MAZIC Duro is a radiopaque, hybrid ceramic, resin-based composite block used for dental restorations. Intended for use by dental professionals in CAD/CAM milling procedures to create inlays, onlays, veneers, and crowns. The device is matched to the patient's tooth shade. It serves as a restorative material for anterior and posterior teeth. Benefits include providing a durable, biocompatible, and aesthetically matched restoration for patients requiring dental repair.
Clinical Evidence
Bench testing only. Performance evaluated via ISO 4049 testing including compressive strength (534.40 MPa), flexural strength (213.30 MPa), elastic modulus (10.17 GPa), radio-opacity, Vickers microhardness, water sorption, and solubility. Biocompatibility testing conducted per ISO 10993-1, including cytotoxicity (ISO 10993-5), genotoxicity (ISO 10993-3), oral mucosa irritation (ISO 10993-10), sensitization (ISO 10993-10), and acute systemic toxicity (ISO 10993-11).
Indicated for the fabrication of inlays, onlays, veneers, and crowns in dental patients.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 17, 2017
Vericom Co., Ltd. Myung-Hwan Oh R&d Director 48 Toegyegongdan 1-gil Chuncheon-si, 200-944 REPUBLIC OF KOREA
Re: K163346
Trade/Device Name: Mazic Duro Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF, POW Dated: December 28, 2016 Received: December 30, 2016
Dear Myung-Hwan Oh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Susan Runno, DDS, MA
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
Device Name MAZICTM Duro
Indications for Use (Describe)
· Fabrication of inlays, onlays, veneers and crowns
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services
Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(K) SUMMARY
# MAZIC™Duro
#### Date:
## I. SUBMITTER
Vericom Co., Ltd. 48 Toegyegongdan 1-gil, Chuncheon-si, Gangwon-do 200-944, Korea Tel: +82-31-441-2881 Fax: +82-31-441-2883
Contact Name: Myung-Hwan Oh Email: omh@vericom.co.kr
### II. DEVICE
| Name of Device: | MAZIC™ Duro |
|----------------------|----------------------------|
| Regulation Name: | Tooth shade resin material |
| Classification Name: | Tooth shade resin material |
| Regulatory Class: | 2 |
| Product Code: | EBF |
#### III. Predicate Device
K133824, CERASMART, GC AMERICA INC.
# IV. DEVICE DESCRIPTION
MAZIC™Duro, a dental restorative material, is a radiopaque, block type, hybrid ceramic that is intended for the fabrication of inlays, onlays, veneers and crowns. The device can be matched to the patient's tooth shade and is intended for use in dental CAD/CAM milling procedures.
## V. INDICATIONS FOR USE
· Fabrication of inlays, onlays, veneers and crowns
{4}------------------------------------------------
| | | Subject Device | Predicate Device | Summary |
|---------------|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | | MAZIC™ Duro | CERASMART | - |
| Manufacturer | | Vericom Co., Ltd. | GC AMERICA INC. | - |
| 510(k) Number | | New | K133824 | - |
| Intended user | | Dental professionals | Dental professionals | - |
| Similarity | Description | MAZIC™ Duro, a dental<br>restorative material, is a<br>radiopaque, block type,<br>hybrid ceramic that is<br>intended for the<br>fabrication of inlays,<br>onlays, veneers and<br>crowns. The device can<br>be matched to the<br>patient's tooth shade<br>and is intended for use<br>in dental CAD/CAM<br>milling procedures. | CERASMART is a pre-<br>cured composite block<br>for milling CAD/CAM<br>indirect restorations.<br>The milled device is used<br>for the restorations of<br>both anterior and<br>posterior teeth. | Description<br>of<br>subject device<br>and predicate<br>device is very<br>similar; pre-curd<br>composite block<br>for milling by<br>dental<br>CAD/CAM |
| | Indications for<br>Use | Fabrication of inlays,<br>onlays, veneers and<br>crowns | The product is indicated<br>for inlays, onlays,<br>veneers and full crown<br>restorations, including<br>crowns on implants | Both device are<br>used for inlays,<br>onlays, veneers<br>and crowns. |
| | Design | Pre-cured block | Pre-cured block | Both<br>devices<br>are provided in<br>pre-curd block |
| | Chemical<br>composition | Resin base | Resin base | Both<br>devices<br>are resin based<br>composite<br>block. |
| | Physical<br>and<br>performance<br>properties | ISO 4049<br>-Compressive strength:<br>534.40 MPa<br>-Flexural Strength:<br>213.30 MPa<br>-Elastic modulus: 10.17<br>GPa<br>-Radio-opacity: 2.4 step<br>wedge<br>-Vicker'smicrohardness:<br>85.28 HV0.2<br>-Water sorption: 15.51<br>µg/cm2<br>-Solubility: 1.90 µg/cm2 | ISO 4049<br>-Compressive strength:<br>501.76 MPa<br>-Flexural Strength:<br>207.91 MPa<br>-Elastic modulus: 9.15<br>GPa<br>-Radio-opacity : 2.6 step<br>wedge<br>-Vicker'smicrohardness:<br>70.97 HV0.2<br>-Water sorption: 20.67<br>µg/cm3<br>-Solubility : 2.10 µg/cm2 | Physical<br>and<br>performance<br>properties<br>of<br>the subjective<br>device<br>is<br>substantial<br>equivalent<br>to<br>predicate<br>device. |
| | Biocompatibil<br>ity | -Cytotoxicity(Agar<br>diffusion)<br>-Acute systemic<br>toxicity(Oral) | Device is biocompatible<br>when used as directed<br>by dental professionals<br>per ISO 10993-1.<br>-Acute systemic<br>toxicity(Oral) | Both<br>devices<br>are<br>biocompatible. |
| | | -Oral mucosa irritation<br>-Sensitization(GPMT)<br>-Genotoxicity(Ames) | | |
| | Sterilization | Non-sterile | Non-sterile | Same |
| | Storage<br>condition | Avoid exposure to the<br>sunlight.<br>Store at room<br>temperature(2-27°C). | Recommended for<br>optimal performance,<br>store at room<br>temperature(4-25°C /<br>39.2-77.0°F) away from<br>direct sunlight and high<br>humidity. | Both devices<br>are<br>recommended<br>at room<br>temperature<br>away from<br>sunlight. |
| | Shelf Life | 5 years | 5 years | Same |
| Difference | Shade | 30 shades(A1 HT, A2 HT,<br>A3 HT, B1 HT, B2 HT, B3<br>HT, C1 HT, C2 HT, C3 HT,<br>D2 HT, D3 HT, E HT, A1<br>LT, A2 LT, A3 LT, A3.5 LT,<br>A4 LT, B1 LT, B2 LT, B3<br>LT, B4 LT, C1 LT, C2 LT,<br>C3 LT, C4 LT, D2 LT, D3<br>LT, D4 LT, W1 LT, W2 LT) | 33 shades(A1 HT, A2 HT,<br>A3 HT, A3.5 HT, A4 HT,<br>B1 HT, B2 HT, B3 HT, B4<br>HT, C1 HT, C2 HT, C3<br>HT, C4 HT, D2 HT, D3 HT,<br>D4 HT, A1 LT, A2 LT, A3<br>LT, A3.5 LT, A4 LT, B1 LT,<br>B2 LT, B3 LT, B4 LT,<br>C1 LT, C2 LT, C3 LT, C4<br>LT, D2 LT, D3 LT, D4 LT,<br>BL) | Shade is an<br>option that is<br>selected<br>according to<br>patient's teeth<br>color. |
| | Size | -Rectangular type:<br>[12]10.2*12.2*15,<br>[14L]14.5*14.5*18<br>-Disc type:<br>Φ98*8T/10T/12T/14T,<br>Φ100*8T/10T/12T/14T,<br>Φ95*8T/10T/12T/14T | 3 sizes(12, 14, 14L) | Size is an option<br>that is selected<br>according to<br>patient's<br>application<br>region. The disc<br>type of subject<br>device provide<br>several<br>restorations |
# VI. COMPARISON OF TECHNILOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
{5}------------------------------------------------
#### VII. NON-CLINICAL PERFORMANCE TESTING
MAZIC™ Duro and the predicate device, CERASMART were tested for the nonclinical performance including compressive strength, flexural strength, elastic modulus, radio-opacity, vicker's micro hardness, water sorption, and solubility.
All of test results obtained from these tests indicated that MAZIC™ Duro is substantially equivalent to the predicate device, CERASMART, in terms of physical properties and performance.
The biocompatibility evaluation for the MAZIC™ Duro was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1:Evaluation and Testing''' May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1 : Evaluation and Testing Within a Risk Management Process" as recognized by FDA. The device of testing included the following tests:
{6}------------------------------------------------
- Cytotoxicity(Agar diffusion), ISO 10993-5[2009] ·
- Genotoxicity(Ames), ISO 10993-3[2003] •
- Oral mucous Irritation, ISO 10993-10[2010]
- Sensitization(GPMT), ISO 10993-10[2010]
- . Acute systemic toxicity, ISO 10993-11[2006]
The results of test indicated that MAZIC™ Duro is considered to be biocompatible and therefore equivalent with its predicate device, CERASMART, that considered biocompatibility per ISO 10993-1.
#### VIII. SUBSTANTIAL EQUIVALENCE DISCUSSION
The subject device is similar to predicate device in terms of indication for use, design, chemical composition, physical properties and performance etc. The subject device consists of resin base composite block, as is the predicate device. The physical properties and performance of subject device are not significantly different from the predicate device, as indicated by non-clinical performance testing.
The variety of shades and sizes are different between subject device and predicate device.
#### IX. CONCLUSIONS
The indications and technological characteristics of MAZIC™ Duro are very similar to predicate device. Therefore, MAZIC™ Duro is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.