The APVolution System is used to fabricate prostheses for missing/damaged teeth. The APVolution pressable ceramic ingots are pressed onto metal alloy (APVolution M) or zirconia (APVolution Z) frames by dental technicians to fabricate ceramic restorations. APVolution M and APVolution Z ceramic layering powders and/or pastes are used to layer or build up the pressed ceramic to final tooth morphology. APVolution S ceramic ingots are used to fabricate all-ceramic restorations. AP stains and glaze are used with the APVolution ceramic ingots, powders and/or pastes during the fabrication process to achieve the correct shade.
Device Story
APVolution System comprises silica glass ceramic materials (ingots, powders, pastes, stains, glazes) for dental restoration fabrication. Dental technicians use pressable ingots (APVolution M, Z, S) to create restorations on metal alloy or zirconia frames; layering powders/pastes build final tooth morphology; stains/glazes achieve aesthetic shade matching. Used in dental laboratories; output is a custom-fabricated dental prosthesis. Benefits include restoration of tooth function and aesthetics for patients with missing or damaged teeth.
Clinical Evidence
Bench testing only. Compliance with ISO 6872 (Dentistry - Ceramic Materials) demonstrated via physical/chemical property testing (flexural strength, thermal expansion, solubility, radioactivity). Biocompatibility confirmed via ISO 10993-5 cytotoxicity testing (L929 cell monolayer, 7-day extraction). No clinical data presented.
Technological Characteristics
Silica glass ceramic materials. Components: ingots (Type I/II, Class 1a and Type II, Class 4b per ISO 6872), powders, pastes, stains, glazes. Fabrication via pressing and layering. Non-sterile, single-use. Compliant with ISO 6872 and ISO 7405.
Indications for Use
Indicated for dental technicians to fabricate ceramic restorations, including all-ceramic, ceramic-metal, or ceramic-zirconia prostheses, for patients with missing or damaged teeth.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 5, 2017
Aesthetic-Press, LLC % Pamela Papineau President Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, Massachusetts 01432
Re: K163228
Trade/Device Name: ApVolution System Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: Class II Product Code: EIH Dated: April 25, 2017 Received: May 2, 2017
Dear Pamela Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mary S. Runner -A
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
APVolution ᠪြို့နေရွာTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: 19MB498! 8910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K163228
Device Name
APVolution System
Indications for Use (Describe)
The APVolution System is used to fabricate prostheses for missing/damaged teeth. The APVolution pressable ingots are pressed onto metal alloy (APVolution M) or zirconia (APVolution Z) frames by dental technicians to fabricate ceramic restorations. APVolution M and APVolution Z ceramic layering powders and/or pastes are used to layer or build up the pressed ceramic to final tooth morphology. APVolution S ceramic ingots are used to create all-ceramic restorations. AP Stains and Glaze are used during the fabrication process to achieve the correct shade.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif; font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif; font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Appendix A - 510(k) Summary (revised) | K163228 |
|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Date: | April 25, 2017 |
| Owner's Name: | Aesthetic-Press, LLC |
| Address: | 6613 State Road 54<br>New Port Richey, FL 34653 |
| Telephone: | 415-692-5372 |
| Fax: | 415-723-7075 |
| Contact Person: | Pamela Papineau, RAC (US, EU, CAN)<br>Delphi Medical Device Consulting, Inc.<br>5 Whitcomb Avenue<br>Ayer, MA 01432 |
| Telephone: | (978) 772-3552 |
| Fax: | (978) 796-5460 |
| Subject Device: | |
| Trade Name: | APVolution System |
| Common Name: | Dental Ceramic |
| Product Code: | EIH |
| FDA Regulation: | 21 CFR 872.6660 - Porcelain Powder for Clinical Use |
| Device Classification: | Class II |
| Predicate Device: | |
| Product Name: | DCceram System |
| Common Name: | Dental Ceramic |
| Product Code: | EIH |
| FDA Regulation: | 21 CFR 872.6660 - Porcelain Powder for Clinical Use |
| Device Classification: | Class II |
| Premarket Notification: | K141400 |
| Indications for Use | |
| The APVolution System is used to fabricate prostheses for missing/damaged teeth. The | |
The APVolution aged teeth licate pro 18/ da pressable ceramic ingots are pressed onto metal alloy (APVolution M) or zirconia (APVolution Z) frames by dental technicians to fabricate ceramic restorations. APVolution M and APVolution Z ceramic layering powders and/or pastes are used to layer or build up the pressed ceramic to final tooth morphology. APVolution S ceramic ingots are used to create all-ceramic restorations. AP Stains and Glaze are used during the fabrication process to achieve the correct shade.
#### Device Description:
The APVolution System consists of several silica glass ceramic materials used to create allceramic, ceramic-metal or ceramic-zirconia dental restorations. The APVolution System includes
{4}------------------------------------------------
three ceramic ingots for use in the fabrication of all-ceramic (APVolution S), press-to-metal (APVolution M) or press-to-zirconia (APVolution Z) restorations. The system also includes powders and pastes for use in build-up and shaping of the final restoration, as well as shades and glazes for use in achieving acceptable aesthetics. The APVolution System components are listed below in the Substantial Equivalence summary table.
### Substantial Equivalence
The Aesthetic-Press APVolution System is substantially equivalent to the Ceramay DCceram System (K141400). Substantial equivalence, which is summarized below, is based on indications for use and device physical and technological characteristics.
| | Aesthetic-Press APVolution System<br>(current submission) | Ceramay DCceram System<br>(K141400) |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common/Usual Name | Dental Ceramic | Dental Ceramic |
| Device Class | Class II | Class II |
| Product Code / Regulation | EIH / 21 CFR 872.6660 | EIH / 21 CFR 872.6660 |
| Regulation Name | Porcelain Powder for Clinical Use | Porcelain Powder for Clinical Use |
| Prescription Use | Rx Only | Rx Only |
| Indications for Use Statement | The APVolution System is used to<br>fabricate prostheses for<br>missing/damaged teeth. The<br>APVolution pressable ceramic ingots<br>are pressed onto metal alloy<br>(APVolution M) or zirconia<br>(APVolution Z) frames by dental<br>technicians to fabricate ceramic<br>restorations. APVolution M and<br>APVolution Z ceramic layering<br>powders and/or pastes are used to<br>layer or build up the pressed ceramic<br>to final tooth morphology.<br>APVolution S ceramic ingots are used<br>to fabricate all-ceramic restorations.<br>AP stains and glaze are used with the<br>APVolution ceramic ingots, powders<br>and/or pastes during the fabrication<br>process to achieve the correct shade. | The DCceram System is used to<br>fabricate prostheses for<br>missing/damaged teeth. The DCceram<br>pressable ceramic ingots are pressed<br>onto metal alloy (APVolution M) or<br>zirconia (APVolution Z) frames by<br>dental technicians to fabricate ceramic<br>restorations. DCceram 12.5 and<br>DCceram 9.2 ceramic layering<br>porcelains and shades are used to layer<br>or build up the pressed ceramic to final<br>tooth morphology and shade on the<br>particular framework (DCceram 9.2<br>zirconia substructures; DCceram 12.5<br>metal substructures). The DCceram<br>conceptPress ceramic ingots are used to<br>fabricate all-ceramic restorations.<br>DCceram conceptArt stains and shades<br>may be used with DCceram<br>conceptPress or DCceram 9.2 ceramics. |
| System Components | APVolution M ceramic ingots<br>APVolution Z ceramic ingots<br>APVolution S ceramic ingots<br>APVolution M porcelain powders<br>APVolution Z porcelain powders<br>APVolution M opaque ceramic paste<br>AP stains & glazes | DCceram 9.2 ceramic ingots<br>DCceram 12.5 ceramic ingots<br>conceptPress ceramic ingots<br>DCceram 9.2 liner/layering porcelain<br>DCceram 12.5 liner/layering porcelain<br>DCceram 12.5 opaque ceramic paste<br>conceptArt stains, shades & glazes |
| Range of Vita® Tooth Shades | A1 - D4 | A1 - D4 |
{5}------------------------------------------------
| | Aesthetic-Press APVolution System<br>(current submission) | Ceramay DCceram System<br>(K141400) |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| ISO 6872 Type I/II, Class 1a<br>Materials | APVolution M & APVolution<br>ceramic ingots | DCceram 9.2 & DCceram 12.5<br>ceramic ingots |
| | APVolution M & APVolution Z<br>porcelain powders | DCceram 9.2 & DCceram 12.5<br>porcelain powders |
| | APVolution M opaque ceramic paste<br>AP stains & glazes | DCceram 12.5 opaque ceramic paste<br>conceptArt stains, shades & glazes |
| ISO 6872 Type II, Class 4b<br>Materials | APVolution S ceramic ingots | conceptPress ceramic ingots |
| | | |
| ISO 7405 Biocompatibility | ISO 10993-5 cytotoxicity test: L929<br>cell monolayer proliferation/inhibition<br>during contact with extended (7 day)<br>extraction using DMSO in complete<br>cell culture medium | Meets requirements of ISO 7405<br>Dentistry – Evaluation of<br>Biocompatibility of Medical Devices<br>Used in Dentistry |
| ISO 6872 Performance Testing | • Identification of Types & Classes<br>of System Components<br>• Uniformity<br>• Freedom from Extraneous<br>Materials<br>• Mixing & Condensation Properties<br>(Class I Ceramics)<br>• Physical & Chemical Properties<br>• Flexural Strength<br>• Linear Thermal Expansion<br>Coefficient<br>• Glass Transition Temperature<br>• Chemical Solubility<br>• Radioactivity | Meets performance testing<br>requirements of ISO 6872 Dentistry –<br>Dental Ceramics |
| Sterile Device | No | No |
| Prescription Only | Yes | Yes |
| Single Use | Yes | Yes |
## Performance Testing:
The APVolution System components have been tested to confirm compliance with ISO 6872 Dentistry - Ceramic Materials. This Abbreviated 510(k) submission includes a Declaration of Conformity to the performance testing methods and acceptance criteria defined in ISO 6872. The specific inspections and tests included in this declaration are listed in the table above. This submission also includes biocompatibility test reports to demonstrate that the APVolution System components meet the biocompatibility requirements described in ISO 7405:2008 Dentistry -Evaluation of biocompatibility of medical devices used in dentistry. Biocompatibility testing consisted of ISO 10993-5 cytotoxicity testing where the proliferation or inhibition of an L929 cell monolayer was evaluated following direct contact with an extended-time (7 day) extract of the system components using DMSO in complete culture medium. The APVolution components met all requirements of the ISO 6872 and ISO 7405 standards.
## Conclusion:
The Aesthetic-Press APVolution System has been demonstrated to be substantially equivalent to the predicate device as demonstrated through comparison of the system indications for use,
{6}------------------------------------------------
description of system components, biocompatibility testing in accordance with ISO 7405, performance testing in accordance with ISO 6872, and device labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.