Single Use Micro Electrode (Brass with gold plated shaft); Single Use Micro Electrode (Stainless Steel shaft)

K163442 · Modern Medical Equipment Manufacturing Limited · GEI · Sep 28, 2017 · General, Plastic Surgery

Device Facts

Record IDK163442
Device NameSingle Use Micro Electrode (Brass with gold plated shaft); Single Use Micro Electrode (Stainless Steel shaft)
ApplicantModern Medical Equipment Manufacturing Limited
Product CodeGEI · General, Plastic Surgery
Decision DateSep 28, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The electrode is used to cut and / or coaqulate soft tissues by means of high frequency electrical current during an electrosurgical procedure

Device Story

Monopolar active electrode; consists of conductive tip, insulated shaft, and conductive post. Tip configurations include micro needles, scalpel blade, ball, broadbent, wire, and various loops. Insulated with Polyolefin or PTFE. Plugs into standard electrosurgical pencil; receives high-frequency current from compatible generator. Used in conjunction with patient grounding pad. Operated by clinicians during electrosurgical procedures to cut or coagulate soft tissue. Benefits include precise tissue management via various tip geometries.

Clinical Evidence

Bench testing only. Comparative performance testing conducted on porcine tissue (kidney, liver, muscle) in both cut and coagulation modes. Results demonstrated no significant difference in thermal effect or tissue stickiness compared to predicate devices. Biocompatibility testing (intracutaneous reactivity, in vitro cytotoxicity, skin sensitization) confirmed no biological effects.

Technological Characteristics

Monopolar electrode; conductive tip (various geometries), insulated shaft (Polyolefin/PTFE), conductive post (2.36mm or 1.6mm). Energy source: high-frequency electrosurgical generator. Electrical safety: IEC 60601-1:2005, IEC 60601-2-2:2009. Limited contact duration (<24 hours).

Indications for Use

Indicated for cutting and/or coagulating soft tissues during electrosurgical procedures using high-frequency electrical current.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three faces in profile, one behind the other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the figure in a circular pattern. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 28, 2017 Modern Medical Equipment Manufacturing Limited Mr. Samson Woo Quality Assurance Manager 35-41 Tai Lin Pai Rd., Unit F, 5th Fl. Gold King Ind. Bldg. Kwai Chung, N.T. Hong Kong, China Re: K163442 Trade/Device Name: Single Use Micro Electrode (Brass with gold plated shaft and Stainless Steel shaft) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 14, 2017 Received: August 3, 2017 Dear Mr. Woo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 {1}------------------------------------------------ CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K163442 Device Name Single Use Micro Electrode Indications for Use (Describe) The electrode is used to cut and / or coagulate soft tissues by means of high frequency electrical current during an electrosurgical procedure X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | 1. 510(k) Owner | | |----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Name: | Modern Medical Equipment Manufacturing Limited | | Address: | Modern Medical Equipment Mfg., Ltd.<br>35-41 Tai Lin Pai Rd., Unit F,<br>5th Fl., Gold King Ind. Bldg<br>Kwai Chung, N.T. Hong Kong, China | | Telephone: | 86 (768) 8897 8999 | | Fax: | 86 (768) 8897 8666 | | Contact person: | Samson Woo | | Date of preparation: | 26 July, 2017 | | 2. Device | | | Name of Device: | Single Use Micro Electrode | | Trade or proprietary name: | Single Use Micro Electrode | | Common or usual name: | Single Use Micro Electrode | | Classification name: | Electrosurgical, Cutting & Coagulation & Accessories | | Classification Panel: | General & Plastic Surgery | | Product Code: | GEI | | Class: | II | # 3. Device description The Single Use Micro Electrodes are the monopolar active device which consist of a conductive electrode tip, an insulated shaft and a conductive post. The conductive electrode tip configuration includes micro needles, scalpel blade, ball, broadbent, wire, triangle loop, diamond loop and round loop. The conductive electrode tip is the conductive portion of the electrode which delivers high frequency current onto a soft tissue for cutting and coagulation during an electro-surgery The insulated shaft is the mid portion of the electrode. It is insulated with non-conductive material included Polyolefin and /or PTFE Shrink Wrap, overmold, in order to prevent accidental conduction of the electrosurgical current from this point to the patient. The conductive post is used to pluq the electrode into the nozzle of an electrosurgical pencil from which it receives the electrical current. The diameter of the conductive post is available in 2.36mm and 1.6mm. This device is used in conjunction with electrosurgical pencil, a compatible monopolar electrosurgical generator and patient grounding pad during electrosurgical procedure. {4}------------------------------------------------ #### 4. Intended use The electrode is used to cut and / or coaqulate soft tissues by means of high frequency electrical current during an electrosurgical procedure # 5. Predicate device Ellman's Ace-Tip Electrodes with 510(k) number K071343. and Unimed's Needle Electrodes with 510(k) number K944265 # 6. Technological characteristic The proposed Single Use Micro Electrode has equivalent construction and performance as the predicate devices. Different electrode tip configuration accommodate various procedural conditions. There are no new questions raised regarding to effectiveness and safety. # 7. Substantial Equivalence The technological characteristic in both proposed devices and predicate devices are identical. Moreover, performance testing data verified that operation result by using proposed device and predicate device shall be the same. # 8. Performance Testing Data Validation and Verification testing was performed on the device and its packaging. Operation Tests include comparative testing was performed on porcine tissue: kidney, liver, and muscle for both Cut mode and Coagulation mode. It demonstrates that the thermal effect and tissue stickiness on the tip of electrode are no significant difference between proposed device and predicate device. # 9. Safety and Biocompatibility testing data Electrical Safety complies with IEC60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) IEC60601-2-2: 2009 (Fifth Ed) The proposed device is a external device that contact tissue/bone/dentin for a limited contact duration (less than 24 hours). No biological effect resulted in below three tests - · Intracutaneous Reactivity - In vitro Cytotoxicity . - · Skin sensitization # 10. Conclusion Based on comparing technological characteristic and performance testing data, we believe that the proposed device, Single Use Micro Electrodes are substantial equivalence to predicate devices (K091672 and K944265).
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...