← Product Code [MBI](/productcode/MBI) · K163258

# CrossFT Knotless Suture Anchor with Disposable Driver (K163258)

_Conmed Corporation · MBI · Jan 25, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K163258

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Jan 25, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The CrossFT™ Knotless Suture Anchor with Disposable Driver is intended to reattach soft tissue to bone in orthopedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the sutured to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriative immobilization, throughout the healing period.

## Device Story

Sterile, single-use knotless suture anchor system; used in arthroscopic or open orthopedic surgery to reattach soft tissue to bone. Device consists of PEEK anchor preloaded on disposable driver. Surgeon inserts anchor into bone; suture/tape limbs are tensioned and secured to stabilize soft tissue during healing. System includes bone punch, tap, and drill bit. Provides mechanical fixation of ligaments, tendons, or joint capsules. Benefits include stabilization of damaged tissue to facilitate healing.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance testing included reliability, ultimate fixation strength, cyclic testing, biocompatibility, pyrogenicity, sterilization, shelf-life, and user validation. Bacterial endotoxin testing met established limits.

## Technological Characteristics

Materials: PEEK. Dimensions: 4.0 mm, 4.75 mm, 5.5 mm widths. Components: Suture anchor, disposable driver (two-piece handle/grip), bone punch, bone tap, drill bit. Features: Suture cleats for tensioning, threader for suture/tape loading. Sterilization: Ethylene Oxide (EtO). Single-use.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- CrossFT™ Suture Anchor with Disposable Driver ([K091549](/device/K091549.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 25, 2017

ConMed Corporation Ms. Nyrobia Freeman Regulatory Affairs Specialist 525 French Road Utica, New York 13502

Re: K163258

Trade/Device Name: CrossFT10 Knotless Suture Anchor with Disposable Driver Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: December 29, 2016 Received: January 5, 2017

Dear Ms. Freeman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K163258

Device Name

CrossFTTM Knotless Suture Anchor with Disposable Driver

Indications for Use (Describe)

The CrossFT™ Knotless Anchor with Disposable Driver is intended to reattach soft tissue to bone in orthopedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the sutured to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriative immobilization, throughout the healing period.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size:16px">❌</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside on the left. To the right of the square is the word "CONMED" in black, with the word "CORPORATION" in smaller, blue letters underneath.

# 510(k) SUMMARY

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K163258.

#### I. SUBMITTER

CONMED Corporation 11311 Concept Blvd Largo, Florida 33773

Phone: 727-399-5416 Fax: 727-399-5264

Contact Person: Nyrobia Freeman Date Prepared: November 18, 2016

#### II. DEVICE NAME

| Device Name:         | CrossFT™ Knotless Suture Anchor with Disposable Driver |
|----------------------|--------------------------------------------------------|
| Common Name:         | Nonabsorbable Suture Anchor                            |
| Classification Name: | Fastener, Fixation, Non-degradable, Soft-Tissue        |
| Regulatory Class:    | Class II, per 21 CFR Part 888. 3040                    |
| Product Codes:       | MBI                                                    |

#### III. PREDICATE/ LEGALLY MARKET DEVICE

| Device Name:  | CrossFTTM Suture Anchor with Disposable Drive |
|---------------|-----------------------------------------------|
| Company Name: | CONMED Linvatec                               |
| 510(k) #:     | K091549                                       |

#### IV. DEVICE DESCRIPTION

The CrossFT™ Knotless Suture Anchors with Disposable Driver are sterile, single use devices. The CrossFT™ Knotless Suture Anchors are manufactured from PolyEtherEtherKetone (PEEK™). The anchors are provided sterile, single use and preloaded on a disposable driver. The anchors are available in three sizes and nine configurations. (See table 1)

| Catalog Number | Description                                                                           |
|----------------|---------------------------------------------------------------------------------------|
| CFK-40         | CrossFTTM Knotless 4.0 mm Suture Anchor                                               |
| CFK-40TA       | CrossFTTM Knotless 4.0 mm Suture Anchor with one 2 mm Hi-Fi             ® Tape (Blue) |

Table 1: CrossFT™ Knotless Suture Anchors with Disposable Driver sizes and configurations

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CONMED Corporation. The logo consists of a blue square with a white abstract shape on the left, and the word "CONMED" in black capital letters on the right. Below the word "CONMED" is the word "CORPORATION" in smaller black capital letters, underlined by a thin blue line. The logo is clean and professional, and the colors are simple and effective.

| CFK-40TB   | CrossFTTM Knotless 4.0 mm Suture Anchor with one 2 mm Hi-Fi® Tape (White/Black) |
|------------|---------------------------------------------------------------------------------|
| CFK-475S   | CrossFTTM Knotless 4.75 mm Suture Anchor                                        |
| CFK-475STA | CrossFTTM Knotless 4.75 mm Suture Anchor with one 2 mm Hi-Fi Tape (Blue)        |
| CFK-475STB | CrossFTTM Knotless 4.75 mm Suture Anchor with one 2 mm Hi-Fi Tape (White/Black) |
| CFK-55S    | CrossFTTM Knotless 5.5 mm Suture Anchor                                         |
| CFK-55STA  | CrossFTTM Knotless 5.5 mm Suture Anchor with one 2 mm Hi-Fi® Tape (Blue)        |
| CFK-55STB  | CrossFTTM Knotless 5.5 mm Suture Anchor with one 2 mm Hi-Fi® Tape (White/Black) |

#### V. INTENDED USE/ INDICATIONS FOR USE

The CrossFT™ Knotless Suture Anchor with Disposable Driver is intended to reattach soft tissue to bone in orthopedic surgical procedures.

The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

#### COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The CrossFTTM Knotless Suture Anchor with Disposable Driver is substantially equivalent in design, materials, intended use, principles of operation, and technical characteristics to the predicate CrossFT™ Suture Anchors with Disposable Driver and raises no new issues of safety or effectiveness.

The similarities and difference between the predicate and proposed Nonabsorbable Suture Anchors are the following-

|                                                 | CrossFT     TM Knotless Suture Anchor with<br>Disposable Driver<br>Proposed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | ConMed Linvatec Soft Tissue to Bone System<br>(CrossFT     TM Suture Anchor with Disposable Driver)<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                    | CONMED Corporation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | ConMed Linvatec                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 510k                                            | Pending                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K091549                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                                    | MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | MBI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification                                  | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use                                    | The CrossFT     TM Knotless Suture Anchor is<br>intended to reattach soft tissue to bone in<br>orthopedic surgical procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | The CrossFT     TM Suture Anchor is intended to<br>reattach soft tissue to bone in orthopedic surgical<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indication for Use                              | The device may be used in either arthroscopic<br>or open surgical procedures. After the suture is<br>anchored to the bone, it may be used to reattach<br>soft tissue, such as ligaments, tendons, or joint<br>capsules to the bone. The suture anchor system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | The device may be used in either arthroscopic or<br>open surgical procedures. After the suture is<br>anchored to the bone, it may be used to reattach<br>soft tissue, such as ligaments, tendons, or joint<br>capsules to the bone. The suture anchor system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                 | CrossFTTM Knotless Suture Anchor with<br>Disposable Driver<br>Proposed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | ConMed Linvatec Soft Tissue to Bone System<br>(CrossFTTM Suture Anchor with Disposable Driver)<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Contraindications                               | thereby stabilizes the damaged soft tissue, in<br>conjunction with appropriate postoperative<br>immobilization, throughout the healing period.<br>*Pathological conditions of bone which would<br>adversely affect the CrossFT Knotless Suture<br>Anchor.<br>*Pathological conditions in soft tissue to be<br>repaired or reconstructed which would<br>adversely affect suture fixation.<br>*Physical conditions that would eliminate, or<br>tend to eliminate adequate implant support or<br>retard healing.<br>*Conditions which tend to limit the patient's<br>ability or willingness to restrict activities or<br>follow directions during the healing period.<br>*Attachment of artificial ligaments or other<br>implants.<br>*Foreign body sensitivity, known or suspected<br>allergies to implant and/or instrument materials.<br>*This device is not cleared for screw<br>attachment or fixation to the posterior elements<br>(pedicles) of the cervical, thoracic or lumbar<br>spine. | thereby stabilizes the damaged soft tissue, in<br>conjunction with appropriate postoperative<br>immobilization, throughout the healing period.<br>*Pathological conditions of bone which would<br>adversely affect the CrossFT Knotless Suture<br>Anchor.<br>*Pathological conditions in soft tissue to be<br>repaired or reconstructed which would adversely<br>affect suture fixation.<br>*Physical conditions that would eliminate, or tend<br>to eliminate adequate implant support or retard<br>healing.<br>*Conditions which tend to limit the patient's ability<br>or willingness to restrict activities or follow<br>directions during the healing period.<br>*Attachment of artificial ligaments or other<br>implants.<br>*Foreign body sensitivity, known or suspected<br>allergies to implant and/or instrument materials.<br>*This device is not cleared for screw attachment or<br>fixation to the posterior elements (pedicles) of the<br>cervical, thoracic or lumbar spine. |
| Handle                                          | Two pieces that consist of a handle and grip                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | One piece handle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Anchor<br>Dimensions                            | 4.0 mm, 4.75 mm and 5.5 mm in width<br>14.5" inches in length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 4.5mm, 5.5mm and 6.5mm in width<br>9.5" inches in length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Suture and/or<br>Suture Tapes<br>configurations | CrossFTTM Knotless 4.0 mm Suture Anchor<br>CrossFTTM Knotless 4.0 mm Suture Anchor<br>with one 2 mm Hi-Fi Tape (White/Black)<br>CrossFTTM Knotless 4.75 mm Suture Anchor<br>CrossFTTM Knotless 4.75 mm Suture Anchor<br>with one 2 mm Hi-Fi Tape (Blue)<br>CrossFTTM Knotless 4.75 mm Suture Anchor<br>with one 2 mm Hi-Fi Tape (White/Black)<br>CrossFTTM Knotless 5.5 mm Suture Anchor<br>CrossFTTM Knotless 5.5 mm Suture Anchor<br>with one 2 mm Hi-Fi Tape (Blue)<br>CrossFTTM Knotless 5.5 mm Suture Anchor<br>with one 2 mm Hi-Fi Tape (White/Black)                                                                                                                                                                                                                                                                                                                                                                                                                                    | Cross FT Anchor with two #2 Hi-Fi Suture, 4.5mm<br>Cross FT Anchor with three #2 Hi-Fi Suture,<br>4.5mm<br>Cross FT Anchor with two # 2 Hi-Fi Suture,<br>5.5mm<br>Cross FT Anchor with three # 2 Hi-Fi Suture,<br>5.5mm<br>Cross FT Anchor with two #2 Hi-Fi Suture,<br>6.5mm<br>Cross FT Anchor with two #2 Hi-Fi Suture,<br>6.5mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Threader                                        | Used to thread Hi-Fi® Tape limbs or Hi-Fi® #2<br>suture limbs through the anchor and driver<br><br>For the 4.0 mm anchor, a maximum of two (2)<br>Hi-Fi® Tape limbs can be loaded.<br><br>For the 4.75 mm and 5.5 mm anchors, a<br>maximum of two (2) Hi-Fi® Tape limbs plus<br>two (2) Hi-Fi® #2 suture limbs, or six (6) Hi-<br>Fi® #2 suture limbs can be loaded.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Cleat                                           | Available with two (2) suture cleats. The first<br>cleat is part of the handle design and located on<br>top of the handle. This cleat is used to wrap the<br>preloaded suture and/or suture type stands<br>around for packaging purposes. The second                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The suture cleat is used to hold the suture<br>strands while tensioning the anchor and for<br>packaging purpose                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                 | CrossFTTM Knotless Suture Anchor with<br>Disposable Driver<br>Proposed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | ConMed Linvatec Soft Tissue to Bone System<br>(CrossFTTM Suture Anchor with Disposable Driver)<br>Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                 | cleat is grey in color and has slotted indents on<br>each side to wrap the suture strands around. The<br>second cleat is also used to maintain tension on<br>the suture strands while the anchor is driven<br>into the bone.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Packaging                                       | Pouch<br>Tray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Pouch<br>PETG retaining clip<br>Tyvek and Nylon/PE Film pouch                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Single Use/<br>Reusable                         | Sterile, Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Sterile, Single-Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Sterilization                                   | Eto                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Eto                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Components                                      | Bone Punch<br>Bone Tap<br>Drill Bit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Bone Punch<br>Bone Tap                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Shelf-Life                                      | 18 months                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 5 year                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside, followed by the word "CONMED" in black, bold letters. Below the word "CONMED" is the word "CORPORATION" in smaller, blue letters with a blue line above it.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Conmed Corporation. The logo consists of a blue square with a white curved shape inside on the left. To the right of the square is the word "CONMED" in black, with a blue line underneath, and the word "CORPORATION" in black below the line.

# PERFORMANCE DATA

Testing has been completed to demonstrate that the CrossFT™ Knotless Suture Anchor with Disposable Driver performs as intended and is substantially equivalent to the predicate device. The bacterial endotoxin testing was conducted and met the endotoxin limits.

Completed testing includes the following:

- . Reliability
- Packaging .
- Ultimate Fixation Strength ●
- Cyclic ●
- Sterilization ●
- Pyrogenicity
- Verification Testing ●
- . Transportation
- Biocompatibility ●
- User Validation ●
- Shelf-life ●

# VIII. CONCLUSION

The CrossFTTM Knotless Suture Anchor with Disposable Driver is either substantially equivalent or identical in design, materials, intended use, principles of operation, and technical characteristics to the predicate CrossFT™ Suture Anchor with Disposable Driver. Based upon the findings of our performance testing, the differences present no new issues of safety and efficacy and the CrossFT™ Knotless Suture Anchor with Disposable Driver is substantially equivalent to the CrossFT™ Suture Anchor with Disposable Driver (K091549).

---

**Source:** [https://fda-staging.innolitics.com/device/K163258](https://fda-staging.innolitics.com/device/K163258)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).