The Dual Mode Infrared Thermometer (Model DM-IR200) is intended for the intermittent measurement and monitoring of human body temperatures. It is for non-professional use by consumers at home. The ear canal mode is indicated for use by people above one year old and the skin/forehead mode is indicated for use by people of all ages.
Device Story
Hand-held, battery-powered infrared thermometer; measures thermal infrared energy from ear canal or forehead skin. Used by consumers at home for intermittent body temperature monitoring. Device converts detected infrared radiation into temperature readings displayed on LCD; includes buzzer for alerts. Provides 20-set memory storage. Enables non-professional users to track temperature; aids in identifying fever or monitoring health status at home.
Clinical Evidence
Clinical study of 105 subjects (including infants <1 year, children 1-5 years, and patients >5 years) with >30% febrile subjects. Evaluated clinical bias and repeatability per ASTM E1965. Results: clinical bias ≤ 0.06°C (uncertainty ≤ 0.02°C); clinical repeatability 0.1°C. Demonstrated substantial clinical agreement with predicate devices.
Technological Characteristics
Infrared radiation detection; dual-mode (ear/forehead). Powered by two 1.5V AAA batteries. LCD display with buzzer. Memory capacity: 20 sets. Complies with ASTM E1965, IEC 60601-1, IEC 60601-1-2, ISO 10993-1/5/10, IEC 62304, and ISO 14971. Biocompatible materials per ISO 10993.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature in home settings. Ear canal mode for individuals >1 year old; skin/forehead mode for all ages.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Microlife Dual Mode Thermometer, Model IFR1MJ1 (K113141)
easystem duo, Dual Thermometer, Forehead/Ear, Model BT-021 (K033077)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, with the profiles overlapping each other. The profiles are positioned above a set of three wavy lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the top of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2017
Intelliworks, LLC Dr. Jessica Leonardi R&D Director 4910 Wright Road, Suite 120 Stafford, Texas 77477
Re: K163256
Trade/Device Name: Dual Mode Infrared Thermometer. Model DM-IR200 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: March 2, 2017 Received: March 8, 2017
Dear Dr. Jessica Leonardi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina
Kiang-S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163256
Device Name
Dual Mode Infrared Thermometer, Model DM-IR200
Indications for Use (Describe)
The Dual Mode Infrared Thermometer (Model DM-IR200) is intentittent measurement and monitoring of human body temperatures. It is for non-professional use by consumers at home. The ear canal mode is indicated for use by people above one year old and the skin/forehead mode is indicated for use by people of all ages.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 2. 510(k) Summary K163256
#### 1. Submitter's Information
| Name: | Intelliworks LLC |
|----------|--------------------------------------------------------|
| Address: | 4910 Wright Road, Ste 120<br>Stafford, TX 77477<br>USA |
| Contact: | Dr. Jessica Leonardi, R&D Director |
| Phone: | 832-715-9539 |
| Email: | jess@innovogroups.com |
Date Summary Prepared: November 15, 2016
#### 2. Proposed Device
Name: Dual Mode Infrared Thermometer, Model DM-IR200 Common Name: Ear Canal IR Thermometer; Skin IR Thermometer Classification Name: Clinical Electronic Thermometer Review Panel: General Hospital Product Code: FLL Device Class: II Regulation Number: 21 CFR 880.2910
#### 3. Predicate Devices
Predicate device #1: Microlife Dual Mode Thermometer, Model IFR1MJ1 510(k) Number: K113141
Predicate device #2: easystem duo, Dual Thermometer, Forehead/Ear, Model BT-021 510(k) Number: K033077
## 4. Device Description
The Dual Mode Infrared Thermometer (Model DM-IR200) is a hand-held, non-sterile, reusable, battery powered device designed to measure human body temperature via one of two body sites: the ear canal or skin of the forehead.
The DM-IR200 thermometer utilizes infrared technology to measure the natural thermal infrared energy emitted from the surface of the forehead or from the eardrum tissue.
## 5. Indications for Use
The Dual Mode Infrared Thermometer (Model DM-IR200) is intended for the intermittent measurement and monitoring of human body temperatures. It is for non
{4}------------------------------------------------
professional use by consumers at home. The ear canal mode is indicated for use by people above one year old and the skin/forehead mode is indicated for use by people of all ages.
#### 6. Comparison to Predicate Devices and Substantial Equivalence
The Dual Mode Infrared Thermometer (Model DM-IR200) is substantially equivalent to the predicate devices, which uses the same fundamental technology, is similar in design and has the same intended use.
The proposed device and predicate devices all utilize an infrared technology to measure natural infrared energy emitted from the skin surface of the forehead or from the eardrum tissue.
The proposed device and predicate devices are intended for the intermittent measurement and monitoring of human body temperatures by detecting body temperatures from the forehead or auditory canal. The proposed and predicate devices are designed for nonprofessional use by consumers at home. The intended population of the proposed device is all ages for the forehead mode and above one year old for the ear mode. There is a minor difference with the predicate devices with intended population of all ages.
The proposed and predicate devices mainly differ by the measurement range, memory sets, power supply and storage condition. A side-by-side comparison of the proposed device and the predicate device is included in the 510(k) submission.
| Characteristics | Predicate Device #1 | Predicate Device #2 | Proposed Device | |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------|
| 510(k) Number | K113141 | K033077 | K163256 | |
| Device name | Microlife Dual<br>Mode Thermometer,<br>Model IFR1MJ1 | Easystem duo, Dual<br>Thermometer,<br>Forehead/Ear, Model<br>BT-021 | Dual Mode Infrared<br>Thermometer, Model<br>DM-IR200 | |
| Indications for use | The device is | The device is intended | The device is intended | |
| | intended for the<br>intermittent<br>measurement and | for the intermittent<br>measurement and<br>monitoring of human | for the intermittent<br>measurement and<br>monitoring of human | |
| | monitoring of | body temperature in | body temperatures. It is | |
| | human body | the home. It is | for non-professional use | |
| | temperatures. The | intended for use on | by consumers at home. | |
| | device is indicated | people of all ages. | The ear mode is | |
| | for use by people | | indicated for use by | |
| | above 12 years old<br>to geriatric for ear<br>mode and from<br>newborn to geriatric<br>for forehead mode in<br>the home. | | people above one year<br>old and the forehead<br>mode is indicated for use<br>by people of all ages. | |
| Measurement<br>method | Infrared radiation<br>detection | Infrared radiation<br>detection | Infrared radiation<br>detection | |
| Measurement mode | Dual mode – Ear<br>and Forehead mode | Dual mode – Ear and<br>Forehead mode | Dual mode – Ear and<br>Forehead mode | |
| Measurement<br>range | Ear mode:<br>32.0°F-212.0°F<br>(0.0°C-100.0°C)<br>Forehead mode:<br>93.2°F-107.9°F<br>(34.0°C-42.2°C) | Ear mode:<br>32.0°F-212.0°F<br>(0.0°C-100.0°C)<br>Forehead mode:<br>71.6°F-109.4°F<br>(22.0°C-43.0°C) | Ear mode:<br>32.0°F-212.0°F (0.0°C-<br>100.0°C)<br>Forehead mode:<br>64.4°F-107.6°F (18.0°C-<br>42.0°C) | |
| | Measurement<br>accuracy | ± 0.4°F /± 0.2°C | ± 0.4°F /± 0.2°C | ± 0.4°F / ± 0.2°C |
| | Display resolution | 0.1°F/0.1°C | 0.1°F/0.1°C | 0.1°F/0.1°C |
| | Signal output and<br>display | LCD, Buzzer | LCD | LCD, Buzzer |
| °F/°C switchable | Yes | Yes | Yes | |
| Memory | 12 sets | 10 sets | 20 sets | |
| Power supply | One 3V CR2032<br>battery | One 3V CR2032<br>battery | Two 1.5V AAA batteries | |
| Operating<br>conditions | Temperature:<br>50.0°F-104.0°F<br>(10.0°C-40.0°C)<br>Humidity: 15%-95%<br>RH | Temperature:<br>60.8°F-104.0°F<br>(16.0°C-40.0°C)<br>Humidity: 15%-95%<br>RH | Temperature:<br>50.0°F-104.0°F (10.0°C-<br>40.0°C)<br>Humidity: 15%-95% RH | |
| Storage conditions | Temperature:<br>-13.0°F to +131.0°F<br>(-25.0°C to<br>+55.0°C)<br>Humidity: 15%-95%<br>RH | Temperature:<br>-4.0°F to +122.0°F (-<br>20.0°C to +50.0°C)<br>Humidity: 15%-95%<br>RH | Temperature:<br>-4.0°F to +122.0°F (-<br>20.0°C to +50.0°C)<br>Humidity: 15%-95% RH | |
Comparison of basic technological characteristics
{5}------------------------------------------------
## 7. Performance Summary
{6}------------------------------------------------
The Dual Mode Infrared Thermometer (Model DM-IR200) conforms to applicable standards that include:
ASTM E1965 Standard specification for infrared thermometers for intermittent determination of patient temperature
IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility
ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process
ISO 10993-5 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity
ISO 10993-10 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
IEC 62304 Medical device software - Software life cycle processes
ISO 14971 Medical devices – Application of risk management to medical devices
Guidance document includes the FDA Guidance on the Content of Premarket Notification 510(k) Submissions for Clinical Electronic Thermometers.
#### 8. Clinical Testing
A clinical study was performed to determine the clinical accuracy and to provide comparison with predicate device. The three groups of subjects being tested were: 1) infants under 1 year old, 2) children between 1 and 5 years old, and 3) patients over 5 vears old.
The study included 105 subjects, with 33.3% febrile subjects. For each age group, the ratio of febrile subjects was more than 30%. The study excluded subjects with medical conditions such as inflammation at the measuring sites and subjects using medications known to affect body temperature. From each test site, a total of 105 data sets were collected.
The clinical accuracy of the proposed device was evaluated by determining two kinds of errors in accordance with requirements of ASTM E1965 - clinical bias with stated uncertainty and clinical repeatability. The clinical test results showed that there is substantial clinical agreement and repeatability between the proposed device and
{7}------------------------------------------------
predicate devices. The calculated clinical bias did not exceed 0.06°C, with uncertainty that did not exceed 0.02°C. The clinical repeatability was 0.1°C.
#### 9. Conclusion
The Dual Mode Infrared Thermometer (Model DM-IR200) has the same intended use and similar technological characteristics as the predicate devices. In addition, compliance with applicable voluntary standards, as well as performance testing demonstrate that any differences in technological characteristics do not raise any new questions of safety and/or effectiveness. Therefore, we conclude that the Dual Mode Infrared Thermometer (Model DM-IR200) is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.