K163341 · Siemens Medical Solutions USA, Inc. · JAK · Feb 9, 2017 · Radiology
Device Facts
Record ID
K163341
Device Name
syngo.CT Extended Functionality
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Feb 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
syngo.CT Extended Functionality is intended to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners. An interface shall enable the connection between the syngo.CT Extended Functionality software package and the interconnected CT Scanner system. Result images created with the syngo.CT Extended Functionality software package can be used to assist trained technicians or physicians in diagnosis.
Device Story
Software bundle providing advanced visualization/measurement tools for CT images; operates on syngo-compatible host systems (e.g., syngo.via VB20 or higher). Inputs: DICOM CT data (spiral/sequence scans). Functionality: 1) Vascular assessment; 2) Oncology case preparation; 3) Osteo case assessment (bone mineral content/mid-vertebral slice calculation); 4) Neuro DSA bone subtraction. Outputs: Processed images, measurements, and DICOM key images. Used in clinical settings by physicians/technicians to assist in diagnosis. Does not modify original clinical data. Supports workflow integration via interface with CT scanners; allows combination of different clinical field tools. Benefits: Enhanced visualization and quantitative analysis for specific clinical fields.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing conducted to demonstrate software integration, performance, and risk mitigation per ISO 14971. Testing confirmed that the device performs as intended and meets all software specifications.
Technological Characteristics
Software-only post-processing package. Operates on syngo-compatible platforms (e.g., syngo.via). Connectivity: DICOM standard (PS 3.1–3.20). Standards: IEC 62304 (Software Life Cycle), ISO 14971 (Risk Management), IEC 62366-1 (Usability).
Indications for Use
Indicated for use by trained technicians and physicians to perform qualitative and quantitative measurements and analysis of clinical data acquired and reconstructed by CT scanners for diagnostic purposes.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2017
Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 40 Liberty Blvd., Mail Code 65-1A MALVERN PA 19355
Re: K163341
Trade/Device Name: syngo.CT Extended Functionality Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: November 17, 2016 Received: November 29, 2016
Dear Ms. Mangum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163341
Device Name syngo.CT Extended Functionality
#### Indications for Use (Describe)
syngo.CT Extended Functionality is intended to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners.
An interface shall enable the connection between the syngo.CT Extended Functionality software package and the interconnected CT Scanner system.
Result images created with the syngo.CT Extended Functionality software package can be used to assist trained technicians or physicians in diagnosis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the word "SIEMENS" in a large, sans-serif font. The text is a light blue color. The background is white, which makes the text stand out.
## 510(K) Summary FOR SYNGO.CT EXTENDED FUNCTIONALITY
Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Date Prepared: January 24, 2017
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### l. Submitter
Importer/Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
#### Establishment Registration Number 2240869
#### Manufacturing Site
Siemens Healthcare GmbH Siemensstr. 1 D-91301 Forchheim, Germany
#### Establishment Registration Number 3004977335
#### Contact Person
Kimberly Mangum Regulatory Affairs Specialist Siemens Medical Solutions, Inc. USA 40 Liberty Boulevard Malvern, PA 19355 Phone: (610) 448 - 6477 Fax: (610) 640 - 4481 Email: kimberly.mangum@siemens.com
#### II. Device Name and Classification
510(k) Number: K163341 Product Name: syngo.CT Extended Functionality Propriety Trade Name: syngo.CT Extended Functionality Classification Name: Computed Tomography X-ray System Classification Panel: Radiology 21 CFR §892.1750 CFR Section: Device Class: Class II Product Code: 90JAK
{4}------------------------------------------------
# SIEMENS
## III. Predicate Device
#### Predicate Device:
Trade Name: syngo.CT Vascular Analysis 510(k) Number: K112020 08/18/2011 Clearance Date: Classification Name: System, X-Ray, Tomography, Computed Classification Panel: Radiology CFR Section: 21 CFR § 892.1750 Device Class: Class II Product Code: JAK
#### Predicate Device:
| Trade Name: | syngo.CT Lung CAD |
|-----------------------|-----------------------------------------------------------|
| 510(k) Number: | K143196 |
| Clearance Date: | 05/12/2015 |
| Classification Name: | Lung Computed Tomography System, Computer-Aided Detection |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR § 892.2050 |
| Device Class: | Class II |
| Product Code: | OEB |
#### Predicate Device:
- Osteo CT Option for SOMATOM CT Systems Trade Name: 510(k) Number: K971054 Clearance Date: 06/13/1997 Classification Name: System, X-Ray, Tomography, Computed Classification Panel: Radiology CFR Section: 21 CFR § 892.1750 Device Class: Class II Product Code: JAK Subsequent Product Code: KGI
### Predicate Device:
| Trade Name: | syngo Neuro DSA CT |
|-----------------------|-------------------------------------|
| 510(k) Number: | K053024 |
| Clearance Date: | 11/04/2005 |
| Classification Name: | System, X-Ray, Tomography, Computed |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR § 892.1750 |
| Device Class: | Class II |
| Product Code: | JAK |
### IV. Device Description
syngo.CT Extended Functionality is a software bundle consisting of previously cleared unmodified and modified post-processing applications that offer tools to support special clinical evaluations. syngo.CT Extended Functionality can be used to create advanced visualizations and measurements on clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners.
Depending on the clinical question, the user can select functionality which supports the explicit clinical fields as listed below. The syngo.CT Extended Functionality software
{5}------------------------------------------------
# SIEMENS
package is designed to operate on the most recent version syngo-compatible postprocessing platform, which currently supports the following four tools:
- 1. Preparation of Vascular Case for Reading Physician
- 2. Preparation of Oncology Case for Reading Physician
- 3. Preparation of Osteo Case for Reading Physician
- 4. Preparation of Neuro DSA Bone Subtraction for Reading Physician
The supported functionality can be used on any CT data if basic requirements are met (e.g. spiral or sequence scan, reconstruction kernel). The supported functionality will check to ensure the basic requirements are met and will not allow its execution or will provide a warning or info message to the user if appropriate. This check also allows combination of functionality of different clinical fields, (e.q. a vascular case can be prepared also on Neuro DSA bone subtracted data or on the same case as Lung CAD computation, etc.). Afterwards, any tool can be accessed as long as the data and viewing type allows it. For example, an evaluation of a ROI defined by a contour and two HU thresholds can be used to measure a certain area. No specific sequential workflow is required.
The original clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners will not be modified in any form. The results of the syngo.CT Extended Functionality can be stored as additional DICOM images if needed as kev images or range or images. The subject device syngo.CT Extended Functionality is designed to operate on a syngo compatible host system (e. g. syngo.via VB20 software platform or higher).
### V. Indications for Use
syngo.CT Extended Functionality is intended to provide advanced visualization tools to prepare and process medical images for diagnostic purpose. The software package is designed to support technicians and physicians in qualitative and quantitative measurements and in the analysis of clinical data that was acquired and reconstructed by Computed Tomography (CT) scanners.
An interface shall enable the connection between the syngo.CT Extended Functionality software package and the interconnected CT scanner system. Result images created with the syngo.CT Extended Functionality software package can be used to assist trained technicians or physicians in diagnosis.
#### VI. Comparison of Technological Characteristics with the Predicate Device
This subject device is a bundle software package consisting of previously cleared unmodified and modified software applications that provide advanced visualization and measurement tools. At a high-level, the subject and predicate device are based on the following same/similar technological characteristics:
| Subject Supported<br>Functionality | Predicate Device and Supported<br>Functionality | Comparison Results |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Tools and layouts for vascular<br>assessment | predicate device:<br>K112020 - syngo.CT Vascular Analysis<br>Tools and layouts for vascular<br>assessment | Supports the same<br>unmodified functionality<br>from the predicate device. |
| Display of the findings of the<br>Lung CAD algorithm | predicate device:<br>K143416 - syngo.CT Lung CAD | Supports the same<br>unmodified functionality |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. The word is displayed horizontally and is the main focus of the image.
| Subject Supported<br>Functionality | Predicate Device and Supported<br>Functionality | Comparison Results |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Display of the findings of the Lung CAD<br>algorithm | from the predicate device. |
| Provides a set of routine tools<br>for the assessment of bone<br>mineral content | predicate device:<br>K971054 - Osteo CT Option for<br>SOMATOM CT Systems<br>Provides a set of routine tools for the<br>assessment of bone mineral content | Modified to evaluate<br>images from volume scans<br>and semi-automatically<br>calculate the position and<br>orientation of mid-vertebral<br>slices (MVS). |
| Provides a set of tools for<br>assessment of the cerebral<br>vasculature by removing<br>interfering bone structures | predicate device:<br>K971054<br>Provides a set of tools for assessment of<br>the cerebral vasculature by removing<br>interfering bone structures | Supports the same<br>unmodified functionality<br>from the predicate device. |
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination. This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional) were conducted for syngo.CT Extended Functionality during product development. The modifications described in this Premarket Notification were supported with verification/validation testing. Verification testing for the Osteo feature was conducted to demonstrate successful software integration and performance in accordance to Siemens internal procedure which includes risk identification and mitigation in accordance with ISO 14971. All verification and validation testing has been completed and meets Siemens acceptance criteria. Additionally, all risks for the complete subject device have been identified and mitigated in accordingly.
#### Risk Analvsis
The risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
#### Software Verification and Validation
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.
#### Non-Clinical Testing Summary
Performance tests were conducted to test the functionality of the synqo.CT Extended Functionality. These tests have been performed to test the ability of the included features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence.
#### General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis. which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing.
{7}------------------------------------------------
# SIEMENS
syngo.CT Extended Functionality is designed to fulfill the requirements of the following safety and performance standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Publication<br>Date | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|------------------------------------------|
| 12-300 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set; PS 3.1 – 3.20 | 06/27/2016 | NEMA |
| 13-32 | Software | Medical Device Software -Software Life<br>Cycle Processes; 62304:2006 (1st<br>Edition) | 08/20/2012 | AAMI, ANSI, IEC |
| 5-40 | Software/<br>Informatics | Medical devices - Application of risk<br>management to medical devices; 14971<br>Second Edition 2007-03-01 | 08/20/2012 | ISO |
| 5-95 | General I<br>(QS/RM) | Medical devices - Part 1: Application of<br>usability engineering to medical devices<br>IEC 62366-1:2015 | 06/27/2016 | IEC |
| 19-4 | General II<br>(ES/EMC) | AAMI / ANSI ES60601-1:2005/(R)2012<br>and A1:2012, Medical electrical<br>equipment - part 1: general<br>requirements for basic safety and<br>essential performance (IEC 60601-<br>1:2005, mod) | 07/09/2014 | AAMI, ANSI |
## VIII. Conclusions
This subject device combines four advanced visualization and measurement tools into one software bundle package. The fundamental software technology which is provided within the scope of the advanced tools is already cleared and remains unchanged in comparison to the predicate devices. The Indications for Use for the subject device has been adapted to provide a more specific description of the syngo.CT Extended Functionality realized as software package including a software bundle that consists of previously cleared post-processing software applications. The conducted modifications described in this Premarket Notification were supported with verification and validation testing. The Risk analysis was completed and risk control implemented to mitigate identified hazards.
The predicate devices were cleared based on non-clinical supportive information. The results of these tests demonstrate that the predicate devices are adequate for the intended use. The comparison of technological characteristics, non-clinical performance data, and software validation demonstrates that the subject device is as safe and effective when compared to the predicate device that is currently marketed for the same intended use. For the subject device, syngo.CT Extended Functionality, Siemens used the same testing with the same workflows as used to clear the predicate device. Since both devices were tested using the same methods, Siemens believes that the data generated from the syngo.CT Extended Functionality testing supports a finding of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.