Reveal LINQ Insertable Cardiac Monitor

K163460 · Medtronic, Inc. · MXD · Feb 14, 2017 · Cardiovascular

Device Facts

Record IDK163460
Device NameReveal LINQ Insertable Cardiac Monitor
ApplicantMedtronic, Inc.
Product CodeMXD · Cardiovascular
Decision DateFeb 14, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2

Indications for Use

The Reveal LINQ ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: patients with clinical syndromes or situations at increased risk of cardiac arrhythmias patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia The device has not been tested specifically for pediatric use.

Device Story

The Reveal LINQ ICM is an insertable cardiac monitor that continuously records subcutaneous ECG signals. It functions as an automatically-activated and patient-activated monitoring system. The device detects arrhythmias, including bradycardia, asystole, ventricular tachycardia, and atrial tachyarrhythmias. It also monitors heart rate variability and activity. The system includes the LINQ Mobile Manager application and a patient connector (Model 24967) for data transmission. Used in clinical settings or by patients, the device provides diagnostic data to healthcare providers to assist in clinical decision-making regarding cardiac rhythm management. The device benefits patients by enabling long-term monitoring for transient symptoms that may otherwise go undetected.

Clinical Evidence

No clinical studies were performed for this submission. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), electrical safety, EMC, mechanical testing, and software verification/validation.

Technological Characteristics

Insertable cardiac monitor; titanium can, titanium nitride electrodes, polyurethane/silicone header, parylene coating. Patient connector (Model 24967) uses polycarbonate housing. Features include subcutaneous ECG sensing, bi-directional telemetry, and automated arrhythmia detection algorithms. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).

Indications for Use

Indicated for patients at increased risk of cardiac arrhythmias or experiencing transient symptoms (dizziness, palpitation, syncope, chest pain) suggestive of arrhythmia. Not tested for pediatric use.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures in profile, arranged in a cascading manner. The figures are depicted in a dark color, contrasting with the white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### February 14, 2017 Medtronic Inc. Laura Danielson Principal Regulatory Affairs Specialist 8200 Coral Sea St NE Mounds View, Minnesota 55112 Re: K163460 Trade/Device Name: Reveal LINO Insertable Cardiac Monitor, Model LNO11 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI Dated: December 8. 2016 Received: December 9, 2016 Dear Laura Danielson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely Mude Yellman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K163460 Device Name Reveal LINQ Insertable Cardiac Monitor (Model LNQ11) Indications for Use (Describe) The Reveal LINQ ICM is an insertable automatically-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: · patients with clinical syndromes or situations at increased risk of cardiac arrhythmias · patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia The device has not been tested specifically for pediatric use. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Date Prepared: | 13 February 2017 | |------------------------------------------|-------------------------------------------------------------------------------------------------------| | 510(k) Owner / Address: | Medtronic, Inc.<br>Cardiac Rhythm and Heart Failure<br>8200 Coral Sea Street<br>Mounds View, MN 55112 | | Contact: | Laura L. Danielson<br>Principal Regulatory Affairs Specialist | | Telephone: | (763) 526-2385 | | Fax: | (651) 367-0603 | | E-mail: | laura.l.danielson@medtronic.com | | Trade / Proprietary Name: | Reveal® LINQTM Insertable Cardiac Monitor,<br>Model LNQ11 | | Common Name: | Insertable Cardiac Monitor | | Classification /<br>Classification Name: | Class II<br>Arrhythmia detector and alarm<br>(21 CFR 870.1025) | | Product Code: | DSI | # Summary of Substantial Equivalence The intended use, design, materials and performance of the Reveal LINQ ICM (Model LNQ11) in association with the LINQ Mobile Manager Application (Model MSW001 or MSW002) and patient connector (Model 24965 and 24967) (referred to as the LINQ Mobile Manager system) are substantially equivalent to the following predicate device: - Reveal LINO ICM (Model LNO11) cleared via K160689 on April 22, 2016. ● ## Indications for Use Comparison to the Predicate The indications for use remains unchanged from the predicate and are as follows: The Reveal LINQ ICM is an insertable automatically-activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in the following cases: - patients with clinical syndromes or situations at increased risk of cardiac arrhythmias ● - patients who experience transient symptoms such as dizziness, palpitation, syncope, and ● chest pain that may suggest a cardiac arrhythmia The device has not been tested specifically for pediatric use. {4}------------------------------------------------ # Comparison of Technological Characteristics with Predicate Device The following table contains a comparison of the CareLink SmartSync Device Manager patient connector, Model 24967 to the predicate device (Model 24965) with no differences. | | Reveal LINQ ICM Model<br>LNQ11 (with existing LINQ<br>Mobile Manager Application<br>Models MSW001 or MSW002<br>and existing patient<br>connector Model 24965) | Reveal LINQ ICM Model<br>LNQ11 (with existing LINQ<br>Mobile Manager Application<br>Models MSW001 and<br>MSW002 and additional<br>patient connector Model<br>24967) | |----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Medtronic, Inc. | Medtronic, Inc. | | 510(k) Number | K160809 | KXXXXXX | | Model Number | LNQ11 | LNQ11 | | Intended Use | Same (See Indications for<br>Use Form 3881) | Same (See Indications for<br>Use Form 3881) | | Ambulatory ECG Monitor | Same | Same | | Continually Record ECG<br>Information? | Same | Same | | Record Pre- & post-event | Same | Same | | R Wave Sensing | Same | Same | | Sampling Rate | Same | Same | | Storage Time | Same | Same | | Noise Reversion | Same | Same | | Brady Detection | Same | Same | | Asystole Detection | Same | Same | | Ventricular Tachycardia<br>Detection | Same | Same | | Patient Activated & Auto<br>Activated | Same | Same | | Atrial Tachyarrhythmia<br>Monitoring & Diagnostics | Same | Same | | Day & Night Heart Rate<br>Diagnostics | Same | Same | | Afib Detection | Same | Same | | Heart Rate Variability<br>Algorithm & Diagnostics | Same | Same | | Telemetry | Same | Same | | Activity Monitoring | Same | Same | | Longevity | Same | Same | | Electrode Spacing (inside-<br>to-inside) | Same | Same | | | Reveal LINQ ICM Model<br>LNQ11 (with existing LINQ<br>Mobile Manager Application<br>Models MSW001 or MSW002<br>and existing patient<br>connector Model 24965) | Reveal LINQ ICM Model<br>LNQ11 (with existing LINQ<br>Mobile Manager Application<br>Models MSW001 and<br>MSW002 and additional<br>patient connector Model<br>24967) | | Volume | Same | Same | | Mass | Same | Same | | Episode Storage | Same | Same | | Patient Symptom Mark | Same | Same | | Cardiac Compass | Same | Same | | MRI Compatibility | Same | Same | | Clinician Notification | Same | Same | | Bi-Directional Telemetry | Same | Same | | Detection Algorithms | Same | Same | | CareLink | Same | Same | | Wireless Telemetry | Same | Same | | Patient's CareLink Clinic<br>Name and ID | Yes | Yes | #### Predicate Comparisons {5}------------------------------------------------ The following patient contacting material comparison demonstrates substantial equivalence between patient contacting materials. Information on differences can be found in the paragraph following the table. | Component | Material | Reveal LINQ ICM Model<br>LNQ11 (with existing<br>LINQ Mobile Manager<br>Application Models<br>MSW001 or MSW002 and<br>existing patient<br>connector Model 24965) | Reveal LINQ ICM Model<br>LNQ11 (with existing LINQ<br>Mobile Manager<br>Application Models<br>MSW001 and MSW002 and<br>additional patient<br>connector Model 24967) | |------------|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Can | Titanium | Yes | Yes | | Electrodes | Titanium nitride | Yes | Yes | | Header | Polyurethane,<br>silicone | Yes | Yes | | Coating | Parylene | Yes | Yes | #### Patient Contacting Material Comparison {6}------------------------------------------------ | Component | Material | Reveal LINQ ICM Model<br>LNQ11 (with existing<br>LINQ Mobile Manager<br>Application Models<br>MSW001 or MSW002 and<br>existing patient<br>connector Model 24965) | Reveal LINQ ICM Model<br>LNQ11 (with existing LINQ<br>Mobile Manager<br>Application Models<br>MSW001 and MSW002 and<br>additional patient<br>connector Model 24967) | |-------------------|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Telemetry<br>Head | Amorphous<br>thermal plastics,<br>thermal plastic<br>elastomer,<br>silicone rubber<br>and polyester film | Polycarbonate Lexan | Silicone rubber Rogers<br>Bisco HT-6240, clear | | | | | Polycarbonate Sabic Lexan<br>EXL9330 with 8T9D168<br>white colorant (Medtronic<br>color chip M954239A001) | | | | | Silicone rubber Wacker<br>Elastosil LR 3003/60, color<br>chip SP 50812 (blue) | | | | | Silicone rubber Momentive<br>TSE221 per color chip<br>Pantone Cool Grey 8 button;<br>polyester Autotex V200 clear<br>(back screened) with hard<br>coated top surface and<br>velvet finish graphics | ## Telemetry Head Material The housing material of the CareLink SmartSync Device Manager patient connector is polycarbonate. The use of polycarbonate boosts chemical resistance concerning exposure to repeat cleaning and disinfection. ## Performance Data The following performance data were provided in support of the substantial equivalence determination. ## Biocompatibility Testing The biocompatibility evaluation for the CareLink SmartSync Device Manager, patient connector, Model 24967 was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing for materials used in the Model 24967 included the following tests: - Cytotoxicity ● {7}------------------------------------------------ - . Sensitization - . Irritation The Model 24967 has acute duration of skin contact with the patient, and is not implanted. ### Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the CareLink SmartSync Device Manager, patient connector, Model 24967, associated applications and non-medical mobile platform (i.e. Tablet) with associated cables as necessary for test. The system complies with the IEC 60601-1, standards for safety and the IEC 60601-1-2 third and fourth edition versions of the standard for EMC. ### Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator. ### Mechanical Testing The following is a list of testing performed: - Inspection of the required mechanical design features and function ● - . Workmanship inspection concerning all external surfaces that can cause injury such as sharp edges or pinch points - Product labeling inspection - Forces required to activate controls - Chemical resistance testing for effects of repeat cleaning cycles ● - Environmental and drop testing ● - . Reliability testing of buttons, electrical contacts, user connector insertions, and replaceable or moving mechanical components ### Animal Study There were no formalized animal studies performed for this 510(k) submission. ### Clinical Studies There were no formalized clinical studies performed for this 510(k) submission. ## Conclusions The non-clinical data support the safety of the device and the hardware and software verification and validation demonstrate that the CareLink SmartSvnc Device Manager, patient connector. Model 24967 should perform as intended in the specified use conditions.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...