AQT90 FLEX Myo Test Kit, AQT90 FLEX CKMB Test Kit, AQT90 FLEX
Applicant
Radiometer Medical Aps
Product Code
JHX · Clinical Chemistry
Decision Date
Sep 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1215
Device Class
Class 2
Indications for Use
The AQT90 FLEX analyzer is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoagulant has been added. It is intended for use in point-of-care and laboratory settings.
Device Story
AQT90 FLEX is a cartridge-based immunoassay analyzer; uses time-resolved fluorescence with europium (Eu) chelate labels; processes whole-blood or plasma samples. Modifications include semi-automated system cleaning (using blank cartridges/cleaning solution), updated hematocrit calculation algorithm, new flash lamp, removal of liquid pathway filter, and pre-laser cut cartridge foil. Device operates in point-of-care and laboratory settings; physician/technician-operated. Input: blood collection tube; output: quantitative analyte concentration. Output aids clinicians in rapid diagnosis of myocardial infarction. Benefits: rapid, automated quantitative testing for cardiac markers.
Clinical Evidence
No clinical studies performed. Evidence consists of bench testing including precision/reproducibility (CLSI EP5-A3), linearity (CLSI EP6-A), limit of detection (CLSI EP17-A2), and method comparison (CLSI EP09-A3) against predicate devices. Method comparison showed high correlation (r²=1.00) for both Myo and CKMB tests. Matrix comparison validated interchangeability of EDTA/lithium heparin whole blood and plasma.
Indicated for use as an aid in the diagnosis of myocardial infarction (CKMB Test) and the rapid diagnosis of heart disease, e.g., acute myocardial infarction (Myo Test). Used with EDTA or lithium heparin whole-blood or plasma specimens in point-of-care and laboratory settings.
Regulatory Classification
Identification
A creatine phosphokinase/creatine kinase or isoenzymes test system is a device intended to measure the activity of the enzyme creatine phosphokinase or its isoenzymes (a group of enzymes with similar biological activity) in plasma and serum. Measurements of creatine phosphokinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k163462
B. Purpose for Submission:
Modifications were made to the previously cleared AQT90 FLEX system devices (k112161, k120326) to include a new semi-automated system clean function, a new flash lamp, a redesign of the Test and CAL cartridge housing, and a change in the algorithm to calculate the hematocrit value.
C. Measurand:
Myoglobin
Creatine kinase isoform MB (CKMB)
D. Type of Test:
Quantitative fluorometric immunoassay
E. Applicant:
Radiometer Medical ApS
F. Proprietary and Established Names:
AQT90 Flex Myo Test Kit
AQT90 FLEX CKMB Test Kit
AQT90 FLEX
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| DDR | II | 866.5680, Myoglobin
Immunological Test System | Immunology (82) |
| JHX | II | 862.1215, Creatine
phosphokinase/creatine kinase or | Clinical Chemistry (75) |
{1}
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| | | isoenzymes test system | |
| KHO | I | 862.2560, Fluorometer for clinical use | Clinical Chemistry (75) |
## H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
**AQT90 FLEX Myo Test Kit**
The Myo Test is an in vitro diagnostic assay for the quantitative determination of myoglobin in EDTA or lithium heparin whole-blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is intended for use as an aid in the rapid diagnosis of heart disease, for example, acute myocardial infarction.
**AQT90 FLEX CKMB Test Kit**
The CKMB Test is an in vitro diagnostic assay for the quantitative determination of creatine kinase isoform MB in EDTA or lithium heparin whole-blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is intended for use as an aid in the diagnosis of myocardial infarction.
**AQT90 FLEX**
The AQT90 FLEX analyzer is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoagulant has been added. It is intended for use in point-of-care and laboratory settings.
3. Special conditions for use statement(s):
For in vitro diagnostic use; for prescription use only; for point-of-care use.
4. Special instrument requirements:
Studies were performed using the AQT90 FLEX analyzer.
## I. Device Description (Description of Device Modification):
The AQT90 FLEX Myo Test Kit consists of ten Test Cartridges and one Calibration Cartridge. The components are as follows:
Mouse monoclonaal anti-myoglobin capture and tracer antibodies, approximately 500
{2}
ng/cup. Bovine serum albumin, bovine y'-globulin, and mouse IgG (blocker substances for heterophilic interference), TRIS buffer, and <0.1% sodium azide.
The AQT90 CKMB Test Kit consists of ten Test Cartridges and one Calibration Cartridge. The components are as follows:
Mouse monoclonal anti-CKMB capture and tracer antibodies, approximately 600 ng/cup. Bovine serum albumin, bovine y'-globulin, and mouse IgG (blocker substances for heterophilic interference), TRIS buffer, and <0.1% sodium azide.
The AQT90 FLEX is a cartridge-based immunoassay analyzer, based on time-resolved fluorescence using a europium (Eu) chelate as the fluorescent label and was previously cleared under k112161.
The modifications made to the AQT90 FLEX system in the current submission are outlined below:
- System clean has been updated to implement semi-automated system clean using two new accessories: a blank cartridge and cleaning solution tubes and the needle used in the cup wash has been shortened.
- Algorithm used to determine hematocrit value has been updated.
- Flash lamp in detection system has been changed to a new flash lamp.
- The filter, which is located in the liquid pathway after the cup washing step, has been removed.
- Test cartridge and CAL cartridge foil is now pre-laser cut on one side of the foil instead of both sides.
- Recombinant streptavidin which is chemically identical to the old native streptavidin used is now from a new supplier. Streptavidin is used to bind capture antibody onto the cup surface.
J. Substantial Equivalence Information:
1. Predicate device name(s):
| Predicate name | 510(k) number |
| --- | --- |
| AQT90 FLEX Myo Test | k112161 |
| AQT90 FLEX CKMB Test Kit | k120326 |
| AQT90 FLEX analyzer | k112161 |
{3}
2. Comparison with predicate:
AQT90 FLEX Myo Test Kit:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate device AQT90 FLEX Myo Test (k112161) | Candidate device AQT90 FLEX Myo Test Kit (k163462) |
| Intended use | AQT90 FLEX Myo Test is an in vitro diagnostic assay for the quantitative determination of myoglobin in EDTA or lithium- heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is indicated for use as an aid in the rapid diagnosis of heart disease, e.g. acute myocardial infarction. | Same |
| Principle | Quantitative time-resolved fluorimetric one-step sandwich immunoassay. | Same |
| Traceability | Scripps M0725 | Same |
| Reportable range | 20 to 900 ng/mL | Same |
| Cartridge foil | Test cartridge foil is pre-laser cut on both sides of the foil | Test cartridge foil is pre-laser cut on one side of the foil |
AQT90 FLEX CKMB Test Kit:
| AQT90 FLEX CKMB Test Kit | | |
| --- | --- | --- |
| Item | Predicate device AQT90 FLEX CKMB Test Kit, (k120326) | Candidate device AQT90 FLEX CKMB Test Kit (k163462) |
| Intended use | AQT90 FLEX CKMB Test is an in vitro diagnostic assay for the quantitative determination of creatine kinase isoform MB in EDTA or lithium-heparin whole blood or plasma specimens on the AQT90 FLEX analyzer in point of care and laboratory settings. It is intended for use as an aid in the diagnosis of myocardial infarction. | Same |
{4}
5
| Principle | Quantitative time-resolved fluorimetric one-step sandwich immunoassay | Same |
| --- | --- | --- |
| Traceability | ERM-AD455/IFCC | Same |
| Reportable range | 1.5 to 300 ng/mL | Same |
| Cartridge foil | Test cartridge foil is pre-laser cut on both sides of the foil | Test cartridge foil is pre-laser cut on one side of the foil |
Analyzer
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Predicate device AQT90 FLEX analyzer (k112161) | Candidate device AQT90 FLEX analyzer (k163462) |
| Intended use | The AQT90 FLEX analyzer is an immunoassay instrument based on the quantitative determination of time-resolved fluorescence to estimate the concentrations of clinically relevant markers on whole-blood and plasma specimens to which a relevant anticoagulant has been added. It is intended for use in point of care and laboratory settings. | Same |
| System cleaning | System cleaning using customer prepared cleaning sample and test cup | Semi-automated system cleaning using cleaning tube from cleaning solution tubes and blank cup from blank cartridge |
K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP5-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline—Third Edition
- CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- CLSI EP9-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline—Third Edition
- CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline—Second Edition
{5}
6
L. Test Principle:
The AQT90 FLEX is a cartridge-based immunoassay analyzer, based on time-resolved fluorescence using a europium (Eu) chelate as the fluorescent label. The test receptacles for the assay are test cups, which contain the antibodies used for capture of the analyte, and the Eu chelate labeled antibodies used to trace the captured analyte. The sample is added to the test cup together with assay buffer. The cup is then incubated to allow formation of the immuno-complex, and subsequently washed to remove unbound antibodies and sample material. Finally, the cup is exposed to excitation light, and after a delay the emitted light generated by the fluorescent label is measured by single photon counting. The total count is then compared to an assay calibration curve to obtain a quantitative measurement of the analyte's concentration in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Precision studies were performed for both the Myo Test Kit and the CKMB Test Kit using whole blood and plasma samples on the AQT90 FLEX system in accordance with the CLSI Guideline EP5-A3.
1. Myo Test Kit:
In the whole blood precision study, three lithium heparin whole blood samples with different myoglobin concentrations were measured within one day by performing five runs testing five replicates of each sample for a total of 25 measurements per sample. One analyzer and one Myo Test Kit lot were used. One of the samples was a native specimen (L2), one was a diluted native specimen (L1) and one was spiked with antigen (L3).
Myoglobin whole blood precision results (n=25):
| Sample | Mean (ng/mL) | Within run | | Between Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| L1 | 57 | 1.30 | 2.3 | 0.00 | 0.0 | 1.3 | 2.3 |
| L2 | 92 | 2.57 | 2.8 | 0.93 | 1.0 | 2.73 | 3.0 |
| L3 | 622 | 14.1 | 2.3 | 4.94 | 0.8 | 15.0 | 2.4 |
In the plasma precision study, three lithium heparin plasma pools (two native and one antigen spiked plasma specimen pool) were measured across 20 test days, twice a day in duplicate for a total of 80 measurements per sample. One analyzer and one Myo Test Kit lot were used.
{6}
Myoglobin plasma precision results (n=80):
| Sample | Mean (ng/mL) | Within Run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L1 | 53 | 1.0 | 2.0 | 0.47 | 0.9 | 0.00 | 0.0 | 1.1 | 2.1 |
| L2 | 95 | 1.9 | 2.0 | 0.00 | 0.0 | 0.84 | 0.9 | 2.1 | 2.2 |
| L3 | 586 | 13 | 2.2 | 0.00 | 0.0 | 0.00 | 0.0 | 13 | 2.2 |
2. CKMB Test Kit:
In the whole blood precision study, three lithium-heparin whole blood samples with different CKMB concentrations were measured within one day by performing five runs testing five replicates of each sample for a total of 25 measurements per sample. One analyzer and one CKMB Test Kit lot were used. Level one (L1) was a diluted native sample, level 2 and 3 (L2 and L3) were native samples spiked with antigen.
CKMB whole blood precision results (n=25):
| Sample | Mean (ng/mL) | Within Run | | Between Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| L1 | 2.6 | 0.13 | 4.8 | 0.00 | 0.0 | 0.13 | 4.8 |
| L2 | 14 | 0.5 | 3.5 | 0.49 | 3.4 | 0.70 | 4.9 |
| L3 | 204 | 6.9 | 3.4 | 0.00 | 0.0 | 6.9 | 3.4 |
In the plasma precision study three lithium heparin plasma pools (one native pool, one native pool spiked with patient specimen and one native pool spiked with antigen) were measured across 20 test days, twice a day in duplicate for a total of 80 measurements per sample. One analyzer and one CKMB Test kit lot were used.
CKMB plasma precision results (n=80):
| Sample | Mean (ng/mL) | Within Run | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L1 | 2.3 | 0.08 | 3.5 | 0.026 | 1.1 | 0.00 | 0.0 | 0.084 | 3.7 |
| L2 | 8.4 | 0.21 | 2.5 | 0.083 | 1.0 | 0.088 | 1.0 | 0.24 | 2.8 |
| L3 | 209 | 3.9 | 1.9 | 1.4 | 0.7 | 1.8 | 0.9 | 4.6 | 2.2 |
b. Linearity/assay reportable range:
Linearity studies were conducted for both the Myo Test Kit and the CKMB Test Kit using whole blood and plasma samples on the AQT90 FLEX system according to CLSI guideline EP6-A.
1. Myo Test Kit :
A linearity study was performed with lithium heparin whole blood and lithium heparin plasma samples. A linearity series with a total of 11 sample levels were prepared by serially diluting a high concentration sample with a low concentration
{7}
sample. The samples were measured in replicates of ten using one MyoTest Kit lot and one analyzer during one day. The measured mean myoglobin concentrations for each sample type were as follows:
Whole blood: 12.9, 15.6, 34.0, 70.8, 144.4, 275.9, 467.2, 682.4, 897.5, 1148.6, and 1208.3 ng/mL
Plasma: 15.4, 17.2, 31.4, 59.2, 116.5, 224.4, 375.9, 550.5, 738.5, 1045.8, and 1052.3 ng/mL
The second order fit was significant for whole blood, and the third order fit was significant for plasma. The non-linearity degree (%) was calculated to evaluate linearity. The results showed that the deviation from linearity did not exceed 10% for either whole blood or plasma.
The claimed measuring range for Myo is 20 – 900 ng/mL for whole blood and plasma.
2. CKMB Test Kit:
A linearity study was performed with lithium heparin whole blood and lithium heparin plasma samples. A linearity series with a total of 11 sample levels were prepared by serially diluting a high CKMB concentration sample with a low CKMB concentration sample. The linearity series samples were measured in replicates of 10 using one Test Kit lot and one analyzer during one day. The measured mean CKMB concentrations for each sample type were as follows:
Whole blood: 1.3, 1.7, 9.0, 15.0, 21.7, 50.4, 98.8, 177.6, 252.8, 330.4, and 331.7 ng/mL
Plasma: 1.5, 1.8, 8.5, 13.1, 19.0, 42.5, 90.2, 158.5, 220.1, 289.3, and 291.2 ng/mL
The second and third order fits were not significant for whole blood, thus the whole blood assay was considered to be acceptably linear. The third order fit was significant for plasma, and the non-linearity degree (%) was calculated to evaluate linearity. The results showed that the deviation from linearity did not exceed 10% for plasma.
The claimed measuring range for CKMB is 1.5 – 300 ng/mL for whole blood and plasma.
8
{8}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
AQT90 FLEX Myo Test Kit: The antigen material, M0725 by Scripps, was used to establish the metrological traceability chain of the AQT90 FLEX Myo assay in k112161 and remains unchanged.
AQT90 FLEX CKMB Test Kit: The antigen material, ERM-AD455/IFCC material, was used to establish the metrological traceability chain of the AQT90 FLEX CKMB assay in k120326 and remains unchanged.
d. Detection limit:
Limits of detection studies followed the guidelines in CLSI EP17-A2. For the limit of blank (LoB) and limit of detection (LoD) studies the sponsor set the Type I and II error risks to 5%.
LoB: The LoB study was performed by measuring four blank samples with five replicates on three days using two test kit lots and two AQT90 FLEX analyzers (one lot/analyzer). The total number of measurements per test kit lot across all samples and days was 60 and a nonparametric method was used.
LoD and LoQ: The study was performed with lithium heparin whole blood and lithium heparin plasma samples. Ten samples per matrix were measured using two test kit lots and two AQT90 FLEX analyzers (one lot/analyzer). Each sample was measured in 45 replicates per lot.
The LoD was calculated using the estimated LoB and the SDtotal of the low concentration samples.
To determine the LoQ, a power function was fit to the concentration versus CVtotal plot using data from all samples. For the Myo Test Kit, the intersection of this fit and 10% CVtotal was used to determine the LoQ. For the CKMB Test Kit, the intersection of this fit and the 20% CVtotal was used to determine the LoQ.
The results for the Myo Test Kit are summarized below:
| | Whole Blood (ng/mL) | Plasma (ng/mL) |
| --- | --- | --- |
| LoB | 0.020 | 0.020 |
| LoD | 0.058 | 0.044 |
| LoQ | 0.35 | 0.22 |
The LoB, LoD and LoQ claims for the Myo Test Kit in the package insert are: 0.5 ng/mL, 1.0 ng/mL and 1.0 ng/mL respectively.
{9}
The measuring range of the Myo test is 20 – 900 ng/mL.
The results for the CKMB Test Kit are summarized below:
| | Whole Blood (ng/mL) | Plasma (ng/mL) |
| --- | --- | --- |
| LoB | 0.076 | 0.076 |
| LoD | 0.21 | 0.15 |
| LoQ | 1.0 | 0.21 |
The LoB, LoD and LoQ claims for the CKMB Test Kit in the package insert are: 0.5 ng/mL, 1.0 ng/mL and 1.0 ng/mL respectively.
The measuring range of the CKMB test is 1.5 to 300 ng/mL.
e. Analytical specificity:
Potential interference of the Myo Test Kit with common endogenous and exogenous substances were evaluated in k112161 and the claims are unchanged.
Potential interference of the CKMB Test Kit with common endogenous and exogenous substances were evaluated in k120326 and the claims are unchanged.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
Method comparison studies were conducted in accordance with the CLSI guideline EP09-A3. The studies were conducted in-house. One Myo Test Kit lot and one CKMB Test Kit lot were used for each method comparison study, respectively. Lithium heparin plasma samples were run in duplicate using two candidate AQT90 FLEX analyzers with modified software and modified reagent and two predicate AQT90 FLEX analyzers with unmodified software and the original reagent. The first data set of the duplicate runs was used in the regression analysis.
1. Myo Test Kit: a total of 103 native lithium heparin plasma samples were tested for the method comparison study. The results ranged from 26 to 897 ng/mL. The result of the regression analysis is shown below.
Passing-Bablok regression: y = 1.01x – 0.14, r² = 1.00
2. CKMB Test Kit: a total of 107 lithium heparin plasma samples (including 5 spiked samples) were tested for the method comparison study. The results ranged
10
{10}
from 1.5 to 296 ng/mL. The result of the regression analysis is shown below.
Passing-Bablok regression: y = 0.99x - 0.18, r² = 1.00
b. Matrix comparison:
A matrix comparison study was performed to validate the interchangeable use of four matrices when analyzing Myo and CKMB on the AQT90 FLEX system: EDTA whole blood and plasma and lithium heparin whole blood and plasma. One Myo Test Kit lot and one CKMB Test Kit lot were used in the study.
1. Myo study: A matrix comparison study was performed by testing a total of 127 native sample sets with myoglobin values spanning the measuring range (22 to 892 ng/mL). Passing-Bablok regression analyses were performed and the results of these studies are shown below:
LiHep plasma (y) = 0.99 LiHep WB (x) - 1.0 (n= 125; r²= 1.0)
EDTA plasma (y) = 0.96 EDTA WB (x) - 1.4 (n= 125; r²= 1.0)
EDTA WB (y) = 1.01 LiHep WB (x) + 0.8 (n= 127; r²= 1.0)
EDTA plasma (y) = 0.99 LiHep plasma (x) + 0.2 (n= 124; r²= 1.0)
EDTA WB (y) = 1.03 LiHep plasma (x) + 1.6 (n= 125; r²= 1.0)
2. CKMB study: A matrix comparison study was performed by testing a total of 112 sample sets (105 native and 7 spiked). The sample CKMB concentration spanned the measuring range (1.5 to 297 ng/mL). Passing-Bablok regression analyses were performed and the results of these studies are shown below:
LiHep plasma (y) = 0.99 LiHep WB (x) + 0.01 (n= 106; r²= 1.0)
EDTA plasma (y) = 0.99 EDTA whole blood (x) - 0.02 (n= 104; r²= 1.0)
EDTA WB (y) = 1.02 LiHep WB (x) + 0.01 (n= 103; r²= 1.0)
EDTA plasma (y) = 1.01 LiHep plasma (x) - 0.04 (n= 101; r²= 1.0)
EDTA WB (y) = 1.03 LiHep plasma (x) - 0.02 (n= 104; r²= 1.0)
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
11
{11}
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
The expected values for the Myo test kit were established in k112161 and the claims remain unchanged. The 97.5th percentile was determined to be 75 ng/mL for women and 142 ng/mL for men.
The expected values for the CKMB test kit were established in k120326 and the claims are unchanged. The 97.5th percentile was determined to be 6.9 ng/mL for women and 11 ng/mL for men.
N. Instrument Name:
AQT90 FLEX analyzer
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X
{12}
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
Samples are identified from the barcode on the blood collection tube. The code is read automatically by the analyzer.
4. Specimen Sampling and Handling:
Blood is drawn into a blood collection tube. The blood collection tube is placed directly in the analyzer. A volume of sample is automatically drawn from the tube through its rubber seal. The closed tube is discarded and the used immunoassay cup is discarded into a closed waste bin in the analyzer.
5. Calibration:
The analyzer can be automatically calibrated by means of calibration cartridges at the time of test cartridge lot change.
6. Quality Control:
AQT FLEX LQC Multi-CHECK, Levels 1, 2 and 3 (cleared in k112161).
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not applicable.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.