The Grind Care System is indicated to aid in the evaluation and management of sleep bruxism by reducing the temporalis muscle EMG activity during sleep.
Device Story
Portable EMG, electrical stimulation, and biofeedback device for sleep bruxism management. Sensor adheres to skin over temporalis muscle via disposable GelPad; records EMG activity; detects muscle contractions associated with bruxing. Upon detection, device delivers mild, biphasic, charge-balanced electrical stimulation to relax muscle and inhibit bruxing. Data transfers wirelessly via infrared to Docking Station, then via Bluetooth to mobile app (iOS/Android). Used in home environment by patient. Healthcare provider/patient reviews grinding/stimulation data and diary notes via app. Benefits include reduction of bruxing events and muscle activity. System includes Sensor, Docking Station, USB cable, power adapter, and GelPads.
Clinical Evidence
Clinical study demonstrated that the modified detection algorithm decreased the number of EMG events per hour by 32.45% over baseline. The device showed a statistically significant reduction in the number of EMG events per hour in a representative patient population, consistent with the predicate device performance.
Technological Characteristics
Class II biofeedback device. Materials: Conductive gel/adhesive tripolar electrode. Energy: 70mAh LiPo battery (IEC 62133 compliant), inductive charging. Output: Biphasic symmetrical, 40V max, 7mA max, 500µs pulse width, 230 pulses/sec. Connectivity: Infrared (Sensor to Dock), Bluetooth (Dock to App). Standards: IEC 60601-1, IEC 60601-2-40, IEC 60601-1-6, IEC 60601-1-11, ISO 10993-5/10. Microprocessor: MSP430-F5333/F5338.
Indications for Use
Indicated for patients with sleep bruxism to aid in evaluation and management by reducing temporalis muscle EMG activity during sleep.
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three human profiles facing right, with flowing lines representing movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2017
Sunstar Suisse S.A. % Calley Herzog Senior Consultant Biologics Consulting Group 400 N. Washington St., Suite 100 Alexandria, Virginia 22314
Re: K163448
Trade/Device Name: GrindCare System Regulation Number: 21 CFR 882.5050 Regulation Name: Biofeedback Device Regulatory Class: Class II Product Code: HCC, KZM, NUW Dated: August 3, 2017 Received: August 4, 2017
#### Dear Calley Herzog:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Andrew I. Steen -S
for Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163448
Device Name GrindCare System
#### Indications for Use (Describe)
The Grind Care System is indicated to aid in the evaluation and management of sleep bruxism by reducing the temporalis muscle EMG activity during sleep.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the GrindCare System is provided below.
| Device Common Name: | Biofeedback Device |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | GrindCare System |
| Applicant: | Sunstar Suisse S.A.<br>Route de Pallatex 11, 1163 Etoy<br>Switzerland<br>+41-21-821-0500 |
| Prepared by: | Calley Herzog<br>Senior Consultant<br>Biologics Consulting Group, Inc.<br>cherzog@biologicsconsulting.com<br>Phone: 720-883-3633<br>Fax: 703-548-7457 |
| Date Prepared: | August 31, 2017 |
| Classification Regulation: | 882.5050 - Biofeedback device, Class II |
| Panel / Product Code: | Neurology / HCC |
| Secondary Product Codes: | KZM - Diagnostic Electromyograph<br>NUW - Powered Muscle Stimulator, Dental |
| Predicate Device: | GrindCare, K092675 |
### Indication for Use:
The GrindCare System is indicated to aid in the evaluation and management of sleep bruxism by reducing the temporalis muscle EMG activity during sleep.
# Device Description:
The GrindCare System is a portable electromyographic (EMG), electrical stimulation, and biofeedback device. It consists of a Sensor that is adhered to the skin over the temporalis muscle by means of an adhesive, disposable, single-use GelPad. The System also includes a Docking Station, USB Cable, and Power Adaptor. The Sensor and Docking Station record and store EMG activity data, which is transferred from the Docking Station to the GrindCare Mobile App, which allows the user to review grinding and stimulation data and enter diary notes. The Sensor uses EMG to sense contraction of the temporalis muscle that is associated with bruxing events. In response to the EMG-measured contraction, it delivers mild electrical stimulation that is intended to relax the muscle and inhibit the bruxing event. Bruxism is the occurrence of
{4}------------------------------------------------
(typically sleep) grinding and clenching of the jaw and is one of the most common sleep disorders.
The components of the GrindCare System:
- . Sensor: A self-contained and battery powered component, to be placed on the skin of the patient. It records EMG activity and reacts with electrical stimulation when muscle activity is detected.
- Docking Station: A USB-powered device, where the Sensor is inserted when not in use for charging and to transfer data. Both charging and data transfer is performed wirelessly.
- USB Cable: A USB cable with a micro-USB connector in one end and a standard USB connector at the other end used for power (no data transfer is possible).
- Power Adapter
- . GelPad: GrindCare is to be used with an adhesive, single-use GelPad. One pack of 28 GelPads is provided with the device and additional GelPads can be purchased separately. The GrindCare device is not intended to be used with any other gel pads.
- GrindCare App: An iOS and Android mobile app that allows the user to review grinding and stimulation data and enter diary notes.
### Substantial Equivalence Discussion:
The predicate device is a previous version of the GrindCare System and was cleared in K092675. The predicate device is based on the same principle of operation as the subject device, but uses a modified detection algorithm. The table below provides a comparison of the technological features of the subject device to the predicate device.
| | Proposed Device | Predicate Device |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K092675 |
| Applicant | Sunstar | Medotech A/S |
| Device Name | GrindCare System | GRINDCARE |
| Classification Regulation | 882.5050 | 882.5050 |
| Product Code<br>Subsequent Product<br>Codes | HCC: Biofeedback Device<br>NUW: Powered Muscle<br>Stimulator, Dental<br>KZM: Muscle Monitoring<br>Device, Dental | HCC: Biofeedback Device<br>NUW: Powered Muscle Stimulator,<br>Dental<br>KZM: Muscle Monitoring Device,<br>Dental |
| Number of Output<br>Channels | 1 | 1 |
| Number of Output Modes | 1 | 1 |
| Waveform | Biphasic symmetrical | Biphasic symmetrical |
| | Proposed Device | Predicate Device |
| Shape | Rectangular | Rectangular |
| Placement | On the skin at the temple, over<br>the Temporalis Anterior muscle. | On the skin at the temple, over the<br>Temporalis Anterior muscle. |
| Regulated Current or<br>Regulated Voltage? | Regulated Current | Regulated Current |
| Maximum Output<br>Voltage | 40 V | 40 V |
| Maximum Output<br>Current | 7 mA | 7 mA |
| Pulse Width | $500 \mu s$ | $500 \mu s$ |
| Frequency | 230 pulses/sec | 230 pulses/sec |
| Net Charge | 0 (charge balanced) | 0 (charge balanced) |
| Maximum Phase Charge | $1.75\mu C$ | $1.75\mu C$ |
| Maximum Current<br>Density | $35mA/cm^2$ | $35mA/cm^2$ |
| Maximum Power Density | $0.022 W/cm^2$ | $0.022 W/cm^2$ |
| Electrode | Conductive gel w/adhesive<br>galvanic tripolar electrode | Conductive gel w/adhesive galvanic<br>tripolar electrode |
| Input Impedance | 4.4MOhms | 4.4MOhms |
| Common Mode Rejection<br>Ratio | 90dB | 90dB |
| Power Source / Charging | 70mAh LiPo<br>IEC 62133 compliant Lithium-<br>ion rechargeable battery in the<br>Sensor<br>Inductive (wireless) link for<br>charging the Sensor battery. | 340mAh LiPo<br>Lithium-ion rechargeable battery in<br>the Stimulator and GrindDock<br>Gold spring contacts for charging<br>the Stimulator when docked in the<br>GrindDock. |
| Average Battery Charge<br>Time | 3.5 hours | 3.5 hours |
| Communication | Infrared link for transferring data<br>from Sensor to Docking Station.<br>Bluetooth communication for<br>transferring data from Docking<br>Station to GrindCare App. | Zigbee type radio link for<br>communication between Stimulator<br>and GrindDock |
| Data Storage | Sensor stores only a single<br>session of data. Docking Station<br>stores up to 5 years of data. | Stimulator stores up to 30 sessions<br>of data.<br>No data stored on the GrindDock |
| | Proposed Device | Predicate Device |
| PCB Size | 35x19mm | 45x30mm |
| Microprocessor | Docking Station:<br>MSP430-F5333 processor<br>Sensor:<br>MSP430-F5338 processor | GrindDock:<br>MSP430-F2618 processor<br>Stimulator:<br>MSP430-F2618 processor |
| Viewing the Data | Bluetooth type radio link for<br>communication between Docking<br>Station and GrindCare App on an<br>iOS or Android device. There is<br>no display on the device.<br>Mobile App is used to view the<br>data. | Data transferred by USB to a PC.<br>LCD display on the GrindDock. |
| Grind Threshold | Automatically calculated<br>continuously based on<br>background noise. | Determined by user input before<br>each use. |
{5}------------------------------------------------
{6}------------------------------------------------
## Comparison of Indications:
Predicate device indication statement (K092675):
The GRINDCARE device is indicated to aid in the evaluation and management of nocturnal bruxism by reducing the temporalis muscle EMG activity during sleep.
Subject device indication statement:
The GrindCare System is indicated to aid in the evaluation and management of sleep bruxism by reducing the temporalis muscle EMG activity during sleep.
As shown above, the indication statement of the subject device is equivalent to that of the predicate device. The term "nocturnal bruxism" has been replaced with "sleep bruxism", to reflect the generally accepted terminology used in the scientific community to describe the behavior, as this type of bruxism is linked to the awake/sleep state and not to the time of day/night. The change is a clarification of the indication statement and does not represent a change to the intended use.
### Discussion of Technological Differences:
Power Source / Charging - In the predicate device both the GrindDock and the Stimulator were battery powered. In the subject device the Sensor battery is charged wirelessly. The lithium-ion battery complies with the FDA-recognized standard IEC 62133. Power to the Docking Station comes from a USB adapter or a USB port. Performance testing demonstrated the functionality of the wireless charging.
Communication - The predicate device used a Zigbee type radio link for communication between Stimulator and GrindDock. The subject device uses an infrared link for transferring data from Sensor to Docking Station and Bluetooth communication for transferring data from the Docking Station to the GrindCare App. Performance testing demonstrated the functionality of
{7}------------------------------------------------
communication between the Sensor and the Docking Station and between the Sensor and the GrindCare App.
Data Storage - In the predicate device, the Stimulator stored up to 30 nights of data. In the subject device, the Sensor stores one night of data, but every time it is inserted into the Charger, the data is transferred. The Charger can store data from up to 5 years of typical use. Performance testing demonstrated the functionality of the data storage and transfer.
PCB Size - All input and output circuitry relating to patient contact is unchanged and specifications are the same. Only the circuitry that generates the compliance voltage for the stimulation circuit has been redesigned, to reduce size and improve efficiency. Performance testing demonstrated the functionality of the PCB.
Microprocessor - The microprocessor chosen for the subject device is similar to the processor in the predicate device. There are no differences that have any impact on functionality or performance requirements for the subject device compared to the predicate device. Performance testing demonstrated the functionality of the microprocessor.
Viewing the Data - In the predicate device, the data was viewed on the GrindDock device, on the LCD display or was uploaded to a PC to be viewed in a dedicated program, the GrindCare Manager. In the subject device, the data is viewed on a portable device, based on either the Android operating system or the iOS operating system (iPhones, iPads). Performance testing demonstrated the functionality of the GrindCare App.
Grind Threshold - Like the predicate device, the subject device detects grinds by comparing the amplitude of the signal generated by the user's muscle activity to a baseline, or background noise level. The subject device calculates the threshold differently than the predicate device. This difference in setting the threshold for a grinding event does not change the intended use of the device as a biofeedback system for reducing muscle activity during sleep. For both devices, a grind is defined as measured muscle activity that is quantifiably above background noise. Once a grind is detected, the subject device provides the identical stimulation signal as was provided by the predicate device. Performance testing demonstrated equivalent performance of the grind detection algorithm to the predicate device.
Shelf Life - The 510(k) for the predicate device did not indicate a shelf life, the GelPads are now labeled with a shelf life of 24 months. Performance testing demonstrated that the device meets specifications after 24-months.
| Type | Description of Testing or Standard Referenced |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Verification<br>Testing | Hardware, Algorithm and System Level Testing was completed to verify that the<br>device meets design and performance specifications and is equivalent to the<br>predicate device. |
| Safety | IEC 60601-1: 2005 + CORR1:2006 + CORR.2:2007 + AM1:2012<br>Medical electrical equipment -- Part 1: General Requirements for basic safety and<br>essential performance |
# Performance Data Provided to Establish Substantial Equivalence:
{8}------------------------------------------------
| Type | Description of Testing or Standard Referenced |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | IEC 60601-2-40: 1998 Medical electrical equipment - Part 2-40: Particular<br>requirements for the safety of electromyographs and evoked response equipment |
| | IEC 60601-1-6: 2010 (Third Edition) + A1:2013 Medical electrical equipment Part<br>1-6 General requirements for safety - Collateral Standard: Usability |
| | IEC 60601-1-11: 2010 (First edition) Medical Electrical Equipment - Part 1-11:<br>General Requirements for Basic Safety and Essential Performance – Collateral<br>Standard: Requirements for Medical Electrical Equipment and Medical Electrical<br>System Used in Home Healthcare Environment |
| EMC | IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment - part 1-2: general<br>requirements for basic safety and essential performance - collateral standard:<br>electromagnetic compatibility - requirements and tests. |
| Battery<br>Certification | IEC 62133 Edition 2.0: 2012-12, secondary cells and batteries containing alkaline<br>or other non-acid electrolytes - safety requirements for portable sealed secondary<br>cells, and for batteries made from them, for use in portable applications |
| Wireless<br>Coexistence | Testing to demonstrate wireless coexistence with other FR transmitters common in<br>the home environment. |
| Biocompatibility<br>of Gelpad | ISO 10993-5:2009 biological evaluation of medical devices - part 5: tests for in<br>vitro cytotoxicity. |
| | ISO 10993-10: 2010 biological evaluation of medical devices - part 10: tests for<br>irritation and skin sensitization |
| Shelf Life | Shelf life testing was performed that demonstrates the device continues to meet<br>performance specifications for the 24-month shelf life of the gelpads. |
| Usability<br>Testing | Subjects were observed while completing device and app critical tasks. Use error<br>information (including an evaluation of labeling comprehension) was analyzed.<br>The study demonstrated that users and prescribers were able to perform the tasks<br>associated with use of the GrindCare device with the GrindCare App. |
| Clinical Data | The clinical study demonstrated that the modified detection algorithm decreased<br>the number of EMG events per hour by 32.45% over baseline. |
| | Like the predicate device, the subject device demonstrated a statistically significant<br>reduction on the number of EMG events per hour on a representative patient<br>population. |
# Substantial Equivalence:
The predicate device is a previous version of the GrindCare System and was cleared in K092675. The subject device has the same intended use as the predicate device and delivers the identical stimulation signal. The technical differences between the subject device and the predicate device are supported by performance testing to demonstrate that the differences do no raise new questions of safety or effectiveness. Therefore, the subject device can be found substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.