K163251 · Blue Sky Bio · EIH · Aug 4, 2017 · Dental
Device Facts
Record ID
K163251
Device Name
Life Essence Universal Porcelain System
Applicant
Blue Sky Bio
Product Code
EIH · Dental
Decision Date
Aug 4, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
The Life Essence Universal Porcelain System is a dental porcelain material to be used in conjunction with metal, zirconia, or pressable ceramic framework in the construction of crowns and/or bridgework and veneers.
Device Story
Life Essence Universal Porcelain System is a dental porcelain material used by dental technicians to fabricate crowns, bridges, and veneers. The system includes various components such as pressable ingots, opaque pastes/powders, dentins, incisal powders, stains, and glaze materials. Technicians apply these materials over metal, zirconia, or pressable ceramic frameworks. The primary function is to provide a restorative aesthetic layer that is thermally compatible with the underlying framework to prevent cracking or damage. The device is used in a dental laboratory setting by trained technicians to create custom prosthetics for individual patients. The final restoration is placed by a dentist to restore dental function and aesthetics.
Clinical Evidence
Bench testing only. Testing was conducted in accordance with ISO 6872:2008 (Dentistry-Ceramic Materials) to evaluate flexural strength, solubility, and glass transition temperature. No clinical data was required or provided.
Technological Characteristics
Dental porcelain system consisting of powders, pastes, and liquids. Materials are categorized by class (1-8) per ISO 6872:2008. Key characteristic is the adjustment of the coefficient of thermal expansion to match metal or zirconia frameworks. Components include pressable pellets, opaque pastes/powders, dentins, incisal powders, stains, and glaze materials.
Indications for Use
Indicated for use as a dental porcelain material in conjunction with metal, zirconia, or pressable ceramic frameworks for the construction of crowns, bridgework, and veneers in dental patients.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other. The profiles are connected and form a wing-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 4, 2017
Blue Sky Bio % Angela Blackwell Senior Consultant Blackwell Device Consulting 888 E. Belvedere Rd., Suite 212 Grayslake, Illinois 60030
Re: K163251
Trade/Device Name: Life Essence Universal Porcelain System Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: June 27, 2017 Received: July 5, 2017
Dear Angela Blackwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Andrew I. Steen -S
for Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K163251
Device Name
Life Essence Universal Porcelain System
Indications for Use (Describe)
The Life Essence Universal Porcelain System is a dental porcelain material to be used in conjunction with metal, zirconia, or pressable ceramic framework in the construction of crowns and/or bridgework and veneers.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
#### 510(k) Summary
#### Life Essence Universal Porcelain System
Blue Sky Bio LLC 888 E Belvidere Rd Suite 212 Grayslake. IL 60030
718-376-0422
#### www.blueskybio.com
Contact Person: Albert Zickmann
azickmann@blueskybio.com
Date Prepared: August 4, 2017
| General Information | |
|---------------------|--------------------------------------------|
| Trade Name | Life Essence Universal Porcelain System |
| Common Name | Porcelain powder for dental use |
| Predicate 510(k) | Luminesse Porcelain K140848 |
| Product Code | EIH |
| Regulation No. | 872.6660 Porcelain Powder for Clinical Use |
| Classification | Class II |
#### Device Description:
Life Essence Universal Porcelain is dental porcelain used by dental technicians to create crowns, bridges, and veneers. It consists of three categories: low fusing, and zirconia porcelain. The application, indication, and performance is the same for all three categories; therefore, it is a porcelain system. It includes Pressable Ingots, Opaque Pastes, Opaque Powders, Opacious Dentins, Dentins, Incisal Powders, Stains, Incisal Transluscents, Dentin Modifiers, Correction Powder, Glaze Powder, Glaze Liquid, Spray Glaze, Modeling Liquid, and Opaque Liquid. The dental technician will use various components of the system to create the specific, desired dental prosthetic, for the sole use of individual dental patients. It is for prescription use only.
#### Indications for Use:
The Life Essence Universal Porcelain System is a dental porcelain material to be used in conjunction with metal, zirconia, or pressable ceramic framework in the construction of crowns and/or bridgework and veneers.
#### Primary Predicate Indications for Use:
The Luminesse Porcelain System is a dental porcelain material to be used in conjunction with metal, zirconia, or pressable ceramic framework in the construction of crowns and/or bridgework and veneers.
{4}------------------------------------------------
# Technological Characteristics:
Life Essence Universal Porcelain is used to veneer or press over the appropriate base materials (metal or zirconia). In order to prevent cracking and avoid damage of the restoration due to the thermally-induced stress, the average coefficient of thermal expansion has been adjusted to a coefficient of thermal expansion of the appropriate base materials, which allows fusion to the base materials. This is the main technological characteristic of the dental porcelains in general, and it is shared by the subject device and predicate device. Bench testing (according to ISO 6872:2008 Dentistry-Ceramic Materials) to determine flexural strength, solubility and glass transition temperature was provided. See tables below for the specific product types and what class they are according to the standard.
| Class | Dental Powder | Life Essence Low-Fusing Porcelain Powder |
|---------|----------------------|------------------------------------------|
| 1 | Core ceramic powder | Opaque Powder, Dentin Modifier |
| 2 | Dentin | Dentin Powder, Opacious Dentin |
| 3 | Enamel | Incisal |
| 4 | Neck material | Margin |
| 5 | Transparent Material | Incisal Translucent |
| 6 | Stains | Universal Stain Shades |
| 7 | Add-on material | Correction Powder |
| 8 | Glaze materials | Universal Glaze |
| Type 2, | Core ceramic powder | Pressable Pellets |
| class 1 | | |
| Class | Dental Powder | Life Essence High-Fusing Porcelain Powder |
|-------|----------------------|--------------------------------------------|
| 1 | Core ceramic powder | Opaque Powder |
| 1.1 | Core ceramic paste | Opaque Paste |
| 2 | Dentin | Dentin Powder |
| 2.1 | Dentin Modifier | Opacious Dentin, Dentin Modifier, Gingival |
| 3 | Enamel | Incisal |
| 4 | Neck Material | Margin, Neck Translucent |
| 5 | Transparent material | Incisal Translucent |
| 6 | Stains | Universal Stain Shades |
{5}------------------------------------------------
| 7 | Add-on material | Correction Powder |
|---|-----------------|-------------------|
| 8 | Glaze materials | Universal Glaze |
| Class | Dental Powder | Life Essence Zirconia Porcelain Powder |
|-------|----------------------|----------------------------------------|
| 1 | Core ceramic powder | Dentin Modifier |
| 2 | Dentin | Dentin Powder, Gingival |
| 3 | Enamel | Incisal |
| 4 | Neck material | Margin, Neck Translucent |
| 5 | Transparent Material | Opal Translucent, Incisal Translucent |
| 6 | Stains | Universal Stain Shades |
| 7 | Add-on material | Correction Powder |
| 8 | Glaze materials | Universal Glaze |
## Conclusion:
All components found in Life Essence Universal Porcelain are identical to components in the predicate device system. It has the same technological characteristics, chemical composition, manufacturing process, and same indications for use as the predicate devices. Components of this product have not changed in any way from the predicate that would adversely affect biocompatibility; therefore, it is determined that no biocompatibility testing is necessary for this product.
Life Essence Universal Porcelain, as designed and manufactured, is substantially equivalent to the predicate device in terms of indications for use, chemical composition, manufacturing and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.