K163399 · Boston Scientific Corporation · FAD · Jun 22, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K163399
Device Name
Tria Firm Ureteral Stent
Applicant
Boston Scientific Corporation
Product Code
FAD · Gastroenterology, Urology
Decision Date
Jun 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ureteral Stent is intended to facilitate drainage from the kidney to the bladder via placement endoscopically, fluoroscopically or during an open surgical procedure by a trained physician.
Device Story
Tria™ Firm Ureteral Stent; double-pigtail indwelling catheter; facilitates drainage from kidney to bladder. Constructed from Percuflex polymer using modified extrusion process for ultra-smooth surface topography; designed to minimize accumulation of urine calcium and magnesium salt deposits. Placed by physicians in clinical settings (endoscopic, fluoroscopic, or open surgery). Includes monofilament retrieval lines, straight stent positioner, and pigtail straightener. Provides clinical benefit by reducing salt encrustation during indwelling period. Healthcare providers use device to maintain ureteral patency; output is physical drainage path.
Clinical Evidence
Bench testing only. Evaluated bladder/renal coil length, flow rate, surface roughness, MRI safety, structural integrity (tensile/column strength), and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, muscle implantation, pyrogenicity). In vitro testing demonstrated statistically significant reduction in calcium and magnesium salt accumulation compared to competitive devices. No clinical data provided.
Technological Characteristics
Percuflex polymer construction; double-pigtail design; ultra-smooth surface topography via modified extrusion. Includes monofilament retrieval lines, straight stent positioner, and pigtail straightener. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for patients requiring drainage from the kidney to the bladder, placed by a trained physician via endoscopic, fluoroscopic, or open surgical procedures.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol composed of three stylized human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2017
Boston Scientific Corporation Christine Shoemaker Sr. Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K163399
> Trade/Device Name: Tria™ Firm Ureteral Stent Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: Mav 19, 2017 Received: May 22, 2017
Dear Christine Shoemaker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163399
Device Name Tria™ Firm Ureteral Stent
Indications for Use (Describe)
The Ureteral Stent is intended to facilitate drainage from the kidney to the bladder via placement endoscopically, fluoroscopically or during an open surgical procedure by a trained physician.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### SECTION 5
## Boston Scientific
## 510(k) SUMMARY
## 510(k) Summary for Tria™ Firm Ureteral Stent
#### A. Sponsor
Boston Scientific Corporation Urology and Pelvic Health Division 100 Boston Scientific Way Marlborough, MA 01756
#### B. Contact
Christine Shoemaker Sr. Specialist, Regulatory Affairs 508-683-4214 christine.shoemaker@bsci.com Or Lisa Sullivan Sr. Manager, Regulatory Affairs 508-683-4745 lisa.sullivan@bsci.com
Date Prepared: June 21, 2017
#### C. Device Name
Trade name: Tria™ Firm Ureteral Stent Common/usual name: stent, ureteral Classification Name: FAD - Stent, Ureteral
#### D. Predicate and Reference Devices
Contour™ Ureteral Stent (Boston Scientific) Predicate Trade name: Common/usual name: stent, ureteral Classification Name: FAD - Stent, Ureteral Premarket Notification: Contour™ (Modified Ureteral Indwelling Catheter/Stent) Ureteral Stent, K974541, Sep 02, 1998. The Contour Ureteral Stent has not been the subject of a design-related recall.
PercuflexTM Ureteral Stent (Boston Scientific) Reference 1 Trade name: Common/usual name: stent, ureteral Classification Name: FAD - Stent, Ureteral Premarket Notification: K834468, Apr 20, 1984.
Reference 2 Trade name: AdvantixTM Pancreatic Stent (Boston Scientific) Common/usual name: Biliary Catheters and Accessories Classification Name: FGE - Biliary Catheter and accessory Premarket Notification: K133700, May 14, 2014.
{4}------------------------------------------------
#### SECTION 5
Boston Scientific
#### 510(k) SUMMARY
Reference 3 Trade name: InLay Optima® Ureteral Stent (Bard) Common/usual name: stent, ureteral Classification Name: FAD - Stent, Ureteral Premarket Notification: K022447, Jan 23, 2003.
### E. Device Description
The Tria™ Firm Ureteral Stent is intended to facilitate drainage from the kidney to the bladder via placement endoscopically or fluoroscopically or during an open surgical procedure. It is constructed of the same Percuflex polymer as other stents currently marketed by BSC. Tria Firm utilizes a modified extrusion process to provide ureteral stents with an ultra-smooth surface topography.
The Tria™ Firm stent has a double pigtail design and utilizes the same monofilament retrieval lines that are used on other BSC ureteral stents. Tria Firm is packaged with a standard straight stent positioner and a pigtail straightener that are currently provided with other ureteral stents marketed by Boston Scientific.
The purpose of the Tria™ Firm Ureteral Stent is to provide physicians with a product that is aimed at addressing accumulation of urine salt deposits during indwelling. The proprietary surface technology on both the outside and inside of the stent provides maximum coverage from calcium and magnesium salt deposition.
In vitro testing conducted on the Tria™ Firm Ureteral Stent showed a statistically significant lower level of urine calcium and magnesium salt accumulation on the stent surface compared to competitive devices. Correlation of in vitro data to clinical outcomes have not been established.
#### F. Intended Use
The Ureteral Stent is intended to facilitate drainage from the kidney to the bladder via placement endoscopically, fluoroscopically or during an open surgical procedure by a trained physician.
#### G. Technological Characteristics
The proposed Tria™ Firm Ureteral Stents have the same technological characteristics, fundamental ureteral stent design, and types of materials as the predicate device. A modified extrusion process provides an ultra-smooth surface. The device is packed using a thermoformed, multi-product tray. The lidded tray is then placed in a poly/Tyvek pouch that is heat sealed and labeled. The labeled pouch is placed into a labeled shelf carton along with a DFU.
#### H. Substantial Equivalence
A direct comparison of key characteristics has been performed and demonstrates that the proposed Tria™ Firm Ureteral Stent is substantially equivalent to the predicate device in terms of intended use, technological characteristics, types of
{5}------------------------------------------------
## SECTION 5
materials and performance characteristics. The proposed Tria Firm Ureteral Stent is as safe, as effective, and performs as well as the predicate device.
### I. Performance Testing (Bench Evaluation)
Boston Scientific has conducted performance testing with samples aged at T=0 and 13-months Accelerated Aging in support of the new ureteral stent. Third party testing was performed in support of new claims for 1) minimizing the accumulation of urine calcium and magnesium salts, both with and without the presence of bacteria and 2) the proprietary surface technology on both the outside and inside of the stent providing maximum coverage from calcium and magnesium salt deposition .
The results of the testing listed below support the determination of substantial equivalence.
- Performance testing - bladder coil length, renal coil length/shape, working length, flow rate, surface roughness, MRI safety assessment
- 0 Structural integrity testing - removal force (tensile strength), retrieval line to stent shaft tensile, column strength
- Biological safety testing (biocompatibility) ●
- Sterile barrier testing ●
## J. Biocompatibility Testing
The following testing was performed to support the biocompatibility of the stents:
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
- Acute Systemic Toxicity ●
- Muscle Implantation ●
- Material Mediated Pyrogenicity ●
- Chemical analysis extractables ●
- Risk assessment of potential toxicity .
#### K. Conclusions
Based on the information and data provided in this Traditional 510(k), the TriaTM Firm Ureteral Stents are substantially equivalent in intended use, design, principle of operation, technology, materials, and performance to the predicate Contour Ureteral Stents (K974541) and are safe and effective for their intended use. Tria Firm was also demonstrated to be substantially equivalent for specific properties when compared to other reference devices (BSC Percuflex (K834468), BSC Advantix (K133700), and Bard Inlay Optima (K022447)).
<sup>1</sup> Proteus mirabilis was used as the microbial challenge due to its known urease production.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.