K163450 · Starmed Co., Ltd. · GEI · Mar 8, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163450
Device Name
VIVA combo RF System
Applicant
Starmed Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VIVA combo RF Ablation System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue.
Device Story
VIVA combo RF System is an electrosurgical unit for tissue coagulation and ablation. System components include an RF generator, active electrode, grounding pad, and peristaltic pump for electrode cooling. The generator delivers up to 200 W of RF power at 480 kHz. During operation, the peristaltic pump circulates chilled water through the electrode to prevent overheating; the cooling system is enclosed and does not contact the patient. The generator monitors power, resistance, current, and temperature, with power output limited by software control. Used in clinical settings (percutaneous or intraoperative) by healthcare professionals. The device provides localized tissue heating at the electrode tip to achieve coagulation or ablation, aiding in the treatment of target tissue lesions.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by bench testing, including electrical safety (AAMI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), performance testing (IEC 60601-2-2), software validation, and biocompatibility testing (cytotoxicity, irritation, sensitization, acute systemic toxicity).
Technological Characteristics
Monopolar electrosurgical system. Components: RF generator, active electrode, grounding pad, peristaltic pump, foot pedal. Output: 480 kHz ± 10%, max 200 W @ 50 ohm. Cooling: Closed-loop chilled water system. Sterilization: Ethylene oxide for single-use electrodes. Software: Moderate level of concern. Compliance: AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-2.
Indications for Use
Indicated for percutaneous and intraoperative coagulation and ablation of tissue in patients requiring electrosurgical intervention.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2017
STARmed Co., Ltd. Jun-Young Jung Deputy General Manger B-dong, 4F, 158, Haneulmaeul-ro, IlsanDong-gu, Goyang-si Gyeonggi-do, 10355 Korea
Re: K163450
Trade/Device Name: VIVA Combo RF System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: March 3, 2017 Received: March 7, 2017
Dear Jun-Young Jung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the logo for STARmed. The logo features a stylized orange circle with the word "STAR" in orange, bold letters inside the circle. To the right of the circle, the word "med" is written in gray, sans-serif letters. The overall design is clean and modern, suggesting a company in the medical or healthcare industry.
| | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| | Indications for Use | |
| 510(k) Number (if known) | K163450 | |
| Device Name | VIVA combo RF System | |
| Indications for Use (Describe) | The VIVA combo RF Ablation System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue. | |
Type of Use (Select one or both, as applicable)
| <span> × Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for STARmed. The logo consists of the word "STAR" in orange, with the letters arranged inside of an orange circle. To the right of the word "STAR" is the word "med" in gray. The logo is simple and modern.
B-dong, 4F, 158, Haneulmaeul-ro, IlsanDong-gu, Goyang-si Gyeonggi-do, Republic of Korea
## 510(k) Summary
[As required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR 807.92(a)(a)]
December 6, 2016
#### Submitter's Information [21 CFR 807.92(a)[1]] 2.
- . Name of Sponsor: STARmed Co., Ltd. Address: B-dong, 4F, 158, Haneulmaeul-ro, IlsanDong-gu, Goyang-si, -Gyeonggi-do, Republic of Korea
- Contact Name: Jun-Young Jung Telephone No.: +82 70-4673-8628 -Fax No.: +82 31-816-4546 -
- -Email Address: jjy3412@starmed4u.com
- . Registration Number: In process
- Name of Manufacturer: Same as Sponsor -Address: Same as Sponsor
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
- . Trade Name: VIVA combo RF System . Common Name: Electrosurgical Cutting and Coagulation Device and Accessories
- . Classification:
| Classification Name | Electrosurgical, Cutting & Coagulation & Accessories |
|---------------------------|------------------------------------------------------|
| Classification Regulation | 21 CFR 878.4400 |
| Product Code | GEI |
| Device Class | II |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for STARmed. The logo consists of the word "STARmed" in a stylized font. The "STAR" portion of the name is in orange, while the "med" portion is in gray. To the left of the word "STAR" is an orange circular design that appears to be made of multiple curved lines.
#### Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] 4.
The identified predicate devices within this submission are shown as follow:
### Predicate Device
| • | 510(k) Number: | K042216 |
|---|------------------|----------------------------------------------------------------|
| • | Applicant: | Valleylab |
| • | Regulation Name: | Electrosurgical cutting and coagulation device and accessories |
| • | Trade Name: | Cool-tip™ RF Ablation System |
There are no significant differences between VIVA combo RF System and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
#### 5. Description of the Device [21 CFR 807.92(a)[4]]
The VIVA combo RF System consists of RF generator, active electrode, grounding pad and peristaltic pump for electrode cooling. This device is designed to produce local tissue heating at the tip of the electrodes causing the coagulation and ablation of tissue.
The VIVA combo RF System is capable of delivering up to 200 W of RF power and the available power is limited through software control. This generator monitors the power, resistance, current and temperature.
The active electrodes are a sterile, single-use, hand-held electrosurgical instrument designed for use with VIVA combo RF System. Cooling of the electrode is provided by chilled water which is pumped through the inflow tubing, the electrode and out through the outflow tubing. This is an enclosed system within the electrode and the water is not to be in contact with the patient.
#### Intended Use [21 CFR 807.92(a)(5)] 6.
The VIVA combo RF System is intended for use in percutaneous and intraoperative coagulation and ablation of tissue.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for STARmed. The logo features the word "STAR" in orange, with the word "med" in gray next to it. To the left of the word "STAR" is an orange circle with a break in the upper left.
#### 7. Technological Characteristics [21 CFR 807.92(a)(6)]
The VIVA combo RF System is substantially equivalent to legally marketed predicate device with respect to indications for use and technology characteristics. The table below presents side by side comparisons for each major component of each device:
| Item | Proposed Device | Primary Predicate |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Not known | K042216 |
| Model | VIVA combo RF System | Cool-tip™ RF Ablation<br>System |
| Manufacture | STARmed Co., Ltd. | Valleylab |
| Intended Use | The VIVA combo RF System is<br>intended for use in<br>percutaneous and<br>intraoperative coagulation<br>and ablation of tissue. | The Cool-tip™ RF Ablation<br>System (Generator and<br>Accessories) is intended for<br>use in percutaneous,<br>laparoscopic and<br>intraoperative coagulation<br>and ablation of tissue, such as<br>partial or complete ablation<br>of non-resectable liver<br>lesions. |
| Type of Use | Prescription Use | Prescription Use |
| Component | Electrosurgical unit, active<br>electrode, grounding pad,<br>peristaltic pump and foot<br>pedal | Electrosurgical unit, active<br>electrode, grounding pad,<br>peristaltic pump and foot<br>pedal |
| Electrosurgical Unit | | |
| Output Frequency | $480 kHz \pm 10 \%$ | $480 kHz \pm 10\%$ |
| Drive on Time | Up to 30 minutes | Up to 30 minutes |
| Maximum Power Output | Up to 200 watts @ 50 ohm | Up to 200 watts @ 50 ohm |
| Active Accessory | | |
| Monopolar / Bipolar | Monopolar | Monopolar |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for STARmed. The logo features the word "STAR" in orange, with the word "med" in gray next to it. To the left of the word "STAR" is an orange circular design that is not fully closed.
## Non-clinical Test Summary
- . Electrical Safety, Electromagnetic Compatibility and Performance:
Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- Testing to confirm compliance with the safety requirements of standard AAMI -ES60601-1
- Testing to confirm compliance with EMC requirements of standard IEC 60601-1-2 -
- Testing to confirm compliance with particular requirements of standard IEC 60601-2-2 -
- . Software Validation
The VIVA combo RF System contains MODERATE level of concern software. The software was designed and developed according to a software development process and was verified and validated.
The software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
- Biocompatibility
The following biocompatibility tests were performed:
- Cytotoxicity -
- Irritation/intracutaneous reactivity -
- Sensitization -
- Acute Systemic Toxicity -
- . Sterilization and Shelf Life
The active electrode of VIVA combo RF System is delivered in a sterile state and is intended for single use only. The sterilization method used is ethylene oxide and this device has a shelf life of 3 years.
## Clinical Test Summarv
No clinical studies were considered necessary and performed.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for STARmed. The logo consists of the word "STARmed" in a stylized font. The "STAR" portion of the word is in orange, while the "med" portion is in gray. To the left of the word "STAR" is an orange circular design that appears to be a stylized atom.
B-dong, 4F, 158, Haneulmaeul-ro, IlsanDong-gu, Goyang-si, Gyeonggi-do, Republic of Korea
#### 8. Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
When compared to the predicate devices, the VIVA combo RF System in this submission presented the same in terms of indications for use, components and technology characteristics.
#### 9. Conclusion [21 CFR 807.92(b)(3)]
In according to with the Federal food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification STARmed Co., Ltd. concludes that VIVA combo RF System is substantially equivalent in safety and effectiveness to predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.