← Product Code [EBI](/productcode/EBI) · K163482

# Orthodontic Acrylic 2 (K163482)

_Lang Dental Manufacturing Co., Inc. · EBI · May 31, 2017 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163482

## Device Facts

- **Applicant:** Lang Dental Manufacturing Co., Inc.
- **Product Code:** [EBI](/productcode/EBI.md)
- **Decision Date:** May 31, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K163482 · May 31, 2017 | Orthodontic Acrylic 2 | Lang Dental Manufacturing Co., Inc. | Clinical study of 32 patients | A limited clinical study was conducted to compare biofilm formation on the surface of the Orthodontic Acrylic 2 appliance versus the predicate device. | Biofilm formation; Clinical study; Orthodontic appliance |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Limited clinical study; Comparative clinical study | 32 patients; Sample Size: 32 | Orthodontic Acrylic (predicate device) | Reduction of biofilm formation on the surface of the appliance |

## Indications for Use

Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.)

## Device Story

Orthodontic Acrylic 2 is a two-part, fast-curing, self-cure methacrylate resin system consisting of powder and liquid components. Mixed by dental professionals to fabricate orthodontic appliances like retainers and bite guards. Incorporates Quaternary Ammonium Methacryloxy Silicate (QAMS) into the polymer network to provide antibacterial and antifungal properties, inhibiting adhesion of Candida albicans and reducing Streptococcus mutans and Actinomyces naeslundii. Used in dental clinics; appliance maintenance remains the patient's responsibility. QAMS presence aims to reduce biofilm formation on appliance surfaces, though clinical outcome benefits are not established.

## Clinical Evidence

Limited clinical study of 32 patients showed reduced biofilm formation on appliances compared to non-QAMS predicate. In-vitro studies confirmed reduction in S. mutans, A. naeslundii, and C. albicans biofilm formation over 3 months. Bench testing per ISO 20795-2:2010 confirmed compliance for flexural strength, flexural modulus, fracture toughness, water sorption, and water solubility. Biocompatibility testing included Guinea Pig Maximization, Oral Mucosal Irritation (ISO 10993-10), and cytotoxicity (ISO 10993-5), all meeting requirements.

## Technological Characteristics

Two-part self-cure methacrylate resin system. Contains Quaternary Ammonium Methacryloxy Silicate (QAMS) for antimicrobial properties. Physical/mechanical properties validated against ISO 20795-2:2010. Biocompatibility validated per ISO 10993-1:2009 (cytotoxicity, irritation, sensitization).

## Regulatory Identification

A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.

## Predicate Devices

- Orthodontic Acrylic ([K141439](/device/K141439.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 31, 2017

Lang Dental Manufacturing Co., Inc. Dave Lang President 175 Messner Drive Po Box 969 Wheeling, Illinois 60090

Re: K163482

Trade/Device Name: Orthodontic Acrylic 2 Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: February 27, 2017 Received: March 3, 2017

Dear Dave Lang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/7 description: The image shows the name "Mary S. Runner -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

Device Name Orthodontic Acrylic 2

Indications for Use (Describe)

Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.)

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

| Applicant:                            | Lang Dental Manufacturing Company Incorporated<br>175 Messner Drive<br>P. O. Box 969<br>Wheeling, Illinois 60090-0969 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Person:                       | David Lang<br>Tel: 847-215-6622<br>Fax: 847-215-6678                                                                  |
| Date Prepared:                        | 30 May, 2017                                                                                                          |
| Trade Name:<br>Common Name:           | Orthodontic Acrylic 2<br>Fast Curing Orthodontic Acrylic Resin Powder and Liquid                                      |
| Product Code:<br>Classification/Name: | EBI<br>Denture relining, repairing, or rebasing resin<br>Class II per CFR 872.3760                                    |

### Predicate Devices:

Orthodontic Acrylic 2 is substantially equivalent to the following Lang Dental Manufacturing Company Incorporated's:

Orthodontic Acrylic, K141439

#### Indications for Use:

Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.).

#### Description of Applicant Device:

Orthodontic Acrylic 2 is a fast curing self-cure 2 part system consists of a powder and a liquid. The combination of the powder and liquid is converted into a hard methacrylate finished product.

#### Technological Characteristics:

Orthodontic Acrylic 2 is based upon industry standard chemistry. The Orthodontic Acrylic 2 Liquid contains Quaternary Ammonium Methacryloxy Silicate (QAMS). Since QAMS is copolymerizable, the antibacterial/antifungal properties are independent of loss of surface layer since it is incorporated throughout the entire polymer network. In-vitro testing of Orthodontic Acrylic 2 shows that QAMS inhibits adhesion of Candida albicans and reduces Streptococcus mutans and Actinomyces naselunii for at least 3 months.

{4}------------------------------------------------

Orthodontic Acrylic 2 containing QAMS has been shown in a limited clinical study involving 32 patients to reduce biofilm formation on the surface of the appliance, as compared to Orthodontic Acrylic that does not contain QAMS. In addition, in-vitro studies conducted after three months showed a substantial reduction in S. mutans, A. naeslundii and C. albicans biofilm formation on the device. QAMS in Orthodontic Acrylic 2 aids in keeping the oral appliance clean and is not a substitute for regular cleaning of the appliance by the patient. A reduction in biofilm on the surface of appliance has not been shown to have enhanced clinical outcomes.

The indications for use of Orthodontic Acrylic 2 are the same as those for Orthodontic Acrylic and are summarized in the table below:

### Orthodontic Acrylic

Orthodontic Acrylic is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.).

### Orthodontic Acrylic 2

Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards and bite plates, etc.).

Comparison of the chemical composition of Orthodontic Acrylic 2 to the predicate is provided in the following table:

| Chemical Composition | Orthodontic<br>Acrylic            | Orthodontic<br>Acrylic 2          |
|----------------------|-----------------------------------|-----------------------------------|
| Formulation          | Methacrylate liquid               | Methacrylate liquid               |
| Formulation          | Powder                            | Powder                            |
| Presentation         | 2-part system                     | 2-part system                     |
| Polymerization       | Self-Cured                        | Self-Cured                        |
| Additional Feature   | Anti-bacterial/ Anti-Fungal Agent | Anti-bacterial/ Anti-Fungal Agent |

#### Performance Date:

The following physical/mechanical properties of Orthodontic Acrylic 2 were tested: Physical / Mechanical Property Orthodontic Acrylic 2

| Flexural Strength (ISO 20795-2:2010)  | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for flexural strength  |
|---------------------------------------|--------------------------------------------------------------------------------------------|
| Flexural Modulus (ISO 20795-2:2010)   | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for flexural modulus   |
| Fracture Toughness (ISO 20795-2:2010) | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for fracture toughness |
| Water sorption (ISO 20795-2:2010)     | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for water sorption     |
| Water Solubility ( ISO 20795-2:2010)  | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2-2010 for water solubility   |

{5}------------------------------------------------

# Physical / Mechanical Property

Anti-Bacterial Testing (in vitro)

Anti-fungal testing (in vitro)

Orthodontic Acrylic 2

Orthodontic Acrylic 2 is equivalent to the predicate against S. mutans and A. naeslundii. Orthodontic Acrylic 2 is equivalent to the predicate against C. albicans.

# Biocompatibility:

An evaluation of biocompatibility was conducted using ISO 10993-1:2009 to determine the biological testing requirements for Orthodontic Acrylic 2.

Orthodontic Acrylic 2 was tested for Guinea Pig Maximization Testing and Oral Mucosal Irritation (ISO 10992-10) and cytotoxicity (ISO 10993-5); Orthodontic Acrylic 2 met the requirements for these tests.

# Conclusion:

Side by side comparisons demonstrate that the applicant device is substantially equivalent in safety and effectiveness to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K163482](https://fda-staging.innolitics.com/device/K163482)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
