Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System
K163506 · Vertera, Inc. · MAX · Jun 19, 2017 · Orthopedic
Device Facts
Record ID
K163506
Device Name
Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System
Applicant
Vertera, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jun 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine with accompanying radicular symptoms. Implants are used to facilitate fusion in the lumbar spine (L2 to S1), and are intended to be used with autogenous bone graft and supplemental fixation systems that have been cleared by the FDA for use in the lumbar spine.
Device Story
Coalesce™ Lumbar Interbody Fusion System consists of sterile, single-use PEEK implants; available in various footprints/heights; features integrated porous architecture derived from implant body; includes tantalum markers for radiographic visibility. Used by surgeons in lumbar interbody fusion procedures to facilitate spinal fusion; requires supplemental fixation and autogenous bone graft. Device provides structural support to intervertebral space; porous surface designed to facilitate fusion. Clinical benefit includes stabilization of lumbar spinal column in patients with degenerative disc disease.
Clinical Evidence
Bench testing only. Functional performance testing conducted per ASTM F2077-14 (static/dynamic compression, compression shear), ASTM F2267-04 (subsidence), and ASTM F1877-16 (particle characterization). Results support substantial equivalence.
Technological Characteristics
Materials: PEEK (implant body), Tantalum (markers). Design: Porous architecture integrated into PEEK body (not a coating). Standards: ASTM F2077-14, ASTM F2267-04, ASTM F1877-16. Sterilization: Sterile, single-use. Connectivity: None.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (discogenic back pain with radiographic confirmation) at one or two contiguous lumbar levels (L2-S1) with radicular symptoms. Used with autogenous bone graft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 19, 2017
Vertera, Inc. Mr. Wei Chang Project Leader 739 Trabert Avenue NW Suite F Atlanta, Georgia 30318
Re: K163506
Trade/Device Name: Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: May 23, 2017 Received: Mav 24, 2017
Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163506
Device Name
Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System
#### Indications for Use (Describe)
The Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine with accompanying radicular symptoms. Implants are used to facilitate fusion in the lumbar spine (L2 to S1), and are intended to be used with autogenous bone graft and supplemental fixation systems that have been cleared by the FDA for use in the lumbar spine.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Vertera Spine. The logo consists of a green circle with a white "V" inside, followed by the word "Vertera" in a thin, sans-serif font, also in green. To the right of "Vertera" and rotated 90 degrees is the word "spine", also in a thin, sans-serif font and in green.
# Traditional 510(k) Summary
Date Submitted: June 16, 2017
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92(e).
- A. 510(k) Submitter: Vertera, Inc. Establishment Registration No. 3012232921 739 Trabert Ave. NW. Suite F Atlanta, Georgia 30318 Phone: (571) 758-3783 Fax: (678) 669-1975
- B. Company Contact: Wei (Allen) Chang Project Leader (571) 758-3783 allen.chang@verteraspine.com
C. Device Information: Trade Name(s):
Coalesce™ Lumbar Interbody Fusion System
- Coalesce™-Straight Interbody Fusion Device .
- Coalesce™-Convex Interbody Fusion Device .
- Coalesce™-Crescent Interbody Fusion Device
- Coalesce™-Lateral Interbody Fusion Device 0
- Coalesce™-Anterior Interbody Fusion Device ●
- Coalesce™-Oblique Interbody Fusion Device 0
Common Name: Lumbar Interbody Fusion Device
- Intervertebral Body Fusion Device, Lumbar D. Classification Name: MAX, 888.3080
- ட். Predicate Device(s): Eisertech, LLC, Interbody Cage, K140348, Primary
Vertera, Inc., Hedgehog, K143685, Reference Amendia, Zeus®, K151322, Reference Medtronic, Clydesdale®, K151128, Reference
#### ட் Physical Description:
The proposed devices within the Coalesce™ Lumbar Interbody Fusion System are sterile, single use implant grade polyetheretherketone (PEEK) devices, available in varied footprints and heights, designed for supplemental stabilization of the lumbar spinal column in lumbar intervertebral body fusion procedures.
{4}------------------------------------------------
Each device within the Coalesce™ System is comprised of a continuous body of PEEK formed into the final product shape with a porous architecture on select faces of the implant. The porous architecture is derived directly from the implant body and is not a sintered or otherwise additive coating. In addition to PEEK, the device assembly may contain two or more tantalum markers, depending on footprint, to enable visibility under x-ray in vivo.
- G. Indications for Use:
The Coalesce™ (-Straight, -Convex, -Crescent, -Lateral, -Anterior, or -Oblique) Lumbar Interbody Fusion System is indicated for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies) at one or two contiguous levels of the lumbar spine with accompanying radicular symptoms. Implants are used to facilitate fusion in the lumbar spine (L2 to S1), and are intended to be used with autogenous bone graft and supplemental fixation systems that have been cleared by the FDA for use in the lumbar spine.
- H. Comparison of Technological Characteristics:
The Coalesce™ Lumbar Interbody Fusion System is substantially equivalent in function and intended use to the primary predicate device manufactured by Eisertech, LLC, Interbody Cage (K140348).
- l. Non-Clinical Testing:
Functional performance testing, including static and dynamic compression and compression shear, subsidence, and expulsion, has been conducted per applicable standards as recommended through the FDA guidance document for Intervertebral Body Fusion Device. The standards are (ASTM) F2077-14. Test Methods for Intervertebral Body Fusion Devices, (ASTM) F2267-04, Standard Test Method for Measuring Load Induced Subsidence of an Intervertebral Body Fusion Device Under Static Axial Compression, and (ASTM) F1877-16, Standard Practice for Characterization of Particles. Analysis of the results supports the conclusion that the proposed device is substantially equivalent to the predicate devices.
- J. Conclusion:
Based on comparison to predicate devices and non-clinical performance testing, Vertera, Inc.'s Coalesce™ Lumbar Interbody Fusion System has demonstrated to be substantially equivalent to FDA-cleared legally marketed predicate devices.
> Wei (Allen) Chang Project Leader
Image /page/4/Picture/11 description: The image shows the logo for Vertera Spine. The logo consists of a green circle with a white "V" inside, followed by the word "Vertera" in a thin, green font. To the right of "Vertera" is the word "Spine" in a smaller, light gray font, oriented vertically.
739 Trabert Ave. NW. Suite F Atlanta. GA 30318 Phone: 571-758-3783
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.