← Product Code [DRG](/productcode/DRG) · K163453

# VitalPatch® VitalConnect Platform (K163453)

_Vitalconnect, Inc. · DRG · Nov 2, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163453

## Device Facts

- **Applicant:** Vitalconnect, Inc.
- **Product Code:** [DRG](/productcode/DRG.md)
- **Decision Date:** Nov 2, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This can include heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data are transmitted wirelessly from the VitalConnect Sensor for storage and analysis. The VitalConnect Platform can include the ability to notify healthcare professionals when physiological data fall outside selected parameters. The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the VitalConnect Platform are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.

## Device Story

VitalPatch is a disposable, battery-powered, adhesive wearable biosensor; monitors physiological data including ECG, HR, HRV, R-R interval, respiratory rate, skin temperature, activity, and posture. Sensors collect data from patient torso; data transmitted wirelessly to relay devices and optional secure server for storage and analysis. Used in home and healthcare settings by healthcare professionals to aid diagnosis and treatment. System provides notifications when data fall outside selected parameters. Modification extends wear duration from 96 to 120 hours.

## Clinical Evidence

Bench testing only. Performance testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), battery performance, skin adherence, wear duration, sensor functionality, wireless coexistence, and firmware verification and validation.

## Technological Characteristics

Disposable adhesive patch with integrated sensor module; battery-powered; wireless data transmission. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC). Wear duration: 120 hours.

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- VitalConnect Platform ([K152139](/device/K152139.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. To the right of the blue box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 2, 2017

Vital Connect, Inc. Kevin Potgieter Senior Manager, Regulatory Affairs 900 E. Hamilton Avenue Suite 500 Campbell, California 95008

Re: K163453

Trade/Device Name: VitalPatch® VitalConnect Platform Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter And Receiver Regulatory Class: Class II Product Code: DRG, DSI, MHX Dated: October 12, 2017 Received: October 13, 2017

Dear Kevin Potgieter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

M.A. Willehemen
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K163453

#### Device Name VitalPatch® VitalConnect Platform

#### Indications for Use (Describe)

The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This can include heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data are transmitted wirelessly from the VitalConnect Sensor for storage and analysis. The VitalConnect Platform can include the ability to notify healthcare professionals when physiological data fall outside selected parameters.

The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the VitalConnect Platform are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY - Modification to VitalPatch®

# 510(k) Owner's Name, Address, and Telephone Number

VitalConnect, Inc. 224 Airport Parkway. Suite 300 San Jose, CA 95110 (408) 963-4600

# Contact Person

Kevin Potgieter, MS, RAC Senior Manager, Regulatory Affairs Direct: (408) 963-4603 (408) 963-2828 Fax: Email: kpotgieter@vitalconnect.com

510(k) Number: K163453

Date Prepared: 30 October 2017

Trade Name of Device: VitalPatch® VitalConnect Platform

Common or Usual Name: Wearable Biosensor

## Classification Name:

21 CFR 870.2910 - Transmitters and Receivers, Physiological Signal, Radiofrequency

Product Code: DRG

## Predicate Device(s) VitalConnect Platform by VitalConnect, Inc. (K152139)

## Intended Use

The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings.

## Indications for Use

The VitalConnect Platform is a wireless remote monitoring system intended for use by healthcare professionals for continuous collection of physiological data in home and healthcare settings. This can include heart rate, electrocardiography (ECG), heart rate variability, R-R interval, respiratory rate, skin temperature, activity (including step count), and posture (body position relative to gravity including fall). Data are transmitted wirelessly from the VitalConnect Sensor for storage and analysis. The VitalConnect Platform can include the ability to notify healthcare professionals when physiological data fall outside selected parameters.

{4}------------------------------------------------

The device is intended for use on general care patients who are 18 years of age or older as a general patient monitor, to provide physiological information. The data from the VitalConnect Platform are intended for use by healthcare professionals as an aid to diagnosis and treatment. The device is not intended for use on critical care patients.

#### Device Description

VitalPatch is a wearable biosensor which may be used as part of the VitalConnect Platform. The device is a disposable, battery-powered, adhesive patch containing sensors used to gather patients' physiological data, which are transmitted wirelessly. The modification proposed in this 510(k) increases the wear duration for the VitalPatch wearable biosensor from 96 hours to 120 hours.

#### Technological Characteristics

This 510(k) does not involve any changes to the technological characteristics of the VitalPatch wearable biosensor.

#### Performance Data

The following device performance tests were performed on the VitalPatch wearable biosensor during the early design and development phase:

| Test Description                        | Standard/Guidance Referenced                                  |
|-----------------------------------------|---------------------------------------------------------------|
| Electrical Safety Testing               | IEC 60601-1, 3.1 Ed.                                          |
| Electromagnetic Compatibility Testing   | IEC 60601-1-2, 4th Ed.                                        |
| Battery Performance Testing             | N/A – internal performance testing                            |
| Skin Adherence/Wear Duration            | N/A – internal performance testing                            |
| Wear Performance (Sensor Functionality) | N/A – internal performance testing                            |
| Wireless Coexistence                    | FDA Radio Frequency Wireless<br>Technology in Medical Devices |

The following device performance tests were performed on the VitalPatch wearable biosensor during the commercialization phase, in support of this 510(k) submission for the extension of wear duration from 96 to 120 hours:

| Test Description                     | Standard/Guidance Referenced                                                                          |
|--------------------------------------|-------------------------------------------------------------------------------------------------------|
| Firmware Verification and Validation | FDA Guidance for the Content of<br>Premarket Submissions for Software<br>Contained in Medical Devices |

{5}------------------------------------------------

# Substantial Equivalence

VitalPatch with the 120-hour wear duration has the same intended use, indications for use, and technological characteristics as the predicate device. Thus, VitalPatch is substantially equivalent to the predicate device.

## Conclusions

VitalPatch is substantially equivalent to the predicate device. The modifications described in this 510(k) do not raise different questions of safety and effectiveness.

| Summary of Substantial Equivalence Table |                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                 |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Topic                                    | Predicate Device<br>K152139                                                                                                                                                                                                     | Modified Device<br>K163453                                                                                                                                                                                                      |
| Classification                           | Class II<br>21 CFR 870.2910                                                                                                                                                                                                     | Class II<br>21 CFR 870.2910                                                                                                                                                                                                     |
| Classification<br>Name                   | Transmitters and Receivers,<br>Physiological Signal, Radiofrequency                                                                                                                                                             | Transmitters and Receivers,<br>Physiological Signal, Radiofrequency                                                                                                                                                             |
| Product Code                             | DRG                                                                                                                                                                                                                             | DRG                                                                                                                                                                                                                             |
| Trade Name                               | VitalPatch®                                                                                                                                                                                                                     | VitalPatch®                                                                                                                                                                                                                     |
| Patch<br>Configuration                   | Disposable adhesive patch containing<br>integrated disposable sensor module                                                                                                                                                     | Disposable adhesive patch containing<br>integrated disposable sensor module                                                                                                                                                     |
| Shelf Life                               | 9 months                                                                                                                                                                                                                        | 9 months                                                                                                                                                                                                                        |
| Vital Signs<br>Monitored                 | Heart rate, electrocardiography<br>(ECG), heart rate variability, R-R<br>interval, respiratory rate, skin<br>temperature, activity (including step<br>count), and posture (body position<br>relative to gravity including fall) | Heart rate, electrocardiography<br>(ECG), heart rate variability, R-R<br>interval, respiratory rate, skin<br>temperature, activity (including step<br>count), and posture (body position<br>relative to gravity including fall) |
| Wear<br>Duration                         | 96 hours                                                                                                                                                                                                                        | 120 hours                                                                                                                                                                                                                       |
| Principles of<br>Operation               | Adhesive patches affixed to user's<br>torso allowing for continuous<br>monitoring of wearer's physiological<br>data. Wireless transmission and<br>storage of data via Relay devices and<br>optional Secure Server.              | Adhesive patches affixed to user's<br>torso allowing for continuous<br>monitoring of wearer's physiological<br>data. Wireless transmission and<br>storage of data via Relay devices and<br>optional Secure Server.              |

Summary of Substantial Equivalence Table

---

**Source:** [https://fda-staging.innolitics.com/device/K163453](https://fda-staging.innolitics.com/device/K163453)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
