subject | The fundamental technology of the device<br>remains the same. The enhancements made<br>to the device are so the same technology may<br>be used on the newer version of the iPad<br>tablet. In addition, changes have been made<br>to accommodate the new accessory for<br>imaging the used sponges. |
| 2.2 | Principle<br>of<br>Operation | Using the mobile platform camera, the<br>Triton Sponge System scans blood-<br>containing surgical sponges, counts<br>the sponges, and sends the images to<br>the cloud server for processing. On the<br>server,<br>the<br>image-processing<br>algorithm estimates the Hemoglobin<br>mass loss (sHbL) contained on the<br>sponges and sends results back to the<br>OR. | Same<br>device | as | subject | The work flow for system operation remains<br>unchanged. |
{7}------------------------------------------------
| 2.3 | System | A mobile App, Server, and Algorithm | Same as the subject<br>device | The core components of the system remain<br>unchanged. Change is to add new hardware<br>accessories and updates to system that have<br>been made to accommodate the functioning<br>of the System. |
|-------|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2.5 | Function of the<br>Image Processing<br>Software | App captures images and uses a visual<br>algorithm to estimate Hb mass | Same as the subject<br>device | Although software has been updated to<br>accommodate the new hardware accessories,<br>the core function of the image processing<br>software remains unchanged. |
| 2.6 | Function<br>of<br>Algorithm | Calculates Hb mass and EBL.<br>Requires user input of patient serum<br>Hb value. | Same as the subject<br>device | The function of the Algorithm remains the<br>same. The only change to the Algorithm has<br>been to include the scene normalization<br>feature. Details below (3.4) in this table.<br>Additionally, a Sponge Recognition Algorithm<br>was added, to interface with the IR camera<br>connected to the iPad and detect when a user<br>has presented a sponge for imaging. |
| 2.7 | Function of Server | The remote "cloud" Server computer<br>runs the Server software and<br>algorithm. The Server software<br>coordinates communication between<br>the Algorithm and the App. | Same as the subject<br>device | The function of the Server remains the same.<br>The updates to the Server are incremental. |
| 2.8 | Validated Ranges | Validated range specific to each<br>sponge type as listed in IFU | Same as the subject<br>device | Validated range specific to each sponge type<br>is confirmed using the same process as<br>detailed in the predicate 510(k) K 160338 and<br>the original 510(K) K 130190. |
| 2.9 | App to Server<br>Communication | App communicates wirelessly with<br>Gauss Surgical's cloud-based server | Same as the subject<br>device | No change to the app to Server<br>communication |
| 2.10 | Estimate<br>of | Estimated HbL and estimated EBL | Same as the subject | The display of estimated EBL and HBL errors |
| | Cumulative HBL<br>and EBL Errors | errors are displayed on the App using<br>the Bland-Altman method. | device | is a special requirement for this device<br>category and the method of display remains<br>the same with this update. |
| 3 | New Features | | | |
| 3.1 | Mobile Hardware | iPad Pro (A1584) | iPad 2 (A1395) | Change does not affect indications for use,<br>fundamental scientific technology or<br>operation. Safety and effectiveness are<br>demonstrated through verification and<br>validation testing. Per one of the special<br>controls detailed for this category of devices,<br>the iPad Pro and the Structure Sensor<br>successfully passed EMC testing and<br>wireless coexistence. |
| 3.1.2 | Operating System | Apple, Inc. iOS 10 or greater | Apple, Inc. iOS 5.0 and<br>greater | The Operating System is provided by Apple.<br>iPad Pro includes iOS 10. The system<br>verification and validation confirm that iOS 10<br>with iPad Pro platform is safe and effective. |
| 3.2 | Imaging<br>Accessories | NUI Sensor: Gauss qualified new 3D<br>IR laser depth sensor (NUI sensor) that<br>allows for automatic detection of<br>sponges as well as a touch-free<br>interface with the device to facilitate<br>imaging of the sponge. | Foot Pedals: PageFlip<br>Firefly Pedals and<br>AirTurn BT-105 with 2<br>ATFS-2 Pedals and<br>Pedal Board | A new accessory was qualified with the<br>Triton Sponge System to improve usability of<br>the device. Safety and effectiveness are<br>demonstrated through validation testing.<br>Since this accessory would be used in a<br>hospital setting along with the iPad Pro, it<br>was also tested for and passed EMC and<br>wireless coexistence. In addition, the System<br>Validation Testing confirmed the<br>performance of the entire system to meet the<br>same predetermined acceptance criteria as<br>the predicate system. Therefore, the modified<br>device is substantially equivalent to<br>predicate. |
| | | | | |
| 3.3 | Calibration<br>Placard | Added step to calibrate using a<br>calibration placard | New feature | The Calibration Placard was verified to be<br>designed according to hardware<br>requirements upon successful passing of the<br>Hardware Verification Protocol which called<br>out the dimensional and color requirements<br>for the Placard. System Validation Testing<br>confirmed the performance of the entire<br>System (including Calibration Placard) to<br>meet the same predetermined acceptance<br>criteria as the predicate system. |
| 3.4 | Scene<br>Normalization<br>feature to Hb<br>Algorithm | Added a scene normalization feature<br>to normalize background settings for<br>the sponge. | New feature | This feature was added as an incremental<br>improvement to improve product<br>performance.<br>This feature when used with the Calibration<br>Placard helps standardize each sponge<br>image for ambient lighting conditions.<br>This enhancement to the algorithm does not<br>have any impact to user-facing components<br>of the App. However, using the Calibration<br>Placard is a new step and was included in<br>the Human Factors Usability study. All Users<br>were successfully able to use the Calibration<br>Placard. In addition, the functioning of the<br>System Validation Testing which is<br>implemented with all the System accessories<br>including iPad Pro and Calibration Placard<br>confirmed the performance of the entire<br>system to meet the same predetermined<br>acceptance criteria as the predicate system. |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data was provided in support of the substantial equivalence determination and to demonstrate the Triton Sponge System performs as anticipated for its intended use conditions as detailed below.
#### a. Electromagnetic Compatibility (EMC) and Wireless Coexistence Testing
Electromagnetic Compatibility and Wireless Coexistence Testing was completed for Gauss Surgical's Triton Sponge System.
iPad Pro EMC Testing: The new iPad Pro, A1584, manufactured by Apple, Inc., was tested per the relevant requirements of IEC 60601-1-2:2007 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
Testing demonstrated that the iPad Pro is EMC compatible with the operating room environment. This testing is considered sufficient to demonstrate the electromagnetic compatibility of the iPad Pro in the surgical suite.
The iPad Pro (A1584) was also tested for Wireless Coexistence by confirming the performance of the iPad Pro in the presence of other in-band emitters. The test addressed environmental specifications including co-channel and adjacent channel interference from other devices and users of the RF band to demonstrate that the iPad Pro performs properly in proximity to other RF wireless in-band sources. The iPad Pro was found to maintain essential wireless functionality under all test conditions.
Distance requirements for all interferers are noted in the Instructions for Use. The system was not tested in the presence of MRI, CT, diathermy and electromagnetic security systems such as metal detectors; this is noted in the Instructions for Use.
NUI Sensor EMC testing: The new Structure Sensor (ST01) manufactured by Occipital, Inc., (i.e. "NUI Sensor") was tested per the relevant requirements of IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests.
Testing demonstrated that the Structure Sensor is EMC compatible with the operating room environment. This testing is considered sufficient to demonstrate the electromagnetic compatibility of the iPad Pro in the surgical suite.
Wireless Coexistence testing was not necessary for the Structure Sensor as it has no wireless functionalities or components.
### b. Magnetic Resonance (MR) Compatibility
No testing has been conducted to demonstrate whether the device is MR compatible. The labeling includes a warning that states "The device is MR Unsafe. Do not bring the device into an MR environment. The device must not be used in an MR environment."
#### c. Software Verification and Validation Testing
The software is considered a moderate level of concern (LOC) because inaccurate estimated blood loss may result in consequences to health. All of the elements of software information
{11}------------------------------------------------
corresponding to moderate LOC devices as outlined in FDA's Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 11, 2005) are provided. Documentation describing the software development program are provided.
Software verification testing demonstrated that all specified requirements, including hazard mitigations for the Triton Sponge System have been fulfilled. Testing included verification of the special control to display a correct estimate of the cumulative error associated with estimated blood loss values. Verification demonstrated that this functionality displays correctly and as designed.
Validation testing, including performance and human factors validation, demonstrated that all device specifications conform with user needs and intended uses.
All testing demonstrated the software performs as intended and all software related risks have been adequately mitigated.
#### d. Performance Testing Bench
Bench testing demonstrated that the Triton Sponge performs as intended under anticipated conditions of use. Bench top validation studies were performed to evaluate the accuracy of the device's estimation of hemoglobin mass loss on sponges (sHbL, g) and blood volume loss on sponges (sEBL, ml) in comparison to known Hb mass and blood volume contained on the surgical sponges.
Bench top performance validation testing confirmed the ability of the Triton Sponge to estimate sponge hemoglobin mass loss and sponge blood volume loss by comparing Triton generated results to known quantities of hemoglobin mass and blood contained on surgical sponges. Testing was performed using the new automatic imaging method (NUI Sensor) for imaging and the iPad Pro. This testing was completed for the expected conditions of use as labeled for the product. The bench-top validation performed on the system was the same protocol performed on the predicate product with the addition of testing across sponge illuminations, and used the same clinically justified acceptance criteria cleared for the predicate device.
Testing involved running the software app/system test protocol with sample specimens to ensure all specifications were met.
Briefly, this testing was undertaken by depositing known quantities of blood volume with a known Hb mass on surgical sponges and imaging them with Triton Sponge using the NUI accessory and iPad Pro. As is common in surgical procedures, saline is often also part of the fluid on these sponges and therefore known volumes of saline were added to some samples. The fluid samples represented the clinically-expected ranges and distributions of fluid volume, dilution (by saline), Hb mass, ambient light, sponge illumination, and serum patient Hb. For each sponge type, the Triton Sponge App was used to capture scans of the sponges under the three different ambient lighting conditions and sponge illuminations to confirm the ability of the algorithm to operate consistently across a range of intraoperative ambient illuminance. User-entered volumes input into the app were recorded as well.
The images were then transferred to the server-based hemoglobin algorithm software, which was used to calculate hemoglobin mass (Triton sHbL) for each imaged sponge. The sHbL value obtained via the algorithm for each sponge was then compared to the pre-measured Hb mass (Assay sHbL) of the samples.
{12}------------------------------------------------
A plot of the association between Triton sHbL and Assay sHbL demonstrated a strong positive linear correlation between the two methods of measurement and the predicate across the range of expected intraoperative conditions. Bland-Altman analysis demonstrated acceptable agreement between the sHbL and the pre-determined Hb mass and evaluation of the parameters via the acceptance criteria.
The results of this performance validation testing using Bland-Altman methods inform the lookup tables and resultant "error estimate" values displayed on the user interface each time that the Triton Sponge System is used intraoperatively.
System validation testing of the Triton Sponge followed a similar protocol to performance validation, demonstrating the device met the user requirements under expected conditions of use. A mock surgical case was simulated by reconstituting whole blood samples of known Hb concentration from units of human packed red blood cells and plasma to create various pre-specified blood volumes. Serial dilution with sterile saline vielded sponge blood samples reconstituted to ranges of fluid volume, dilution, and Hb mass representative of a surgical operation. For each dilution level achieved, Triton Sponge was used to capture scans of the surgical sponges in an operating room. User-entered baseline Hb was recorded into the App. System validation testing was performed using the new automatic imaging method (NUI Sensor) and iPad Pro.
The images and corresponding user-entered volumes were then transferred automatically to the server-based software via the App/Server interfaces, and the Triton HB Algorithm automatically calculated hemoglobin mass (Triton sHbL) within each sponge type. The live sHbL returned to the App from the Algorithm and displayed to the user on the screen was then recorded for comparison with the Assay (pre-measured) Hb mass (Assay sHbL) of the reconstituted samples.
Triton sHbL vs. Assay sHbL demonstrated a strong positive linear association between the two methods of measurement across the range of expected intraoperative conditions. Bland-Altman analysis demonstrated acceptable agreement between the sHbL and the predetermined Hb mass and evaluation of the parameters via the acceptance criteria.
In summary, the device was tested within the anticipated use conditions including variable lighting, range of expected patient serum hemoglobin concentrations, range of expected blood volume and the presence of other non-sanguineous fluids, and the resulting data demonstrated acceptable performance.
#### e. Human Factors Testing
The Triton Sponge System software was developed to conform to the Human Interface Guidelines (HIG) as published by Apple for iOS Apps. Additionally, human factors usability testing and analysis validated that the device design (i.e. user interface) and labeling are sufficient for appropriate use by intended users of the Triton Sponge System.
A usability study was conducted to explore critical and frequently used functions associated with the Triton Sponge System in a simulated setting using hardware (iPad Pro) and the validated accessory (NUI Sensor). Both quantitative survey data was collected and analyzed. Participants included personnel who are typically required to track blood loss during surgical procedures. All users were able to successfully complete the tasks per the
{13}------------------------------------------------
protocol pass/fail criteria demonstrating the product meets applicable Human Factors requirements and customer design requirements per product specifications.
Human factors testing validated that the device design and labeling are sufficient for appropriate use by intended users of the device.
#### f. Labeling
Labeling has been provided which includes the special controls as called out in K160338 and in the original de novo petition clearance for the predicate device. Specifically, the labeling includes:
- An appropriate prescription statement as required by 21 CFR 801.109;
- Warnings, cautions and limitations needed for safe use of the device;
- A detailed summary of the performance testing pertinent to the use of the device, including a description of the bias and variance the device exhibited during testing;
- . The validated surgical materials, range of hemoglobin mass, software, hardware, imaging methodologies, and accessories that the device is intended to be used with; and
- EMC and wireless technology instructions and information. ●
### VIII. SUMMARY OF SPECIAL CONTROLS REQUIRED FOR 21 CFR §880.2750
With the granting of Gauss Surgical's de novo petition for the original Pixel 3 System, the FDA established special controls for products cleared under this classification. As summarized in this notification, the Triton Sponge System has met the six (6) special controls as specified and detailed below in Table 6-1 below.
### TABLE 6-1: SPECIAL CONTROLS REQUIRED FOR 12 CFR §880.2750
| Special Control Required | Special Control Met |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Non-clinical performance data must<br>demonstrate that the device performs as<br>intended under anticipated conditions of<br>use. Demonstration of the performance<br>characteristics must include a comparison<br>to a scientifically valid alternative method<br>for measuring deposited hemoglobin mass.<br>The following use conditions must be<br>tested:<br>A. Lighting conditions;<br>B. Range of expected hemoglobin<br>concentrations;<br>C. Range of expected blood volume<br>absorption; and<br>D. Presence of other non-sanguineous<br>fluids (e.g., saline irrigation fluid) | Non-clinical performance testing using the<br>iPad Pro and the NUI sensor was<br>conducted and demonstrated that the<br>device performs as intended under<br>anticipated conditions of use including<br>expected lighting conditions, range of<br>expected hemoglobin values, range of<br>expected blood volume absorption and<br>presence of other non-sanguineous fluids.<br>Performance data under anticipated<br>conditions of use demonstrate that the<br>Triton Sponge System performs as<br>intended. |
| 2. Human factors testing and analysis must<br>validate that the device design and labeling | Human factors testing and analysis<br>demonstrated that the Triton Sponge |
| Special Control Required | Special Control Met |
| are sufficient for appropriate use by<br>intended users of the device. | System, labeling, and the NUI sensor as an<br>accessory to support scanning of images<br>with the Triton Sponge System are<br>sufficient for appropriate use by intended<br>users of the Triton Sponge System. |
| 3. Appropriate analysis and non-clinical<br>testing must validate the electromagnetic<br>compatibility (EMC) and wireless<br>performance of the device. | EMC and wireless performance of the<br>Triton Sponge System with new iPad Pro<br>and accessory NUI Sensor have been<br>validated. |
| 4. Appropriate software verification,<br>validation and hazard analysis must be<br>performed. | Software verification, validation and hazard<br>analysis have been performed. |
| 5. Software display must include an<br>estimate of the cumulative error associated<br>with estimated blood loss values. | An estimate of the cumulative error<br>associated with blood loss values is<br>displayed to the user with each estimated<br>hemoglobin mass and blood loss value.<br>The results of verification testing and Bland-<br>Altman methods inform the look-up tables<br>and resultant values displayed on the user<br>interface each time the Triton Sponge<br>System is used for the estimation of<br>hemoglobin mass and blood volume loss. |
| 6. Labeling must include:<br>A. Warnings, cautions, and limitations<br>needed for safe use of the device;<br>B. A detailed summary of the performance<br>testing pertinent to use of the device,<br>including a description of the bias and<br>variance the device exhibited during<br>testing;<br>C. The validated surgical materials, range<br>of hemoglobin mass, software, hardware,<br>and accessories that the device is intended<br>to be used with; and<br>D. EMC and wireless technology<br>instructions and information. | Labeling includes all details as required by<br>the special controls. |
{14}------------------------------------------------
# IX. CONCLUSIONS
The Triton Sponge introduces no new indication for use and is substantially equivalent in terms of Indications for Use as the identified predicate device.
{15}------------------------------------------------
The Triton Sponge System was verified in its operating environments. Verification and validation activities related to the device modification were performed on the applicant device, and the predetermined acceptance criteria were met in all cases.
From the testing data included in this submission, it can be concluded that the Gauss Surgical Triton Sponge System his substantially equivalent to the currently marketed predicate device in indications for use, design, technological characteristics, mechanism of action and performance. In addition, all special controls as required for products under this classification, image processing devices for estimation of external blood loss, have been addressed. The differences between the subject and predicate device do not raise different types of safety or effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.