# SurgiCube (K163455)

_Surgicube International B.V. · ORC · Sep 28, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163455

## Device Facts

- **Applicant:** Surgicube International B.V.
- **Product Code:** [ORC](/productcode/ORC.md)
- **Decision Date:** Sep 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5070
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

SurgiCube® is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments. The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

## Device Story

SurgiCube is a stand-alone air-handling apparatus for surgical operating rooms. It provides a localized, unidirectional, HEPA-filtered airflow over the surgical site and instruments during ophthalmic procedures. The device uses a mechanical filter system (HEPA efficiency 99.995% for 0.3um particles) to reduce airborne contamination to <5 CFU/m3. The patient is positioned outside the primary surgical area, while the operating site is placed under the downward airflow. A single-use sterile drape attached to side columns maintains the sterile field. The device is operated by the surgical team in a clinical setting. By creating a compact, manageable surgical area, it minimizes contamination risks, potentially improving patient outcomes by reducing infection risk during surgery.

## Clinical Evidence

Clinical study of 26 patients undergoing ophthalmic procedures. Primary endpoints: CFU/m3 and particulate counts at the surgical/instrument site. Results: Average air viables 0.12 CFU/m3 (well below 5 CFU/m3 limit); average particles (≥0.5 µm) 666/m3 (well below 3520/m3). Comparison to background air showed significant improvement: ~470x fewer particles (≥0.5 µm), ~725x fewer particles (≥5.0 µm), and >126x fewer CFU/m3 inside the SurgiCube environment. 95% confidence level achieved.

## Technological Characteristics

Stand-alone air-handling apparatus. Materials: stainless steel, aluminum, safety glass. HEPA filtration (99.995% efficiency at 0.3 µm) using waterproof glass fiber media. Airflow capacity: 353-777 CFM (model dependent). Electrical: 120V AC, 60Hz. Compliance: IEC 60601-1 (electrical safety), FCC/47 CFR 15 (EMC). Sterile accessory: single-use drape (ETO sterilized, 10⁻⁶ SAL).

## Regulatory Identification

Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- Operio ([K153498](/device/K153498.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2017

Surgicube International B.v. Ger Vijfvinkel CEO Seggelant-noord 4 Vierpolders, 3237 MG NL

Re: K163455

Trade/Device Name: Surgicube Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus For A Surgical Operating Room Regulatory Class: Class II Product Code: ORC Dated: August 24, 2017 Received: August 28, 2017

Dear Ger Vijfvinkel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Michael J. Ryan -S

for Tina Kiang, PhD Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K163455

Device Name SurgiCube®

#### Indications for Use (Describe)

SurgiCube® is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-fittered to reduce the presence of particulate matter to reduce the microorganisms to a level of < S CFU per m3 at the surgical site and instruments.

The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

#### Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for SurgiCube International. The logo features a blue square with a white curved line inside, next to the text "SurgiCube" in a blue sans-serif font. Below "SurgiCube" is the word "International" in a smaller, also blue, sans-serif font. The logo is clean and modern, with a focus on the company name.

# 510(k) Summary

# Date the summary was prepared is September 25, 2017

# The submitter of the 510(k) is:

- Ger Vijfvinkel CEO SurgiCube International B.V. Seggelant-Noord 4 3237 MG Vierpolders The Netherlands Phone +31 181 760 660 Email g.vijfvinkel@surgicube.com
# Device subject to this 510(k)

| Trade name:         | SurgiCube®                                           |
|---------------------|------------------------------------------------------|
| Common name:        | Air handling apparatus for surgery                   |
| Classification name | Air-handling apparatus for a surgical operating room |
| Device class        | II                                                   |
| Regulation number   | 21 CFR 878.5070                                      |
| Product code        | ORC                                                  |

## Predicate device

The legally marketed device to which substantial equivalence

is claimed is: Operio K153498 Cleared 07/20/2016

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for "SurgiCube International". The logo consists of a blue square with a white square inside, and a white curved line inside the inner square. To the right of the square is the text "SurgiCube" in a blue sans-serif font, with the word "International" underneath in a smaller font.

## Device description

The SurgiCube® is an apparatus that provides a localized, optimally HEPA filtered, surgical environment to carry out minimally invasive surgery for ophthalmic procedures. It supplies air around the operating surface using uni directional air flow technique. Possible sources of contamination are sidelined. The patient is physically positioned outside the surgical area; just the operating area is in the field. The surgical team can move around the operating surface without interfering with the air. The uni directional flow together with correct surgical draping create a compact and manageable surgical area that will eliminate airborne contamination risks for patients. The SurgiCube® is equipped with a multiple mechanical filter system, including a High Efficiency Particle Air (HEPA) filter with an efficiency of filtering 99,995% of all particles of 0.3um.

## Indications for use

SurgiCube®is a stand-alone device that creates a surgical operating environment with a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m³ at the surgical site and instruments.

The surgical site of the patient is intended to be placed under the air flow which is directed downwards to the surgical site and/or instruments within 32" (80 cm) in length. Device effectiveness may not be reliably detectable at a distance of 32" (80 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

## Summary of the technological characteristics compared to the predicate device

SurgiCube® and the predicate device, Operio, have the same intended use and similar technological characteristics.

#### lt is similar in the following way:

- The positioning to the surgical site of the delivery of the air flow is the same.
- . The surgical indication, ophthalmic surgery is the same.
- The cleaning efficiency of the HEPA filtered clean air over the surgical site is under 5 CFU for both the subject device (0.12 CFU/ m3) and the predicate device (0.4 CFU/ m3).
- The air flow from the subject device is HEPA-filtered and has the same efficiency (99.995%) to reduce the presence of particulate matter and microorganisms at the surqical site.

#### lt differs in the following way:

- The subject device is not a portable device; it is a stand-alone device. ●
- . The subject device doesn't use a sterile shield in front of the air outlet, instead it uses a single use sterile drape attached onto the side columns.
- . The reach of the clean air flow is up to 32" cm from the HEPA filter compared to 47" for the predicate device.
- . The delivered air volume to the surgical site for the subject device is 353-777 CFM compared to 235 CFM for the predicate device.

The above stated differences do not introduce any new hazards and test results can confirm that the subject device is as safe and effective for use. In the hazard analysis possible risks have been mitigated that could be associated with the use of the subject device.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for SurgiCube International. The logo consists of a blue square with a white outline on the left side, with a curved white line inside the square. To the right of the square is the text "SurgiCube" in a blue sans-serif font, with the word "International" in a smaller font below it.

# Comparison table:

| Characteristic                                                  | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Review Panel                                                    | SurgiCube®<br>General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Operio K153498<br>General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indications for use                                             | SurgiCube® is a stand-alone device<br>that creates a surgical operating<br>environment with a directed, non-<br>turbulent flow of air to the surgical site<br>during ophthalmic surgery and to the<br>sterile instruments used during<br>surgery. The air flow from the device is<br>HEPA-filtered to reduce the presence<br>of particulate matter to reduce the<br>microorganisms to a level of < 5 CFU<br>per m³ at the surgical site and<br>instruments.<br><br>The surgical site of the patient is<br>intended to be placed under the air<br>flow which is directed downwards to<br>the surgical site and/or instruments<br>within 32" (80 cm) in length. Device<br>effectiveness may not be reliably<br>detectable at a distance of 32" (80 cm)<br>from the air flow outlet, and<br>effectiveness depreciates beyond this<br>specified area. | Operio is a portable device for use in a<br>surgical operating room that produces<br>a directed, non-turbulent flow of air to<br>the surgical site during ophthalmic<br>surgery and to the sterile instruments<br>used during surgery. The air flow from<br>the device is HEPA-filtered to reduce<br>the presence of particulate matter to<br>reduce the microorganisms to a level of<br>< 5 CFU per m³ at the surgical site and<br>instruments.<br><br>The air flow is intended to be directed<br>parallel to the surgical site and/or<br>instruments within: 20" (50 cm) in<br>width, 47" (120 cm) in length and 15"<br>(40 cm) in height. Device effectiveness<br>may not be reliably detectable at a<br>distance of 47 inches (120 cm) from the<br>air flow outlet, and effectiveness<br>depreciates beyond this specified area. |
| Physical<br>Dimension                                           | 102-150 x 63 x 95 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 24 x 18 x 47-67 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Weight                                                          | 992 – 1763 lbs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 92 lbs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Material                                                        | Stainless steel, aluminium and safety<br>glass                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Aluminium, powder coated steel, ABS<br>plastic, stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Electrical<br>Specifications                                    | 1. Mains power input: 120 V (AC), 60<br>Hz<br>2. Power consumption: 350 VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1. Mains power input: 120 V (AC), 60 Hz<br>2. Power consumption: 290 VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Type of device                                                  | Prescription use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Prescription use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Movability                                                      | SurgiCube® is a stand-alone device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Operio is a portable device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Air filtration                                                  | HEPA filtered air with at least 99.995%<br>efficiency against 0.3 µm particles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | HEPA filtered air with at least 99.995%<br>efficiency against 0.3 µm particles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| HEPA filter media                                               | Waterproof glass fibre                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Nelior membrane media.<br>Folded sheets of nelior.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Air inlet Pre-filter<br>Media                                   | Synthetic fibres and high efficiency<br>water resistant glass fibre                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Polyamide mesh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regular<br>maintenance                                          | HEPA filter to be replaced after 1 year<br>and pre-filters to be replaced after 6<br>months.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | HEPA filter to be replaced after 2000<br>hours/yearly.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Airflow capacity                                                | SC100: 353 CFM<br>SC180: 636 CFM<br>SC200: 706 CFM<br>SC220: 777 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Airflow is 235 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Efficiency                                                      | Bacteria carrying particles at the<br>surgical site is < 0.12 CFU/m³                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Bacteria carrying particles at the<br>surgical site is 0.4 CFU/m3.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Airflow speed                                                   | 0.45 m/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 0.4 m/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Air volume<br>delivered to incision<br>in CFM                   | SC100: 353 CFM<br>SC180: 636 CFM<br>SC200: 706 CFM<br>SC220: 777 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Airflow over surgical site is 235 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Measured average<br>particulate density<br>at surgical incision | 18 particulates of 0.5 µm/f³ (666/m³)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 1.1 particulate of 0.5 µm/f³                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Air flow position                                               | Air flow is provided over the surgical<br>site and instruments.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Air flow is provided over the surgical<br>site and instruments.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterile accessory                                               | Single-use sterile drape                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Single-use sterile shield                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Sterility Assurance<br>(SAL)<br>for sterile<br>drape            | 10⁻⁶ SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 10⁻⁶ SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterilization Method<br>for sterile drape                       | ETO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Gamma irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for SurgiCube International. The logo consists of a blue square with a white curved line inside, next to the text "SurgiCube" in a light blue font. Below "SurgiCube" is the word "International" in a smaller, light blue font.

## Summary of Testing

SurgiCube® fulfils and has been tested against the electromagnetic compatibility requirements. An electromagnetic compatibility test has been carried out at an accredited test laboratory. The report indicates that the SurqiCube® complies to the FCC requlation and 47 CFR 15. The relevant clauses for electrical safety were tested to confirm that the SurgiCube® complies with the relevant electrical requirements of IEC 60601-1 to ensure that no hazardous situation can occur for the patient. The test results have been presented in the submission.

The following non-clinical tests has been conducted: Simulated use test, Particulate counting test, Colony Forming Units counting test, Air velocity test, Air leakage test, Smoke tests and Air cleaning efficiency in ambient area to demonstrate air flow and efficiency.

The bench tests demonstrate that SurgiCube® met all performance and acceptance criteria, and does not introduce considerations for safety and efficacy.

## Summary of clinical testing

Clinical tests with 26 patients were carried out at ophthalmic procedures. Measurements where made near the surgical site and over the instruments where the results demonstrated a statistically significant reduction in CFU levels as well as particulates. Air viables and particles were measured in order to determine the cleanliness of the clean air area at the surgical site and the instrument site when the SurgiCube® was used for air filtering on patients.

The analysis of the data revealed that it is possible to state at a 95% confidence level that

- The average number of air viables is with 0.12 CFU/m³ well below 5 CFU/m³ 1.
- 2. The average number of particles with a size of ≥0.5 µm is with 666 per m³ well below 3520 per m3
- 3. A significant improvement is achieved when comparing the values inside the Surgicube® with the values outside the Surgicube®:

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for SurgiCube International. The logo consists of a blue square with a white curved line inside, next to the text "SurgiCube" in a light blue font. Below "SurgiCube" is the word "International" in a smaller, light blue font. The logo is clean and modern, with a focus on the company's name and international presence.

a. Approximately 470 times less particles of ≥0.5 µm per m³ (313 821 in background vs. 666 inside Surgicube) and approximately 725 times less particles of ≥5.0 µm per m³ (1952 in background vs. 2.69 in Surgicube)

b. More than 126 times less CFU per m³ (15.19 in background vs. 0.12 inside Surgicube)

## Conclusion

Based on a comparison of the performance characteristics, the tests performed, clinical and nonclinical, it is concluded that the subject device SurgiCube® is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K163455](https://fda-staging.innolitics.com/device/K163455)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).