← Product Code [PGW](/productcode/PGW) · K163416

# Fiagon Navigation – PointerTube Straight and PointerTube Keat (K163416)

_Fiagon GmbH · PGW · Feb 16, 2017 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163416

## Device Facts

- **Applicant:** Fiagon GmbH
- **Product Code:** [PGW](/productcode/PGW.md)
- **Decision Date:** Feb 16, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The Fiagon Navigation - PointerTube Straight and the PointerTube Keat are intended as aids for precisely locating anatomical structures in either open or percutaneous procedures. They are indicated for use with the Fiagon Navigation system using electromagnetic navigation. The instruments are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy. Example procedures include, but are not limited to: ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access

## Device Story

Fiagon Navigation PointerTube Straight and PointerTube Keat are 10-time use surgical instruments for ENT procedures; used with Fiagon Navigation system. Each instrument contains an integrated electromagnetic sensor tracked within a low-energy magnetic field generated by the system. Navigation software maps instrument position onto preoperative CT, MRI, or fluoroscopy scans. Used by surgeons in clinical settings to precisely locate anatomical structures during open or percutaneous procedures. Output is real-time visual guidance on a display, assisting surgeons in navigating complex anatomy like paranasal sinuses or skull base. Benefits include improved surgical precision and anatomical orientation.

## Clinical Evidence

Bench testing only. Testing confirmed continued precision, accuracy, functionality, and compatibility with the Fiagon Navigation system following design modifications.

## Technological Characteristics

Electromagnetic navigation instruments; 10-time use; integrated sensor for tracking in low-energy magnetic field; compatible with Fiagon Navigation system; dimensions and tip geometry modified from predicates.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Fiagon Navigation - PointerTube Straight ([K141456](/device/K141456.md))
- Fiagon Navigation - PointerTube Sinus Frontalis ([K141456](/device/K141456.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles a stylized caduceus or a representation of human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 16, 2017

Fiagon Gmbh % Ms. Yarmela Pavlovic Regulatory Counsel Hogan Lovells US LLP 3 Embarcadero Center, Ste. 1500 San Francisco, CA 94111

Re: K163416

Trade/Device Name: Fiagon Navigation - Pointertube Straight And Pointertube Keat Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: PGW Dated: January 18, 2017 Received: January 18, 2017

Dear Ms. Pavlovic:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K163416

### Device Name

Fiagon Navigation - PointerTube Straight and PointerTube Keat

Indications for Use (Describe) Indications for Use (Describe)

The Fiagon Navigation - PointerTube Straight and the PointerTube Keat are intended as aids for precisely locating anatomical structures in either open or percutaneous procedures. They are with the Fiagon Navigation system using electromagnetic navigation.

The instruments are indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to:

ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span style="font-size: 20px;">☑</span> </div> |
|----------------------------------------------|------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div> <span style="font-size: 20px;">☐</span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# SECTION 5 - 510(K) SUMMARY

# 510(k) Summary

## December 5, 2016

#### 1. Submitter Information/ 510(k) Holder

| Submitter: | Fiagon GmbH                |
|------------|----------------------------|
| Address:   | Neuendorfstrasse 23b       |
|            | 16761 Hennigsdorf, Germany |

| Telephone: | +49 3302 201 21 10 |
|------------|--------------------|
| Telefax:   | +49 3302 201 21 15 |

Mr. Dirk Mucha, CTO Contact:

#### 2. Device Information

| Trade Name:     | Fiagon Navigation – PointerTube Straight and PointerTube Kea |
|-----------------|--------------------------------------------------------------|
| Common Name:    | Image guided surgery system                                  |
| Classification: | Class II per 21 CFR 882.4560                                 |
| Device:         | Ear, Nose, and Throat Stereotaxic Instrument                 |
| Product Code:   | PGW                                                          |

#### 3. Purpose of Submission

The purpose of this submission is to gain clearance for modified versions of two previously cleared instruments, Fiagon Navigation - PointerTube Straight (K141456) and PointerTube Sinus Frontalis (K141456). The modified version of the PointerTube Straight has modified olive leaf-shaped tip with a different inner and outer diameter. The tube length remains identical in this case. The Sinus Frontalis has been modified in instrument length as well as diameter to a new instrument - the PointerTube Keat.

#### 4. Predicate Device Information

Fiagon Navigation - PointerTube Straight (K141456) and PointerTube Sinus Frontalis (K141456)

{4}------------------------------------------------

#### ક. Device Description

The Fiagon Navigation - PointerTube Straight and the PointerTube Keat are 10 time use instruments intended to be used with the Fiagon Navigation system.

Each device incorporates a sensor, which is tracked by the navigation system within the lowenergy magnetic field of a field generator (part of the navigation system). The navigation software (part of the navigation system) displays the position of the instruments in preoperative scans (e.g., CT, MRI, fluoroscopy).

#### 6. Intended Use

The Fiagon Navigation - PointerTube Straight and the PointerTube Keat are intended as aids for precisely locating anatomical structures in either open or percutaneous procedures. They are indicated for use with the Fiagon Navigation system using electromagnetic navigation.

The instruments are indicated for any medical condition in which the use of stereotactic surgerv may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery, such as the paranasal sinuses, mastoid anatomy, can be identified relative to a CT or MR based model of the anatomy.

Example procedures include, but are not limited to:

ENT Procedures; Transphenoidal access procedures. Intranasal procedures. Sinus procedures, such as Maxillary antrostomies, Ethmoidectomies, Sphenoidotomies/Sphenoid explorations, Turbinate resections, and Frontal sinusotomies. Skull base procedures for ENT access

#### Comparison of Technological Characteristics 7.

The PointerTube Straight and the PointerTube Keat are modified versions of previously cleared PointerTube Straight (K141456) and PointerTube Sinus Frontalis (K141456) respectively. The reason for this Special 510(k) is to describe changes in the diameter and shape of the tip for the PointerTube Straight and change of diameter, length and tip for the PointerTube Keat. These changes do not raise new issues of safety and effectiveness of the instruments.

{5}------------------------------------------------

#### 8. Performance Data

Testing was performed in order to confirm continued precision and accuracy of the modified devices. Testing was also completed to ensure functionality and compatibility with the Fiagon Navigation system, as well as to confirm that the change in design of the modified instrument does not alter the performance characteristics of the device.

#### 9. Conclusion

Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, the PointerTube Straight and the PointerTube Keat have been shown to be substantially equivalent to the predicate PointerTube Straight and the PointerTube Sinus Frontalis, respectively, and do not present any new issues of safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K163416](https://fda-staging.innolitics.com/device/K163416)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
