Limited clinical study; Comparative clinical study
32 patients; Sample Size: 32
Orthodontic Acrylic (predicate device)
Reduction of biofilm formation on the surface of the appliance
Indications for Use
Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.)
Device Story
Orthodontic Acrylic 2 is a two-part, fast-curing, self-cure methacrylate resin system consisting of powder and liquid components. Mixed by dental professionals to fabricate orthodontic appliances like retainers and bite guards. Incorporates Quaternary Ammonium Methacryloxy Silicate (QAMS) into the polymer network to provide antibacterial and antifungal properties, inhibiting adhesion of Candida albicans and reducing Streptococcus mutans and Actinomyces naeslundii. Used in dental clinics; appliance maintenance remains the patient's responsibility. QAMS presence aims to reduce biofilm formation on appliance surfaces, though clinical outcome benefits are not established.
Clinical Evidence
Limited clinical study of 32 patients showed reduced biofilm formation on appliances compared to non-QAMS predicate. In-vitro studies confirmed reduction in S. mutans, A. naeslundii, and C. albicans biofilm formation over 3 months. Bench testing per ISO 20795-2:2010 confirmed compliance for flexural strength, flexural modulus, fracture toughness, water sorption, and water solubility. Biocompatibility testing included Guinea Pig Maximization, Oral Mucosal Irritation (ISO 10993-10), and cytotoxicity (ISO 10993-5), all meeting requirements.
Technological Characteristics
Two-part self-cure methacrylate resin system. Contains Quaternary Ammonium Methacryloxy Silicate (QAMS) for antimicrobial properties. Physical/mechanical properties validated against ISO 20795-2:2010. Biocompatibility validated per ISO 10993-1:2009 (cytotoxicity, irritation, sensitization).
Indications for Use
Indicated for the fabrication of methacrylate-based orthodontic appliances, including retainers, bite guards, and bite plates.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2017
Lang Dental Manufacturing Co., Inc. Dave Lang President 175 Messner Drive Po Box 969 Wheeling, Illinois 60090
Re: K163482
Trade/Device Name: Orthodontic Acrylic 2 Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: February 27, 2017 Received: March 3, 2017
Dear Dave Lang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/7 description: The image shows the name "Mary S. Runner -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The letters are uniformly sized and spaced, contributing to the legibility of the name.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name Orthodontic Acrylic 2
Indications for Use (Describe)
Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Applicant: | Lang Dental Manufacturing Company Incorporated<br>175 Messner Drive<br>P. O. Box 969<br>Wheeling, Illinois 60090-0969 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | David Lang<br>Tel: 847-215-6622<br>Fax: 847-215-6678 |
| Date Prepared: | 30 May, 2017 |
| Trade Name:<br>Common Name: | Orthodontic Acrylic 2<br>Fast Curing Orthodontic Acrylic Resin Powder and Liquid |
| Product Code:<br>Classification/Name: | EBI<br>Denture relining, repairing, or rebasing resin<br>Class II per CFR 872.3760 |
### Predicate Devices:
Orthodontic Acrylic 2 is substantially equivalent to the following Lang Dental Manufacturing Company Incorporated's:
Orthodontic Acrylic, K141439
#### Indications for Use:
Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.).
#### Description of Applicant Device:
Orthodontic Acrylic 2 is a fast curing self-cure 2 part system consists of a powder and a liquid. The combination of the powder and liquid is converted into a hard methacrylate finished product.
#### Technological Characteristics:
Orthodontic Acrylic 2 is based upon industry standard chemistry. The Orthodontic Acrylic 2 Liquid contains Quaternary Ammonium Methacryloxy Silicate (QAMS). Since QAMS is copolymerizable, the antibacterial/antifungal properties are independent of loss of surface layer since it is incorporated throughout the entire polymer network. In-vitro testing of Orthodontic Acrylic 2 shows that QAMS inhibits adhesion of Candida albicans and reduces Streptococcus mutans and Actinomyces naselunii for at least 3 months.
{4}------------------------------------------------
Orthodontic Acrylic 2 containing QAMS has been shown in a limited clinical study involving 32 patients to reduce biofilm formation on the surface of the appliance, as compared to Orthodontic Acrylic that does not contain QAMS. In addition, in-vitro studies conducted after three months showed a substantial reduction in S. mutans, A. naeslundii and C. albicans biofilm formation on the device. QAMS in Orthodontic Acrylic 2 aids in keeping the oral appliance clean and is not a substitute for regular cleaning of the appliance by the patient. A reduction in biofilm on the surface of appliance has not been shown to have enhanced clinical outcomes.
The indications for use of Orthodontic Acrylic 2 are the same as those for Orthodontic Acrylic and are summarized in the table below:
### Orthodontic Acrylic
Orthodontic Acrylic is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards, and bite plates, etc.).
### Orthodontic Acrylic 2
Orthodontic Acrylic 2 is intended for the fabrication of methacrylate-based orthodontic appliances (such as retainers, bite guards and bite plates, etc.).
Comparison of the chemical composition of Orthodontic Acrylic 2 to the predicate is provided in the following table:
| Chemical Composition | Orthodontic<br>Acrylic | Orthodontic<br>Acrylic 2 |
|----------------------|-----------------------------------|-----------------------------------|
| Formulation | Methacrylate liquid | Methacrylate liquid |
| Formulation | Powder | Powder |
| Presentation | 2-part system | 2-part system |
| Polymerization | Self-Cured | Self-Cured |
| Additional Feature | Anti-bacterial/ Anti-Fungal Agent | Anti-bacterial/ Anti-Fungal Agent |
#### Performance Date:
The following physical/mechanical properties of Orthodontic Acrylic 2 were tested: Physical / Mechanical Property Orthodontic Acrylic 2
| Flexural Strength (ISO 20795-2:2010) | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for flexural strength |
|---------------------------------------|--------------------------------------------------------------------------------------------|
| Flexural Modulus (ISO 20795-2:2010) | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for flexural modulus |
| Fracture Toughness (ISO 20795-2:2010) | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for fracture toughness |
| Water sorption (ISO 20795-2:2010) | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2:2010 for water sorption |
| Water Solubility ( ISO 20795-2:2010) | Orthodontic Acrylic 2 meets the requirements of<br>ISO 20795-2-2010 for water solubility |
{5}------------------------------------------------
# Physical / Mechanical Property
Anti-Bacterial Testing (in vitro)
Anti-fungal testing (in vitro)
Orthodontic Acrylic 2
Orthodontic Acrylic 2 is equivalent to the predicate against S. mutans and A. naeslundii. Orthodontic Acrylic 2 is equivalent to the predicate against C. albicans.
# Biocompatibility:
An evaluation of biocompatibility was conducted using ISO 10993-1:2009 to determine the biological testing requirements for Orthodontic Acrylic 2.
Orthodontic Acrylic 2 was tested for Guinea Pig Maximization Testing and Oral Mucosal Irritation (ISO 10992-10) and cytotoxicity (ISO 10993-5); Orthodontic Acrylic 2 met the requirements for these tests.
# Conclusion:
Side by side comparisons demonstrate that the applicant device is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.