K163401 · Medtech Products, Inc. · LJL · Mar 21, 2017 · General Hospital
Device Facts
Record ID
K163401
Device Name
Nix Ultra Lice Treatment Kit
Applicant
Medtech Products, Inc.
Product Code
LJL · General Hospital
Decision Date
Mar 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5960
Device Class
Class 1
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K163401 · Mar 21, 2017
Nix Ultra Lice Treatment Kit
Medtech Products, Inc.
Post-marketing adverse event reports from European sales history
Post-marketing surveillance data was used to support the safety profile of the device by analyzing adverse event reports collected during routine commercial use.
European Post-Marketing Surveillance; Post-marketing surveillance / Adverse event reporting; Follow-up/Duration: June 2007 to December 2015; Study Period: 2007-2015
Users of the device in Europe; Sample Size: 518,095 units sold
Not applicable for this study
Incidence of adverse events
Indications for Use
Nix® Ultra Lice Treatment Kit is intended to kill and remove head lice and their eggs from adults and children two years of age or older.
Device Story
Kit consists of dimethicone/isoparaffin solution and fine-tooth removal comb; used at home by patients or caregivers. Solution physically coats lice/nits, clogging spiracles to induce suffocation and inhibit water exchange; lubricant facilitates mechanical removal of dead lice/nits via comb. Treatment duration 10 minutes plus combing time. Provides pediculicidal and ovicidal action. Benefits include effective lice/egg elimination, including permethrin- and DDT-resistant strains, without chemical neurotoxicity.
Clinical Evidence
Randomized, investigator-blinded study (n=52) compared Nix Ultra to reference product in subjects aged 2-63. Primary endpoint (Day 10 cure rate) showed 92% success for Nix Ultra vs 96% for reference (p > 0.5). Secondary endpoints at Days 1 and 7 showed comparable efficacy. Adverse events were rare, mild, and temporary. European post-marketing data (518,095 units) reported 5 non-serious AE cases.
Technological Characteristics
Formulated solution containing dimethicone and isoparaffin; mechanical removal via fine-tooth comb. Non-sterile. Biocompatibility per ISO 10993-1 (cytotoxicity, skin/corneal irritation, sensitization).
Indications for Use
Indicated for adults and children 2 years of age or older with head lice infestation.
Regulatory Classification
Identification
The lice removal kit is a comb or comb-like device intended to remove and/or kill lice and nits from head and body hair. It may or may not be battery operated.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol of three stylized human profiles facing right, with flowing lines beneath them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 21, 2017
Medtech Products Inc. c/o Vincent Argiro Prestige Brands Holdings 660 White Plains Road Tarrytown, New York 10591
Re: K163401
Trade/Device Name: Nix® Ultra Lice Treatment Kit Regulation Number: 21 CFR 880.5960 Regulation Name: Lice Removal Kit Regulatory Class: I Product Code: LJL Dated: February 7, 2017 Received: February 8, 2017
Dear Vincent Argiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Tina Kiang
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163401
Device Name Nix® Ultra Lice Treatment Kit
Indications for Use (Describe)
Nix® Ultra Lice Treatment Kit is intended to kill and remove head lice and their eggs from adults and children two years of age or older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
# Nix® Ultra Lice Treatment Kit
### K163401
#### 1. Submitter
| Name & Address: | Medtech Products Inc.<br>660 White Plains Road<br>Tarrytown, NY, 10591 |
|-----------------|------------------------------------------------------------------------|
| Contact: | Jean Boyko, Ph.D. |
| Title: | Sr. Vice President, Science & Technology |
| Phone number: | (914) 524-6894 |
#### 2. Official Correspondent
| Name & Address: | Medtech Products Inc. |
|-----------------|------------------------------|
| | 660 White Plains Road |
| | Tarrytown, NY, 10591 |
| Contact: | Vincent Argiro, RAC |
| Title: | Director, Regulatory Affairs |
| Phone number: | (914) 524-8721 |
| Email: | vargiro@prestigebrands.com |
#### 3. Date Prepared
February 7th, 2017
### 4. Device Identification
| Trade/Proprietary Name: | Nix® Ultra Lice Treatment Kit |
|-------------------------|------------------------------------------------------|
| Common/Usual Name: | Over-the-counter lice elimination product |
| Classification Name: | Lice removal kit |
| Regulation Number: | 21 C.F.R. § 880.5960 |
| Product Code: | LJL [Detectors and removers, lice (including combs)] |
| Device Class: | Class I, 510(k) exempt |
| Classification Panel: | General Hospital and Personal Use Devices |
{4}------------------------------------------------
#### 5. Legally Marketed Predicate Device
Clear™ Lice Egg Remover System (K981147), Product Code LJL
#### 6. Device Description
The Nix® Ultra Lice Treatment Kit consists of two components: Nix® Ultra Lice Killing Solution and Nix® Ultra Lice Removal Comb. The kit also includes an Instructions for Use Leaflet.
Nix® Ultra Lice Killing Solution is a formulated device containing dimethicone and isoparaffin. The product kills lice and their eggs by physically coating the lice and clogging the spiracles through which they breathe and conduct water exchange; the solution also lubricates the hair to facilitate removal of dead lice and nits with the supplied comb. The Nix® Ultra Lice Treatment Kit is intended for over-thecounter use.
#### 7. Indications for Use
Nix® Ultra Lice Treatment Kit is intended to kill and remove head lice and their eggs from adults and children two years of age or older.
#### 8. Summary of the Technological Characteristics
The following table compares the Nix® Ultra Lice Treatment Kit to the predicate device with respect to intended use and technological characteristics. This comparison of the devices provides detailed information demonstrating the basis for the determination of substantial equivalence. The indications between the subject device and the predicate are not identical as the subject device has an additional action of killing lice and is intended for children 2 years and older. These minor differences between the subject and predicate do not raise different question of safety or effectiveness.
| | Proposed Device | Predicate |
|---------------------------------------------------------|----------------------------------|-----------------------------------|
| Trade Name /Device Name | Nix® Ultra Lice<br>Treatment Kit | Clear™ Lice Egg<br>Remover System |
| 510(k) Number | TBD | K981147 |
| REGULATORY CLASSIFICATION | | |
| Regulatory Class | Class I | Class I |
| Name of Generic Device Type | Lice Removal Kit | Lice Removal Kit |
| Regulation | 21 CFR § 880.5960 | 21 CFR § 880.5960 |
| Product Code | LJL | LJL |
| Applicable Performance<br>Standards or Special Controls | N.A. | N.A. |
| | | Table 5-1: Comparison of Characteristics |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| DEVICE DESCRIPTION | | |
| Intended Use | To remove lice and lice eggs, from head hair. | To remove lice and lice eggs, from head hair. |
| Indications for Use | Nix® Ultra Lice Treatment Kit is intended to kill and remove head lice and their eggs from adults and children two years of age and older. | To remove lice eggs from hair |
| OTC or Rx | OTC | OTC |
| Technology | Dimethicone and isoparaffin solution physically coats the lice/nits and clogs the spiracles through which they breathe and exchange moisture; the solution also lubricates the hair to facilitate removal of dead lice and nits with the supplied comb. | A lubricant gel that facilitates the combing out of lice eggs (nits) from the hair and a comb to remove lice and nits. |
| Energy | Mechanical | Mechanical |
| Presentation of Kit | Convenience kit consisting of two components: Nix Ultra Lice Killing Solution and Nix Ultra Lice Removal Comb | Convenience kit consisting of two components: a lubricating gel and a glass-reinforced nylon fine tooth comb |
| Mode of Action | Suffocation and interference with water exchange and removal by combing | Removal by combing |
| Duration of treatments | 10 minutes for solution, plus comb out time | 3 minutes for solution, plus comb out time |
| Target Population | 2 years of age and | Minimum age not |
| | older | specified in available materials |
| Multiple Use Device | Yes (bottle contains multiple applications) | Yes (bottle contains multiple applications) |
| Sterile Device | No | No |
| Comb included | Yes | Yes |
| Gel/Solution included | Yes | Yes |
| Removal of nits | Yes | Yes |
| Ovicidal | Yes | No |
| Removal of lice | Yes | Yes |
| Pediculicidal | Yes | No |
| Use environment | Home | Home |
| Anatomical site of use | Hair/Head | Hair/Head |
{5}------------------------------------------------
{6}------------------------------------------------
#### 9. Non-Clinical Performance Data
A series of in-vitro studies of Nix® Ultra Lice Killing Solution were conducted in adult lice, nits, and lice genetically confirmed to be permethrin- and DDT-resistant (BR-HL strain). In each of these investigations, the study subjects (i.e., adult lice, nits, or resistant lice) were exposed to the test articles for a duration matching the products' instructions for use, and then subjects were washed and periodically observed for signs of life. The results of these tests clearly demonstrate the claimed performance of Nix® Ultra Lice Killing Solution against adult lice, nits and resistant lice. The mortality rates calculated in each test appear in Table 5-2 below.
| Table 5-2: In-vitro Mortality Rates | | |
|-------------------------------------|--|--|
|-------------------------------------|--|--|
| Study Subject | Test Article | Mortality Rate (%) |
|------------------------------------|---------------------------------|--------------------|
| Nits (louse eggs) | Nix Ultra Lice Killing Solution | 97 |
| | Drug comparator | 24 |
| | Control (water) | 10 |
| Adult Lice | Nix Ultra Lice Killing Solution | 100 |
| | Drug comparator | 36.67 |
| | Control (water) | 1 |
| Permethrin- and DDT-resistant lice | Nix Ultra Lice Killing Solution | 100 |
| | Drug comparator | 43.3 |
| | Control (water) | 1.7 |
Additionally, biocompatibility testing of Nix® Ultra Lice Killing Solution was conducted in accordance with the FDA guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". This testing included evaluations of cytotoxicity, skin irritation (both in humans and rabbits), corneal irritation and sensitization. The results demonstrate that the formulation is not cytotoxic. Furthermore, the product
{7}------------------------------------------------
was non-irritating in the human patch test, caused negligible irritation in rabbits, is non-significantly or weakly sensitizing and can be considered only a mild corneal irritant.
The results of the non-clinical testing support the substantial equivalence of the Lice Treatment Kit as compared to the predicate device.
#### 10. Clinical Performance Data
A randomized, controlled clinical study supports the substantial equivalence of the Nix® Ultra Lice Treatment Kit. There is also post-marketing adverse event data from the product's European sales history.
#### Clinical Study (Ref# NUHL001-10/15)
A randomized, controlled, investigator- and assessor-blinded study was conducted to demonstrate the claimed performance of Nix® Ultra head lice lotion and comb (also called Nix Ultra™ 2-in-1 Lice Treatment) in subjects ranging in age from 2 to 63 years old with confirmed head lice infestation. (Note: Nix Ultra head lice lotion is identical in composition to Nix Ultra® Lice Killing Solution.) Assessments were conducted at baseline, one day after treatment, one week later (at which point a second treatment was given) and on Day 10. In this study, Nix Ultra™ met all pre-determined endpoints. The results are presented below in Table 5-3.
| Primary Endpoint | Reference | Test | P-Value |
|-----------------------------------------|-------------|-------------|----------|
| Day 10 - All baseline infestations | | | |
| No | 24 (96%) | 23 (92%) | >0.50001 |
| Yes | 1 (4%) | 2 (8%) | |
| Secondary Endpoint | | | |
| Day 10 - For Mild and Moderate Baseline | | | |
| Infestation | | | |
| No | 19 (100%) | 16 (94.12%) | 0.47221 |
| Yes | 0 (0%) | 1 (5.88%) | |
| Day 7 - All baseline infestations | | | |
| No | 22 (88%) | 23 (88.46%) | >0.50001 |
| Yes | 3 (12%) | 3 (11.54%) | |
| Day 7 - For Mild and Moderate Baseline | | | |
| Infestation | | | |
| No | 19 (100%) | 17 (94.44%) | 0.48651 |
| Yes | 0 (0%) | 1 (5.56%) | |
| Day 1 - All baseline infestations | | | |
| No | 25 (96.15%) | 24 (92.31%) | >0.50001 |
| Yes | 1 (3.85%) | 2 (7.69%) | |
| Day 1 - For Mild and Moderate Baseline | | | |
| Infestation | | | |
| No | 19 (100%) | 17 (94.44%) | 0.48651 |
| Yes | 0 (0%) | 1 (5.56%) | |
Table 5-3: Analysis of Cure rate on Days- 10, 7 and 1 using Chi-Square Test (all baseline infestations)
No Significant difference between Reference and Test; T (Test Product) = Nix Ultra "" Lotion & Comb,
Note: The exact p-value is reported due to the small number of responses (less than five)
{8}------------------------------------------------
Both products were considered well tolerated. There was one adverse event reported during the course of the study, dryness and redness behind the right ear, for a subject randomized to the test product (Nix Ultra); this event was judged by the investigator to be mild in severity and possibly related to the test product.
#### Post-Marketing Adverse Event Data
Finally, post-marketing adverse event (AE) data supplied by the product's European marketer, Oystershell Laboratories, described only five cases involving thirteen patients in relation to 518,095 units sold during the period from June 2007 to December 2015. These AEs involved scalp and ear irritation, skin redness, desquamation, and swollen ears or eyes. All AEs were classified by Oystershell Laboratories as non-serious and temporary.
In conclusion, the results of the clinical testing and European post-marketing data demonstrate the ability of the Nix® Ultra Lice Treatment Kit to kill and remove head lice without presenting a significant risk to the end user and therefore support the substantial equivalence of the device as compared to the predicate.
#### 11. Statement of Substantial Equivalence
The Nix® Ultra Lice Treatment Kit has the same intended use as the following legally marketed lice removal device: the Clear™ Lice Egg Remover System. The non-clinical and clinical data provided in this submission demonstrate that the Nix® Ultra Lice Treatment Kit is as safe and as effective as the predicate device identified in this submission.
The Nix® Ultra Lice Treatment Kit is therefore substantially equivalent to the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.